Penumbra, Inc. v. RapidPulse, Inc.

Penumbra, Inc. v. RapidPulse, Inc. (§ II.E.3) IPR2021-01466, Paper 34 (March 10, 2023) (designated: November 15, 2023) [AIA – clarifying the distinction between pre-AIA 35 U.S.C. § 102(e) and AIA 35 U.S.C. § 102(d) and holding that the requirement in Dynamic Drinkware, LLC v. Nat’l Graphics, Inc., 800 F.3d 1375, 1378 (Fed. Cir. 2015) that under pre-AIA 35 U.S.C. § 102(e), a reference patent’s claims must have written description support in its provisional application in order to be entitled to the filing date of the provisional application, does not apply to AIA 35 U.S.C. § 102(d)]Ptab Trials

Abrir fonte

Texto completo

Trials@uspto.gov Paper 34
Tel: 571-272-7822 Entered: March 10, 2023
UNITED STATES PATENT AND TRADEMARK OFFICE
BEFORE THE PATENT TRIAL AND APPEAL BOARD
PENUMBRA, INC.,
Petitioner,
v.
RAPIDPULSE, INC.,
Patent Owner.
IPR2021-01466
Patent 10,531,883 B1
Before JASON W. MELVIN, DAVID COTTA, and JAMIE T. WISZ,
Administrative Patent Judges.
WISZ, Administrative Patent Judge.
JUDGMENT
Final Written Decision
Determining All Claims Unpatentable
35 U.S.C. § 318(a)
IPR2021-01466
Patent 10,531,883 B1
2
I. INTRODUCTION
A. BACKGROUND
Penumbra, Inc. (“Petitioner”) filed a Petition (Paper 2, “Pet.”)
requesting an inter partes review of claims 1–18 of U.S. Patent No.
10,531,883 B1 (Ex. 1001, “the ’883 patent”). Petitioner supported its
Petition with the Declarations of Brian Brown and Ian Ross, MD (Exs. 1005,
1023). RapidPulse, Inc. (“Patent Owner”) filed a Preliminary Response
(Paper 6, “Prelim. Resp.”) along with the Declarations of Ender Finol, PhD,
and Michael Chen, MD (Exs. 2001, 2003). Upon consideration of the
Petition, the Preliminary Response, and the preliminary evidence of record,
we determined that Petitioner had demonstrated a reasonable likelihood that
it would prevail with respect to at least one of the challenged claims of the
’883 patent (Paper 7, “Institution Decision” or “Inst. Dec.”). Thus, we
instituted review with respect to all of the challenged claims.
Patent Owner filed a Patent Owner’s Response (Paper 14,
“PO Resp.”) with additional Declarations from Dr. Finol and Dr. Chen
(Exs. 2014, 2015). Petitioner filed a Reply (Paper 22, “Pet. Reply”),
supported by the Supplemental Declaration of Brian Brown (Ex. 1030).
Patent Owner filed a Sur-Reply (Paper 27, “PO Sur-Reply”).
An oral hearing was held on December 14, 2022, and a transcript of
the hearing is included in the record (Paper 33, “Tr.”).
We have jurisdiction under 35 U.S.C. § 6. After considering the
parties’ arguments and supporting evidence, we conclude that Petitioner has
proven by a preponderance of the evidence that claims 1–18 of the ’883
patent are unpatentable. 35 U.S.C. § 316(e).
IPR2021-01466
Patent 10,531,883 B1
3
B. REAL PARTIES-IN-INTEREST
Petitioner identifies Penumbra, Inc. as the real party-in-interest.
Pet. 1. Patent Owner identities RapidPulse, Inc. as the real party-in-interest.
Paper 4, 1.
C. RELATED PROCEEDINGS
The parties do not identify any other proceedings involving the ’883
patent.
D. THE ’883 PATENT
The ’883 patent, titled “Aspiration Thrombectomy System and
Methods for Thrombus Removal with Aspiration Catheter,” is directed to an
aspiration thrombectomy system that may be used for treatment of ischemic
strokes caused by a blood clot that blocks or plugs a blood vessel in the
brain. Ex 1001, code (54), 1:28–3:6. We reproduce Figure 47 from the ’883
patent below.
IPR2021-01466
Patent 10,531,883 B1
4
Figure 47 illustrates an embodiment of the aspiration thrombectomy
system 600 of the ’883 patent that operates in a Rapid Onset Aspiration
Repeater (“ROAR”) mode. Ex. 1001, 33:22–24; 35:19–22; Fig. 47. The
system includes a vacuum source 610 fluidically connected to the input of a
controllable vacuum valve 620, which in turn is fluidically connected to the
vacuum input of a manifold 630. Id. at 35:22–28. A vent fluid source 640
containing a vent liquid 642 is fluidically connected to a controllable vent
valve 650. Id. at 35:29–31. A ROAR catheter 660, configured to operate in
relatively small vessels, “defines a working lumen 662 fluidically
connecting a distal end 664 thereof to a proximal manifold connector
assembly 670 at a proximal end of the ROAR catheter 660.” Id. at 35:55–
58.
Operation of the aspiration thrombectomy system 600 occurs through
a controller 700 that “selectively opens and closes the vacuum and vent
valves 620, 650 such that, when the vacuum valve 620 is opened, the
vacuum source 610 is fluidically connected to the liquid column in the
lumen 662 and, when the vent valve 650 is opened, vent liquid 642 is
fluidically connected to the liquid column in the lumen 662.” Ex. 1001,
36:39–45. Controller 700 controls the timing of these valves to “change a
level of vacuum at the distal end 664 and prevent the distal portion of the
liquid column in the lumen 662 from exiting the distal end 664—
substantially no forward flow.” Id. at 36:45–49.
During a ROAR cycle, “suction is applied to the clot by rapidly
opening a valve, causing a rapid rise in vacuum pressure.” Id. at 40:32–34.
The source of suction is then turned off and a vent fluid source is rapidly
opened, which “relieves the vacuum present in the catheter 660, which again
IPR2021-01466
Patent 10,531,883 B1
5
rapidly changes the pressure applied to the clot.” Id. at 40:34–37. In some
embodiments of the ROAR cycle, both the vacuum valve 620 and the vent
valve 650 are closed at the same time before the vent valve is opened. Id. at
6:11–17, 43:17–44:10, Tables 1, 4, 5. “The vent valve 650 is then rapidly
closed and the vacuum valve 620 is rapidly opened.” Id. at 40:37–38. “This
cycle is repeated multiple times per second.” Id. at 40:39.
According to the ’883 patent, the “ROAR effect overcomes the static
friction of a clot that is fixed or ‘stuck’ on the catheter tip while under
constant suction” by “provid[ing] an oscillating/alternating displacement that
causes the clot to ‘shuttle’ back and forth to overcome static frictional
force.” Id. at 7:9–14. With ROAR, “there is a morcellation of the clot that
overcomes different clot morphologies as well as overriding volume and
diameter constraints of the small, fixed luminal volume dictated by the
micro-anatomic environment.” Id. at 7:14–18.
E. ILLUSTRATIVE CLAIMS
Petitioner challenges claims 1–18 of the ’883 patent. Claim 1, which
is an independent claim, is illustrative of the challenged claims, and is
reproduced below:
1. [Pre] A clot removal system, comprising:
[a] a catheter having a distal end and defining a lumen filled
with a liquid column having a proximal portion and a distal
portion;
[b] a controllable vacuum valve;
[c] a vacuum source fluidically connected to the vacuum valve;
[d] a controllable vent valve having a vent liquid input;
IPR2021-01466
Patent 10,531,883 B1
6
[e] a vent fluid source containing a vent liquid and fluidically
connected to the vent valve to retain the vent liquid at the
vent fluid input;
[f] a manifold connected to the catheter, to the vacuum valve,
and to the vent valve, the manifold fluidically connecting the
proximal portion of the liquid column in the lumen: to the
vacuum source through the vacuum valve; and to the vent
fluid source through the vent valve;
[g] a controller connected to the vacuum valve and the vent
valve and configured to selectively open and close the
vacuum valve and the vent valve such that: responsive to
opening the vacuum valve, the vacuum source is fluidically
connected to the liquid column in the lumen; and responsive
to opening the vent valve, the vent fluid source is fluidically
connected to the liquid column in the lumen,
[h] the controller configured to cyclically open and close the
vacuum valve and the vent valve to: change a level of
vacuum at the distal end; and
[i] prevent forward flow of the distal portion out from the distal
end during each cycle.
Ex. 1001, 55:5–35 (annotated with bracketed letters). Challenged claims 2–
17 depend from claim 1, either directly or indirectly. Challenged claim 18 is
independent and is similar to claim 1 except for the last few limitations. The
underlined text below highlights the additions to claim 18:
. . . [18.h] the controller configured to cyclically open and close
the vacuum valve and the vent valve in a repeated cycle
comprising a double-closed state in which the vacuum valve
is closed and the vent valve is closed to: change a level of
vacuum at the distal end; and
[18.i] prevent forward flow of the distal portion out from the
distal end during each cycle, and
IPR2021-01466
Patent 10,531,883 B1
7
[18.j] a time of the double-closed state is not greater than
approximately 30 ms.
Id. at 56:40–57:7.
F. INSTITUTED GROUNDS OF UNPATENTABILITY
We instituted trial to determine whether claims 1–18 of the ’883
patent are unpatentable based on the following grounds:
Ground Claims Challenged 35 U.S.C. §1 Reference(s)/Basis
1 1–12, 15, 16, 18 103 Teigen,2 Grey3
2 2, 3, 6–9, 18 103 Teigen, Grey,
Rubenstein4
3 13, 14 103 Teigen, Grey, Yang5
4 17 103 Teigen, Grey, Matteo6
1 The Leahy-Smith America Invents Act (“AIA”), Pub. L. No. 112–29, 125
Stat. 284 (2011), amended 35 U.S.C. § 103, effective March 16, 2013.
Because the ’883 patent has an effective filing date after March 16, 2013,
the AIA version of § 103 applies.
2 Teigen et al., US 11,096,712 B2, issued Aug. 24, 2021 (Ex. 1007,
“Teigen”).
3 Grey et al., WO 2014/151209 A1, published Sept. 25, 2014 (Ex. 1010,
“Grey”).
4 Morton K. Rubenstein, US 3,955,574, issued May 11, 1976 (Ex. 1011,
“Rubenstein”).
5 Yang et al., US 2017/0238953 A1, published Aug. 24, 2017 (Ex, 1012,
“Yang”).
6 Joseph Matteo, US 2012/0138833 A1, published June 7, 2012 (Ex. 1013,
“Matteo”).
IPR2021-01466
Patent 10,531,883 B1
8
II. ANALYSIS
A. PRINCIPLES OF LAW
To prevail in its challenges to the patentability of all claims of the
’883 patent, Petitioner must demonstrate by a preponderance of the evidence
that the claims are unpatentable. 35 U.S.C. § 316(e) (2018); 37 C.F.R.
§ 42.1(d) (2019). “In an [inter partes review], the petitioner has the burden
from the onset to show with particularity why the patent it challenges is
unpatentable.” Harmonic Inc. v. Avid. Tech., Inc., 815 F.3d 1356, 1363
(Fed. Cir. 2016); see also 35 U.S.C. § 312(a)(3) (requiring inter partes
review petitions to identify “with particularity . . . the evidence that supports
the grounds for the challenge to each claim”). That burden of persuasion
never shifts to Patent Owner. Dynamic Drinkware, LLC v. Nat’l Graphics,
Inc., 800 F.3d 1375, 1378 (Fed. Cir. 2015); see also In re Magnum Oil Tools
Int’l, Ltd., 829 F.3d 1364, 1375–78 (Fed. Cir. 2016) (discussing the burden
of proof in inter partes review).
B. PERSON OF ORDINARY SKILL IN THE ART
We consider the asserted grounds of unpatentability in view of the
understanding of a person of ordinary skill in the art (“POSA” or
“POSITA”), and thus begin with the level of ordinary skill in the art. The
level of ordinary skill in the art is “a prism or lens through which . . . the
Board views the prior art and claimed invention” to prevent hindsight bias.
Okajima v. Bourdeau, 261 F.3d 1350, 1355 (Fed. Cir. 2001).
The parties dispute the level of ordinary skill in the art. Petitioner,
relying on the testimony of Mr. Brown, contends that a person of ordinary
skill in the art as of the relevant date would be either:
IPR2021-01466
Patent 10,531,883 B1
9
(1) a person with a B.S. in engineering or an equivalent field,
with two to four years of academic or industry experience in
aspiration embolectomy and thrombectomy or comparable
industry experience who would, where necessary or desired,
work or consult with others, including a physician, to develop
embolectomy or thrombectomy devices; or (2) an interventional
neurosurgeon with at least three years of experience developing
and/or using medical devices in embolectomy or thrombectomy
procedures, and who would, where necessary, work or consult
with others, including an engineer, to develop such a medical
device.
Pet. 15 (citing Ex. 1005 ¶ 48). Petitioner also contends that, “[a] person with
less education but more relevant practical experience, or more relevant
education but less practical experience, may also meet this standard.” Id.
Patent Owner contends that neither of Petitioner’s proposed
definitions of a POSITA reflect the level of ordinary skill because
“individuals with a B.S. degree in an undefined engineering discipline or an
interventional neurosurgeon with no engineering training would lack the
necessary background in fluid mechanics.” PO Resp. 6 (citing Ex. 2014
¶ 43). Instead, Patent Owner, relying on the testimony of Dr. Finol, asserts
that a POSITA:
would have a B.S. in mechanical engineering with at least three
years of experience in designing systems for minimally
invasive intravascular procedures for treating occluded vessels
or the equivalent and in consultation with medical professionals
such as interventional neurosurgeons.
Id. (citing Ex. 2014 ¶ 44).
In determining the level of ordinary skill, various factors may be
considered, including “(1) the educational level of the inventor; (2) type of
problems encountered in the art; (3) prior art solutions to those problems;
IPR2021-01466
Patent 10,531,883 B1
10
(4) rapidity with which innovations are made; (5) sophistication of the
technology; and (6) educational level of active workers in the field.” Daiichi
Sankyo Co. v. Apotex, Inc., 501 F.3d 1254, 1256 (Fed. Cir. 2007). These
factors are “merely a guide,” and the weight or significance ascribed to each
depends on the particular case. Id.
Based on our consideration of the full record, we find that, overall,
Patent Owner’s definition is more specifically tailored to the claimed subject
matter of the ’883 patent. We also find that, a person with a biomedical
engineering degree could have the required background in fluid mechanics.
See Ex. 1031, 45:21–46:7. Therefore, we find that a person of ordinary skill
in the art would have a B.S. in mechanical or biomedical engineering with at
least three years of experience in designing systems for minimally invasive
intravascular procedures for treating occluded vessels or the equivalent and
in consultation with medical professionals such as interventional
neurosurgeons.
