HAWAII ADMINISTRATIVE RULES
TITLE 11
DEPARTMENT OF HEALTH
CHAPTER 45
RADIATION CONTROL
Subchapter 1 	General Provisions
§11-45-1
§11-45-2
§11-45-3
§11-45-4
§11-45-5
§11-45-6
§11-45-7
§11-45-8
§11-45-9
§11-45-10
§11-45-11
§11-45-12
§11-45-13
§11-45-14
§11-45-15
§11-45-16
Purpose and scope
Definitions
Exemptions
Inspections
Proceedings before director
General requirements
Additional requirements
Reserved
Reserved
Reserved
Prohibited uses
Reserved
Units of exposure and dose
Units of radioactivity
Penalties and injunctions
Severability
Subchapter 2 	License
§11-45-17
§11-45-18
§11-45-19
§11-45-20
§11-45-21
§11-45-22
§11-45-23
§11-45-24
§11-45-25
§11-45-26
§11-45-27
§11-45-28
§11-45-29
§11-45-30
§11-45-31
Purpose and scope
Exemptions
Reserved
Reserved
Reserved
Application for license
Issuance of license
Renewal of license
Report of changes
Approval not implied
Assembly and transfer
Reserved
Reserved
Out-of-state source of radiation
Prohibitions
Subchapter 3 	Shielding Evaluation and Area Radiation Survey
§11-45-32 	Purpose and scope
45-1 	65
Complete printed PDF page 1
§11-45-33
§11-45-34
§11-45-35
Shielding evaluation
Area radiation survey
Reserved
Subchapter 4 	Standards 	for Protection Against Radiation
§11-45-36
§11-45-37
§11-45-38
§11-45-39
§11-45-40
§11-45-41
§11-45-42
§11-45-43
§11-45-44
§11-45-45
§11-45-46
§11-45-47
§11-45-48
§11-45-49
§11-45-50
§11-45-51
§11-45-52
§11-45-53
§11-45-54
§11-45-55
§11-45-56
§11-45-57
§11-45-58
§11-45-59
§11-45-60
§11-45-61
§11-45-62
§11-45-63
§11-45-64
§11-45-65
§11-45-66
§11-45-67
§11-45-68
§11-45-69
§11-45-70
§11-45-71
§11-45-72
§11-45-73
§11-45-74
§11-45-75
Purpose and scope
Exemptions
Implementation
Radiation protection programs
Occupational dose limits 	for adults
Compliance with requirements for summation of external
and internal doses
Determination of external dose from airborne
radioactive material
Determination of internal exposure
Determination of prior occupational dose
Planned special exposures
Occupational dose limits 	for minors.
Dose to an embryo/fetus
Dose limits 	for individual members of the public
Compliance with dose limits 	for individual members of
the public
Reserved
Reserved
Testing for leakage or contamination of sealed sources
surveys and monitoring
Conditions requiring individual monitoring of external
and internal occupational dose
Control of access to high radiation areas
Control of access to very high radiation areas
Control of access to very high radiation areas -
irradiators
Use of process or other engineering controls
Use of other controls
Use of individual 	respiratory protection equipment
Security of stored sources of radiation
Control of sources of radiation not in storage
Caution signs
Posting requirements
Exceptions to posting requirements
Labeling containers and radiation machines
Exemptions to labeling requirements
Reserved
Reserved
Waste disposal
Method for obtaining approval of proposed disposal
procedures
Disposal by release into sanitary sewerage
Treatment or disposal by incineration
Reserved
Transfer for disposal and manifests
45-2 6
Complete printed PDF page 2
§11-45-76
§11-45-77
§11-45-78
§11-45-79
§11-45-80
§11-45-81
§11-45-82
§11-45-83
§11-45-84
§11-45-85
§11-45-86
§11-45-87
§11-45-88
§11-45-89
§11-45-90
§11-45-91
§11-45-92
§11-45-93
§11-45-94
§11-45-95
§11-45-96
§11-45-97
Compliance with environmental and health protection
regulations
Units and quantities on records
Records of radiation protection programs
Records of surveys
Records of tests 	for leakage or contamination of
sealed sources
Records of prior occupational dose
Records of planned special exposures
Records of individual monitoring results
Records of dose to individual members of the public
Records of waste disposal
Records of testing entry control devices 	for very high
radiation areas
Form of records
Reports of stolen, 	or lost, 	or missing sources of
radiation
Notification of incidents
Reports of exposures, 	radiation levels, 	and
concentrations of radioactive material exceeding the
limits
Reports of planned special exposures
Reserved
Reserved
Reserved
Notifications and reports 	to individuals
Reports of leaking or contaminated sealed sources
Vacating premises
Subchapter 5 	Industrial Radiographic Operations
§11-45-98
§11-45-99
§11-45-100
§11-45-101
§11-45-102
§11-45-103
§11-45-104
§11-45-105
§11-45-106
§11-45-107
§11-45-108
§11-45-109
§11-45-110
§11-45-111
§11-45-112
§11-45-113
§11-45-114
Purpose and scope
Exemptions
Storage precautions
Radiation survey instruments
Utilization logs
Inspection and maintenance
Permanent radiographic installations
Reserved
Training and testing of radiographic personnel
Operating and emergency procedures
Personnel monitoring control
Security
Posting
Radiation surveys and survey records
Documents and records 	required at temporary jobsites
Specific requirements 	for 	radiographic personnel
performing industrial radiography
Special requirements and exemptions 	for cabinet
radiography
45-3 i...:...
Complete printed PDF page 3
Subchapter 6 	X-Rays in the Healing Arts
§11-45-115
§11-45-116
§11-45-117
§11-45-118
§11-45-119
§11-45-120
§11-45-121
§11-45-122
§11-45-123
§11-45-124
§11-45-125
Purpose and scope
General requirements
All diagnostic x-ray systems
Fluoroscopic x-ray systems
Radiographic systems other than fluoroscopic, 	dental
intraoral, 	or computed tomography systems
Intraoral dental 	radiographic systems
Computed tomography x-ray systems
Bone densitometry systems
Reserved
Reserved
Mammography x-ray systems and facilities
Subchapter 7 	Radionuclides in the Healing Arts
§11-45-126
§11-45-127
§11-45-128
§11-45-129
§11-45-130
§11-45-131
§11-45-132
§11-45-133
§11-45-134
§11-45-135
§11-45-136
§11-45-137
§11-45-138
§11-45-140
§11-45-141
§11-45-142
§11-45-143
§11-45-144
§11-45-145
§11-45-146
§11-45-147
§11-45-148
§11-45-149
§11-45-150
§11-45-151
§11-45-152
§11-45-153
§11-45-154
§11-45-155
Purpose and scope
Exemptions
Notifications
ALARA program
Radiation safety officer
Radiation safety committee
Statement of authorities and responsibilities
Supervision
Visiting authorized user
Mobile nuclear medicine service administrative
requirements
Procedures 	for human administrations
Records, 	notifications, 	and reports of
misadministrations
Suppliers
Possession, 	use, 	calibration, 	and check of dose
calibrators
Calibration and check of survey instruments
Assay of radiopharmaceutical dosages
Authorization for calibration and reference sources
Possession of sealed sources and brachytherapy sources
Syringe shields
Syringe labels
Vial shields
Vial shield labels
Surveys for contamination and ambient 	radiation dose
rate
Release of patients containing radiopharmaceuticals or
permanent implants
Mobile nuclear medicine service technical requirements
Storage of volatiles and gases
Decay-in-storage
Reserved
Specific requirements 	for the use of
radiopharmaceuticals 	for uptake, 	dilution, 	or
excretion studies
45-4
Complete printed PDF page 4
§11-45-156
§11-45-157
§11-45-158
§11-45-159
§11-45-160
§11-45-161
§11-45-162
§11-45-163
§11-45-164
§11-45-165
§11-45-166
§11-45-167
§11-45-168
§11-45-169
§11-45-170
§11-45-171
Specific requirements 	for the use of
radiopharmaceuticals, 	generators, 	and reagent kits
for imaging and localization studies
Specific requirements 	for the use of
radiopharmaceuticals 	for therapy
Specific requirements 	for the use of sealed sources
for diagnosis
Specific requirements 	for the use of sources 	for
brachytherapy
Reserved
Reserved
Reserved
Reserved
Reserved
Training for radiation safety officer
Training for experienced radiation safety officer
Training for uptake, 	dilution, 	or excretion studies
Training for imaging and localization studies
Training for experienced authorized users
Physician training in a 	three-month program
Recentness of training
Subchapter 8 	Analytical X-Ray Equipment
§11-45-172
§11-45-173
§11-45-174
§11-45-175
§11-45-176
§11-45-177
Purpose and scope
Exemptions
Equipment requirements
Area requirements
Operating requirements
Personnel 	requirements
Subchapter 9 	Particle Accelerators
§11-45-178
§11-45-179
§11-45-180
§11-45-181
§11-45-182
§11-45-183
§11-45-184
§11-45-185
§11-45-186
§11-45-187
Purpose and scope
General requirements
Reserved
Limitations
Shielding and safety design requirements
Particle accelerator controls and interlock systems
Warning devices
Operating procedures
Radiation monitoring requirements
Ventilation systems
Subchapter 10 Notices, 	Instructions, 	Reports, 	and
Inspections
§11-45-188
§11-45-189
§11-45-190
§11-45-191
Purpose and scope
Posting of notices to workers
Instructions to workers
Notifications and reports to individuals
45-5
Complete printed PDF page 5
§11-45-192
§11-45-193
§11-45-194
§11-45-195
§11-45-196
§11-45-197
§11-45-198
§11-45-199
Presence of representatives of licensees and workers
during inspection
Consultation with workers during inspections
Requests by workers 	for inspections
Inspections not warranted; 	informal review
Reserved
Reserved
Reserved
Reserved
Subchapter 11 Health Physics Services and Medical 	Physics
Services
§11-45-200
§11-45-201
§11-45-202
§11-45-203
§11-45-204
Purpose and scope
General requirements
Records
Use of medical physics services
Reserved
Subchapter 12 Transportation of Radioactive Material
§11-45-205
§11-45-206
Purpose and scope
General requirements
Subchapter 13 Wireline Services Operations and
Subsurface Tracer Studies
§11-45-207
§11-45-208
§11-45-209
§11-45-210
§11-45-211
§11-45-212
§11-45-213
§11-45-214
§11-45-215
§11-45-216
§11-45-217
§11-45-218
§11-45-219
§11-45-220
§11-45-221
§11-45-222
§11-45-223
§11-45-224
§11-45-225
§11-45-226
§11-45-227
§11-45-228
§11-45-229
Purpose and scope
Exemptions
Prohibition
Limits on levels of radiation
Storage precautions
Transport precautions
Radiation survey instruments
Leak testing of sealed sources
Quarterly inventory
Utilization records
Design, 	performance, 	and certification criteria for
sealed sources used in downhole operations
Labeling
Inspection and maintenance
Reserved
Training requirements
Operating and emergency procedures
Personnel monitoring
Security
Handling tools
Subsurface tracer studies
Particle accelerators
Radiation surveys
Documents and records 	required at field stations
45-6 2 ,, r-
D :J
Complete printed PDF page 6
§11-45-2
§11-45-230
§11-45-231
Documents and records 	required at temporary jobsites
Notification of incidents, 	abandonment, 	and lost
sources
Subchapter 14 Therapeutic Radiation Machines
§11-45-232
§11-45-233
§11-45-234
§11-45-235
§11-45-236
§11-45-237
§11-45-238
§11-45-239
§11-45-240
§11-45-241
Purpose and scope
General administrative requirements 	for 	facilities
using therapeutic radiation machines
General technical requirements 	for 	facilities using
therapeutic radiation machines
Quality management program
Therapeutic radiation machines of less than five
hundred kilovolts
Reserved
Therapeutic radiation machines - 	photon therapy
systems 	(five hundred kilovolts and above) 	and
electron therapy systems 	(five hundred kilo electron
volts and above)
Reserved
Calibration and check of survey instruments
Shielding and safety design requirements
SUBCHAPTER 1
GENERAL PROVISIONS
§11-45-1 	Purpose and scope. 	This chapter sets minimum
standards 	for all persons and facilities who receive, 	possess,
use, 	transfer, 	own, 	or acquire any source of radiation, 	all
persons who inst~ll and_se7vice so~rces of radirA.~Qn~ &nsg~l
persons who provide radiation services. 	[Eff ,~uv 	1' I 	9(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	21-
11 (21), 	321-71)
§11-45-2 	Definitions. 	As used in this chapter:
"Absorbed dose" means the energy imparted by ionizing
radiation per unit mass of irradiated material. 	The units of
absorbed dose are the gray 	(Gy) 	and the rad.