Although we agree with, and have primarily adopted Patent Owner’s
definition for an ordinarily skilled artisan in this proceeding, our analysis
and conclusions herein would not change even under Petitioner’s definition.
The parties also do not indicate that the outcome would differ based on the
definition of the person of ordinary skill in the art. See Pet. 15; PO Resp. 6;
Pet. Reply 1.
C. CLAIM CONSTRUCTION
Having defined the ordinarily skilled artisan, we now turn to claim
construction. In this inter partes review, claim terms are construed using the
same claim construction standard that would be used to construe the claim in
IPR2021-01466
Patent 10,531,883 B1
11
a civil action under 35 U.S.C. § 282(b). 37 C.F.R. § 42.100(b). Under this
claim construction standard, claim terms are given their ordinary and
customary meaning as would have been understood by one of ordinary skill
in the art at the time of the invention. See id; Phillips v. AWH Corp., 415
F.3d 1303, 1313 (Fed. Cir. 2005) (en banc). A patentee may define a claim
term in a manner that differs from its ordinary and customary meaning;
however, any special definitions must be set forth in the specification with
reasonable clarity, deliberateness, and precision. See In re Paulsen, 30 F.3d
1475, 1480 (Fed. Cir. 1994); see also Martek Biosciences Corp. v.
Nutrinova, Inc., 579 F.3d 1363, 1380 (Fed. Cir. 2009) (“When a patentee
explicitly defines a claim term in the patent specification, the patentee’s
definition controls”); 3M Innovative Properties Co. v. Avery Dennison
Corp., 350 F.3d 1365, 1371 (Fed. Cir. 2003) (“a definition of a claim term in
the specification will prevail over a term’s ordinary meaning if the patentee
has acted as his own lexicographer and clearly set forth a different
definition”).
1. “the change in the level of vacuum at the distal end is greater than
approximately [15]/[20]/[25] inHg and no greater than approximately
[50]/[30]/[20] ms”
Petitioner provides proposed constructions for the terms, “the change
in the level of vacuum at the distal end is greater than approximately
[15]/[20]/[25] inHg and no greater than approximately [50]/[30]/[20] ms.”
Pet. 21–23. Petitioner proposes that these terms be construed to mean that
“the change in the level of vacuum from low to high or from high to low at
the distal end is greater than approximately [15]/[20]/[25] inHg and the time
for that change is no greater than approximately [50]/[30]/[20] ms.” Id. at
IPR2021-01466
Patent 10,531,883 B1
12
21 (citing Ex. 1005 ¶ 82). Patent Owner does not contest this proposed
construction. Ex. 2014 ¶ 40. We find that this proposed construction is
consistent with the specification of the ’833 patent and consistent with how a
POSA would understand the term at the time of the invention. Ex. 1001,
38:57–39:5; Ex. 1005 ¶¶ 81–82. Therefore, we adopt Petitioner’s proposed
construction for these terms.
2. “forward flow”
In the Institution Decision, we found that the term “forward flow”
required construction. Inst. Dec. 26. We construed the term to mean, “an
amount of liquid in the lumen . . . that exits the distal end . . . in the distal
direction” and “is defined as greater than 6 microliters of fluid.” Id. (citing
Ex. 1001, 36:18–22). Petitioner agrees with this construction, but Patent
Owner does not. See Pet. Reply 1–2; PO Resp. 30–33; PO Sur-Reply 3.
Patent Owner contends that the phrase, “prevent forward flow of the distal
portion out of the distal end during each cycle,” means “controlling the
distal movement of the fluid column such that the distal portion of the fluid
column is prevented from fully exiting the distal end of the catheter during
each cycle.” PO Resp. 30; 33 (citing Ex. 2014 ¶ 89). We address the
parties’ positions below.
a. Patent Owner’s Position
Patent Owner disagrees with the Board’s construction of “forward
flow” in the Institution Decision because it contends that the ’883 patent:
does not teach that the state of no forward flow is limited to
when at least a portion of the distal portion of the fluid column
is prevented from exiting the distal end, but rather that the state
IPR2021-01466
Patent 10,531,883 B1
13
of no forward flow is when the distal portion of the fluid
column is prevented from exiting the distal end of the catheter.
PO Resp. 30–31 (citing Ex. 2014 ¶ 86). Patent Owner asserts that a
POSITA would have understood that “forward flow is prevented so long as
the full extent of the distal portion of the fluid column is prevented from
exiting the catheter.” Id. at 31 (citing Ex. 2014 ¶ 86).
Patent Owner further asserts that the’883 patent defines the distal
portion of the fluid column as follows:
depending on the context used with respect to the catheter 660,
the proximal and distal portions of the liquid column can be a
given amount (e.g., less than 20 microliters or less than 5
microliters), can be a given length (e.g., a few mm or cm) or it
can be an instance of the column that is approximated by using
statistical flow analyses.
PO Resp. 31 (citing Ex. 1001, 36:4–10). Thus, according to Patent Owner,
the distal portion of the fluid column can be 20 microliters of fluid in the
fluid column at the distal end of the catheter. Id. (citing Ex. 2014 ¶ 87).
Patent Owner asserts that a person of ordinary skill would adopt its
construction because the purpose of preventing forward flow is to prevent
the fluid within the catheter from pushing away or breaking up the clot. Id.
at 32–33. Patent Owner further contends that the Board’s construction of
“forward flow” is improper because it imports limitations from a single
embodiment into the claims. PO Sur-Reply 2–3.
b. Petitioner’s Position
Petitioner counters that the Board’s proposed construction of “forward
flow” is supported by the ’883 patent specification and is consistent with the
language of the claims. Pet. Reply 2 (citing Ex. 1030 ¶¶ 13–15). Petitioner
IPR2021-01466
Patent 10,531,883 B1
14
also asserts that Patent Owner’s proposed construction is ambiguous,
confusing, and unsupported. Id. at 3–4. According to Petitioner, Patent
Owner’s construction “effectively eliminates the concept of preventing
forward flow” because “it would permit any arbitrarily selected amount of
fluid to leave the catheter,” which could displace the thrombus a distance
away from the catheter in to the patient’s vasculature. Id. at 4 (citing
Ex. 1030 ¶¶ 17–19; Ex. 1001, 2:65–3:4, 37:45–54). Petitioner also contends
that Dr. Finol’s testimony regarding “forward flow” does not support Patent
Owner’s proposed construction. Id. at 4–5 (citing Ex. 1031, 110:2–6,
112:13–17).
c. Construction of “forward flow”
We agree with Petitioner’s proposed construction of “forward flow”
as adopted by the Board in the Institution Decision. The ’883 patent
explicitly defines this term, stating:
The term “forward flow” is used herein to define an amount of
liquid in the lumen 662 that exits the distal end 664 in a distal
direction. Forward flow is defined as greater than 6
microliters of fluid (approximately 1 mm of catheter length of
ID 0.071″=5.7 μL).
Ex. 1001, 36:18–23 (emphasis added). Moreover, in discussing problems
with the prior art systems, the ’883 patent states that “[w]hen positive
pressure exists at the distal end of the lumen, liquid from inside the lumen
exits out from the distal end of the catheter in a distal direction. This is
referred to as forward flow.” Id. at 2:62–65. In the ’883 patent, the patentee
acted as its own lexicographer as evidenced by usage of language such as “is
used herein to define,” “is defined,” and “is referred to as,” to discuss
“forward flow.” We find that the definition of “forward flow” in the ’883
IPR2021-01466
Patent 10,531,883 B1
15
patent specification would provide a person of ordinary skill with
“reasonable clarity, deliberateness, and precision” as to the meaning of the
term. See Paulsen, 30 F.3d at 1480; see also Ex. 1030 ¶¶ 13–14; Ex. 2001
¶¶ 53, 62. Thus, we construe the term, “forward flow,” to mean “an amount
of liquid in the lumen that exits the distal end in the distal direction that is
greater than 6 microliters.”
We disagree with Patent Owner’s contention that this construction
imports limitations from a single embodiment into the claims. As explained
above, the ’883 patent clearly defines the term “forward flow” and this
definition is not limited to a single embodiment. In fact, the ’883 patent
indicates that, while “[f]orward flow is defined as greater than 6 microliters
of fluid . . . [l]ess forward flow is also included in this definition.” Ex. 1001,
36:21–24. The ’883 patent specifies that “the amount of forward flow can
be restricted to no greater than 2 microliters or, in a particularly beneficial
embodiment, forward flow is approximately zero microliters.” Id. at 36:24–
29. This language indicates that our construction encompasses multiple
embodiments from the ’883 patent.
We also disagree with Patent’s Owner’s proposed construction of
“forward flow” because it is not supported by the ’883 patent specification.
Patent Owner contends that “there is ‘no forward flow’ until the full extent
of the distal portion” exits the distal end and thus forward flow is prevented
when only less than “the full extent of the distal portion” exits the distal end.
PO Resp. 30–31 (emphasis omitted). To adopt Patent Owner’s proposed
construction, we would have to essentially ignore the ’883 patent’s express
definition of this term. Further, there is no support in the’883 patent
specification for “full extent” or “fully exiting.” We also agree with
IPR2021-01466
Patent 10,531,883 B1
16
Petitioner that Patent Owner’s construction effectively eliminates the
concept of preventing forward flow because, as Patent Owner acknowledges,
it could allow, for example, 20 milliliters of fluid to exit the catheter, which
could displace the thrombus into the patient’s vasculature, a problem that the
’883 patent seeks to avoid. See PO Resp. 31 (citing Ex. 2014 ¶ 87).
We are also not persuaded by Patent Owner’s proposed construction
due to the inconsistent positions of its Declarant, Dr. Finol, regarding the
meaning of “forward flow.” In his first Declaration, Dr. Finol appeared to
agree with the Board’s construction of the term. See Ex. 2001 ¶¶ 53 (“[a]
POSITA would understand that, according to fundamental principles of fluid
mechanics, to achieve forward flow out from the distal end of the catheter as
defined in the ’883 patent (i.e., more than 6 μl of fluid out from the distal
end) . . . .”); 62 (“the ‘883 Patent defines “forward flow” as a volume of
liquid greater than about 6 μl, so ‘preventing forward flow’ allows flow up
to 6 μl or less.”).
By contrast, in Dr. Finol’s second Declaration, he adopts Patent
Owner’s new construction of the term “forward flow.” Compare Ex. 2001
¶¶ 53, 62, with Ex. 2014 ¶¶ 86–89. Dr. Finol’s discussion of “forward flow”
in his second Declaration appears to be nearly identical to the claim
construction discussion in Patent Owner’s Response, without any additional
analysis. Compare Ex. 2014 ¶¶ 86–89, with PO Resp. 30–33. Furthermore,
during his deposition, Dr. Finol appears to have provided a different
definition of “forward flow” in stating:
the definition of forward flow is such that the entire distal
portion of the column of fluid exits the distal end of the
catheter, [and] is then drawn back into the catheter during
each cycle. That’s how you would prevent forward flow.
IPR2021-01466
Patent 10,531,883 B1
17
Ex. 1031, 110:2–6 (emphasis added).
If 20 microliters leave the distal end of the catheter during one
phase of the cycle, as long as those 20 microliters are
withdrawn back into the catheter during another phase of the
cycle before the cycle ends, then we have essentially prevented
forward flow.
Id. at 112:13–17. Dr. Finol’s opinion that preventing “forward flow” allows
any amount of liquid to flow forward out the distal end so long as that same
amount of liquid is eventually drawn back in is inconsistent with the goal of
the ’883 patent to prevent the thrombus from flowing downstream. Because
Dr. Finol’s testimony regarding the construction of “forward flow” is
inconsistent, we find his opinions on this topic to be less credible and we do
not find that a POSITA would construe the term as he suggests in his second
Declaration.
D. THE ’883 PATENT PRIORITY DATE
The application that issued as the ’883 patent was filed on July 18,
2019, and claims priority to U.S. Provisional App. Nos. 62/750,011
(“the ’011 provisional), filed October 24, 2018, and 62/701,086 (“the ’086
provisional”), filed July 20, 2018. Ex 1001, codes (22), (60).
The parties dispute whether the claims of the ’883 patent are entitled
to the filing date of the ’011 provisional or whether they are only entitled to
the actual filing date of the ’883 patent.7 See Pet. 8; Pet. Reply 6–8;
7 We note that Patent Owner does not contend that the claims of the ’883
patent are entitled to the filing date of the ’086 provisional. See PO Resp.
26; Ex. 2014 ¶ 74. We also note that Patent Owner does not contend that
claims 10–12 of the ’883 patent are entitled to the benefit of either
provisional application. See PO Resp. 36, n.5.
IPR2021-01466
Patent 10,531,883 B1
18
PO Resp. 29–36, PO Sur-Reply 2–3. In particular, the parties disagree as to
whether the ’011 provisional describes and enables the limitation of
“prevent[ing] forward flow.” Id. After considering all the cited evidence of
record on the priority issue, we find that the evidence favors Petitioner’s
position. As explained below, we find that the challenged claims are not
sufficiently supported by the written description of the ’011 provisional
application and are only entitled to a priority date of July 18, 2019.
1. Principles of Law
Although the petitioner has the ultimate burden of persuasion to prove
unpatentability, a patent owner must demonstrate entitlement to a priority
date when the patent owner relies on that priority date to overcome an
anticipation or obviousness argument. See Dynamic Drinkware, 800 F.3d at
1379–80 (discussing burdens in inter partes review to show entitlement to
provisional filing dates). “[A] patent’s claims are not entitled to an earlier
priority date because the patentee claims priority. . . . Rather, for a patent’s
claims to be entitled to an earlier priority date, the patentee must
demonstrate that the claims meet the requirements of 35 U.S.C. § 120,”
which include fulfilling the requirements of § 112. In re NTP, Inc., 654 F.3d
1268, 1276–77 (Fed. Cir. 2011) (citations omitted)); see also Research
Corp. Techs., Inc. v. Microsoft Corp., 627 F.3d 859, 870–71 (Fed. Cir.
2010); Tech. Licensing Corp. v. Videotek, Inc., 545 F.3d 1316, 1327–29
(Fed. Cir. 2008).
“[T]he hallmark of written description is disclosure,” and “the test [for
satisfaction of the written description requirement] requires an objective
inquiry into the four corners of the specification from the perspective of a
IPR2021-01466
Patent 10,531,883 B1
19
person of ordinary skill in the art.” Ariad Pharms., Inc. v. Eli Lilly & Co.,
598 F.3d 1336, 1351 (Fed. Cir. 2010) (en banc). The written description
“must ‘clearly allow persons of ordinary skill in the art to recognize that [the
inventor] invented what is claimed,’” and “reasonably” convey to those
skilled in the art “that the inventor had possession of the claimed subject
matter as of the filing date.” Id. (alteration in original, internal citations
omitted). Stated another way, “one skilled in the art, reading the original
disclosure, must immediately discern the limitation at issue in the claims.”
Purdue Pharma L.P. v. Faulding Inc., 230 F.3d 1320, 1323 (Fed. Cir. 2000).
“Whether a claim satisfies the written description requirement is a question
of fact.” Nalpropion Pharms., Inc. v. Actavis Labs. FL, Inc., 934 F.3d 1344,
1348 (Fed. Cir. 2019) (citing Ariad, 598 F.3d at 1351).
2. Patent Owner’s Position
Patent Owner argues that the ’011 provisional application provides
support for “prevent[ing] forward flow” under its proposed construction of
that term. PO Resp. 33–35. Patent Owner points to several portions of the
’011 provisional as support for this claim element. Id. First, Patent Owner
cites to the ’011 provisional disclosure which describes an aspiration
thrombectomy system “that temporarily halts the vacuum at the distal end of
the aspiration catheter, pushes the thrombus distally out of the lumen, and
then re-applies the vacuum.” Id. at 33 (citing Ex. 1003 ¶ 15). Patent Owner
also cites to disclosure in the ’011 provisional which states that “[i]t is
desirable to have a minimal amount of momentum transfer from the fluid
column to the stuck clot to unstick the clot sufficiently so that the next
vacuum cycle macerates the clot against the distal end of the catheter 410
IPR2021-01466
Patent 10,531,883 B1
20
and causes it to enter the lumen of the catheter 410 and be removed from the
vessel” and that “[t]he exiting movement of fluid is limited to a specific
volume and/or pressure and is automatically and precisely controlled” to
prevent the “possibility of overshooting.” Id. (citing Ex. 1003 ¶¶ 16, 138).
According to Patent Owner, “[a] POSITA would have understood that a
‘minimal amount of momentum transfer from the fluid column’ means that
some, but only a minimal—e.g., ‘no’—amount of a distal portion of the fluid
column exits the distal end of the catheter.” Id. at 33–34 (citing Ex. 2014
¶ 90).
Patent Owner further contends that the ’011 provisional application
discloses multiple examples of liquid flow exiting the distal end of the
catheter such that no forward flow occurs, such as 1 microliter or 0.5 mm of
column shift. PO Resp. 34 (citing Ex. 1003 ¶ 91 (providing ranges of
approximately “0.001 ml to approximately 1.0 ml,” “0.5 mm to 30 mm,” and
“0.5 mm to 15 mm”)). According to Patent Owner, such specific
embodiments of “substantially no forward flow” provide evidence that a
POSITA would have understood that the inventors of the ’011 provisional
“had possession of a system including controlling the opening and closing of
the vacuum and vent valves to control the distal movement of the fluid
column such that the distal portion of the fluid column is prevented from
fully exiting the distal end of the catheter.” Id. (citing Ex. 2014 ¶ 91).
Patent Owner also contends that Petitioner’s arguments regarding
Grey, one of the references relied upon for the challenges presented in the
Petition, confirm that a person of skill in the art would have understood that
the inventors of the ’011 provisional application had possession of the
“prevent[ing] forward flow” limitation. PO Resp. 35. According to Patent
IPR2021-01466
Patent 10,531,883 B1
21
Owner, Petitioner argues that Grey discloses “preventing forward flow”
without the “more definitive teachings” in the ’011 provisional application.
Id.
3. Petitioner’s Position
Petitioner counters that the ’011 provisional application does not
describe or enable the claim limitation of “prevent forward flow” under any
party’s construction. Pet. Reply 6. Petitioner contends that the ’011
provisional “expressly relied on forward (distal) movement of fluid to create
distal movement of the clot that is stuck.” Id. (citing Ex. 1003 ¶ 126).
According to Petitioner, the ’883 patent explicitly distinguishes the
description of the forward-flow mode of operation in the ’011 provisional
from the claimed requirement of preventing forward flow. Id. at 6–7 (citing
Ex. 1001, 30:58–60, 35:19–20). Petitioner also contends that Patent Owner
does not point to any support from the ’011 provisional application that
limits the amount of exit flow to below 6 microliters, as defined by the ’883
patent specification, or even to 20 microliters. Id. at 7.
4. Analysis of Priority
We find that the disclosure of the ’011 provisional application is not
sufficient to provide adequate written description support for the limitation
of “prevent[ing] forward flow.”
As discussed above, we construed “forward flow” to mean “an
amount of liquid in the lumen that exits the distal end in the distal direction
that is greater than 6 microliters.” Thus, for the ’011 provisional to provide
support for this limitation, the provisional must demonstrate that the
inventors had possession of preventing greater than 6 microliters of fluid out
IPR2021-01466
Patent 10,531,883 B1
22
of the distal end of the catheter. While Patent Owner points to disclosure
from the ’011 provisional application which describes “a minimal amount of
momentum transfer from the fluid column to the stuck clot to unstick the
clot sufficiently” and “[t]he exiting movement of fluid [being] limited to a
specific volume and/or pressure and is automatically and precisely
controlled” to prevent the “possibility of overshooting,” it would not convey
to one of ordinary skill in the art that the inventors were in possession of a
device that limited forward flow to no greater than 6 microliters of fluid.
PO Resp. 33 (citing Ex. 1003 ¶¶ 16, 138).
Patent Owner also contends that the ’011 provisional application
provides “multiple examples of liquid flow exiting the distal end of the
catheter such that no forward flow occurs, such as 1 microliter or 0.5 mm of
column shift.” PO Resp. 34 (citing Ex. 1003 ¶ 91). However, the disclosure
cited by Patent Owner provides ranges of liquid flow of “approximately
0.001 ml to approximately 1.0 ml, in particular, approximately 0.1 ml to
approximately 0.5 ml” and exemplary length of shift distances of
“approximately 0.5 mm to approximately 30 mm, in particular,
approximately 0.5 mm to approximately 15 mm.” Ex. 1003 ¶ 91. There is
nothing in the disclosure of the ’011 provisional application that specifically
directs the person of ordinary skill to the portions of these ranges which fall
within the ’883 patent’s definition of no “forward flow” (i.e., no greater than
6 microliters of fluid). In fact, the majority of the cited ranges are greater
than 6 microliters, going all the way up to 1,000 microliters, such that there
would be forward flow in these examples. See Ex. 1030 ¶ 29; Purdue
Pharma, 230 F.3d at 1326–27 (“[O]ne cannot disclose a forest in the original
application, and then later pick a tree out of the forest and say here is my
IPR2021-01466
Patent 10,531,883 B1
23
invention. In order to satisfy the written description requirement, the blaze
marks directing the skilled artisan to that tree must be in the originally filed
disclosure.”); Fujikawa v. Wattanasin, 93 F.3d 1559, 1571 (Fed. Cir. 1996).
Further, we note that the ’883 patent differentiates between
embodiments that allow forward flow with those in which no forward flow
occurs. Ex. 1001, 30:59–60; 35:19–22; Pet. Reply 6–8. For example, the
’883 patent describes an embodiment in which the “momentum causes a
small amount of fluid to move through the distal portion of the catheter []
and create a small distal movement of the clot.” Ex. 1001, 30:47–50. This
same embodiment is disclosed in the ’011 provisional application. Compare
Ex. 1003 ¶ 126 with Ex. 1001, 30:16–57. The ’883 patent describes this
disclosure as an “exemplary embodiment” in which the “flow of fluid
forward . . . is intentional, which is in contrast to other exemplary
embodiments herein where substantially no forward flow occurs.” Ex. 1001,
30:58–60; see also, id. at 35:19–20 (“Turning now to embodiments . . .
without the forward flow”).8 This further demonstrates that the ’011
provisional application does not support the limitation of “prevent[ing]
forward flow.”
8 Patent Owner contends that the statement in the ’883 patent regarding the
“exemplary embodiment” of “forward flow” is referring to an embodiment
in which the volume of the liquid drawn out from the lumen of the catheter
is “approximately 0.2 ml.” PO Resp. 40–41 (citing Ex. 1001, 30:33–36;
Ex. 1003 ¶ 126; Ex. 2014 ¶ 96). Even if this is correct, the ’011
provisional application does not include the embodiments that the ’883
patent specifically characterizes as being “without the forward flow.”
Pet. Reply 6–7; Ex. 1005 ¶ 68; Ex. 1030 ¶¶ 26–27.
IPR2021-01466
Patent 10,531,883 B1
24
We are also not persuaded by Patent Owner’s argument about Grey.
PO Resp. 35. Although Patent Owner contends that Grey provides less
disclosure than the ’011 provisional with respect to “preventing forward
flow,” Patent Owner’s argument conflates the written description analysis
with an obviousness analysis. See Ariad, 598 F.3d at 1352 (“[A] description
that merely renders the invention obvious does not satisfy” the written
description requirement.).
Thus, we find that the ’011 provisional application does not “clearly
allow persons of ordinary skill in the art to recognize that [the inventors]
invented what is claimed.” In re Gosteli, 872 F.2d 1008, 1012 (Fed. Cir.
1989); Lockwood v. American Airlines, Inc., 107 F.3d 1565, 1572 (Fed. Cir.
1997) (to fulfill the written description requirement, a patent specification
“must describe an invention and do so in sufficient detail that one skilled in
the art can clearly conclude that the inventor invented the claimed
invention”).
We also note that the ’011 provisional application does not contain
adequate support for the limitation of “prevent[ing] forward flow” even
under Patent Owner’s proposed construction. While the ’011 provisional
discloses “controlling the distal movement of the fluid column” (Ex. 1003
¶¶ 15, 16), this does not support Patent Owner’s assertion that the ’011
provisional teaches that “the distal portion of the fluid column is prevented
from fully exiting the distal end.” PO Resp. 34. For example, as discussed
above, the ’011 provisional application discloses permitting distal flow of up
to 1,000 microliters, which would likely exceed the amount that could be
characterized as the full distal portion. Ex. 1003 ¶ 91; Ex. 1030 ¶ 29.
IPR2021-01466
Patent 10,531,883 B1
25
Accordingly, we find that the ’883 patent is only entitled to a priority
date of July 18, 2019.
E. ASSERTED OBVIOUSNESS OF CLAIMS 1–12, 15, 16, AND 18 BY TEIGEN
IN VIEW OF GREY
Petitioner contends that the subject matter of claims 1–12, 15, 16, and
18 of the ’883 patent are obvious over the disclosure of Teigen in view of
Grey. Pet. 23–63. Patent Owner disputes Petitioner’s contentions. PO
Resp. 7–45; PO Sur-Reply 3–17.
1. Principles of Law
A claim is unpatentable under 35 U.S.C. § 103 if the differences
between the subject matter sought to be patented and the prior art are such
that the subject matter, as a whole, would have been obvious at the time the
invention was made to a person having ordinary skill in the art to which said
subject matter pertains. KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 406
(2007). The question of obviousness is resolved on the basis of underlying
factual determinations, including: (1) the scope and content of the prior art;
(2) any differences between the claimed subject matter and the prior art;
(3) the level of skill in the art; and (4) when in evidence, objective evidence
of non-obviousness, i.e., secondary considerations. See Graham v. John
Deere Co., 383 U.S. 1, 17–18 (1966).
“The combination of familiar elements according to known methods
is likely to be obvious when it does no more than yield predictable results.”
KSR, 550 U.S. at 416. “[W]hen the question is whether a patent claiming
the combination of elements of prior art is obvious,” the answer depends on
“whether the improvement is more than the predictable use of prior art
IPR2021-01466
Patent 10,531,883 B1
26
elements according to their established functions.” Id. at 417. It is generally
obvious to those skilled in the art to substitute one known equivalent for
another. See In re Omeprazole Patent Litigation, 483 F.3d 1364, 1374 (Fed.
Cir. 2007) (“[T]his court finds no . . . error in [the] conclusion that it would
have been obvious to one skilled in the art to substitute one ARC [alkaline
reactive compound] for another.”); see also Wm. Wrigley Jr. Co. v. Cadbury
Adams USA LLC, 683 F.3d 1356, 1364 (Fed. Cir. 2012) (the combination of
known elements and substitution of one well known agent for another is
obvious).
When there is a range disclosed in the prior art, and the claimed
invention falls within that range, or the disclosed range overlaps with the
claimed range, there is a presumption of obviousness. Iron Grip Barbell Co.
v. USA Sports, Inc., 392 F.3d 1317, 1322 (Fed. Cir. 2004); In re Peterson,
315 F.3d 1325, 1329 (Fed. Cir. 2003) (“In cases involving overlapping
ranges, we and our predecessor court have consistently held that even a
slight overlap in range establishes a prima facie case of obviousness.”); see
also In re Geisler, 116 F.3d 1465, 1469 (Fed. Cir. 1997) (noting that a
claimed range overlaps with a prior art range if the two ranges share a
common endpoint (e.g., claim range of 50–100 Å overlaps with prior art
range of 100–600 Å)). This presumption may be rebutted by showing the
criticality of the claimed range, that the prior art taught away from the
claimed range, or that the parameter was not recognized as “result-
effective.” E.I. DuPont de Nemours & Company v. Synvina C.V., 904 F.3d
996, 1006 (Fed. Cir. 2018); Iron Grip, 392 F.3d at 1322.
IPR2021-01466
Patent 10,531,883 B1
27
2. Teigen (Ex. 1007)
Teigen, titled “Aspiration Thrombectomy System and Methods for
Thrombus Removal with Aspiration Catheter,” is directed to devices and
methods for controlling clot removal from a patient’s vasculature by
aspiration thrombectomy. Ex. 1007, code (54), 1:17–21. Teigen discloses a
vacuum aspiration control system for use with a vacuum source and an
aspiration catheter. Id., code (57).
Teigen further discloses that its vacuum aspiration system comprises a
connecting tube, on-off valve, sensing unit, controller, fluid introduction
unit, three-point junction, and fluid injection valve for pulsed aspiration.
Ex. 1007, 3:60–4:17, 14:61–18:7. Teigen discloses that the controller is
configured to receive signals from pressure sensors in the system, and opens
and closes the vacuum and fluid injection valves to “clear occlusions in an
aspiration catheter or to facilitate the aspiration of clot that are large or
otherwise hard to aspirate.” Id. at 3:60–4:17, 14:61–18:7. Teigen teaches
that the pressure pulse can create a force impulse that breaks static friction
momentarily, allowing a lower dynamic friction to ingest thrombus. Id. at
14:61–15:9. Teigen also discloses that, in one example, “pressure is
generated by introducing fluid into the catheter, where the fluid is at a
pressure between full vacuum and ambient pressure, at ambient pressure, at
systolic pressure, or above systolic pressure.” Id. at 15:37–40.