"Accelerator-produced material" means any material made
radioactive by a particle accelerator.
"Accessible surface" means the external surface of the
enclosure or housing of an x-ray system component provided by the
manufacturer.
"Activity" means the rate of disintegration or
transformation or decay of radioactive material. 	The units of
activity are the becquerel 	(Bq) 	and the curie 	(Ci).
"Added filtration" means any filtration which is in addition
to 	the inherent filtration.
"Address of use" means the building or buildings that are
identified on the license and where radioactive material may be
45-7
6
Complete printed PDF page 7
§11-45-2
produced, 	prepared, 	received, 	used, 	or stored.
"Adult" means an individual eighteen or more years of age.
"Air kerma" 	(K) 	means the kinetic energy released in air by
ionizing radiation. 	Kerma is determined as the quotient of dE by
dM, 	where dE is the sum of the initial kinetic energies of all the
charged ionizing particles liberated by uncharged ionizing
particles in air of mass dM. 	The SI unit of air kerma is 	joule
per kilogram and the special name for the unit of kerma is the
gray 	(Gy).
"Airborne radioactive material" means any radioactive
material dispersed in the air in the form of dusts, 	fumes,
particulates, 	mists, 	vapors, 	or gases.
"Airborne radioactivity area" means a 	room, 	enclosure, 	or
area in which airborne radioactive materials exist in
concentrations:
(1) 	In excess of the derived air concentrations 	(DACs)
specified in Appendix B, 	Table I, 	of subchapter 4, 	or
(2) 	To such a degree that an individual present in the
area without respiratory protective equipment could
exceed, 	during the hours an individual is present in a
week, 	an intake of 0.6 percent of the annual limit on
intake 	(ALI) 	or twelve DAC-hours.
"Aluminum equivalent" means the thickness of type 1100
aluminum alloy affording the same attenuation, 	under specified
conditions, 	as the material in question.
"Analytical x-ray equipment" means equipment used for x-ray
diffraction or fluorescence analysis.
"Analytical x-ray system" means a group of components
utilizing x or gamma rays to determine the elemental composition
or to examine the microstructure of materials.
"Annual limit on intake" 	(ALI) 	means the derived limit for
the amount of radioactive material taken into the body of an adult
worker by inhalation or ingestion in a 	year. 	ALI 	is the smaller
value of intake of a 	given radionuclide in a 	year by the reference
man that would result in a committed effective dose equivalent of
five 	rems 	(0.05 	sievert) 	or a committed dose equivalent of fifty
rems 	(0.5 	sievert) 	to any individual organ or tissue. 	ALI values
for intake by ingestion and by inhalation of selected
radionuclides are given in Table I, 	Columns 1 and 2, 	of Appendix B
of subchapter 4.
"Applicator" means a 	structure which determines the extent
of the treatment field at a 	given distance from the virtual
source.
"Area of use" means a portion of a physical structure that
has been set aside for the purpose of producing, 	preparing,
receiving, 	using, 	or storing radioactive material.
"As low as is reasonably achievable" 	(ALARA) 	means making
every reasonable effort to maintain exposures to radiation as 	far
below the dose limits in these rules as is practical, 	consistent
with the purpose for which the licensed activity is undertaken,
taking into account the state of technology, 	the economics of
improvements in relation to state of technology, 	the economics of
improvements in relation to benefits to the public health and
safety, 	and other societal and socioeconomic considerations, 	and
45-8
26
Complete printed PDF page 8
§11-45-2
in relation to utilization of nuclear energy and licensed sources
of radiation in the public interest.
"Assembler" means any person engaged in the business of
assembling, 	replacing, 	or installing one or more components into
an x-ray system or subsystem. 	The term includes the owner of an
x-ray system or his or her employee or agent who assembles
components into an x-ray system that is subsequently used to
provide professional or commercial services.
"Attenuation block" means a block or stack of type 1100
aluminum alloy or aluminum alloy having equivalent attenuation
with dimensions twenty centimeters by twenty centimeters by 3.8
centimeters, 	of type 1100 aluminum alloy or other materials having
equivalent attenuation.
"Automatic exposure control" means a device which
automatically controls one or more technique factors 	in order to
obtain at a preselected location(s) 	a 	required quantity of
radiation. 	(Includes devices such as phototimers and ion
chambers.)
"Background radiation" means 	radiation from cosmic sources;
naturally-occurring radioactive materials, 	including radon, 	except
as a decay product of source or special nuclear material, 	and
including global 	fallout as it exists in the environment from the
testing of nuclear explosive devices. 	"Background radiation" does
not include sources of radiation from radioactive materials
regulated by the department.
"Beam axis" means a 	line from the source through the centers
of the x-ray fields.
"Beam-limiting device" means a device which provides a means
to 	restrict the dimensions of the x-ray field.
"Beam scattering filter" means a 	filter used in order to
scatter a beam of electrons.
"Becquerel" 	(Bq) 	means the SI unit of activity. 	One
becquerel is equal to one disintegration or transformation per
second 	(s- 1 ).
"Bioassay" means the determination of kinds, 	quantities or
concentrations, 	and, 	in some cases, 	the locations of radioactive
material in the human body, 	whether by direct measurement, 	in vivo
counting, 	or by analysis and evaluation of materials excreted or
removed from the human body. 	For purposes of these 	rules,
"radiobioassay" is an equivalent term.
"Byproduct material" means:
(1) 	Any radioactive material, 	except special nuclear
material, 	yielded in or made radioactive by exposure
to the radiation incident to the process of producing
or utilizing special nuclear material; 	and
(2) 	The tailings or wastes produced by the extraction or
concentration of uranium or thorium from ore processed
primarily for its source material content, 	including
discrete surface wastes 	resulting from uranium or
thorium solution extraction processes. 	Underground
ore bodies depleted by these solution extraction
operations do not constitute "byproduct material"
within this definition.
"C-Arm x-ray system" means an x-ray system in which the
45-9
265
Complete printed PDF page 9
§11-45-2
image receptor and x-ray tube housing assembly are connected by a
common mechanical support system in order to maintain a desired
spatial relationship. 	This system is designed to allow a change
in the projection of the beam through the patient without a change
in the position of the patient.
"Cabinet radiography" means industrial radiography conducted
in an enclosure or cabinet shielded so that radiation levels at
every location on the exterior meet the limitations specified in
subchapter 4 of these rules.
"Cabinet x-ray system" means an x-ray system with the x-ray
tube installed in an enclosure independent of existing archi-
tectural structures except the floor on which it may be placed.
The cabinet x-ray system is intended to contain at least that
portion of a material being irradiated, 	provide radiation
attenuation, 	and exclude personnel from its interior during
generation of radiation. 	Included are all x-ray systems designed
primarily for the inspection of carry-on baggage at airline,
railroad, 	and bus terminals, 	and in similar facilities. 	An x-ray
tube used within a 	shielded part of a building, 	or x-ray equipment
which may temporarily or occasionally incorporate portable
shielding, 	is not considered a 	cabinet x-ray system.
"Calendar quarter" means not less than twelve consecutive
weeks nor more than fourteen consecutive weeks. 	The first
calendar quarter of each year shall begin in January and
subsequent calendar quarters 	shall be so arranged such that no day
is included in more than one calendar quarter and no day in any
one year is omitted from inclusion within a 	calendar quarter. 	No
licensee shall change the method observed by him of determining
calendar quarters for purposes of this chapter except at the
beginning of a 	calendar year.
"Calibration" means the determination of:
(1) 	The response or reading of an instrument relative to a
series of known radiation values over the range of the
instrument; 	or
(2) 	The strength of a 	source of radiation relative to a
standard.
"Carrier" means a person engaged in the transportation of
passengers or property by land or water as a common, 	contract, 	or
private carrier, 	or by civil aircraft.
"Central axis of the beam" means a 	line passing through the
virtual source and the center of the plane figure 	formed by the
edge of the first beam limiting device.
"Cephalometric device" means a device intended for the
radiographic visualization and measurement of the dimensions of
the human head.
"Certified cabinet x-ray system" means an x-ray system which
has been certified in accordance with 21 C.F.R. 	§1010.2 as being
manufactured and assembled pursuant to the provisions of 21 C.F.R.
§1020.40.
"Certified components" means components of x-ray systems
which are subject to regulations promulgated under Public Law 90-
602, 	Radiation Control 	for Health and Safety Act of 1968.
"Certified system" means any x-ray system which has one or
more certified component(s).
45-10 265
Complete printed PDF page 10
§11-45-2
"C.F.R." means Code of Federal Regulations.
"Changeable filters" means any filter, 	exclusive of inherent
filtration, 	which can be removed from the useful beam through any
electronic, 	mechanical, 	or physical process.
"Chelating agent" means amine polycarboxylic acids,
hydroxycarboxylic acids, 	gluconic acid, 	and polycarboxylic acids.
"Class" means a 	classification scheme for inhaled material
according to its rate of clearance from the pulmonary region of
the lung. 	Materials are classified as D, 	w, 	or Y, 	which applies
to a 	range of clearance half-times: 	for Class D, 	Days, 	of less
than ten days, 	for Class W, 	Weeks, 	from ten to one hundred days,
and for Class Y, 	Years, 	of greater than one hundred days. 	For
purposes of these rules, 	"lung class" and "inhalation class" 	are
equivalent terms.
"Coefficient of variation" means the ratio of the standard
deviation to the mean value of a population of observations.
"Collective dose" means the sum of the individual doses
received in a 	given period of time by a 	specified population from
exposure to a 	specified source of radiation.
"Collimator" means a device used to limit the size, 	shape,
and direction of the primary radiation beam.
"Committed dose equivalent" 	(HT,sol 	means the dose equivalent
to organs or tissues of reference 	(T) 	that will be received from
an intake of radioactive material by an individual during the
fifty-year period following the intake.
"Committed effective dose equivalent" 	(HE, 50 ) 	is the sum of
the products of the weighting factors 	applicable to each of the
body organs or tissues that are irradiated and the committed dose
equivalent to each of these organs or tissues 	(HE,so = I: wT,HT,sol.
"Computed tomography" 	(CT) 	means the production of a
tomogram by the acquisition and computer processing of x-ray
transmission data.
"Contrast scale" means the change in the linear attenuation
coefficient per CTN relative to water, 	that is:
where:
cs= (CTN) X - 	(CTN) w
Linear attenuation coefficient of the material of
interest.
μw = 	Linear attenuation coefficient of water.
(CTN)x = CTN of the material of interest.
(CTN)w = CTN of water.
"Control panel" means that part of the x-ray control upon
which are mounted the switches, 	knobs, 	push buttons, 	and other
hardware necessary for setting the technique factors.
"Controlled area" means an area, 	outside of a 	restricted
area but inside the site boundary, 	access to which can be limited
by the licensee for any reason.
"Cooling curve" means the graphical relationship between
heat units stored and cooling time.
45-11
265
Complete printed PDF page 11
§11-45-2
"CT conditions of operation" means all selectable parameters
governing the operation of a 	CT x-ray system including, 	but not
limited to, 	nominal tomographic section thickness, 	filtration, 	and
the technique factors.
"CT gantry" means the tube housing assemblies, 	beam-limiting
devices, 	detectors, 	and the supporting structures and frames which
hold these components.
"Curie" 	(Ci) 	means a unit of quantity of radioactivity. 	One
curie is that quantity of radioactive material which decays at the
rate of 3.7 x 10 10 transformations per second 	(tps).
"Dead-man switch" means a switch so constructed that a
circuit closing contact can be maintained only by continuous
pressure on the switch by the operator.
"Declared pregnant woman" means a woman who has voluntarily
informed her employer, 	in writing, 	of her pregnancy.
"Deep dose equivalent" 	(Hctl, 	which applies to external whole
body exposure, 	means the dose equivalent at a 	tissue depth of one
centimeter 	(one-thousand milligrams per square centimeter).
"Department" means the department of health.
"Derived air concentration" 	(DAC) 	means the concentration of
a 	given radionuclide in air which, 	if breathed by the reference
man for a working year of two thousand hours under conditions of
light work, 	results in an intake of one ALI. 	For purposes of
these 	rules, 	the condition of light work is an inhalation rate of
1.2 cubic meters of air per hour for two thousand hours in a 	year.
DAC values are given in Table I, 	Column 3, 	of Appendix B of
subchapter 4.
"Derived air concentration-hour" 	(DAC-hour) 	means the
product of the concentration of radioactive material in air,
expressed as a 	fraction or multiple of the derived air
concentration for each radionuclide, 	and the time of exposure to
that radionuclide, 	in hours. 	A licensee may take two thousand
DAC-hours to represent one ALI, 	equivalent to a committed
effective dose equivalent of five 	rems 	(0.05 sievert).
"Detector" 	(See "Radiation detector") .
"Diagnostic clinical procedures manual" means a 	collection
of written procedures that describes each method 	(and other
instructions and precautions) 	by which the licensee performs
diagnostic clinical procedures; 	where each diagnostic clinical
procedure has been approved by the authorized user and includes
the pharmaceutical, 	dosage, 	and route of administration.
"Diagnostic source assembly" means the tube housing assembly
with a beam-limiting device attached.
"Diagnostic x-ray imaging system" means an assemblage of
components 	for the generation, 	emission and reception of x-rays
and the transformation, 	storage and visual display of the
resultant x-ray image.
"Diagnostic x-ray system" means an x-ray system designed for
irradiation of any part of the human or animal body for the
purpose of diagnosis or visualization.
"Direct scattered radiation" means that scattered radiation
which has been deviated in direction only by materials irradiated
by the useful beam 	(See "Scattered radiation").
"Director" means the director of the department of health,
45-12
265
Complete printed PDF page 12
§11-45-2
State of Hawaii, 	or a duly authorized agent.
"Dose" is a 	generic term that means absorbed dose, 	dose
equivalent, 	effective dose equivalent, 	committed dose equivalent,
committed effective dose equivalent, 	or total effective dose
equivalent. 	For purposes of these rules, 	"radiation dose" 	is an
equivalent term.
"Dose equivalent" 	(HT) 	means the product of the absorbed
dose in tissue, 	quality factor, 	and all other necessary modifying
factors 	at the location of interest. 	The units of dose equivalent
are the sievert 	(Sv) 	and rem.
"Dose limits" 	means the permissible upper bounds of
radiation doses established in accordance with these rules. 	For
purposes of these rules, 	"limits" is an equivalent term.
"Dose monitoring system" means a 	system of devices 	for the
detection, 	measurement, 	and display of quantities of radiation.
"Dose monitor unit" means a 	unit response from the dose
monitoring system from which the absorbed dose can be calculated.
"Dose profile" means the dose as a 	function of position
along a 	line.
"Dosimetry processor" means person that processes and
evaluates individual monitoring devices in order to determine the
radiation dose delivered to the monitoring devices.
"Effective dose equivalent" 	(HE) 	means the sum of the
products of the dose equivalent to each organ or tissue 	(HT) 	and
the weighting factor 	(wT) 	applicable to each of the body organs or
tissues 	that are irradiated 	(HE= L wTHTl.
"Elemental area" means the smallest area within a 	tomogram
for which the x-ray attenuation properties of a body are depicted.
(See also 	"Picture element").
"Embryo/fetus" means the developing human organism from
conception until the time of birth.
"Entrance or access point" means any opening through which
an individual or extremity of an individual could gain access to
radiation areas or to licensed radioactive materials. 	This
includes entry or exit portals of sufficient size to permit human
entry, 	irrespective of their intended use.
"Entrance exposure rate" means the exposure free 	in air per
unit time at the point where the center of the useful beam enters
the patient.
"Existing equipment" means therapy systems subject to
subchapter 14 which were manufactured on or before January 1,
1985.
"Explosive material" means any chemical compound, 	mixture,
or device which produces a 	substantial instantaneous release of
gas and heat spontaneously or by contact with sparks or flame.
"Exposure" means, 	when used as a verb, 	being exposed to
ionizing radiation or to radioactive material. 	When used as a
noun, 	"exposure" means the quotient of dQ by dm where 	"dQ" 	is the
absolute value of the total charge of the ions of one sign
produced in air when all the electrons 	(negatrons and positrons)
liberated by photons in a volume element of air having mass 	"dm"
are completely stopped in air. 	The special unit of exposure is
the roentgen 	(R). 	One roentgen is equal to 2.58 x 10- 4 coulombs
per kilogram of air.
45-13 	265
Complete printed PDF page 13
§11-45-2
"Exposure rate" means the exposure per unit of time, 	such as
roentgen per minute and milliroentgen per hour.
"External dose" means that portion of the dose equivalent
received from any source of radiation outside the body.
"Extremity" means hand, 	elbow, 	arm below the elbow, 	foot,
knee, 	and leg below the knee.
"Eye dose equivalent" means the external dose equivalent to
the lens of the eye at a 	tissue depth of 0.3 centimeter 	(three
hundred milligrams per centimeter squared).
"Facility" means the location at which one or more devices
or sources are installed and/or located within a building, 	office,
vehicle, 	or under one roof.
"Fail-safe characteristics" mean a design feature which
causes beam port shutters to close, 	or otherwise prevents
emergence of the primary beam, 	upon the failure of a 	safety or
warning device.
"Field emission equipment" means equipment which uses an
x-ray tube in which electron emission from the cathode is due
solely to the action of an electric field.
"Field-flattening filter" means a 	filter used to provide
dose uniformity over the area of a useful beam of x-rays at a
specified depth.
"Field size" means the dimensions along the major axes of an
area in a plane perpendicular to the specified direction of the
beam of incident radiation at the normal treatment distance and
defined by the intersection of the major axes and the 	fifty
percent isodose line. 	Material 	shall be placed in the beam such
that dose maximum is produced at the normal treatment distance
when field size is being determined.
"Filter" means material placed in the useful beam to absorb
preferentially selected radiations.
"Fissile material" means any special nuclear material
consisting of or containing one or more fissile 	radionuclides.
Fissile radionuclides are plutonium-238, 	plutonium-239, 	plutonium-
241, 	uranium-233, 	and uranium-235. 	Neither natural nor depleted
uranium is 	fissile material.
"Fluoroscopic imaging assembly" means a 	subsystem in which
x-ray photons produce a 	fluoroscopic image. 	It includes the image
receptor(s) 	such as the image intensifier and spot-film device,
electrical interlocks, 	if any, 	and structural material providing
linkage between the image receptor and diagnostic source assembly.
"Focal spot 	(actual)" means the area on the anode of the
x-ray tube bombarded by the electrons accelerated from the cathode
and from which the useful beam originates.
"Gantry" means that part of the system supporting and
allowing possible movements of the radiation head.
"General purpose radiographic x-ray system" means any
radiographic x-ray system which, 	by design, 	is not limited to
radiographic examination of specific anatomical 	regions.
"Giga-" 	(G) 	means one billion times when used in conjunction
with a 	specified unit.
"Gonad shield" means a protective barrier for the testes or
ovaries.
"Gray" 	(Gy) 	means the SI unit of absorbed dose. 	One gray is
45-14
265
Complete printed PDF page 14
equal to an absorbed dose of one 	joule per kilogram 	(one hundred
rads).
§11-45-2
"Half-value layer" means the thickness of specified material
which attenuates the beam of radiation to an extent such that the
exposure rate is 	reduced to one-half of its original value. 	In
this definition, 	the contribution of all scattered radiation,
other than any which might be present initially in the beam
concerned, 	is deemed to be excluded.
"Hazardous waste" means those wastes designated as hazardous
by U.S. 	Environmental Protection Agency regulations in 40 C.F.R.
Part 261.
"Healing arts" means the medical, 	dental, 	chiropractic,
podiatric, 	and veterinary professions.
"Healing arts screening" means the testing of human beings
using x-ray machines 	for the detection or evaluation of health
indications when such tests are not specifically and individually
ordered by a 	licensed practitioner of the healing arts legally
authorized to prescribe such x-ray tests 	for the purpose of
diagnosis or treatment.
"High radiation area" means an area, 	accessible to
individuals, 	in which radiation levels could result in an
individual receiving a dose equivalent in excess of 0.1 rem 	(one
millisievert) 	in one hour at thirty centimeters from any source of
radiation or from any surface that the radiation penetrates. 	For
purposes of these rules, 	rooms or areas in which diagnostic x-ray
systems are used for healing arts purposes are not considered high
radiation areas.
"HRS" means Hawaii Revised Statutes.
"Human use" means 	the internal or external administration of
radiation or radioactive material to human beings.
"HVL" See "Half-value layer".
"Image intensifier" means a device, 	installed in its
housing, 	which instantaneously converts an x-ray pattern into a
corresponding light image of higher energy density.
"Image receptor" means any device, 	such as a 	fluorescent
screen or radiographic film, 	which transforms incident x-ray
photons either into a visible image or into another form which can
be made into a visible image by further transformations.
"Image receptor support" means, 	for mammographic systems,
that part of the system designed to support the image receptor
during mammography.
"Individual" means any human being.
"Individual monitoring" means the assessment of:
(1) 	Dose equivalent by the use of individual monitoring
devices or by the use of survey data; 	or
(2) 	Committed effective dose equivalent by bioassay or by
determination of the time-weighted air concentrations
to which an individual has been exposed, 	that is, 	DAC-
hours.
"Individual monitoring devices" 	means devices designed to
be worn by a 	single individual 	for the assessment of dose
equivalent. 	For purposes of these rules, 	individual monitoring
equipment and personnel monitoring equipment are equivalent terms.
Examples of individual monitoring devices are film badges,
45-15
Complete printed PDF page 15
§11-45-2
thermoluminescent dosimeters 	(TLDs), 	pocket ionization chambers,
and personal air sampling devices.
"Industrial radiography" means the examination of the
macroscopic structure of materials by nondestructive methods using
sources of ionizing radiation to produce radiographic images.
"Inherent filtration" means the filtration of the useful
beam provided by the permanently installed components of the tube
housing assembly.
"Inspection" means an official examination or observation
including but not limited to, 	tests, 	surveys, 	and monitoring to
determine compliance with rules, 	orders, 	requirements, 	and
conditions of the department.
"Interlock" means a 	device arranged or connected such that
the occurrence of an event or condition is required before a
second event or condition can occur or continue to occur.
"Internal dose" means that portion of the dose equivalent
received from radioactive material taken into the body.
"Interruption of irradiation" means the stopping of
irradiation with the possibility of continuing irradiation without
resetting of operating conditions at the control panel.
"Ionizing radiation" means any radiation consisting of
charged particles having sufficient kinetic energy to produce
ionization by collision, 	or uncharged particles which can liberate
directly ionizing particles or can initiate a nuclear
transformation, 	or a mixture of both charged and uncharged
particles.
"Irradiation" means the exposure of matter to ionizing
radiation.
"Isocenter" means a 	fixed point in space located at the
center of the smallest sphere through which the central axis of
the beams passes in all conditions.
"Kilo-" 	(k) 	means one thousand times when used in
conjunction with a 	specified unit.
"Kilovolts peak" 	(See "Peak tube potential").
"kV" means 	kilovolts.
"kVp" 	(See "Peak tube potential").
"kWs" means 	kilowatt second.
"Landfill" means a 	disposal 	facility or part of a 	facility
at which solid waste is permanently placed in or on land and which
is not a 	landspreading facility.
"Landspreading facility" means a 	facility that applies
sludges or other solid wastes onto or incorporates solid waste
into the soil surface at greater than vegetative utilization and
soil conditioner and/or immobilization rates.
"Lead equivalent" means the thickness of lead affording the
same attenuation, 	under specified conditions, 	as the material in
question.
"Leakage radiation" means 	radiation emanating from the
diagnostic or therapeutic source assembly except for:
(1) 	The useful beam; 	and
(2) 	Radiation produced when the exposure switch or timer
is not activated.
"Leakage technique factors" means the technique factors
associated with the diagnostic source assembly which are used in
45-16
265
Complete printed PDF page 16
§11-45-2
measuring leakage radiation. 	They are defined as 	follows:
(1) 	For diagnostic source assemblies intended for
capacitor energy storage equipment, 	the maximum-rated
peak tube potential and the maximum-rated number of
exposures in an hour for operation at the
maximum-rated peak tube potential with the quantity of
charge per exposure being ten millicoulombs, 	i.e., 	ten
milliampere seconds, 	or the minimum obtainable from
the unit, 	whichever is larger.
(2) 	For diagnostic source assemblies intended for 	field
emission equipment rated for pulsed operation, 	the
maximum-rated peak tube potential and the
maximum-rated number of x-ray pulses in an hour for
operation at the maximum-rated peak tube potential.
(3) 	For all other diagnostic source assemblies, 	the
maximum-rated peak tube potential and the
maximum-rated continuous tube current for the
maximum-rated peak tube potential.