In some embodiments of Teigen, “the controller may modulate a
vacuum valve [] and a pressure valve [], whereby the closing of the vacuum
valve [] and the opening of the pressure valve [] may result in a relative
increase in pressure at a distal tip of an aspiration catheter.” Ex. 1007,
17:45–49. Further, “[a]lternatively, the opening of the vacuum valve [] and
IPR2021-01466
Patent 10,531,883 B1
28
the closing of the pressure valve [] may result in a relative decrease in
pressure at the distal tip of the aspiration catheter [].” Id. at 17:50–52.
3. Teigen Priority Date
Teigen issued on August 24, 2021, from U.S. Patent Application No.
16/988,556 (“the ’556 application), filed August 7, 2020. Ex. 1007, codes
(21), (22), (45). The ’556 application is a continuation of
PCT/US2019/043095 (“the ’095 PCT application”), filed on July 23, 2019.
Id. at code (63). The ’095 PCT application claims the benefit of priority to
U.S. Provisional Application Nos. 62/778,708 (“the ’708 provisional”), filed
December 12, 2018, and 62/702,804 (“the ’804 provisional”), filed July 24,
2018. Id. at code (60).
Petitioner contends that Teigen qualifies as prior art under 35 U.S.C.
§§ 102(a)(2) and 102(d)(2) because the relevant subject matter of Teigen
relied on in the Petition is fully supported by both provisional applications
such that Teigen has a priority date of July 24, 2018 (i.e., the filing date of
the ’804 provisional). Pet. 16. Patent Owner contends that Petitioner failed
to establish that Teigen is prior art. PO Resp. 7–26; PO Sur-Reply 3–9.
a. Parties’ Positions
Patent Owner contends that Teigen is not prior art because it is not
entitled to claim priority to either of its provisional applications. PO Resp.
7–26. Specifically, Patent Owner asserts that, under Dynamic Drinkware, in
order to establish that Teigen is entitled to the priority benefit of its
provisional applications, Petitioner has to show that (1) at least one claim of
Teigen is supported by the Teigen provisionals, and (2) the portions of
Teigen relied on to argue obviousness are supported by the provisional
IPR2021-01466
Patent 10,531,883 B1
29
applications. Id. at 9 (citing Dynamic Drinkware, 800 F.3d at 1378–79).
According to Patent Owner, Petitioner fails the first part of the Dynamic
Drinkware analysis because the ’804 provisional does not provide written
description support for one claim of Teigen. Id.
Patent Owner also contends that the Petition failed to include any
arguments that portions of Teigen relied on to show obviousness of the
challenged claims are supported by either provisional because it only
provided bare cites to its Expert Declaration in violation of 37 C.F.R.
§ 42.6(a)(3). PO Sur-Reply 4–5.
Petitioner counters that the first part of the Dynamic Drinkware
analysis is limited to pre-AIA patents and does not apply to Teigen.9
Pet. Rep. 8–9. Petitioner also contends that it has demonstrated that the
subject matter of Teigen relied on in the Petition is supported by the Teigen
provisionals. Id. at 9 (citing Pet. 15–18, 25–63; Ex. 1005 ¶¶ 86, 104–248,
290–312; Ex. 1030 ¶ 31).
b. Analysis
First, we address the issue of whether Petitioner is required to
establish that at least one claim of Teigen is supported by the Teigen
provisionals in order to establish that Teigen is prior art to the ’883 patent.
For the reasons discussed below, we find that, under AIA §§ 102(a)(2) and
102(d), there is no need to evaluate whether any claim of a reference patent
9 Alternatively, Petitioner provides arguments as to how the provisional
applications provide support for claim 1 of Teigen. Pet. 16 (citing
Ex. 1005 ¶¶ 84–95); Pet. Reply 8–13.
IPR2021-01466
Patent 10,531,883 B1
30
document is actually entitled to priority when applying such a reference
patent as prior art.
As discussed above, we find that the priority date of the ’883 patent is
July 18, 2019 (and its earliest effective filing date is July 20, 2018).
Ex. 1001, codes (22), (60). Because these dates are after March 16, 2013,
the AIA version of § 102 applies in this proceeding. 35 U.S.C. § 100 (note).
Thus, we evaluate the prior art status of Teigen under AIA 35 U.S.C.
§§ 102(a)(2) and (d).
Section 102(a)(2) states:
A person shall be entitled to a patent unless— . . .
(2) the claimed invention was described in a patent issued under
section 151, or in an application for patent published or deemed
published under section 122(b), in which the patent or
application, as the case may be, names another inventor and
was effectively filed before the effective filing date of the
claimed invention.
35 U.S.C. § 102(a)(2) (emphasis added).
Section 102(d) states:
For purposes of determining whether a patent or application for
patent is prior art to a claimed invention under subsection
(a)(2), such patent or application shall be considered to have
been effectively filed, with respect to any subject matter
described in the patent or application—
(1) if paragraph (2) does not apply, as of the actual filing date of
the patent or the application for patent; or
(2) if the patent or application for patent is entitled to claim a
right of priority under section 119, 365(a), 365(b), 386(a), or
386(b), or to claim the benefit of an earlier filing date under
section 120, 121, 365(c), or 386(c) based upon 1 or more prior
filed applications for patent, as of the filing date of the earliest
IPR2021-01466
Patent 10,531,883 B1
31
such application that describes the subject matter.
35 U.S.C. § 102(d) (emphases added).
AIA § 102 draws a distinction between actually being entitled to a
right of priority to, or the benefit of, a prior-filed application according to the
definition of “effective filing date” of a claimed invention in AIA 35 U.S.C.
§ 100(i)(1)(B), and merely being entitled to claim a right of priority to, or
the benefit of, a prior-filed application for prior-art purposes according to the
use of “effectively filed” in AIA 35 U.S.C. § 102(d). See 35 U.S.C.
§§ 100(i)(1)(B), 102(a)(2), 102(d)(2); MPEP § 2154.01(b).
USPTO guidance also indicates that AIA § 102 prior-art reference
effective-filing-date determinations differ from pre-AIA law. For example,
a 2018 Memorandum to the Patent Examining Corps states, in the context of
prior-art determinations, that “[t]he requirement that one of the claims in the
patent [document] . . . be supported by the written description of the
provisional application in compliance with pre-AIA 35 U.S.C. § 112, first
paragraph (or AIA 35 U.S.C. § 112(a)), is not applicable when examining an
application subject to the first inventor to file provisions of the AIA.” Robert
W. Bahr, Memorandum re: Critical Reference Date Under pre-AIA 35
U.S.C. §102(e) (Apr. 5, 2018), available at http://www.uspto.gov/sites/
default/files/documents/dynamic_memo_05apr2018_0.pdf (emphasis added)
(citing Dynamic Drinkware, 800 F.3d at 1381 n.2; 157 Cong. Rec. 3423
(2011) (explaining the distinction between being entitled to claim priority or
benefit in AIA 35 U.S.C. § 102(d) and actually being entitled to priority or
benefit under 35 U.S.C. § 119, 120, or 365)). This Memorandum further
indicates:
IPR2021-01466
Patent 10,531,883 B1
32
The date subject matter was “effectively filed” under AIA 35
U.S.C. § 102(d) for purposes of considering a U.S. patent, a
U.S. patent application publication, or an international
application publication as prior art under AIA 35 U.S.C.
§ 102(a)(2) is discussed in MPEP § 2154.01(b), and is not
affected by Dynamic Drinkware or [Amgen v. Sanofi, 872 F.3d
1367, 1380 (Fed. Cir. 2017)] and is not being modified by this
memorandum.
Id. at 2. MPEP § 2154.01(b) explicitly states, as a result of the distinction
discussed above, in application of the AIA version of § 102, “the question of
whether a patent or published application is actually entitled to priority or
benefit with respect to any of its claims is not at issue in determining the
date the patent or published application was ‘effectively filed’ for prior art
purposes.” MPEP § 2154.01(b) (emphases added).
Based on the above, we determine that, for prior-art determinations
under AIA § 102, “there is no need to evaluate whether any claim of [a
reference] patent document is actually entitled to priority or benefit under 35
U.S.C. 119, 120.” MPEP § 2154.01(b) (emphasis added). Rather, under the
AIA, a reference patent document need only meet the “ministerial
requirements” of §§ 119 and 120, and the provisional or other earlier
application(s) to which the reference patent document claims a right of
priority must “describe[] the subject matter” relied upon in the reference
patent document as prior art. 35 U.S.C. § 102(d)(2).
As discussed in the Institution Decision, we find that Teigen meets the
procedural requirements for claiming priority to the ’095 PCT application.
Inst. Dec. 14–15. Teigen also appears to meet the ministerial requirements
for claiming priority to the provisional applications because it properly
claims the benefit of these applications, the ’095 PCT application was filed
IPR2021-01466
Patent 10,531,883 B1
33
co-pending with the provisional applications, and the applications share
common inventors. Ex. 1007, codes (22), (60), (63), (72); Ex.1008, 2;
Ex. 1009, 2; Ex. 2005, codes (22), (30), (72). Patent Owner does not dispute
that Teigen meets the “ministerial requirements” of § 119 at this stage of the
proceeding. Because we find that the Dynamic Drinkware analysis does not
apply here, we agree with Petitioner and Patent Owner, that we need not
reach the issue of whether a claim from Teigen is supported by either Teigen
provisional application. PO Sur-Reply 4; Pet. Reply 8–9.
This determination is consistent with other recent, nonprecedential
PTAB decisions. See, e.g., Apple Inc. v. Telefonaktiebolaget LM Ericsson,
No. IPR2022-00341, Paper 10, at 14–22 (PTAB Sept. 14, 2022) (finding
explicitly that Petitioner did not need to conduct a Dynamic Drinkware
analysis under AIA § 102(d) based on statutory language, legislative history,
USPTO Guidance on Dynamic Drinkware, and the MPEP). We find that the
Forescout Technologies case cited by Patent Owner is inapposite because, in
that case, the Board found that the petitioner failed to show that the
provisional application provided sufficient support for the subject matter
relied upon in the reference patent. See Forescout Technologies, Inc. v.
Fortinet, Inc., No. IPR2021-01328, Paper 12, at 9–10 (PTAB Jan. 27, 2022).
Forescout Technologies did not address whether the petitioner had to show
that the provisional application provided written description support for at
least one claim of the reference patent under Dynamic Drinkware. Id.
With regard to Patent Owner’s contention that the Petition failed to
include any arguments that the portions of Teigen relied on to show
obviousness of the challenged claims were supported by either provisional,
as explained in our Institution Decision, the Petition states that the relevant
IPR2021-01466
Patent 10,531,883 B1
34
subject matter relied on in Teigen is fully supported by both provisional
applications. Inst. Dec. 15; Pet. 16 (citing Ex. 1005 ¶¶ 84–95). The Petition
also cites to portions of Teigen it relies on and provides parallel citations to
both Teigen provisionals. Pet. 16–18. Further, in discussing each claim
limitation, the Petition cites to the Brown Declaration, which contains
specific citations to the supporting disclosure from the provisional
applications. See, e.g., Pet. 23–63 (citing Ex. 1005 ¶¶ 112–247). We do not
find this to be an improper incorporation by reference.
As recognized by the Board and the Federal Circuit, the prohibition
against incorporation by reference set forth in 37 C.F.R. § 42.6(a)(3)
“minimizes the chance that an argument would be overlooked and eliminates
abuses that arise from incorporation and combination,” and is intended to
prevent the situation where “the Board would be forced to ‘play archeologist
with the record’ for arguments that might have been made outside the
parties’ briefing.” 3M Co. v. Evergreen Adhesives, Inc., 860 F. App’x 724,
725 (Fed. Cir. 2021). Here, both Patent Owner and the Board were provided
sufficient notice of Petitioner’s reliance on the cited portions of the Brown
Declaration concerning support in the provisional applications for Teigen’s
disclosure. As such, there was no need to “play archeologist with the
record.”
Petitioner thus sufficiently shows that the Teigen provisional
applications describe the subject matter relied on in Teigen.10 Our
10 Because we agree with Petitioner’s proposed construction of the term
“forward flow” and find that the ’883 patent is only entitled to a priority date
of July 18, 2019, we agree with Patent Owner that “Teigen’s priority to [the
’804 provisional] does not matter. PO Sur-Reply 3.
IPR2021-01466
Patent 10,531,883 B1
35
obviousness analysis below also provides citations to the provisional
applications, showing how they describe the relied on subject matter from
Teigen for each limitation.
4. Grey (Ex. 1010)
Grey, titled “Dynamic Aspiration Methods and Systems,” is directed
to “an aspiration system” including “a catheter[], an aspiration source
providing a negative pressure in the catheter, and a device which repetitively
changes the negative pressure to produce a dynamic suction force at a distal
end of the catheter.” Ex 1010, codes (54), (57), 5:1–4. Grey discloses that
“[n]egative pressure and negative pressure oscillations are delivered by a
fluid medium,” and that, in some embodiments, a “second aspiration fluid
may be provided at a positive pressure.” Id. at 3:22–24, 8:24–25. Grey
teaches that “dynamic aspiration as taught [] may cause facture and break-up
of a clot applying smaller suction forces which are repetitively varied to
induce material fatigue.” Id. at 10:17–19.
Grey also discloses that “[o]ne very serious risk is the escape and
travel of occlusive material from the initial site of occlusion to a distal
portion of the lumen or to another lumen branching therefrom.” Ex. 1010,
7:10–12. According to Grey, “[p]roviding a non-zero suction force in the
catheter at all times reduces the risk of occlusive material breaking free of a
clot and traveling distally in the vessel.” Id. at 13:7–11.
Grey was published on September 25, 2014 and Petitioner contends
that it qualifies as prior art at least under §102(a)(1) because it was published
prior to the effective filing date of the ’883 Patent. Ex. 1010, code (43);
IPR2021-01466
Patent 10,531,883 B1
36
Pet. 18. Patent Owner does not contest that Grey is prior art to the ’883
patent.
5. Claim 1
Petitioner, with supporting testimony from Mr. Brown, provides
evidence as to where Teigen and Grey disclose the limitations of claim 1 of
the ’883 patent. Pet. 25–41; Ex. 1005 ¶¶ 104–160. Patent Owner argues
that Petitioner did not establish a motivation to combine Teigen and Grey.
PO Resp. 43–45. We address the claim limitations and Patent Owner’s
arguments below.
a. Claim 1 [1Pre]–[1h]
We find that Teigen alone or in combination with Grey discloses
limitations [1Pre]–[1h] of claim 1. Patent Owner does not contest that these
claim elements are disclosed by Teigen or Grey. See generally PO Resp.
Teigen discloses “[a] clot removal system” because it describes
“devices and methods for controlling clot removal from a patient’s
vasculature by aspiration thrombectomy.” Ex. 1007, 1:17–21; Pet. 26 (citing
Ex. 1005 ¶¶ 112–115 (citing, e.g., Ex. 1008 ¶¶ 1, 5–7; Ex. 1009 ¶¶ 2, 6–9)).