"License" means a written authorization issued by the
department in accordance with the rules adopted by the department,
when used without reference to the U.S. 	Nuclear Regulatory
Commission, 	agreement state, 	or other state.
"Licensee" means any person who is issued a license by the
department and is legally obligated to have a 	license with the
department pursuant to these rules.
"Light field" means that area of the intersection of the
light beam from the beam-limiting device and one of the set of
planes parallel to and including the plane of the image receptor,
whose perimeter is the locus of points at which the illumination
is one-fourth of the maximum in the intersection.
"Limits" 	(See "Dose limits").
"Line-voltage regulation" means the difference between the
no-load and the load line potentials expressed as a percent of the
load line potential. 	It is calculated using the following
equation: 	Percent line-voltage regulation= 100 	(Vn-Vl)/Vl where
Vn = No-load line potential and Vl = Load line potential.
"Local components" mean part of an analytical x-ray system
and include areas that are struck by x-rays such as 	radiation
source housings, 	port and shutter assemblies, 	collimators, 	sample
holders, 	cameras, 	goniometers, 	detectors, 	and shielding, 	but do
not include power supplies, 	transformers, 	amplifiers, 	readout
devices, 	and control panels.
"Lost or missing licensed material" means licensed material
whose location is unknown. 	This definition includes licensed
material that has been shipped but has not reached its planned
destination and whose location cannot be readily traced in the
transportation system.
"mA" means milliampere.
"mAs" means milliampere second.
"Major processor" means a 	user processing, 	handling, 	or
manufacturing radioactive material exceeding Type A quantities as
unsealed sources or material, 	or exceeding four times Type B
quantities as sealed sources, 	but does not include nuclear
medicine programs, 	universities, 	industrial 	radiographers, 	or
45-17 265 .!...!..
Complete printed PDF page 17
§11-45-2
small industrial programs. 	Type A and B quantities are defined in
10 C.F.R. 	§71.4.
"Maximum line current" means the root-mean-square current in
the supply line of an x-ray machine operating at its maximum
rating.
"Mega-" 	(M) 	means a million times when used in conjunction
with a 	specified unit.
"Member of the public" means an individual in a 	controlled
or unrestricted area. 	However, 	an individual is not a member of
the public during any period in which the individual receives an
occupational dose.
"Micro-" 	(μ) 	means one-millionth of when used in conjunction
with a 	specified unit.
"Milli-" 	(m) 	means one-thousandth of when used in
conjunction with a 	specified unit.
"Minor" means an individual less than eighteen years of age.
"Misadministration" means the administration of:
(1) 	External beam radiation therapy involving the wrong
patient, 	wrong treatment modality, 	or wrong treatment
site; 	when the treatment consists of three or fewer
fractions 	and the calculated total administered dose
differs 	from the total prescribed dose by more than
ten percent of the total prescribed dose; 	when the
calculated weekly administered dose is thirty percent
greater than the weekly prescribed dose; 	or when the
calculated total administered dose differs 	from the
total prescribed dose by more than twenty percent of
the total prescribed dose;
(2) 	A therapeutic radiopharmaceutical dosage involving the
wrong patient, 	wrong radiopharmaceutical, 	or wrong
route of administration, 	or when the administered
dosage differs 	from the prescribed dosage by more than
twenty percent of the prescribed dosage;
(3) 	A gamma stereotactic radiosurgery radiation dose
involving the wrong patient or wrong treatment site,
or when the calculated total administered dose differs
from the total prescribed dose by more than ten
percent of the total prescribed dose;
(4) 	A teletherapy radiation dose involving the wrong
patient, 	wrong mode of treatment, 	or wrong treatment
site, 	or when the treatment consists of three or fewer
fractions 	and the calculated total administered dose
differs from the total prescribed dose by more than
ten percent of the total prescribed dose, 	or when the
calculated weekly administered dose is thirty percent
greater than the weekly prescribed dose, 	or when the
calculated total administered dose differs 	from the
total prescribed dose by more than twenty percent of
the total prescribed dose;
(5) 	A brachytherapy radiation dose involving the wrong
patient, 	wrong radionuclide, 	or wrong treatment site
(excluding, 	for permanent implants, 	seeds that were
implanted in the correct site but migrated outside the
treatment site), 	or involving a 	sealed source that is
45-18
Complete printed PDF page 18
§11-45-2
leaking, 	or when, 	for a 	temporary implant, 	one or more
sealed sources are not removed upon completion of the
procedure, 	or when the calculated administered dose to
the treatment site differs from the prescribed dose by
more than twenty percent of the prescribed dose;
(6) 	A diagnostic radiopharmaceutical dosage, 	both
involving the wrong patient, 	wrong
radiopharmaceutical, 	wrong route of administration, 	or
when the administered dosage differs by more than
twenty percent of the prescribed dosage, 	and when the
dose to the patient exceeds 	five 	rems 	(fifty
millisieverts) 	effective dose equivalent or fifty rems
(five hundred millisieverts) 	dose equivalent to any
individual organ.
"Mobile nuclear medicine service" means the transportation
and medical use of radioactive material.
"Mobile x-ray equipment" 	(See "X-ray equipment").
"Monitoring" means the measurement of radiation, 	radioactive
material concentrations, 	surface area activities or quantities of
radioactive material and the use of the results of these
measurements to evaluate potential exposures and doses. 	For
purposes of these rules, 	radiation monitoring and radiation
protection monitoring are equivalent terms.
"Moving beam therapy" means 	radiation therapy with relative
displacement of the useful beam and the patient during
irradiation. 	It includes arc therapy, 	skip therapy, 	and
rotational therapy.
"Multiple tomogram system" means a computed tomography x-ray
system which obtains x-ray transmission data simultaneously during
a 	single scan to produce more than one tomogram.
"Nano-" 	(n) 	means one-billionth of when used in conjunction
with a 	specified unit.
"NARM" means any naturally-occurring and/or accelerator-
produced radioactive material. 	It does not include byproduct,
source, 	or special nuclear material.
"Natural radioactivity" means 	radioactivity of naturally
occurring nuclides.
"New equipment" means systems subject to subchapter 14 which
were manufactured after January 1, 	1985.
"Noise" means the standard deviation of the fluctuations 	in
CTN expressed as a percentage of the attenuation coefficient of
water. 	Its estimate 	(Snl 	is calculated using the following
expression:
where:
100 X CSX S
μw
CS 	Contrast scale
μw = 	Linear attenuation coefficient of water.
s 	= 	Estimated standard deviation of the CTN of picture
elements in a 	specified area of the CT image.
"Nominal tomographic section thickness" means the full width
45-19
6
Complete printed PDF page 19
§11-45-2
at half-maximum of the sensitivity profile taken at the center of
the cross-sectional volume over which x-ray transmission data are
collected.
"Nonstochastic effect" means a health effect, 	the severity
of which varies with the dose and for which a 	threshold is
believed to exist. 	Radiation-induced cataract formation is an
example of a nonstochastic effect. 	For purposes of these rules, 	a
"deterministic effect" is an equivalent term.
"Normal operating procedures" mean step-by-step instructions
necessary to accomplish the analysis. 	These procedures shall
include sample insertion and manipulation, 	equipment alignment,
routine maintenance by the licensee, 	and data recording
procedures, 	which are related to radiation safety.
"Normal treatment distance" means:
(1) 	For electron irradiation, 	the virtual source to
surface distance along the central axis of the useful
beam as specified by the manufacturer for the
applicator; 	and
(2) 	For x-ray irradiation, 	the virtual source to isocenter
distance along the central axis of the useful beam.
For non-isocentric equipment, 	this distance shall be
that specified by the manufacturer.
"Occupational dose" means the dose received by an individual
in a 	restricted area or in the course of employment in which the
individual's assigned duties involve exposure to sources of
radiation, 	whether in the possession of the licensee, 	or other
person. 	Occupational dose does not include dose received: 	from
background radiation, 	as a patient from medical practices, 	from
voluntary participation in medical research programs, 	or as a
member of the public.
"Open-beam configuration" means an analytical x-ray system
in which an individual could accidentally place some part of his
body in the primary beam path during normal operation.
"Package" means the packaging together with its radioactive
contents as presented for transport.
"Particle accelerator" means any machine capable of
accelerating electrons, 	protons, 	deuterons, 	or other charged
particles in a vacuum and of discharging the resultant particulate
or other radiation into a medium at energies usually in excess of
one megaelectron volt.
"Patient" means an individual or animal subjected to healing
arts examination, 	diagnosis, 	or treatment.
"PBL" 	(See "Positive Beam Limitation").
"Peak tube potential" means the maximum value of the
potential difference across the x-ray tube during an exposure.
"Permanent radiographic installation" means an installation
or structure designed or intended for radiography and in which
radiography is regularly performed.
"Person" means any individual, 	partnership, 	firm,
association, 	public or private corporation, 	trust estate or any
other legal entity.
"Personal supervision" means guidance and instruction
provided to a 	radiographer trainee by a 	radiographer instructor
who is present at the site, 	in visual contact with the trainee
45-20
265
Complete printed PDF page 20
while the trainee is using sources of radiation, 	and in such
proximity that immediate assistance can be given if required.
"Personnel monitoring equipment" 	(See 	"Individual
monitoring devices").
"Phantom" means a volume of material behaving in a manner
similar to tissue with respect to the attenuation and scattering
of radiation. 	This requires that both the atomic number 	(Z) 	and
the density of the material be similar to that of tissue.
§11-45-2
"Picture element" means an elemental area of a 	tomogram.
"Pico-" 	(p) 	means one-trillionth of when used in conjunction
with a 	specified unit.
"PID" 	(See "Position indicating device").
"Planned special exposure" means an infrequent exposure to
radiation, 	separate from and in addition to the annual
occupational dose limits.
"Portable x-ray equipment" 	(See "X-ray equipment").
"Position indicating device" means a device on dental x-ray
equipment used to indicate the beam position and to establish a
definite source-surface 	(skin) 	distance. 	It may or may not
incorporate or serve as a beam-limiting device.
"Positive beam limitation" means the automatic or semi-
automatic adjustment of an x-ray beam to the selected image
receptor size, 	whereby exposures cannot be made without such
adjustment.
"Prescribed dosage" means the quantity of
radiopharmaceutical activity as documented:
(1) 	In a written directive; 	or
(2) 	Either in the diagnostic clinical procedures manual or
in any appropriate record in accordance with the
directions of the authorized user for diagnostic
procedures.
"Prescribed dose" means:
(1) 	The total dose and dose per fraction as documented in
the written directive for external beam radiotherapy
which is an estimation from measured data from a
specified therapeutic radiation machine using
assumptions that are clinically acceptable for that
treatment technique and historically consistent with
the clinical calculations previously used for patients
treated with the same clinical technique; 	or
(2) 	The total dose as documented in the written directive
for gamma stereotactic radiosurgery; 	or
(3) 	The total dose and dose per fraction as documented in
the written directive for teletherapy; 	or
(4) 	Either the total source strength and exposure time, 	or
the total dose, 	as documented in the written directive
for brachytherapy.
"Primary beam" means 	radiation which passes through an
aperture of the source housing by a direct path from the x-ray
tube or a 	radioactive source located in the radiation source
housing.
"Primary protective barrier" See "Protective barrier".
"Protective apron" means an apron made of radiation
absorbing materials used to reduce radiation exposure.
45-21
Complete printed PDF page 21
§11-45-2
"Protective barrier" means a barrier of radiation absorbing
material(s) 	used to 	reduce radiation exposure. 	The types of
protective barriers are as 	follows:
( 1) 	"Primary protective barrier" means the material,
excluding filters, 	placed in the useful beam.
(2) 	"Secondary protective barrier" means the material
which attenuates stray radiation.
"Protective glove" means a 	glove made of radiation absorbing
materials used to reduce radiation exposure.
"Public dose" means the dose received by a member of the
public from exposure to sources of radiation either within a
licensee's controlled area or in unrestricted areas. 	It does not
include occupational dose, 	dose received from background
radiation, 	dose received as a patient from medical practices, 	or
dose from voluntary participation in medical research programs.
"Qualified health physicist" means a physicist licensed to
provide health physics services.
"Qualified medical physicist" means a physicist licensed to
provide medical physics services.
"Quality factor" 	(Q) 	means the modifying factor that is used
to derive dose equivalent from absorbed dose.
"Quarter" means a period of time equal to one-fourth of the
year observed by the licensee, 	approximately thirteen consecutive
weeks, 	providing that the beginning of the first quarter in a 	year
coincides with the starting date of the year and that no day is
omitted or duplicated in consecutive quarters.
"Rad" means the special unit of absorbed dose. 	One rad is
equal to an absorbed dose of one hundred erg per gram or 0.01
joule per kilogram 	(0.01 gray).
"Radiation" means alpha particles, 	beta particles, 	gamma
rays, 	x-rays, 	neutrons, 	high-speed electrons, 	high-speed protons,
and other particles capable of producing ions. 	For purposes of
these rules, 	ionizing radiation is an equivalent term. 	Radiation,
as used in these rules, 	does not include non-ionizing radiation,
such as radiowaves or microwaves, 	visible, 	infrared, 	or
ultraviolet light.
"Radiation area" means any area, 	accessible to individuals,
in which radiation levels could result in an individual receiving
a dose equivalent in excess of 0.005 rem 	(0.05 millisievert) 	in
one hour at thirty centimeters from the source of radiation or
from any surface that the radiation penetrates.
"Radiation detector" means a device which in the presence of
radiation provides a 	signal or other indication suitable for use
in measuring one or more quantities of incident radiation.
"Radiation head" means the structure from which the useful
beam emerges.
"Radiation machine" means any device capable of producing
radiation except those devices with radioactive material as 	the
only source of radiation.
"Radiation safety officer" means an individual who has 	the
knowledge and responsibility to apply appropriate radiation
protection regulations.
"Radiation therapy simulation system" means a 	radiographic
or fluoroscopic x-ray system intended for localizing the volume to
45-22
265
Complete printed PDF page 22
be exposed during radiation therapy and confirming the position
and size of the therapeutic irradiation field.
§11-45-2
"Radioactive material" means any solid, 	liquid, 	or gas which
emits radiation spontaneously.
"Radioactivity" means the transformation of unstable atomic
nuclei by the emission of radiation.
"Radiobioassay" 	(See "Bioassay").
"Radiograph" means an image receptor on which the image is
created directly or indirectly by an x-ray pattern and results in
a permanent record.
"Radiographer" means any individual in industrial
radiography who performs or personally supervises industrial
radiographic operations, 	or any individual in the healing arts who
practices radiography on human patients as specified in chapter
11-44.
"Radiographer instructor" means any radiographer who has
been authorized by the department to provide on-the-job training
to 	radiographer trainees, 	relative to industrial radiography.
"Radiographer trainee" means any individual who, 	under the
personal supervision of a 	radiographer instructor, 	uses sources of
radiation, 	related handling tools, 	or radiation survey instruments
during the course of his instruction, 	relative to industrial
radiography.
"Radiographic imaging system" means any system whereby a
permanent or temporary image is recorded on an image receptor by
the action of ionizing radiation.
"Radiographic personnel" means any radiographer,
radiographer instructor, 	or radiographer trainee, 	relative to
industrial radiography.
"Rating" means the operating limits as specified by the
component manufacturer.
"Recordable event" means the administration of:
(1) 	An external beam radiation therapy dose when the
calculated weekly administered dose is 	fifteen percent
greater than the weekly prescribed dose;
(2) 	A radiopharmaceutical or radiation without a written
directive where a written directive is required;
(3) 	A radiopharmaceutical or radiation where a written
directive is required without daily recording of each
administered radiopharmaceutical dosage or radiation
dose in the appropriate record;
(4) 	A therapeutic radiopharmaceutical dosage when the
administered dosage differs 	from the prescribed dosage
by more than ten percent of the prescribed dosage;
(5) 	A teletherapy radiation dose when the calculated
weekly administered dose is fifteen percent greater
than the weekly prescribed dose; 	or
(6) 	A brachytherapy radiation dose when the calculated
administered dose differs 	from the prescribed dose by
more than ten percent of the prescribed dose.
"Recording" means producing a permanent form of an image
resulting from x-ray photons.
"Reference man" means a hypothetical aggregation of human
physical and physiological characteristics determined by
45-23
Complete printed PDF page 23
§11-45-2
international consensus. 	These characteristics may be used by
researchers and public health workers 	to standardize results of
experiments and to relate biological insult to a common base.
"Reference plane" means a plane which is displaced from and
parallel to the tomographic plane.
"Regulations of the U.S. 	Department of Transportation" means
the regulations in 49 C.F.R. 	Parts 100-189.
"Rem" means the special unit of any of the quantities
expressed as dose equivalent. 	The dose equivalent in rem is equal
to the absorbed dose in rad multiplied by the quality factor 	(one
rem= 0.01 sievert).
"Research and development" means:
(1) 	Theoretical analysis, 	exploration, 	or experimentation;
or
(2) 	The extension of investigative findings and theories
of a 	scientific or technical nature into practical
application for experimental and demonstration
purposes, 	including the experimental production and
testing of models, 	devices, 	equipment, 	materials, 	and
processes. 	Research and development does not include
the internal or external administration of radiation
or radioactive material to human beings.
"Residential location" means any area where structures in
which people lodge or live are located, 	and the grounds on which
such structures are located including, 	but not limited to, 	houses,
apartments, 	condominiums, 	and garages.
"Respiratory protective equipment" means an apparatus, 	such
as a 	respirator, 	used to 	reduce an individual's intake of airborne
radioactive materials.
"Restricted area" means any area access 	to which is limited
by the licensee for purposes of protection of individuals against
undue risks 	from exposure to sources of radiation. 	A restricted
area does not include areas used as 	residential quarters, 	but
separate rooms in a 	residential building may be set apart as a
restricted area.
"Roentgen" 	(R) 	means the special unit of exposure. 	One
roentgen 	(R) 	equals 2. 58 x 10- 4 coulombs per kilogram of air. 	(See
"Exposure") .
"Sanitary sewerage" means a 	system of public sewers 	for
carrying off waste water and refuse, 	but excluding sewage
treatment facilities, 	septic tanks, 	and leach fields owned or
operated by the licensee.
"Scan" means the complete process of collecting x-ray
transmission data 	for the production of a 	tomogram. 	Data can be
collected simultaneously during a single scan for the production
of one or more tomograms.
"Scan increment" means the amount of relative displacement
of the patient with respect to the CT x-ray system between
successive scans measured along the direction of such
displacement.
"Scan sequence" means a preselected set of two or more scans
performed consecutively under preselected CT conditions of
operation.
"Scan time" means the period of time between the beginning
45-24
Complete printed PDF page 24
§11-45-2
and end of x-ray transmission data accumulation for a 	single scan.
"Scattered radiation" means radiation that, 	during passage
through matter, 	has been deviated in direction. 	(See "Direct
scattered radiation").
"Sealed source" means radioactive material that is
permanently bonded or fixed in a 	capsule or matrix designed to
prevent release and dispersal of the radioactive material under
the most severe conditions which are likely to be encountered in
normal use and handling.
"Secondary protective barrier" 	(See "Protective barrier").
"Shadow tray" means a device attached to the 	radiation head
to support auxiliary beam limiting material.
"Shallow dose equivalent" 	(Hsi, 	which applies to the
external exposure of the skin or an extremity, 	means the dose
equivalent at a 	tissue depth of 0.007 centimeter 	(seven milligrams
per centimeter squared) 	averaged over an area of one square
centimeter.
"Shielded-room radiography" means industrial 	radiography
conducted in a 	room shielded so that radiation levels at every
location on the exterior meet the limitations specified in
subchapter 4 of these rules.
"Shutter" means a device attached to the tube housing
assembly which can intercept the entire cross sectional area of
the useful beam and which has a 	lead equivalency not less than
that of the tube housing assembly.
"SI" means an abbreviation of the International System of
Units.
"SID" 	(See 	"Source-image receptor distance").
"Sievert" 	(Sv) 	means the SI unit of any of the quantities
expressed as dose equivalent. 	The dose equivalent in sievert is
equal to the absorbed dose in gray multiplied by the quality
factor 	(one sievert = one hundred rem).
"Single tomogram system" means a 	CT x-ray system which
obtains x-ray transmission data during a 	scan to produce a 	single
tomogram.
"Site boundary" means that line beyond which the land or
property is not owned, 	leased, 	or otherwise controlled by the
licensee.
"Source" means the focal 	spot of the x-ray tube.
"Source-image receptor distance" means 	the distance 	from the
source to the center of the input surface of the image receptor.
"Source material" means:
(1) 	Uranium or thorium, 	or any combination thereof, 	in any
physical or chemical form; 	or
(2) 	Ores that contain by weight one-twentieth of one
percent 	(0.05 percent) 	or more of uranium, 	thorium or
any combination of uranium and thorium. 	Source
material does not include special nuclear material.
"Source material milling" means any activity that 	results in
the production of byproduct material as defined by definition 	(2)
of byproduct material.
"Source of radiation" means any radioactive material or any
device or equipment emitting, 	or capable of producing radiation.
"Special 	form radioactive material" means 	radioactive
45-25
265
Complete printed PDF page 25
§11-45-2
material which satisfies the following conditions:
(1) 	It is either a single solid piece or is contained in a
sealed capsule that can be opened only by destroying
the capsule;
(2) 	The piece or capsule has at least one dimension not
less than five millimeters 	(0.197 inch); 	and
(3) 	It satisfies the test requirements specified by the
U.S. 	Nuclear Regulatory Commission. 	A special form
encapsulation designed in accordance with the U.S.
Nuclear Regulatory Commission requirements in effect
on June 30, 	1983, 	and constructed before July 1, 	1985,
may continue to be used. 	A special form encapsulation
either designed or constructed after June 30, 	1985,
shall meet requirements of this definition applicable
at the time of its design or construction.
"Special nuclear material" means:
(1) 	Plutonium, 	uranium-233, 	uranium enriched in the
isotope 233 or in the isotope 235, 	and any other
material that the U.S. 	Nuclear Regulatory Commission,
pursuant to the provisions of section 51 of the Atomic
Energy Act of 1954, 	as amended, 	determines to be
special nuclear material, 	but does not include source
material; 	or
(2) 	Any material artificially enriched by any of the
foregoing but does not include source material.
"Special nuclear material in quantities not sufficient to
form a 	critical mass" means uranium enriched in the isotope U-235
in quantities not exceeding three hundred fifty grams of contained
U-235; 	uranium-233 in quantities not exceeding two hundred grams;
plutonium in quantities not exceeding two hundred grams; 	or any
combination of them in accordance with the following formula: 	For
each kind of special nuclear material, 	determine the ratio between
the quantity of that special nuclear material and the quantity
specified above for the same kind of special nuclear material.
The sum of such ratios 	for all of the kinds of special nuclear
material in combination shall not exceed one. 	For example, 	the
following quantities in combination would not exceed the
limitation and are within the formula:
175(grams contained U-235)+ 50(grams U-233)+50(grams 	Pu) 	1
350 	200 	200
"Specific activity" of a 	radionuclide means the
radioactivity of a 	radionuclide per unit mass of that nuclide.
The specific activity of a material in which the radionuclide is
essentially uniformly distributed is the radioactivity per unit
mass of the material.
"Spot film" means a 	radiograph which is made during a
fluoroscopic examination to permanently record conditions which
exist during that fluoroscopic procedure.
"Spot-film device" means a device intended to transport
and/or position a 	radiographic image receptor between the x-ray
source and fluoroscopic image receptor. 	It includes 	a 	device
intended to hold a 	cassette over the input end of an image
45-26 265
Complete printed PDF page 26
intensifier for the purpose of making a 	radiograph.
"SSD" means the distance between the source and the skin
entrance plane of the patient.
"Stationary beam therapy" means 	radiation therapy without
relative displacement of the useful beam and the patient during
irradiation.
§11-45-2
"Stationary x-ray equipment" 	(See "X-ray equipment").
"Stochastic effect" means a health effect that occurs
randomly and for which the probability of the effect occurring,
rather than its severity, 	is assumed to be a 	linear function of
dose without threshold. 	Hereditary effects and cancer incidence
are examples of stochastic effects. 	For purposes of these rules,
probabilistic effect is an equivalent term.
"Storage" means a 	condition in which a device or source is
not being used for an extended period of time, 	and has been made
inoperable.
"Storage area" means any location, 	facility, 	or vehicle
which is used to store, 	to transport, 	or to secure an x-ray system
when it is not in use and which is locked or has a physical
barrier to prevent accidental exposure, 	tampering with, 	or
unauthorized removal of the device.
"Stray radiation" means the sum of leakage and scattered
radiation.
"Survey" means an evaluation of the radiological conditions
and potential hazards incident to the production, 	use, 	transfer,
release, 	disposal, 	or presence of sources of radiation. 	When
appropriate, 	such evaluation includes, 	but is not limited to,
tests, 	physical examinations, 	and measurements of levels of
radiation or concentrations of radioactive material present.