Petitioner demonstrates that Teigen’s system, explicitly or inherently,
contains a catheter having a “distal end and defining a lumen with a liquid
column having a proximal portion and a distal portion.” Ex. 1007, 3:33–59,
6:6–21; Pet. 27 (citing Ex. 1005 ¶¶ 116–19 (citing Ex. 1008 ¶¶ 10, 11, 20;
Ex. 1009 ¶¶ 12, 13, 22)). Petitioner shows that Teigen, with the knowledge
of a person of ordinary skill, discloses the components described in
limitations [1b]–[1g]; these include: a controllable vacuum valve and
vacuum source that are fluidically connected (Pet. 27–30 (citing Ex. 1005
IPR2021-01466
Patent 10,531,883 B1
37
¶¶ 120–129 (citing Ex. 1008 ¶¶ 12, 18–20, 57, 61, 63, Figs. 9, 11; Ex. 1009
¶¶ 12–14, 21–22, 59, 63, 65, Figs. 9, 11))); a vent fluid source, containing a
vent liquid, that is fluidically connected to a vent valve to retain the vent
liquid at a vent fluid input (Pet. 31–33 (citing Ex. 1005 ¶¶ 130–135 (citing
Ex. 1008 ¶¶ 63–65, Fig. 13; Ex. 1009 ¶¶ 65–67, Fig. 13))); a manifold that
fluidically connects the proximal portion of the catheter/lumen to the
vacuum and vent portions (Pet. 33–35 (citing Ex. 1005 ¶¶ 136–139 (citing
Ex. 1008 ¶¶ 10–12, 13, 17, 19–20, 57, 61–65, Figs. 9, 11, 13; Ex. 1009
¶¶ 12–15, 19–22, 59, 63–69, Figs. 9, 11, 13))); and a controller connected to
and configured to open and close the valves to fluidically connect them to
the liquid column (Pet. 35–37 (citing Ex. 1005 ¶¶ 140–144 (citing Ex. 1008
¶¶ 10–12, 13, 17, 19–20, 57, 61–65, Figs. 9, 11, 13; Ex. 1009 ¶¶ 12–15, 19–
22, 59, 63–69, Figs. 9, 11, 13))). Petitioner further describes how Teigen, in
view of the knowledge of a person of ordinary skill, teaches that its
controller cyclically opens and closes valves to change the level of vacuum
at the distal end. See Pet. 37–40 (citing Ex. 1005 ¶¶ 145–149 (citing
Ex. 1008 ¶¶ 61–67; Ex. 1009 ¶¶ 63–67)).11
b. [1i] “prevent forward flow of the distal portion out from the distal
end.”
Teigen discloses a controller configured to cyclically open and close
its vacuum and fluid injection valves to change a level of vacuum at the
11 Alternatively, we also find persuasive Petitioner’s arguments that Teigen
in view of Grey would teach limitation [1h] because of Grey’s disclosure
of the “dynamic suction force at the distal end of the catheter.” Pet. 39–
40 (citing Ex. 1010, 4:25–30; Ex. 1005 ¶¶ 150–153). We also find that a
person of ordinary skill would have been motivated to combine Teigen and
Grey for the reasons we discuss infra.
IPR2021-01466
Patent 10,531,883 B1
38
distal end of the catheter during each extraction cycle in order to more
effectively digest blood clots at the distal end of the aspiration catheter. See
Pet. 37–38; Ex. 1005 ¶¶ 145–149, 155 (citing Ex. 1007, 3:60–4:26, 5:4–22,
6:6–21; 11:1–25, 14:61–18:22, 23:28–33; Ex. 1008 ¶¶ 12, 13, 17, 20, 57,
61–65; Ex. 1009 ¶¶ 14, 15, 19, 22, 59, 63–69). Teigen also discloses that the
fluid introduced into the catheter from the fluid source can be “at a pressure
between full vacuum and ambient pressure,” such as to partially relieve the
vacuum in the catheter. Pet. 40 (citing Ex. 1005 ¶ 155; Ex. 1007, 15:34–43);
Ex. 1008 ¶ 65; Ex. 1009 ¶ 67.
We further find that, based on the disclosure in Teigen, one of
ordinary skill in the art would understand that the level of vacuum and
amount of fluid injected out of the distal end of the catheter could be
adjusted in order to more effectively remove a clot. Grey describes and
seeks to solve the same problem as the ’883 patent—positive pressure
causing forward clot movement. Grey teaches the “very serious risk [of] the
escape and travel of occlusive material from the initial site of occlusion to a
distal portion of the lumen or to another lumen branching therefrom.”
Ex. 1010, 7:10–12. To minimize this risk, Grey teaches that “the magnitude
of the baseline pressure is preferably maintained above zero (i.e., a vacuum
is maintained)” and that “[p]roviding a non-zero suction force in the catheter
at all times reduces the risk of occlusive material breaking free of a clot and
traveling distally in the vessel.” Id. at 13:7–11. Both parties and their
declarants acknowledge that a device’s ability to prevent clots from traveling
distally was a known goal and would have been a motivating factor to
improve these types of devices for those in the art at the time of filing. See
Pet. 40–41; Ex. 1005 ¶¶ 158–159; Ex. 2001 ¶¶ 24–26; PO Resp. 32–34;
IPR2021-01466
Patent 10,531,883 B1
39
Ex. 2014 ¶¶ 89–90. Accordingly, we find that Grey provides a person of
ordinary skill in the art with the motivation to operate aspiration devices in a
way that would prevent liquid from exiting the distal end of the catheter—
i.e. preventing forward flow.
Based on these disclosures, we find that it would have been obvious to
a person of ordinary skill to operate the controller of Teigen in such a way as
to prevent portions of the clot from breaking free and traveling distally in the
vessel as Grey teaches. In other words, it would have been obvious to
operate Teigen’s controller in a way that would prevent forward flow. This
could be done, for example, by introducing the fluid into the catheter “at a
pressure between full vacuum and ambient pressure,” thereby providing a
non-zero suction force, as disclosed in Teigen and Grey. Ex. 1007, 15:37–
40; Ex. 1008 ¶¶ 63–65; Ex. 1010, 13:7–11; Ex. 1005 ¶¶ 156–160. A person
of ordinary skill would have understood this to mean preventing “greater
than 6 microliters” from exiting the distal end because, as Patent Owner’s
Declarant, Dr. Finol, explains, “[a] POSITA would have understood that
there would be no flow ‘out from the distal end’ at any time if a ‘non-zero
suction force’ was provided at ‘all times in the catheter’ as taught by Grey.”
Ex. 2001 ¶ 82 (emphasis added).
Patent Owner asserts that Petitioner fails to establish a motivation to
combine Teigen with Grey because Petitioner’s arguments are unsupported,
cursory, and fail[] to address the technical realities of the two references it
attempts to combine.” PO Resp. 43–45. According to Patent Owner,
Petitioner provides a list of alleged benefits of Grey and concludes that a
“POSA would recognize that Teigen could be configured to achieve the
benefits as taught by Grey,” while obviousness is dependent on what a
IPR2021-01466
Patent 10,531,883 B1
40
POSITA would do. Id. at 44 (citing Belden Inc. v. Berk–Tek LLC, 805 F.3d
1064, 1073 (Fed. Cir. 2015)). Patent Owner asserts that a person of mere
ordinary skill would not have been motivated to make the proposed
combination. Id. For example, Patent Owner contends that Petitioner
provides no evidence or explanation to support its statement that “combining
the teachings of Teigen and Grey would have been well within the skill of a
POSA because it is nothing more than combining known vacuum aspiration
technologies described in these references to perform their intended
functions with described benefits and predictable results.” Id. at 44 (citing
Pet. 24).
Patent Owner further contends that “Grey unambiguously teaches the
necessity of providing a non-zero suction force at all times” and that a
POSITA:
would understand it would be far more efficient, much simpler,
and present much less risk of causing deadly distal embolus
migration, to maintain a non-zero suction force in the catheter
at all times as taught by Grey if the suction is always “on” - i.e.,
a vacuum valve does not cycle to a “closed” position in which
the suction is stopped.
PO Resp. 45 (citing Ex. 2014 ¶ 161). According to Patent Owner, Petitioner
provides no evidence or arguments that the proposed modifications that
introduce flow out of the distal end of the catheter would have been within
the level of skill of Petitioner’s proposed definition of a POSITA–i.e.,
individuals with no specific training in mechanical engineering or fluid
mechanics. Id.
We find that Petitioner has provided sufficient evidence of motivation
to combine the disclosures of Teigen and Grey and that it would have been
IPR2021-01466
Patent 10,531,883 B1
41
well within a POSITA’s level of skill in the art to modify Teigen’s controller
to implement Grey’s teachings. See Pet. 23–25, 41–42; Ex. 1005 ¶¶ 104–
111; Ex. 1030 ¶ 63. We agree with Petitioner and Mr. Brown that a POSA
would have been motivated to combine the teachings of Teigen and Grey
because they “address similar problems related to conventional aspiration
thrombectomy and share the common objective of improving aspiration
efficiency.” Pet. 23 (citing Ex. 1005 ¶¶ 104–111). We also agree that both
Teigen and Grey “teach similar solutions involving a liquid source to
temporarily vent/relieve vacuum within the catheter to create an accelerated
suction force to efficiently remove blood clots from a patient’s blood vessel
into the device.” Id. Teigen and Grey both discuss issues with clogged
catheters and the limitations of static aspiration technologies for controlled
clot aspiration. Id. at 24 (citing Ex. 1007, 1:57–2:23, 2:46–52, 3:12–32,
14:61–15:33; Ex. 1010, 1:7–9, 1:15–2:3). Both references also describe a
similar approach of creating a dynamic aspiration force to more effectively
clear clots from a clogged catheter. Id. at 24 (citing Ex. 1007, 2:24–64,
3:12–48, 14:61–18:7; Ex. 1010, 2:24–3:24). Further, we are persuaded by
the testimony of Mr. Brown that:
The aspiration cycle frequencies described by Grey are
consistent with and improve on the aspiration cycles that
Teigen discloses to provide effective cyclical aspiration
thrombectomy. Using the pressure pulse profiles and constant
non-zero suction force provided by Grey on Teigen’s patients
would help to reduce the risk of complications from aspiration
thrombectomy, including clogged catheters and forward flow of
material out of the distal end of the catheter into the patient’s
blood stream, as explicitly taught by Grey.
IPR2021-01466
Patent 10,531,883 B1
42
Ex. 1005 ¶ 108 (citing Ex. 1007, 14:61–18:22; Ex. 1008 ¶¶ 63–66; Ex. 1009
¶¶ 65–69; Ex. 1010, 9:13–25, 10:28–11:3) (internal citations omitted).
For the reasons discussed above, we agree with Petitioner and Mr.
Brown that “[c]ombining the teachings of Teigen and Grey would have been
well within the skill of a POSA because it is nothing more than combining
known vacuum aspiration technologies described in these references to
perform their intended functions with described benefits and predictable
results.” Pet. 24 (citing KSR, 550 U.S. at 401; Ex. 1005 ¶ 107 (“A POSA
would understand that Teigen and Grey are each directed to a cyclical
aspiration thrombectomy system and related methods using a source of both
positive and negative pressure to subject a blood clot to dynamic suction
forces and more efficiently aspirate blood clots.”)). We also find that a
POSA would have had a reasonable expectation of success in combining the
disclosures of Teigen and Grey because “it would have been well within a
POSA’s level of skill to modify the logic of Teigen’s controller to
implement Grey’s teachings, and/or to incorporate well-known mechanical
components, if needed, to carry out the functionality described in Grey.”
Pet. 25 (citing Ex. 1005 ¶ 109 (“any modification, if necessary, would
simply add additional control logic to Teigen’s system for modulating the
pressures contained within the aspiration system, in accordance with the
preferable pressure profiles as taught in Grey.”)).
We are not persuaded by Patent Owner’s argument that one of skill in
the art would not have combined the teachings of Teigen and Grey because
it would have been more simple and efficient to always have a non-zero
suction force in the catheter at all times, as taught by Grey, such that the
suction is always “on” – i.e., a vacuum valve does not cycle to a “closed”
IPR2021-01466
Patent 10,531,883 B1
43
position in which the suction is stopped. PO Resp. 45. First, Teigen
discloses that the pressure generated by introducing fluid into the catheter
can be “at a pressure between full vacuum and ambient pressure,” which is a
non-positive pressure. Ex. 1007, 15:37–39; Pet Reply 14; Ex. 1030 ¶ 56.
Second, we are persuaded by the testimony of Dr. Finol and Mr. Brown that
closing the vacuum valve does not necessarily lead to positive pressure or
forward flow, particularly when the tip of the catheter is obstructed.
Ex. 1030 ¶¶ 56–61; Ex. 1031, 91:10–15, 94:2–13; Ex. 2014 ¶ 63. Further,
because we have adopted a definition of a POSA as one who would have
training in mechanical engineering and fluid mechanics, along with
experience in designing systems for minimally invasive intravascular
procedures for treating occluded vessels, applying the teachings of Grey to
provide a non-zero suction force in the catheter of Teigen would have been
well within the level of skill in the art.
Upon review of parties’ arguments and supporting evidence, we
determine that Petitioner sufficiently explains why a POSA would have
found it obvious to implement Grey’s disclosure of “a non-zero suction
force” in the system of Teigen to prevent forward flow. Therefore, we
determine that Petitioner has demonstrated by a preponderance of the
evidence that each limitation of claim 1 is taught or suggested by the
combination of Teigen and Grey and further that the skilled artisan would
have had reason to make the suggested combination with a reasonable
expectation of success.
IPR2021-01466
Patent 10,531,883 B1
44
c. Objective indicia of non-obvious
In determining whether a claim is obvious in light of the prior art, we
also consider any relevant evidence of secondary considerations of non-
obviousness. See Graham, 383 U.S. at 17. Notwithstanding what the
teachings of the prior art would have suggested to one of ordinary skill in the
art at the time of the invention, the totality of the evidence submitted,
including objective evidence of non-obviousness, may lead to a conclusion
that the challenged claims would not have been obvious to one of ordinary
skill. In re Piasecki, 745 F.2d 1468, 1471–72 (Fed. Cir. 1984).
Patent Owner presents evidence of long-felt need, failure of others,
and skepticism. See PO Resp. 62–66. Patent Owner asserts that the ’883
patent addressed a long-felt and persistent need for improved endovascular
thrombectomy devices because, for example, “the ROAR system has nearly
a 300% better full-ingestion rate than conventional aspiration systems.” Id.
at 63. Patent Owner contends that “[p]rior attempts to overcome the
problems plaguing thrombectomy devices failed to significantly meet
physicians’ needs and improve patient outcomes.” Id. at 64–65. Patent
Owner also asserts that physicians performing clot removal procedures were
skeptical about the aspiration thrombectomy devices and preferred to use
stent-retriever treatment alone or aspiration treatment alone. Id. at 65–66
(citing Ex. 2017, 3; Ex. 2015 ¶ 36).