"Target" means that part of a 	radiation head which by design
intercepts a beam of accelerated particles with subsequent
emission of other radiation.
"Technique factors" means the following conditions of
operations:
(1)
(2)
(3)
( 4)
(5)
For capacitor energy storage equipment, 	peak tube
potential in kV and quantity of charge in mAs;
For field emission equipment rated for pulsed
operation, 	peak tube potential in kV and number of
x-ray pulses;
For CT x-ray systems designed for pulsed operation,
peak tube potential in kV, 	scan time in seconds, 	and
either tube current in mA, 	x-ray pulse width in
seconds, 	and the number of x-ray pulses per scan, 	or
the product of tube current, 	x-ray pulse width, 	and
the number of x-ray pulses in mAs;
For CT x-ray systems not designed for pulsed
operation, 	peak tube potential in kV, 	and either tube
current in mA and scan time in seconds, 	or the product
of tube current and exposure time in mAs and the scan
time when the scan time and exposure time are
equivalent; 	and
For all other equipment, 	peak tube potential in
kV, 	and either tube current in mA and exposure
time in seconds, 	or the product of tube current
45-27 265
Complete printed PDF page 27
§11-45-2
and exposure time in mAs.
"Temporary job site" means any location where industrial
radiography is performed other than the location(s) 	listed on the
license.
"Termination of irradiation" means the stopping of
irradiation in a 	fashion which shall not permit continuance of
irradiation without the resetting of operating conditions at the
control panel.
"Test" means the process of verifying compliance with an
applicable provisions of this chapter.
"Tomogram" means the depiction of the x-ray attenuation
properties of a 	section through a body.
"Tomographic plane" means that geometric plane which is
identified as corresponding to the output tomogram.
"Tomographic section" means the volume of an object whose x-
ray attenuation properties are imaged in a 	tomogram.
"Total effective dose equivalent" 	(TEDE) 	means the sum of
the deep dose equivalent for external exposures and the committed
effective dose equivalent for internal exposures.
"Transport index" means the dimensionless number, 	rounded up
to the first decimal place, 	placed on the label of a package to
designate the degree of control to be exercised by the carrier
during transportation. 	The transport index is the number
expressing the maximum radiation level in millirem per hour at one
meter from the external surface of the package.
"Tube" means an x-ray tube, 	unless otherwise specified.
"Tube housing assembly" means the tube housing with tube
installed. 	It includes high-voltage and/or filament transformers
and other appropriate elements when such are contained within the
tube housing.
"Tube rating chart" means the set of curves which specify
the rated limits of operation of the tube in terms of the
technique factors.
"Type A quantity" means a 	quantity of radioactive material,
the aggregate radioactivity of which does not exceed A1 for
special form radioactive material or A2 for normal 	form
radioactive material.
"Uncontrolled area" means any area access 	to which is not
controlled by the licensee for purposes of protection of
individuals from exposure to radiation and radioactive material,
and any area used for 	residential quarters.
"Unrefined and unprocessed ore" means ore in its natural
form prior to any processing, 	such as grinding, 	roasting,
beneficiating, 	or refining.
"Unrestricted area" means an area, 	access to which is
neither limited nor controlled by the licensee. 	For purposes of
these rules, 	"uncontrolled area" 	is an equivalent term.
"Useful beam" means the radiation emanating from the tube
housing port or the radiation head and passing through the
aperture of the beam limiting device when the exposure controls
are in a mode to cause the system to produce radiation.
"Variable-aperture beam-limiting device" means a
beam-limiting device which has capacity for stepless adjustment of
the x-ray field size at a 	given SID.
45-28
6
Complete printed PDF page 28
§11-45-2
"Very high radiation area" means an area, 	accessible to
individuals, 	in which radiation levels could result in an
individual receiving an absorbed dose in excess of five hundred
rads 	(five grays) 	in one hour at one meter from a 	source of
radiation or from any surface that the radiation penetrates. 	(At
very high doses 	received at high dose rates, 	units of absorbed
dose, 	rad and gray, 	are appropriate, 	rather than units of dose
equivalent, 	rem and sievert.)
"Virtual source" means a point from which radiation appears
to originate.
"Visible area" means that portion of the input surface of
the image receptor over which incident x-ray photons are producing
a visible image.
"Waste" means those low-level 	radioactive wastes that are
acceptable for disposal in a land disposal facility. 	For the
purposes of this definition, 	low-level waste has the same meaning
as in the Low-Level Radioactive Waste Policy Act, 	P.L. 	96-573, 	as
amended by P.L. 	99-240, 	effective January 15, 	1986; 	that is,
radioactive waste 	(a) 	not classified as high-level radioactive
waste, 	spent nuclear fuel, 	or byproduct material as defined in
Section lle. (2) 	of the Atomic Energy Act 	(uranium or thorium
tailings and waste) 	and 	(b) 	classified as low-level radioactive
waste consistent with existing law and in accordance with 	(a) 	by
the U.S. 	Nuclear Regulatory Commission.
"Waste handling licensees" mean persons licensed to 	receive
and store radioactive wastes before disposal.
"Week" means seven consecutive days starting on Sunday.
"Weighting factor" wT for an organ or tissue 	(Tl 	means the
proportion of the risk of stochastic effects resulting from
irradiation of that organ or tissue to the total 	risk of
stochastic effects when the whole body is irradiated uniformly.
For calculating the effective dose equivalent, 	the values of wT
are:
ORGAN DOSE WEIGHTING
organ or
Tissue
Gonads
Breast
Red bone marrow
Lung
Thyroid
Bone surfaces
Remainder
Whole Body
FACTORS
0.25
0.15
0.12
0.12
0.03
0.03
0. 30a
a 0.30 results 	from 0.06 for each of five
"remainder" organs, 	excluding the skin and
the lens of the eye, 	that receive the
highest doses.
45-29
Complete printed PDF page 29
§11-45-2
b For the purpose of weighting the
external whole body dose, 	for adding it to
the internal dose, 	a single weighting
factor, 	wT = 1.0, 	has been specified. 	The
use of other weighting factors 	for
external exposure shall be approved on a
case-by-case basis until such time as
specific guidance is issued.
"Whole body" means, 	for purposes of external exposure, 	head,
trunk including male gonads, 	arms above the elbow, 	or legs above
the knee.
"Worker" means an individual engaged in work licensed and
controlled by a 	licensee, 	but does not include the licensee.
"Working level" 	(WL) 	means any combination of short-lived
radon daughters in one liter of air that shall result in the
ultimate emission of 1.3 x 10 5 mega electron volts of potential
alpha particle energy. 	The short-lived radon daughters are -- 	for
radon-222: 	polonium-218, 	lead-214, 	bismuth-214, 	and polonium-214;
and for radon-220: 	polonium-216, 	lead-212, 	bismuth-212, 	and
polonium-212.
"Working level month" 	(WLM) 	means an exposure to one working
level 	for one hundred seventy hours -- two thousand working hours
per year divided by twelve months per year is approximately equal
to one hundred seventy hours per month.
"Written directive" means an order in writing for a 	specific
patient, 	dated and signed by an authorized user before the
administration of a 	radiopharmaceutical or radiation, 	except as
specified in 	(6) 	containing the following information:
(1) 	For external beam radiotherapy, 	total dose, 	dose per
fraction, 	treatment site and overall treatment period;
or
(2) 	For a 	therapeutic administration of a
radiopharmaceutical; 	the radiopharmaceutical, 	dosage,
and route of administration; 	or
(3) 	For gamma stereotactic radiosurgery: 	target
coordinates, 	collimator size, 	plug pattern, 	and total
dose; 	or
(4) 	For teletherapy: 	the total dose, 	dose per fraction,
treatment site, 	and overall treatment period; 	or
(5) 	For high-dose-rate remote afterloading brachytherapy:
the 	radionuclide, 	treatment site, 	and total dose; 	or
(6) 	For all other brachytherapy, 	before implantation: 	the
radionuclide, 	number of sources, 	and source strengths;
and after implantation but before completion of the
procedure: 	the radionuclide, 	treatment site, 	and total
source strength and exposure time 	(or, 	equivalently,
the total dose).
"X-ray exposure control" means a device, 	switch, 	button or
other similar means by which an operator initiates and/or
terminates the 	radiation exposure. 	The x-ray exposure control may
include such associated equipment such as 	timers and back-up
timers.
"X-ray equipment" means an x-ray system, 	subsystem, 	or
45-30
265
Complete printed PDF page 30
§11-45-4
thereof. 	Types of x-ray equipment are as follows:	component
(1) 	"Mobile x-ray equipment" means x-ray equipment mounted
on a permanent base with wheels and/or casters 	for
moving while completely assembled.
(2)
(3)
"Portable x-ray equipment" means x-ray equipment
designed to be hand-carried.
"Stationary x-ray equipment" means x-ray equipment
which is installed in a 	fixed location.
"X-ray field" means 	that area of the intersection of the
useful beam and any one of the set of planes parallel to and
including the plane of the image receptor, 	whose perimeter is the
locus of points at which the exposure rate is one-fourth of the
maximum in the intersection.
"X-ray high-voltage generator" means a device which
transforms electrical energy from the potential supplied by the
x-ray control to the tube operating potential. 	The device may
also include means 	for transforming alternating current to direct
current, 	filament transformers 	for the x-ray tube(s), 	high-voltage
switches, 	electrical protective devices, 	and other appropriate
elements.
"X-ray system" means an assemblage of components 	for the
controlled production of x-rays. 	It includes minimally an x-ray
high-voltage generator, 	an x-ray control, 	a 	tube housing assembly,
a beam-limiting device, 	and the necessary supporting structures.
Additional components which function with the system are
considered integral parts of the system.
"X-ray tube" means any electron tube which is designed to be
used primarily for the production of x-rays.
"Year" means the period of time beginning in January used to
determine compliance with the provisions of these rules. 	The
licensee may change the starting date of the year used to
determine compliance provided that the change is made at the
beginning of the year and~~\t, no da~ is omitted or duplicated in
consecutive years. 	[Eff 	NUV 	12 	,pag 	l 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, ~tI-11(21), 	321-71)
§11-45-3 	Exemptions. 	(a) 	Sources of radiation licensed by
the U.S. 	Nuclear Regulatory Commission are exempt from the
requirements of this chapter to the extent they are regulated by
the U.S. 	Nuclear Regulatory Commission.
(b) 	Nothing in this chapter shall be construed to limit
the kind and amount of radiation that may be intentionally applied
to a person for diagnostic or therapeutic purposes by, 	or under
the direction of, 	persons licensed as physicians, 	dentists,
chiropractors, 	podiatrists, 	or veterinarians in the State provided
their sources of radiation are licensed in accordance with
subchapter 2 an<\1<rfflfO.f~WJA~nJ:he applicable provisions of this
chapter. 	[Eff 	1,uv 	l l 	1:::,::,::, 	l 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11 (21), 	321-71)
§11-45-4 	Inspections. 	(a) 	The department shall inspect
and examine facilities with sources of radiation in order to
45-31 	265
Complete printed PDF page 31
§11-45-4
determine compliance with this chapter. 	The frequency of
inspections shall be determined by the director.
(b) 	If such inspection indicates that the facility is not
in compliance with this chapter, 	the owner, 	operator, 	or user
shall be so notified in writing, 	with specification of any
deficiencies.
(c) 	Persons conducting inspections relative to this
chapter shall have been trained in health physics and radiation
protection, 	the performance testing of radiation machines, 	and the
operation of instruments utilized for radiation detection and
measurement, 	performance standards testing of radiation machines,
and quality control measures. 	Persons who have completed training
shall be cer~ified as competent b-':' ~he directo\J ".l\"J;,h~ ~epartment
shall determine the extent of training. 	[Eff f74 lJ i/ 	1 l 	1900 	l
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	t21.-
ll (21), 	321-71)
Historical Note: 	§11-45-4 is based substantially upon §11-40-15.
[Eff 11/5/81]
§11-45-5 	Proceedings before director. 	(a) 	When the
director has grounds to believe that anyone has violated any
provision of these rules or of any order of the director, 	the
director may notify the alleged violator or violators of the
alleged violation order and may require that the alleged violator
appear before the director at a 	time and place specified in the
notice, 	and answer the allegations. 	The notice and order shall
comply with the chapter 91, 	HRS, 	procedures 	for contested cases.
(b) 	The director may issue an order for immediate action
to protect the public health from an imminent and substantial
danger. 	The director shall provide an opportunity for a 	hearing
within twenty-four hours after service of the order. 	After a
hearing pursuant to this subsection, 	the director may affirm,
modify, 	or rescind the order as appropriate. 	The director may
institute a 	civil action in any court of appropriate 	jurisdiction
for the enforcement~?f any order issued pursuant to this
subsection. 	[Eff 	,~ov 	1 2 10QO l 	(Auth: 	HRS §§321-10, 	321-
11, 	321-20, 	321-71) 	(Iinp: ~HR~"§§'321-1, 	321-11(21), 	321-20, 	321-
71)
§11-45-6 	General requirements. 	(a) 	Each licensee shall
maintain records showing the receipt, 	transfer, 	and disposal of
all sources of radiation. 	Additional record requirements are
specified in these rules.
(b) 	Each licensee shall afford the department at all
reasonable times entrance to private or public property for the
purpose of inspecting and investigating conditions 	relative to
sources of radiation, 	physically examining sources of radiation,
the premises, 	and the facilities wherein such sources of radiation
are used or stored.
(c) 	Each licensee shall make available to the department
inspection records maintained pursuant to these rules. 	Copies of
45-32 	265
Complete printed PDF page 32
§11-45-11
such records 	shall be submitted to the department upon request.
(d) 	Each licensee shall perform upon instructions from the
department, 	or shall utilize a qualified medical physicist, 	or
shall permit the department to perform, 	such reasonable tests as
the department deems appropriate or necessary including, 	but not
limited to, 	tests of:
(1) 	Sources of radiation;
(2) 	Facilities wherein sources of radiation are used or
stored;
(3) 	Radiation detection and monitoring instruments; 	and
(4) 	Other equipment and devices used in connection with
utilization or storage of licensed sources of
radiation.
(e) 	Upon receipt of an inspection report rendered by the
department, 	each licensee shall complete corrective action to meet
the requirements of this chapter in the time period specified by
the department. 	If additional time is necessary to complete
corrective action, 	a written request shall be submitted to the
department for 	review and determination before the date specified
for completion of corrective action. 	[Eff 	f\f 	1 ? 1 	g 	](Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp :1 • 	·§§'.!21 	, 	321-
11 (21), 	321-71)
§11-45-7 	Additional requirements. 	The department may
impose upon any licensee or person such requirements in addition
to those established in these rules as it deems appropriate or
necessary to min\,· 	9FIW. er to public health and safety or
property. 	[Eff 	1\1 	J. t, 1900 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS 	§321-1, 	~ii-11 (21), 	321-71)
§11-45-8 	Reserved.
§11-45-9 	Reserved.
§11-45-10 	Reserved.
§11-45-11 	Prohibited uses. 	(a) 	The following sources of
radiation shall not be used:
(1) 	A hand-held fluoroscopic screen using x-ray equipment
unless it has been listed in the Registry of Sealed
Source and Devices or accepted for certification by
the U.S. 	Department of Health and Human Services.
(2) 	A direct view fluoroscopic unit without image
intensification.
(3) 	A shoe-fitting fluoroscopic device.
(4) 	Ionizing radiation from sources of radiation for
demonstration purposes which violate the exposure
limitations and the as-low-as-is-reasonably-achievable
(ALARA) 	principle specified in subchapter 4.
45-33 	265
Complete printed PDF page 33
§11-45-11
(5) 	Except for sources of radiation exempted from a
license, 	sources of radiation which have no license.
[Eff N	10V 1 	1000 	l 	(Auth: 	HRS §§321-10, 	321-11,
321-7 	1""' (Irtlp 	~~§321-1, 	321-11 (21), 	321-71)
§11-45-12 	Reserved.
§11-45-13 	Units of exposure and dose. 	(a) 	As used in
these rules, 	the unit of exposure is the coulomb per kilogram
(C/kg). 	One roentgen is equal to 2.58 x 10- 4 coulomb per kilogram
of air.
(b)
(1)
(2)
(3)
(4)
As used in these rules, 	the units of dose are:
Gray 	(Gy) 	is the SI unit of absorbed dose. 	One gray
is equal to an absorbed dose of one 	joule per kilogram
(one hundred rads).
Rad is the special unit of absorbed dose. 	One rad is
equal to an absorbed dose of one hundred erg per gram
or 0.01 	joule per kilogram 	(0.01 gray).
Rem is the special unit of any of the quantities
expressed as dose equivalent. 	The dose equivalent in
rem is equal to the absorbed dose in rad multiplied by
the quality factor 	(one rem= 0.01 sievert).
Sievert is the SI unit of any of the quantities
expressed as dose equivalent. 	The dose equivalent in
sievert is equal to the absorbed dose in gray
multiplied by the quality factor 	(one sievert = one
hundred rems) .
(c) 	As used in these rules, 	the quality factors 	for
converting absorbed dose to dose equivalent are shown in Appendix
A of this subchapter.
(d) 	If it is more convenient to measure the neutron
fluence 	rate than to determine the neutron dose equivalent rate in
sievert per hour or rem per hour, 	as provided in subsection 	(c),
one rem 	(0.01 sievert) 	of neutron radiation of unknown energies
may, 	for purposes of these rules, 	be assumed to 	result from a
total fluence 	of twenty five million neutrons per square
centimeter incident upon the body. 	If sufficient information
exists to estimate the approximate energy distribution of the
neutrons, 	the licensee may use the fluence 	rate per unit dose
equivalent or the appropriate Q value from Appendix B of this
subchapter to convert a measured tissue dose in gray or rad to
dose equivalent in sievert or rem.
(e) 	The exhibits at the end of chapter 11-45 entitled
"Appendix A of Subchapter 1, 	Quality Factors and Absorbed Dose
Equivalencies 	(2/2/93)" and "Appendix B of Subchapter 1, 	Mean
Quality Factors, 	Q, 	and Fluence Per Unit Dose Equivalent for
Monoeiy~<f-\f ,.l'Jeutrons 	(2/2/93)" are made a part of this section.
[Eff 	NUV 	1 l 	1999 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
45-34 265
Complete printed PDF page 34
§11-45-18
§11-45-14 	Units of radioactivity. 	For purposes of these
rules, 	activity is expressed in the SI unit of becquerel 	(Bq) 	or
in the special unit of curie 	(Ci), 	or their multiples, 	or
disintegrations or transformations per unit of time:
(1) 	One becquerel 	(Bq) 	= one disintegration or
transformation per second 	(s- 1 ).
(2) 	One curie 	(Ci) 	= 3. 7 x 10 10 disintegrations 	or
transformations per second = 3. 7 x 10 10 becquerel 	(Bq)
=_2.22 x 10 12 dtfffi)e~r'tl,ti9-r.s or transformations per
minute. 	[Eff 	1\JUV 	J. l, 1~99 	] 	(Auth: 	HRS §§321-
10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21),
321-71)
§11-45-15 	Penalties and injunctions. 	Any person who
violates any provisions of this chapter is subject to the
penalties pursuant to §321-18, 	HRS, 	the remedies pursuant to §321-
20, 	HES,,.,\jnd HJ.~WW~ions pursuant to §603-23, 	HRS.
[Eff Nut~ 	1 l 	1::1::;~ 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
Historical Note: 	§11-45-15 is based substantially upon §11-40-
18. 	[Eff 11/5/81]
§11-45-16 	Severability. 	If any provisions of this chapter,
or its application to any person or circumstances, 	is held
invalid, 	the application of such provision to other persons or
circumstances, 	and the ri.~,~~~d~r~o~gtbhs chapter, 	shall not be
affected thereby. 	[Eff 	NUV 	l ~ 	I 9:;; 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	121-11(21), 	321-71)
Historical Note: 	§11-45-16 is based substantially upon §11-40-
19. 	[Eff 11/5/81]
SUBCHAPTER 2
License
§11-45-17 	Purpose and scope. 	(a) 	This subchapter
establishes 	requirements 	for the licensing of sources of
radiation, 	the licensing of persons providing radiation services,
and the licensing of persons transporting or delivering
radioactive material to a carrier for transport.
(b) 	In addition to the requirements of this 	subchapter, 	all
licensees are sub~ect to the applicapAAVPrfvAsions of other
subchapters of this chapter. 	[Eff 	NU · 	1 :I. 1909 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-r 	32'1-11 (21), 	321-71)
§11-45-18 	Exemptions. 	(a) 	The following radiation
machines are exempt from the requirements of this 	subchapter:
(1) 	Electronic equipment that produces radiation
45-35 	6
Complete printed PDF page 35
§11-45-18
incidental to its operation for other purposes is
exempt from the license and notification requirements
of this subchapter, 	provided that the dose equivalent
rate averaged over an area of ten square centimeters
does not exceed 0.5 millirem 	(five microsieverts) 	per
hour at five centimeters from any accessible surface
of such equipment;
(2) 	Radiation machines while in transit or storage
incident thereto; 	and
(3) 	Domestic television receivers;
(b) 	The following source material is exempt from the
requirements of this subchapter:
(1) 	Any chemical mixture, 	compound, 	solution, 	or alloy in
which the source material is by weight less than one-
twentieth of one percent 	(0.05 percent) 	of the
mixture, 	compound, 	solution, 	or alloy;
(2) 	Unrefined and unprocessed ore containing source
material;
(3) 	Any quantities of thorium contained in:
(A) 	Incandescent gas mantles;
(B) 	Vacuum tubes;
(C) 	Welding rods;
(D) 	Electric lamps 	for illuminating purposes
provided that each lamp does not contain more
than fifty milligrams of thorium;
(E) 	Germicidal lamps, 	sunlamps, 	and lamps 	for
outdoor or industrial lighting provided that
each lamp does not contain more than two grams
of thorium;
(F) 	Rare earth metals and compounds, 	mixtures, 	and
products containing not more than 0.25 percent
by weight thorium, 	uranium, 	or any combination
of these; 	and
(G) 	Personnel neutron dosimeters, 	provided that each
dosimeter does not contain more than fifty
milligrams of thorium;
(4) 	Source material contained in the following products:
(A) 	Glazed ceramic tableware, 	provided that the
glaze contains not more than twenty percent by
weight source material;
(B) 	Glassware containing not more than ten percent
by weight source material, 	but not including
commercially manufactured glass brick, 	pane
glass, 	ceramic tile, 	or other glass or ceramic
used in construction;
(C) 	Glass enamel or glass enamel 	frit containing not
more than ten percent by weight source material
imported or ordered for importation into the
United States, 	or initially distributed by
manufacturers in the United States, 	before July
25, 	1983; 	and
(D) 	Piezoelectric ceramic containing not more than
two percent by weight source material;
(5) 	Photographic film, 	negatives, 	and prints containing
45-36 	65
Complete printed PDF page 36
§11-45-18
uranium or thorium;
(6) 	Any finished product or part 	fabricated of, 	or
containing, 	tungsten-thorium or magnesium-thorium
alloys, 	provided that the thorium content of the alloy
does not exceed four percent by weight and that this
exemption shall not be deemed to authorize the
chemical, 	physical, 	or metallurgical treatment or
processing of any such product or part;
(7) 	Uranium contained in counterweights installed in
aircraft, 	rockets, 	projectiles, 	and missiles, 	or
stored or handled in connection with installation or
removal of such counterweights, 	provided that:
( 8)
(9)
(10)
(A) 	The counterweights are manufactured in
(B)
(C)
(D)
accordance with a 	specific license issued by the
U.S. 	Nuclear Regulatory Commission, 	authorizing
distribution by the licensee pursuant to 10
C.F.R. 	Part 40;
Each counterweight has been impressed with the
following legend clearly legible through any
plating or other covering: 	"DEPLETED URANIUM";
Each counterweight is durably and legibly
labeled or marked with the identification of the
manufacturer and the statement: 	"UNAUTHORIZED
ALTERATIONS PROHIBITED"; 	and
This exemption shall not be deemed to authorize
the chemical, 	physical, 	or metallurgical
treatment or processing of any such
counterweights other than repair or restoration
of any plating or other covering;
Natural or depleted uranium metal used as shielding
constituting part of any shipping container, 	provided
that:
(A)
(B)
The shipping container is conspicuously and
legibly impressed with the legend "CAUTION -
RADIOACTIVE SHIELDING - 	URANIUM"; 	and
The uranium metal is encased in mild steel or
equally fire 	resistant metal of minimum wall
thickness of one eighth inch 	(3.2 millimeters);
Thorium contained in finished optical lenses, 	provided
that each lens does not contain more than thirty
percent by weight of thorium, 	and that this exemption
shall not be deemed to authorize either:
(A) 	The shaping, 	grinding, 	or polishing of such lens
or manufacturing processes other than the
assembly of such lens into optical systems and
(B)
devices without any alteration of the lens; 	or
The receipt, 	possession, 	use, 	or transfer of
thorium contained in contact lenses, 	or in
spectacles, 	or in eyepieces in binoculars or
other optical instruments;
Uranium contained in detector heads 	for use in fire
detection units, 	provided that each detector head
contains not more than 0.005 microcuries of uranium;
and
45-37
2 6 5
Complete printed PDF page 37
§11-45-18
(11)
(12)
(c)
material is
(1)
(2)
(3)
Thorium contained in any finished aircraft engine part
containing nickel-thoria alloy, 	provided that:
(A) 	The thorium is dispersed in the nickel-thoria
alloy in the form of finely divided thoria
(thorium dioxide), 	and
(B) 	The thorium content in the nickel-thoria alloy
does not exceed four percent by weight.