Objective evidence of nonobviousness is relevant only if there is a
nexus between this evidence and the claimed invention. Fox Factory, Inc. v.
SRAM, LLC, 944 F.3d 1366, 1373 (Fed. Cir. 2019). A presumption of nexus
applies if the asserted objective evidence “is tied to a specific product and
that product ‘embodies the claimed features, and is coextensive with them.’”
IPR2021-01466
Patent 10,531,883 B1
45
Id. (quoting Polaris Indus., Inc. v. Artic Cat, Inc., 882 F.3d 1056, 1072 (Fed.
Cir. 2018)). To the extent that a presumption of nexus does not apply,
Patent Owner may still prove nexus “by showing that the evidence of
secondary considerations is the ‘direct result of the unique characteristics of
the claimed invention.’” Id. (quoting In re Huang, 100 F.3d 135, 140 (Fed.
Cir. 1996)).
Petitioner argues, and we agree, that Patent Owner fails to
demonstrate a nexus. Pet. Reply 32. Although Patent Owner asserts that the
“claimed ROAR system,” allegedly embodied by its RapidPulse product, has
a nearly 300% better full-ingestion rate than conventional aspiration systems
and that the ‘883 patent invention “addressed a long-felt need for an
effective treatment for patients with ischemic stroke,” Patent Owner
provides no description of the RapidPulse product compared to the claims of
the ’883 patent. PO Resp. 62–64. Absent this information, Patent Owner
does not establish the required nexus connecting the ’883 patent claims to
the RapidPulse product. Id. at 63–64. Also, we find that dynamic aspiration
was known in the art and agree with Petitioner that much of Patent Owner’s
objective indicia are equally applicable to those known techniques.
Pet. Reply 32; Ex. 2014 ¶ 32; Ex. 1030 ¶¶ 99–100; Ex. 1031, 25:6–14,
40:12–41:17.
For these reasons, we are not persuaded by Patent Owner’s arguments
that failure of others, long-felt need, and skepticism weigh toward the non-
obviousness of the claimed subject matter.
IPR2021-01466
Patent 10,531,883 B1
46
d. Conclusions as to obviousness of claim 1 over Teigen and Grey
In sum, we find that the combination of Teigen and Grey teaches or
suggests each and every element of claim 1. We find that an ordinarily
skilled artisan would have been motivated to combine Teigen and Grey, and
would have had a reasonable expectation of success in achieving the claimed
invention. On this record, we also find that the evidence of secondary
considerations of non-obviousness is weak, at best. As discussed above, we
find that Patent Owner has not established the requisite nexus between the
challenged claims and any of the asserted secondary considerations. We are
therefore unable to accord them any substantial weight. Fox Factory, 944
F.3d at 1373.
Thus, after carefully considering the arguments and evidence, we
determine that Petitioner has shown by a preponderance of evidence that
claim 1 of the ’883 patent would have been obvious over Teigen and Grey.
6. Claims 2, 3, 6–9, and 18: “Double-Closed State”
Claim 2 recites the clot removal system of claim 1 “wherein the
controller is configured to cyclically open and close the vacuum valve and
the vent valve in a repeated cycle comprising a double-closed state in which
the vacuum valve is closed and the vent valve is closed.” Ex. 1001, 55:36–
40. Claims 3 and 18 further define that the “time of the double-closed state
is no greater than approximately 30 ms.” Id. at 56:40–57:7. Claim 6 recites
a cycle that comprises a vacuum-only state, a first double-closed state, a
vent-only state, and then a second double-closed state. Id. at 55:50–60.
Claims 7–9 recite the timing of opening/closing the vents and periods of the
cycle. Id. at 55:61–56:2.
IPR2021-01466
Patent 10,531,883 B1
47
Petitioner, with supporting testimony from Mr. Brown, provides
evidence as to where Teigen and Grey disclose the limitations of dependent
claims 2, 3, 6–9, and 18 of the ’833 patent. Pet. 42–46, 49–54, 61–63.
Patent Owner asserts that the Petition fails to establish obviousness of the
“double-closed state” claims. PO Resp. 46–57. We review the parties’
positions below.
a. Petitioner’s Position
To support the contention that Teigen discloses a “double-closed
state,” Petitioner cites to claim 2 of Teigen which recites “a repeated cycle
comprising a closed state in which the vacuum valve is closed and the vent
valve is closed.” Pet. 42 (citing Ex. 1005 ¶¶ 162–165; Ex. 1007, claim 2).
Petitioner also cites to Teigen’s teaching of sequences for opening and
closing the vacuum and vent valves including where “both valves are closed
for at least short periods of time.” Id. at 43 (citing Ex 1007,16:51–18:7).
Petitioner further argues that, “[t]o the extent that Teigen does not
explicitly disclose a double-closed state in which the vacuum valve and vent
valve are both closed, a POSA would have found a double-closed state an
obvious manner of implementing the dynamic aspiration sequence disclosed
in Teigen.” Pet. 43 (citing Ex 1005 ¶¶ 163–164). According to Petitioner,
“[a] POSA would have found it obvious, in actual execution, to conduct the
valve operation in serial to yield effective pulsing, e.g. closing the vacuum
valve first and then opening the vent valve to generate the positive pulse”
because “[t]his would have been an efficient way to build a positive pressure
pulse.” Id. (citing Ex. 1005 ¶164). Petitioner contends that, “if the vacuum
valve remained open during this period, some of the positive pulse
IPR2021-01466
Patent 10,531,883 B1
48
introduced by opening the vent valve would be reduced by the vacuum.” Id.
Thus, according to Petitioner, “[b]ecause of the nature of serial operation, it
would be obvious to have a finite delay (e.g. for confirming the state of the
valve) between the closing of the vacuum valve and the opening of the vent
valve, which leads to at least a brief period where both the vacuum valve and
the vent valve are closed, i.e. the double-closed state.” Id. at 43–44.
b. Patent Owner’s Position
Patent Owner contends that neither claim 2 of Teigen nor the quote
cited from Teigen at column 17, lines 5–7 have written description support
in the provisional applications. PO Resp. 46–48. Patent Owner also argues
that Petitioner’s theories regarding obviousness of claim 2 are inconsistent
with Petitioner’s arguments for combining the disclosures of Teigen and
Grey for claim 1 because “more effectively generating a positive pressure
pulse . . . directly vitiates Petitioner’s argument that a ‘POSA would have
been motivated in view of Grey to operate the dynamic aspiration system of
[Teigen] in a manner that maintains a ‘non-zero suction force’ to prevent
forward flow of the distal portion out from the distal end . . . [to] reduce the
risk of a clot traveling distally, as taught in Grey’ for claim 1.” PO Resp.
48–49 (citing Pet. 43; Ex. 2014 ¶ 160). According to Patent Owner,
“introduction of a ‘double-closed’ state would cut off the vacuum source
such that the ‘positive pressure pulse’ produced during the subsequent open
state of the vent valve would eliminate the ‘non-zero suction force at all
times in the catheter’ required by the combination of Teigen [] and Grey.”
Id. at 49 (citing Ex. 2014 ¶ 160).
IPR2021-01466
Patent 10,531,883 B1
49
Patent Owner also argues that Petitioner did not establish in the
Petition that the portions of Teigen relied on in its analysis of the “double-
closed state” claims were prior art and that our consideration of Petitioner’s
Reply would violate Patent Owner’s notice rights under 37 C.F.R. § 42.23(b)
and the Administrative Procedure Act. PO Sur-Reply 9–10.
c. Analysis
We agree with Patent Owner that neither of Teigen’s provisional
applications include the disclosure from Teigen at column 17, lines 5–7 or
claim 2. Therefore, we will not consider these disclosures in our analysis of
the “double-closed state” claim limitations. However, we also find that
Teigen and its provisionals contemplate the steps of (1) closing the vacuum
valve, (2) opening the saline valve, (3) closing the saline valve, and
(4) opening the vacuum valve, in a sequence that would result in a double-
closed state, or at least render such a double-closed state to be an obvious
manner of implementing the dynamic aspiration sequence disclosed therein.
See Ex. 1005 ¶¶ 162–165 (citing Ex. 1007, 16:51–18:7; Ex. 1008 ¶¶ 63–65;
Ex. 1009 ¶¶ 65, 67, 70); Ex. 1030 ¶¶ 65–74.
We agree with Petitioner that the Teigen provisionals “disclose that
the vacuum valve and fluid injection valve can each be selectively
opened/close when necessary, e.g., to cut off the vacuum and injection of
fluid in the catheter.” Pet. Reply 17–18 (citing Ex. 1008 ¶¶ 63, 65, 69;
Ex. 1009 ¶¶ 65, 67, 70). For example, the ’804 provisional application
states:
…the system will close the on-off valve to the pump and open
the valve 296 to the saline injection unit 290. An influx or
bolus of saline will partially relieve the vacuum. After a short
IPR2021-01466
Patent 10,531,883 B1
50
period of time the system will close valve 296 and reopen the
on-off valve to reestablish the aspiration vacuum.
Ex. 1008 ¶ 65 (emphasis added); see also Ex. 1008 ¶ 64 (the cycle entails
“applying a vacuum, followed by reducing that vacuum with saline…”
(emphasis added); Ex. 1008 ¶ 69 (“While aspiration is cycled off, saline
injection is cycled on until aspiration is turned back on”) (emphasis added).
Thus, Teigen and its provisionals disclose serial valve operation in sequence,
resulting in periods of double-closed states (or at least rendering such
double-closed states to be an obvious manner of implementing the
sequences). Ex. 1030 ¶ 74.
On the record before us during Institution, we expressed some
skepticism as to whether a POSA would have combined the disclosures of
Teigen and Grey to implement a “double-closed” state as recited in the
claims. Inst. Dec. 34–35. On that record, we believed that Petitioner’s
arguments appeared to be in tension with its arguments regarding the
combination of Teigen and Grey for claim 1 because we thought that a
“double-closed state” would eliminate the “non-zero” suction force in the
catheter. Id. However, upon further development of the record, we find that
a “double-closed state” would not, in fact, eliminate the “non-zero” suction
force. We now understand that, even when the vacuum vent is closed and
the saline valve is open, some level of vacuum remains in the catheter in the
absence of an injection of sufficient vent fluid. Ex. 1030 ¶¶ 58–60; Ex. 2014
¶¶ 62; 63 (“Moreover, a POSITA also would understand that merely opening
and closing the vacuum and pressure valves to deliver ‘saline pulsated
aspiration’ as discussed in Teigen-Prov1 would not inherently produce a
positive pressure pulse sufficient to overcome the mean arterial pressure
IPR2021-01466
Patent 10,531,883 B1
51
necessary for ‘flow…out from the distal end of the catheter.’”). We also
credit the testimony of Mr. Brown that, if the saline valve is closed during
that time (i.e., a “double-closed state”), there would be even less vacuum
reduction because the saline injection is stopped such that “a double-closed
state does not inherently increase the risk of creating forward flow.”
Ex. 1030 ¶¶ 61, 75.
Accordingly, we find that a POSITA would not be dissuaded from
closing the vent valve and vacuum valves for a brief period during the
transition of various valve configurations disclosed in Teigen and that such a
“double-closed state” is compatible with Grey’s disclosure of using a “non-
zero suction force at all times.”12 A POSA would understand that merely
switching a vacuum off does not automatically eliminate the negative
pressure created by the vacuum. See Ex. 1023 ¶¶ 18–20; Ex. 1010, 11:16–
28, 13:4–11; Ex. 1008 ¶ 65 (“the system will close the on-off valve to the
pump and open the valve 296 to the saline injection unit 290. An influx or
bolus of saline will partially relieve the vacuum.”); Ex. 1009 ¶ 67; Ex. 1031,
74:3–75:8; Ex. 1030 ¶ 75; Pet. 44; Ex. 1005 ¶ 167. Thus, a person of
ordinary skill in the art would understand that a “double-closed state” is an
obvious way to implement the combination of Teigen and Grey and would
have a reasonable expectation of success in doing so, rendering claim 2
obvious.
12 We do not find this position to be inconsistent with Petitioner’s argument
regarding a “positive pressure pulse” because we credit the testimony of
Mr. Brown that a “positive pressure pulse” does not mean “positive
pressure,” but, rather, means a positive pressure differential (i.e., an
increase in pressure). Ex. 1030 ¶ 75, n.11; Ex. 1007, 15:13–23.
IPR2021-01466
Patent 10,531,883 B1
52
We also find that claims 3 and 18 are obvious for similar reasons.
First, we agree with Petitioner that “[a] POSA would have found it
beneficial and obvious to minimize the double-closed state for efficient
pulsed aspiration” because doing so would “optimize the efficiency of
aspiration while limiting the risks of forward flow out of the distal end of the
catheter.” Pet. 45–46; Ex. 1005 ¶¶ 167–174. Grey teaches that “[d]ifferent
frequencies of cyclic aspiration may be used to improve clot clearance” and
discloses an embodiment in which the “frequencies may be at least 50 Hz.”
Ex. 1010, 10:28–31. We credit the testimony of Mr. Brown that a POSA
would understand that there is in inverse relationship between frequency
(e.g. the measurement of Hertz (Hz)) and cycle period and that “[a] cyclic
aspiration frequency of ‘at least 50 Hz’ corresponds to an aspiration period
of ‘no greater than 20 ms’ (1/(50 Hz) = 0.02 second = 20 millisecond[s]).”
Pet. 45 (citing Ex. 1005 ¶ 169). Thus, a POSA would understand that the
double-closed state portion of the cycle (like any portion of the cycle),
would be less than 20 milliseconds. Ex. 1005 ¶ 169.
We also find that a POSA would have been motivated to experiment
with the duration of the double-closed state in order to optimize the
efficiency of the aspiration while limiting the risks of forward flow out of
the distal end of the catheter to arrive at a double-closed state of “no greater
than 30 ms” as claims 3 and 18 recite. Ex. 1005 ¶¶ 170–174; E.I DuPont De
NeMours, 904 F.3d at 1006 (“where the general conditions of a claim are
disclosed in the prior art, it is not inventive to discover the optimum or
workable ranges by routine experimentation.”) (citation omitted); Peterson,
315 F.3d at 1329.
IPR2021-01466
Patent 10,531,883 B1
53
Claim 6 recites a repeated cycle with two “double-closed states.”
Ex. 1001, 55:50–60. Teigen discloses a vacuum-only state, in which only
the vacuum is on. See Ex. 1007, 15:10–33; Ex. 1008 ¶¶ 64–65;
Ex. 1005 ¶ 190. As discussed above, Teigen in view of Grey discloses a
“double-closed state” when serially turning off/on its vacuum and vent
valves. We also find that it would have been obvious to have a “vent-only
state.” Teigen discloses a “vent-only state” because “[w]hen an extraction
cycle is initiated, the vacuum on-off valve between the catheter and the
aspiration source is closed and the pressure in the aspiration catheter is
increased.” Ex. 1007, 15:10–33; Ex. 1005 ¶¶ 177–180 (citing Ex. 1008
¶¶ 20, 63–65). We credit the testimony of Mr. Brown that, a POSA “would
understand that closing the valve to the vacuum source and then increasing
the pressure in the aspiration catheter” is a disclosure of “a vent-only state.”
Ex. 1005 ¶¶ 178–179; 1007, 16:51–17:7. Because the “double-closed state”
exists when changing between vacuum-only and vent-only states, the same
operation would have been done when changing from vent-only to vacuum-
only. We find that a POSA’s motivation to include one “double-closed
state” would have applied to a second double-closed state as well given
Teigen and Grey’s teachings that the aspiration process is repeated or
cycled. Pet. 50–51; Ex. 1005 ¶¶ 195–196.
Claim 7, which depends from claim 6, recites that the time between
opening and closing the vent “is between approximately 10 ms and
approximately 50 ms.” Ex. 1001, 55:61–64. As discussed above with
respect to claim 3, Grey discloses that an entire aspiration cycle occurs
within 20 ms. Ex. 1010, 9:13–25, 10:28–11:3; Ex. 1005 ¶¶ 185–186. We
are persuaded by Mr. Brown’s testimony that:
IPR2021-01466
Patent 10,531,883 B1
54
Because the vent state, i.e. between an opening of the vent
valve and a closing of the vent valve, can only last for a time
shorter than the full aspiration period for a cyclic aspiration, a
POSA would have understood that a cyclic aspiration frequency
of “at least 50 Hz” inherently discloses a time of the vent state
(i.e. the time between opening and closing the vent valve) of no
greater than 20 ms, or that a cyclic aspiration frequency of “at
least 10 Hz” inherently discloses a time of the vent-only state of
no greater than 100 ms.
Ex. 1005 ¶ 200 (citing Ex. 1010, 9:13–25, 10:28–11:3). We also find that it
would have been obvious to a POSA to “experiment with the tim[ing]” to
come to “the claimed range through routine optimization.” Pet. 52 (citing
Ex. 1005 ¶ 202; E.I. DuPont De NeMours, 904 F.3d at 1006; Ex. 1023 ¶ 20).
Claims 8 and 9 depend from claim 6 and limit “a period of the cycle”
to between “approximately 6 Hz and approximately 16 Hz” and
“approximately 8 Hz and 12 Hz,” respectively. Ex. 1001, 55:65–56:2.
Teigen teaches that the cyclical aspiration’s “frequency with which the on-
off valve opens and closes may be predetermined or responsive to pressure
sensor data.” Ex. 1007, 15:10–33, 15:44–64. Grey discloses cyclical
aspiration frequencies of “above 1 Hz,” “at least 10 Hz,” and specifically
“approximately 6.3 Hz,” among others. Ex. 1010, 9:13–25, 10:28–11:3,
11:29–31. We agree with Petitioner and Mr. Brown that a POSA would
have been motivated to optimize these disclosures and arrive at the claimed
invention. Pet. 53–54.
Thus, after carefully considering the arguments and evidence, we
determine that Petitioner has shown by a preponderance of evidence that
IPR2021-01466
Patent 10,531,883 B1
55
claims 2, 3, 6–9, and 18 of the ’883 patent would have been obvious over
Teigen and Grey.13
7. Claims 4, 5, 15, and 16
Petitioner, with supporting testimony from Mr. Brown, provides
evidence as to where Teigen and Grey disclose the limitations of dependent
claims 4, 5, 15, and 16 of the ’833 patent. See Pet. 46–49, 58–61. Other
than Patent Owner’s arguments regarding the motivation to combine Teigen
and Grey, Patent Owner does not contest that Teigen and Grey disclose the
limitations of these claims. We review the evidence presented by Petitioner
at pages 46–49 and 58–61 of the Petition, which we adopt.
Claim 4 recites the system of claim 1, “wherein the controller is
configured to cyclically open and close the vacuum valve and the vent valve
in a repeated cycle comprising a vent-only state in which the vacuum valve
is closed and the vent valve is open.” Ex. 1001, 55:43–47. We find that
13 We disagree with Patent Owner that Petitioner did not establish in the
Petition that the portions of Teigen relied on in its analysis of the “double-
closed state” claims were prior art. The Petition cited to the portions of
Teigen it relied on to show obviousness of these claims and cited to the
Brown Declaration for the corresponding cross-cites to the Teigen
provisional applications. See, e.g., Pet. 42–44 (citing Ex. 1005 ¶¶ 162–
165). The Brown Declaration explains why a double-closed state would
have been obvious from the disclosure of Teigen and Grey and cites to the
same portions of the Teigen provisional applications that Petitioner cites to
in the Petitioner Reply. See Ex. 1005 ¶¶ 162–165 (citing Ex. 1008 ¶¶ 63,
65; Ex. 1009 ¶¶ 65, 67, 70). Therefore, we find that Petitioner’s Reply
does not include new arguments or evidence with regard to the Teigen
provisional applications. Any other additional material in the Petitioner
Reply was responsive to arguments made in the Patent Owner Preliminary
Response or the Institution Decision.
IPR2021-01466
Patent 10,531,883 B1
56
Teigen and Grey disclose the “repeated cycle” limitation for the same
reasons discussed for claim 1. Pet. 46 (citing Ex. 1007, claim 4). We also
find that Teigen discloses a “vent-only state” for the same reasons discussed
for claim 6 above.
Claim 5 limits claim 4’s “vent-only state” to “no greater than 50 ms.”
Ex. 1001, 55:48–49. Teigen discloses that the “time of the vent-only state is
based on a predetermined frequency.” Ex. 1007, claim 5; Ex. 1008 ¶ 64.
We agree with Petitioner that “a POSA would optimize the time of the vent-
only state in view of Grey,” for example, because of the risk of a clot
breaking free and traveling distally. Pet. 48; Ex. 1005 ¶ 184. Petitioner also
argues that, “[b]ecause the vent-only state only lasts for a time shorter than
the full aspiration period for a cyclic aspiration, a POSA would have
understood that a cyclic aspiration frequency of ‘at least 50 Hz’ discloses a
time of the vent-only state of no greater than 20 ms.” Pet. 48–49 (citing
1005 ¶¶ 185–186). We also credit the testimony of Mr. Brown that a POSA
would have been motivated to optimize the parameters such that “a time of
vent-only state” is a result effective parameter. Id. at 49 (citing Ex. 1005
¶¶ 187–188; E.I. DuPont De NeMours, 904 F.3d at 1006).
Claim 15 depends from claim 1 and recites a “repeated cycle” that
“prevent[s] forward flow of the distal portion out from the distal end during
each cycle by regulating timing of the vent valve.” Ex. 1001, 56:28–32.
Petitioner asserts that Teigen alone or Teigen in view of Grey disclose this
limitation for the reasons discussed for claim 1. Pet. 59 (citing Ex. 1007,
claim 11). We agree with Petitioner and Mr. Brown that a POSA would
have found it obvious to prevent forward flow by regulating the timing of
the vent valve because “a POSA would have recognized when the timing of
IPR2021-01466
Patent 10,531,883 B1
57
the vent valve is not regulated, there will be a significant likelihood of
undesirable forward flow of the distal portion out from the distal end.”
Pet. 59–60 (citing Ex. 1005 ¶¶ 226–227; Ex. 1023 ¶ 20).
Claim 16 recites “[t]he system according to claim 1, wherein the
controller is configured to cyclically open and close the vacuum valve and
the vent valve to retain a level of pressure at the distal end at less than
physiological pressure.” Ex. 1001, 56:33–36. Petitioner asserts that:
A POSA would have understood “prevent forward flow of the
distal portion out from the distal end during each cycle” and the
teachings of Grey—“[p]roviding a non-zero suction force” at
the distal end of the catheter—to mean that the pressure at the
distal end of the catheter remains at less than physiological
pressure because if the pressure were to exceed the
physiological pressure at the distal end, a non-zero suction force
no longer exists and flow direction changes.
Pet. 60–61 (citing Ex. 1005 ¶¶ 233–234; Ex. 1010, 13:4–11; Ex. 1023 ¶ 19;
Ex. 1007, 15:34–43). We are persuaded by Petitioner and Mr. Brown’s
position. Because the non-zero suction force would not exist if the distal
pressure were to exceed physiological pressure, we find that a POSA would
have found it obvious to prevent forward flow by operating Teigen’s
invention with Grey’s “non-zero suction force” at less than physiological
pressure.
Thus, after carefully considering the arguments and evidence, we
determine that Petitioner has shown by a preponderance of evidence that
claims 4, 5, 15, and 16 of the ’883 patent would have been obvious over
Teigen and Grey.
IPR2021-01466
Patent 10,531,883 B1
58
8. Claims 10–12: “delta vacuum-time claims”
Claim 10 recites the system of claim 1, “wherein the change in the
level of vacuum at the distal end is greater than approximately 15 inHg and
no greater than approximately 50 ms.” Ex. 1001, 56:3–6. Claims 11 and 12
are similar but recite different pressures and times: “20 inHg” and “30 ms”
in claim 11 and “25 inHg” and “20 ms” in claim 12. Id. at 56:7–15. As
discussed above, we have construed these claim terms to mean that “the
change in the level of vacuum from low to high or from high to low at the
distal end is greater than approximately [15]/[20]/[25] inHg and the time for
that change is no greater than approximately [50]/[30]/[20] ms.”
a. Petitioner’s Position
Petitioner contends that Teigen discloses a change in the level of
vacuum at the distal end of the catheter and the use of pressure sensors to
measure pressure differentials. Pet. 54–55; Ex. 1007, 11:1–12:9. Petitioner
also contends that:
a POSA could determine the change in the level of vacuum near
the distal end of the catheter based on the change in the level of
vacuum near the proximal end of the catheter using well-
established principles of fluid dynamics, and would be
motivated to do so given the importance of avoiding forward
flow that was known to a POSA at the time.
Pet. 55 (citing Ex. 1005 ¶ 211; Ex. 1023 ¶¶ 17–20).
Petitioner further contends that Grey discloses that “the realized
pressure differential (i.e. the magnitude of the difference between the
maximum and minimum pressure achieved at the catheter tip) may be
between 0 and -100 inHg”, or “[i]n some embodiments... between -5
and -20 inHg.” Pet. 55 (citing Ex. 1010, 13:4–11). Petitioner asserts that
IPR2021-01466
Patent 10,531,883 B1
59
these disclosed pressures at the distal end overlap with the claimed ranges of
“greater than approximately 15/20/25 inHg” of “the change in the level of
vacuum at the distal end.” Pet. Reply 29–30; Pet. 55 (citing Ex. 1005
¶ 213). Petitioner also contends that a “POSA would have recognized that
the pressure differential is a result effective variable for safe and efficient
thrombus extraction.” Pet. 55–56 (citing Ex. 1005 ¶ 213; Ex. 1023 ¶ 20;
Ex. 1010, 10:32–37). For example, “a POSA would have understood that a
high pressure differential generally speeds up the material fatigue of a clot as
taught in Grey and would thus choose to apply a pressure differential above
a minimal threshold value to facilitate clot extraction.” Id. at 55–56 (citing
Ex. 1005 ¶ 213).
With regard to timing, Petitioner contends that Grey discloses that
“[d]ifferent frequencies of cyclic aspiration may be used to improve clot
clearance” and “[f]or some embodiments higher frequencies are more
effective, particularly frequencies above 1 Hz. . . . frequencies may be at
least 10 Hz and . . . frequencies may be at least 50 Hz.” Pet. 56 (citing
Ex. 1010, 10:28–31). Petitioner also cites to Grey’s disclosure that, for a
given frequency, “the rate of change of suction/vacuum pressure in the
catheter and thus the total change of suction/vacuum pressure in the catheter
can be selectively varied,” which allows “control [of] both the frequency and
amplitude of aspiration individually.” Id. (citing Ex. 1010, 11:16–12:13).
Petitioner also reiterates that “a frequency of at least 10 Hz corresponds to a
period of less than 100 ms; a frequency of at least 50 Hz corresponds to a
period of less than 20 ms.” Id. According to Petitioner, “[i]t would have
been obvious for a POSITA to apply the high-pressure differential disclosed
IPR2021-01466
Patent 10,531,883 B1
60
at a high frequency to optimize ‘treatment efficacy’ as directed by Grey.”
Pet. Reply 31 (citing Ex. 1030 ¶ 95).
Petitioner further argues that “[a] POSA would have understood that
the aspiration frequency imposes an inherent limitation to the rate of
change.” Pet. 56 (citing Ex. 1005 ¶ 214). Therefore, according to
Petitioner, “the time for the change in the level of vacuum from low to high
or from high to low” must be shorter than the aspiration period, e.g. “no
greater than 20 ms.” Id. at 56–57 (citing Ex. 1005 ¶ 214). Lastly, according
to Petitioner, the ’883 patent does not teach any particular effect or criticality
of the claimed range. Id. at 58.
b. Patent Owner’s Position
Patent Owner contends that Petitioner does not establish that Teigen
and Grey teach the recited combination of change in the level of vacuum
over time, but, rather, only address each aspect in isolation. PO Resp. 58;
PO Sur-Reply 24. According to Patent Owner, Petitioner could not
demonstrate the recited combination of the change in vacuum over time
because “Grey explicitly teaches that [the change in vacuum] decreases with
increasing frequency such that at 50 Hz or even 10 Hz the [change in
vacuum] would be less than 15 inHg.” PO Resp. 60 (citing Ex. 2014 ¶ 165).
Patent Owner contends that Example 1 is the only disclosure in Grey that
explicitly describes combinations of vacuum pressure differentials together
with frequency (i.e., time) in cyclic aspiration; however, in this example, the
vacuum pressure differentials decreased as the frequency increased, such
that at the 6.3 Hz maximum frequency, the pressure differentials decreased
to only -13.9 inHg. Id. at 60–61 (citing Ex. 1010, 22:6–25:4; Ex. 2014
IPR2021-01466
Patent 10,531,883 B1
61
¶ 166). According to Patent Owner, Grey teaches that increasing the
frequency decreases the vacuum differential linearly such that a POSA
“would understand that increasing the frequency of Grey to 10 Hz or 50 Hz
would result in a [change in the vacuum] far less than the -13.9 inHg that
Grey achieved at only 6.3 Hz.” Id. Therefore, Patent Owner asserts that a
POSA would not have understood Grey to disclose any of the combinations
of change in vacuum over time recited in the claims. Id. at 62 (citing
Ex. 2014 ¶ 168).
Patent Owner also argues that Petitioner failed to demonstrate priority
to the Teigen provisional applications for the “delta vacuum/time” claims.
PO Sur-Reply 2.
c. Analysis
Teigen discloses a change in the level of vacuum at the distal end of
the catheter.14 Ex. 1007, 11:1–12:9. Grey discloses that “the realized
pressure differential (i.e., the magnitude of the difference between the
maximum and minimum pressure achieved at the catheter tip) may be
between 0 and -100 inHg” or “[i]n some embodiments . . . between -5
and -20 inHg.” Ex. 1010, 13:4–11. These disclosed pressures at the distal
end encompass and/or overlap with the claimed range of “greater than
approximately [15]/[20]/[25] in Hg” of “the change in the level of vacuum at
14 We disagree with Patent Owner that Petitioner failed to demonstrate
priority to the Teigen provisional applications for the “delta vacuum/time”
claims. The Petition cites to paragraphs 209 to 221 of the Brown
Declaration in support of its arguments that Teigen (in combination with
Grey), rendered these claims obvious. Pet. 54–58. The cited paragraphs of
the Brown Declaration cite to the portions of the Teigen provisional
applications that disclose the challenged claim limitations.
IPR2021-01466
Patent 10,531,883 B1
62
the distal end.” Ex. 1005 ¶ 213. With regard to timing, we agree with
Petitioner and Mr. Brown that Grey teaches cycle times as low as 20 ms
based on operating at 50 Hz. Pet. 56 (citing Ex. 1010, 10:28–31; Ex. 1005
¶ 214); see also Pet. 45 (citing Ex. 1005 ¶ 169). This cycle time of less than
20 ms falls within the scope of the claimed timing of “no greater than
approximately [50/30/20] ms.” Pet. Reply 30 (citing Ex. 1030 ¶ 94;
Ex. 1005 ¶¶ 214, 219). The question before us then is whether Petitioner
sufficiently connects the pressure differentials disclosed in Grey with the
separately disclosed frequencies in Grey such that the claimed combinations
would be rendered obvious to one of ordinary skill in the art. We find that
Petitioner has done so.
Grey discloses that “[d]ifferent frequencies of cyclic aspiration may
be used to improve clot clearance in comparison to static aspiration” and
“higher frequencies are more effective.” Ex. 1010, 10:28–29. And we are
persuaded by the testimony of Mr. Brown that “a POSA would have
understood that a high-pressure differential generally speeds up the material
fatigue taught in Grey.” Ex. 1005 ¶ 213. Accordingly, we agree with Mr.
Brown that “[i]t would have been obvious for a person of skill in the art to
apply the high-pressure differential disclosed at a high frequency to optimize
‘treatment efficacy’ as directed by Grey.” Id. ¶ 213; Ex. 1030 ¶ 95.
Patent Owner asserts that Example 1 of Grey does not disclose the
claimed combination of pressure differential and timing; however, this is just
an example and does not represent all embodiments of Grey. PO Resp. 60–
61. Further, Grey expressly teaches, in the context of this example, use of a
higher-pressure differential that increases both pressure differential and the
rate of pressure change. Ex. 1030 ¶ 97; Ex. 1010, 12:1–3 (When “second
IPR2021-01466
Patent 10,531,883 B1
63
aspirating medium in Figure 3C operated at a positive pressure . . . the rate
of pressure change would increase due to the increased pressure
differential between aspiration mediums.”). It is also notable that Figure
3C, cited by Patent Owner, depicts manual valve operation, while a POSITA
would understand that a higher frequency with an improved rate of change
could be accomplished using automatically controlled valves as disclosed in
Teigen. Ex. 1010, 22:11, 23:1–9; Ex. 1030 ¶ 98.
We credit the testimony of Mr. Brown that, “[b]ecause “Grey teaches
both a higher pressure differential and a higher frequency are preferable, it
would have been obvious for a person of skill in the art to apply the high-
pressure differential disclosed at a high frequency to optimize ‘treatment
efficacy’ as directed by Grey.” Ex. 1030 ¶ 95 (citing Ex. 1010, 10:22–29,
11:18–20). Also, because Grey discloses ranges that overlap with the
claimed ranges of the change in the level of vacuum over time and Petitioner
has not presented any evidence of criticality to these particular claimed
ranges, we find these claims to be obvious. See E.I. DuPont de NeMours,
904 F.3d at 1006 (explaining that prior art ranges that overlap with a
claimed range create “a presumption of obviousness,” which may be
rebutted if the patentee comes forward with evidence showing, inter alia,
that the claimed invention achieves unexpected results). Accordingly, we
find that Petitioner establishes that claims 10–12 would have been obvious
over Teigen and Grey by a preponderance of the evidence.
F. ASSERTED OBVIOUSSNESS OF CLAIMS 2, 3, 6–9, AND 18 IN VIEW OF
TEIGEN, GREY, AND RUBENSTEIN
Petitioner contends that the subject matter of claims 2, 3, 6–9, and 18
of the ’883 patent are obvious over the disclosure of Teigen in view of Grey
IPR2021-01466
Patent 10,531,883 B1
64
and Rubenstein. Pet. 63–69. Patent Owner disputes Petitioner’s
contentions. PO Resp. 50–57. Because we find that Petitioner has shown by
a preponderance of evidence that claims 2, 3, 6–9, and 18 of the ’883 patent
would have been obvious over Teigen and Grey, we need not reach this
ground. See SAS Inst. Inc. v. Iancu, 138 S. Ct. 1348, 1359 (2018) (holding
that a petitioner “is entitled to a final written decision addressing all of the
claims it has challenged”); Bos. Sci. Scimed, Inc. v. Cook Grp. Inc., 809
F. App’x 984, 990 (Fed. Cir. 2020) (non-precedential) (recognizing that the
“Board need not address issues that are not necessary to the resolution of the
proceeding” and, thus, agreeing that the Board has “discretion to decline to
decide additional instituted grounds once the petitioner has prevailed on all
its challenged claims”).
G. ASSERTED OBVIOUSNESS OF CLAIMS 13 AND 14 OVER TEIGEN IN VIEW
OF GREY AND YANG
Petitioner contends that the subject matter of claims 13 and 14 of the
’883 patent are obvious over the disclosures of Teigen in view of Grey and
Yang. Pet. 69–73. Other than the arguments regarding the combination of
Teigen and Grey discussed above, Patent Owner does not dispute
Petitioner’s contentions. See generally PO Resp. We review the evidence
presented by Petitioner at pages 69–73 of the Petition, which we adopt.
1. Yang (Ex. 1012)
Yang, titled “Telescoping Neurovascular Catheter with Active Distal
Tip,” is directed to a telescoping catheter for “distal neurovascular aspiration
or access” including a control for advancing the distal section into the
patient’s vasculature. Ex 1012, codes (54), (57), Figs. 1, 3A, 3B. The
IPR2021-01466
Patent 10,531,883 B1
65
catheter can connect to “a controller for applying intermittent vacuum to the
lumen” using combinations of vacuum pulses and “spaces of neutral
pressure,” higher negative pressure or lower negative pressure. Id. ¶ 14,
Figs. 14, 25A–C. Yang also discloses ranges of catheter inner diameters.
Id. ¶ 68.
Yang was published on August 4, 2017 and Petitioner contends that it
qualifies as prior art at least under §102(a)(1) because it was published prior
to the effective filing date of the ’883 Patent. Ex. 1012, code (43); Pet. 19.
Patent Owner does not contest that Yang is prior art to the ’883 patent.
2. Analysis
Claim 13 of the ’883 patent provides “[t]he system according to claim
1, wherein: the lumen has an internal diameter of between approximately
0.038" and approximately 0.106"; and the controller is configured to
cyclically open and close the vacuum valve and the vent valve at a frequency
of between 2 and 16 Hz.” Ex. 1001, 56:16–21. Claim 14 is nearly identical
but recites “an internal diameter of between approximately 0.068'' and
approximately 0.088''” Id. at 56:22–27.
Petitioner describes Teigen’s disclosures related to its catheter’s
internal lumen and how to switch between different catheter sizes. Pet. 71
(citing Ex. 1007, 3:33–59, 6:6–21; 12:31–39, 14:15–27; Ex. 1005 ¶¶ 295–
297; Ex. 1023 ¶¶ 13–14). Petitioner summarizes that “Yang is directed to a
telescoping catheter for ‘distal neurovascular aspiration or access.’” Id.
Petitioner demonstrates that Yang discloses the inner diameter of the
catheter to be “between about 0.030 inches and about 0.112 inches.” Id. at
71–72 (citing Ex. 1012 ¶ 68; Ex. 1005 ¶¶ 291–292). Petitioner then
IPR2021-01466
Patent 10,531,883 B1
66
incorporates its discussions from claim 8 regarding frequency here.
Petitioner also provides arguments regarding motivation to combine Teigen,
Grey, and Yang, which we find persuasive. Pet. 69–70.
Thus, after carefully considering the arguments and evidence, we
determine that Petitioner has shown by a preponderance of evidence that
claims 13 and 14 of the ’883 patent would have been obvious over Teigen,
Grey, and Yang.
H. ASSERTED OBVIOUSNESS OF CLAIM 17 OVER TEIGEN IN VIEW OF GREY
AND MATTEO
Petitioner contends that the subject matter of claim 17 of the ’883
patent is obvious over the disclosures of Teigen in view of Grey and Matteo.
Pet. 73–78. Other than the arguments regarding the combination of Teigen
and Grey discussed above, Patent Owner does not dispute Petitioner’s
contentions. See generally PO Resp. We review the evidence presented by
Petitioner at pages 73–78 of the Petition, which we adopt.
1. Matteo (Ex. 1013)
Matteo, titled “Multi-Function Eccentrically Actuated Microvalves
and Micropumps,” is directed to a microfluidics system with “one
cylindrical cam controlling several actuator balls and several microvalves”
for use in various chemical, biological, and biomedical applications,
including “medical and scientific instrumentation.” Ex. 1013, codes (65),
(57) ¶¶ 6, 12, 29, Figs. 11, 12. Matteo discloses that “[c]am-driven pinch-
style microvalves are useful for serving as on/off valve devices for a
microfluidic system.” Id. ¶ 10. Matteo teaches that “[a] cam-driven actuator
activates a microvalve by pressing on the elastomeric layer, deforming the
IPR2021-01466
Patent 10,531,883 B1
67
elastomeric layer so that it meets a second layer at a location within the
channel, thereby either partially or completely obstructing the flow of liquid
through the channel at that location, i.e. ‘pinching’ the channel.” Id. at code
(57).
Matteo was published on June 7, 2012 and Petitioner contends that it
qualifies as prior art at least under §102(a)(1) because it was published prior
to the effective filing date of the ’883 Patent. Ex. 1013, 1; Pet. 20. Patent
Owner does not contest that Matteo is prior art to the ’883 patent.
2. Analysis
Claim 17 provides “[t]he system according to claim 1, which further
comprises a shaft and the vacuum valve and the vent valve are mounted
together on the shaft.” Ex. 1001, 56:37–39.
Petitioner demonstrates that Matteo discloses microfluidics systems
“to control multiple microvalves” and that an exemplary embodiment
includes “one cylindrical cam controlling several actuator balls and several
microvalves.” Pet. 75 (citing Ex. 1013, Abstract, ¶¶ 6, 12, 29, Figs. 11, 12).
Matteo further discloses that “[c]am-driven pinch-style microvalves are
useful for serving as on/off valve devices for a microfluidic system.” Id. at
75–76 (citing Ex. 1013 ¶ 10). Petitioner also contends that Matteo teaches
that “a single cam, controlled by a single position-control mechanism, is
able to control multiple microvalves.” Id. at 78 (citing Ex. 1013, Abstract,
¶¶ 10, 11, Figs. 11, 12). We credit the testimony of Mr. Brown that, a POSA
“would therefore recognize the disclosures of the ‘single cam, controlled by
a single position-control mechanism’ in Matteo to be a shaft with mounted
valves because the cam disclosed in Matteo is used to move the cam
IPR2021-01466
Patent 10,531,883 B1
68
actuators in and out of position to open and close their respective valves.”
Ex. 1005 ¶ 311. Petitioner, with supporting testimony from Mr. Brown
concludes that, “[i]t would have been obvious for a POSA and required no
undue experimentation to implement the valve operations and controls
taught by Teigen and Grey with the cam-driven valves disclosed in Matteo
to achieve the desired dynamic aspiration.” Id. (citing Ex. 1005 ¶ 311).
Petitioner also asserts several reasons why it would have been obvious
to combine Teigen, Grey, and Matteo, which we find persuasive. Pet. 73–
75. Upon review of the arguments and supporting evidence, we determine
that Petitioner persuasively explains why a POSA would have found it an
obvious design choice, when implementing the combination of Teigen and
Grey, to place the vacuum and vent valves together on a shaft, as taught in
Matteo. See Pet. 73–75. Thus, after carefully considering the arguments
and evidence, we determine that Petitioner has shown by a preponderance of
evidence that claim 17 of the ’883 patent would have been obvious over
Teigen, Grey, and Matteo.
III. CONCLUSION15
Petitioner establishes by a preponderance of the evidence that claims
1–18 of the ’883 patent are unpatentable as follows.
15 Should Patent Owner wish to pursue amendment of the challenged claims
in a reissue or reexamination proceeding subsequent to the issuance of this
decision, we draw Patent Owner’s attention to the April 2019 Notice
Regarding Options for Amendments by Patent Owner Through Reissue or
Reexamination During a Pending AIA Trial Proceeding. See 84 Fed. Reg.
16,654 (Apr. 22, 2019). If Patent Owner chooses to file a reissue
application or a request for reexamination of the challenged patent, we
remind Patent Owner of its continuing obligation to notify the Board of
IPR2021-01466
Patent 10,531,883 B1
69
IV. ORDER
In consideration of the foregoing, it is hereby:
ORDERED that claims 1–18 of the ’883 patent have been proven to
be unpatentable; and
FURTHER ORDERED that because this is a Final Written Decision,
parties to the proceeding seeking judicial review of the Decision must
comply with the notice and service requirements of 37 C.F.R. § 90.2.
any such related matters in updated mandatory notices. See 37 C.F.R.
§ 42.8(a)(3), (b)(2).
Claims 35
U.S.C. §
Reference(s)/Basis Claims
Shown
Unpatentable
Claims
Not Shown
Unpatentable
1–12, 15,
16, 18
103(a) Teigen, Grey 1–12, 15, 16,
18
13, 14 103(a) Teigen, Grey,
Yang
13, 14
17 103(a) Teigen, Grey,
Matteo
17
Overall
Outcome
1–18
IPR2021-01466
Patent 10,531,883 B1
70
For PETITIONER:
Eliot D. Williams
Jeremy Taylor
BAKER BOTTS L.L.P.
eliot.williams@bakerbotts.com
jeremy.taylor@bakerbotts.com
For PATENT OWNER:
Lori a. Gordon
Paul T. Parker
PERKINS COIE LLP
gordon-ptab@perkinscoie.com