Source material licensed by the U.S. 	Nuclear
Regulatory Commission.
The following radioactive material other than source
exempt from the requirements of this subchapter:
Products containing radioactive material introduced in
concentrations not in excess of those listed in
Appendix A of this subchapter;
Radioactive material in individual quantities each of
which does not exceed the applicable quantity set
forth in Appendix B of this subchapter;
Timepieces or hands or dials containing not more than
the following specified quantities of radioactive
material and not exceeding the following specified
radiation dose rate:
(A) 	Twenty-five millicuries 	(nine hundred twenty-
five megabecquerels) 	of tritium per timepiece;
(B) 	Five millicuries 	(one hundred eighty-five
megabecquerels) 	of tritium per hand;
(C) 	Fifteen millicuries 	(five hundred fifty-five
megabecquerels) 	of tritium per dial 	(bezels when
used shall be considered as part of the dial);
(D) 	one hundred microcuries 	(3.7 megabecquerels) 	of
promethium-147 per watch or two hundred
microcuries 	(7.4 megabecquerels) 	of promethium-
147 per any other timepiece;
(E) 	Twenty microcuries 	(0.74 megabecquerels) 	of
promethium-147 per watch hand or forty
microcuries 	(1.48 megabecquerels) 	of promethium-
147 per other timepiece hand;
(F) 	Sixty microcuries 	(2.22 megabecquerels) 	of
promethium-147 per watch dial or one hundred
twenty microcuries 	(4.44 megabecquerels) 	of
promethium-147 per other timepiece dial 	(bezels
when used shall be considered as part of the
dial);
(G) 	The radiation dose rate from hands and dials
containing promethium-147 shall not exceed, 	when
measured through fifty milligrams per square
centimeter of absorber;
(H) 	For wrist watches, 	0.1 millirad 	(one microgray)
per hour at ten centimeters from any surface;
(I) 	For pocket watches, 	0.1 millirad 	(one microgray)
per hour at one centimeter from any surface;
(J) 	For any other timepiece, 	0.2 millirad 	(two
microgray) 	per hour at ten centimeters from any
surface; 	and
(K) 	One microcurie 	(thirty-seven kilobecquerels) 	of
45-38 26
Complete printed PDF page 38
§11-45-18
radium-226 per timepiece in timepieces acquired
before the effective date of this chapter;
(4) 	Lock illuminators containing not more than fifteen
millicuries 	(five hundred fifty-five megabecquerels)
of tritium or not more than two millicuries 	(seventy-
four megabecquerels) 	of promethium-147 installed in
automobile locks. 	The radiation dose rate from each
lock illuminator containing promethium-147 shall not
exceed one millirad 	(ten microgray) 	per hour at one
centimeter from any surface when measured through
fifty milligrams per square centimeter of absorber;
(5) 	Precision balances containing not more than one
millicurie 	(thirty-seven megabecquerels) 	of tritium
per balance or not more than 0.5 millicurie 	(18.5
megabecquerels) 	of tritium per balance part;
(6) 	Automobile shift quadrants containing not more than
twenty-five millicuries 	(nine hundred twenty-five
megabecquerels) 	of tritium;
(7) 	Marine compasses containing not more than seven
hundred fifty millicuries 	(27.8 gigabecquerels) 	of
tritium gas and other marine navigational instruments
containing not more than two hundred fifty millicuries
(9.25 gigabecquerels) 	of tritium gas;
(8) 	Thermostat dials and pointers containing not more than
twenty-five millicuries 	(nine hundred twenty-five
megabecquerels) 	of tritium per thermostat;
(9) 	Electron tubes; 	provided, 	that each tube does not
contain more than one of the following specified
quantities of radioactive material:
(A) 	One hundred fifty millicuries 	(5.55
gigabecquerels) 	of tritium per microwave
receiver protector tube or ten millicuries
(three hundred seventy megabecquerels) 	of
tritium per any other electron tube;
(B) 	One microcurie 	(thirty-seven kilobecquerels) 	of
cobalt-60;
(C) 	Five microcuries 	(one hundred eighty-five
kilobecquerels) 	of nickel-63;
(D) 	Thirty microcuries 	(1.11 megabecquerels) 	of
krypton-85;
(E) 	Five microcuries 	(one hundred eighty-five
kilobecquerels) 	of cesium-137; 	and
(F) 	Thirty microcuries 	(1.11 megabecquerels) 	of
promethium-147;
And provided further, 	that the 	radiation dose
rate from each electron tube containing
radioactive material shall not exceed
one millirad 	(ten microgray) 	per hour at one
centimeter from any surface when measured
through seven milligrams per square centimeter
of absorber;
(10) 	Ionizing radiation measuring instruments containing,
for purposes of internal calibration or
standardization, 	one or more sources of radioactive
45-39
Complete printed PDF page 39
§11-45-18
material, 	provided that:
(A) 	Each source contains no more than one exempt
quantity set forth in Appendix B of this
subchapter; 	and
(B) 	Each instrument contains no more than ten exempt
quantities. 	For purposes of this requirement,
an instrument's source(s) 	may contain either one
or different types of radionuclides and an
individual exempt quantity may be composed of
fractional parts of one or more of the exempt
quantities in Appendix B of this subchapter,
provided that the sum of such fractions 	shall
not exceed unity; 	and
(C) 	0.05 microcuries 	(1.85 kilobecquerels) 	of
americium-241;
(11) 	Spark gap irradiators containing not more than one
microcuries 	(thirty-seven kilobecquerels) 	of cobalt-60
per spark gap irradiator for use in electrically
ignited fuel 	oil burners having a 	firing 	rate of at
least three gallons 	(11.4 liters) 	per hour;
(12) 	Self-luminous products containing radioactive
material:
(A) 	Tritium, 	krypton-85 or promethium-147 in self-
luminous products manufactured, 	processed,
produced, 	imported, 	or transferred in accordance
with a 	specific license issued by the U.S.
Nuclear Regulatory Commission pursuant to 10
C.F.R. 	§32.22, 	which authorizes the transfer of
the product to persons who are exempt from
regulatory requirements; 	and
(B) 	Articles containing less than 0.1 microcuries
(3.7 	kilobecquerels) 	of radium-226 which were
acquired before the effective date of this
chapter;
(13) 	Radioactive material in gas and aerosol detectors
designed to protect life or property from fires 	and
airborne hazards provided that detectors containing
radioactive material shall have been manufactured,
imported, 	or transferred in accordance with a 	specific
license issued by the U.S. 	Nuclear Regulatory
Commission pursuant to 10 C.F.R. 	§32.26, 	which
authorizes the transfer of the detectors to persons
who are exempt from regulatory requirements;
(14) 	Synthetic plastic resins containing scandium-46 which
are designed for sand consolidation.
(15) 	Radioactive material other than source material
licensed by the U.S. 	Nuclear Regulatory Commission.
(d) 	The following persons are exempt from the requirements
of this 	subchapter:
(1) 	Common and contract carriers, 	freight 	forwarders, 	and
warehousemen which are subject to the requirements of
the U.S. 	Department of Transportation in 49 C.F.R.
Parts 170 through 189 or the U.S. 	Postal Service in
the Postal Service Manual 	(Domestic Mail Manual),
45-40 	6
Complete printed PDF page 40
(2)
{e)
"Appendix
"Appendix
a part 	of
§§321-10,
§11-45-22
Section 124.3 incorporated by reference, 	39 C.F.R.
§111.11 	(1974), 	and the U.S. 	Postal Service are exempt
from the requirements of this subchapter to the extent
that they transport or store radioactive material in
the regular course of their carriage for others or
storage incident thereto.
Any person who delivers to a 	carrier for transport a
package containing radioactive material having a
specific activity not greater than 0.002 microcuries
per gram {seventy-four becquerels/gram), 	or a package
containing no more than a Type A quantity of
radioactive material which contains no 	fissile
radioactive material.
The exhibits at the end of chapter 11-45 entitled
A of Subchapter 2, 	Exempt Concentrations 	(2/2/93)" 	and
B of Subchapter 2, 	EXtifffi/2\J:F-~· 	• 9 (2/2/93) ", 	are made
this section. 	[Eff 	1,UV 	1 '1 	] 	{Auth: 	HRS
321-11, 	321-71) 	{Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-19 	Reserved.
§11-45-20 	Reserved.
§11-45-21 	Reserved.
§11-45-22 	Application for license. 	(a) 	All persons
purchasing or acquiring any radiation machine, 	which is not exempt
pursuant to section 11-45-18 after the effective date of these
rules, 	shall apply for a 	license with the department at least
sixty days before purchasing, 	acquiring, 	or operating any
radiation machine.
(b) 	All persons purchasing or acquiring any radioactive
material, 	which is not exempt pursuant to section 11-45-18 after
the effective date of these 	rules, 	shall apply for a license with
the department at least sixty days before purchasing or acquiring,
storing, 	manufacturing, 	using, 	or handling any radioactive
material.
(c) 	Any person who is engaged in the business of selling,
leasing, 	installing, 	or offering to install radiation machines,
engaged in the business of furnishing or offering to furnish
radiation machine servicing or services, 	engaged in the business
of furnishing radiation services for 	radioactive materials,
engaged in the business of providing health physics services, 	or
engaged in the business of providing medical physics services in
this State shall apply for a 	license of such services with the
department at least sixty days before furnishing or offering to
furnish any such services.
(d) 	Any person, 	other than those who are exempt under
section 11-45-18, 	who is engaged in transporting or delivering
radioactive material to a 	carrier for transport shall apply for a
45-41 65
Complete printed PDF page 41
§11-45-22
license with the department at least sixty days before furnishing
such services.
(e) 	Application for a license shall be completed on forms
furnished by the department and shall contain all the information
required by the form and accompanying instructions. 	Information
includes:
(1)
(2)
(3)
(4)
(5)
(6)
(7)
( 8)
(9)
Applicant identification;
Type of facility or service;
Purpose of source of radiation use;
Location of radiation facility or service;
Listing of each radiation machine according to the
manufacturer's name, 	model number, 	serial number, 	and
date of manufacture of the control assembly;
Listing of radioactive materials and maximum activity;
Current curriculum vitae and copies of board
certification for applicants of health physics
services or medical physics services.
Identification of the radiation machine installer; 	and
Such other information deemed as 	the director may
request.
(f) 	The director shall not act upon or consider any
incomplete application for a 	license. 	An application shall be
deemed complete only when all 	required and requested information,
including the application form, 	have been submitted.
(g) 	Every application shall be signed by the applicant and
shall 	constitute an acknowledgment and agreement that the
applicant shall comply with all the conditions of the license and
this chapter.
(h) 	The director may require the submission of additional
information after the application has been submitted, 	and may
ensure that, 	if an application is incomplete or otherwise
deficient, 	processing of the application shall not be completed
until 	such time as the applicant has submitted all required
information or otherwise corrected the deficiency.
(i) 	The failure of the director to act on a completed
application within sixty days of the receipt of such application,
shall be deemed a 	grant of such application; 	provided that the
applicant acts consistently with the application process.
(j) 	The application for a 	license shall be accompanied by
a non-refundable, 	initial license fee per facility payable to the
department. 	The initial license fee 	schedule according to type of
facility or services is as 	follows:
(1) 	Medical x-ray facility with eight or more x-ray units
- 	$150.
(2) 	Medical x-ray facility with more than four but less
than eight x-ray units - 	$100.
(3) 	Medical x-ray facility with four or less x-ray units -
$50.
(4) 	Chiropractic x-ray facility - 	$50.
(5) 	Dental x-ray facility with more than four x-ray units
- 	$50.
(6) 	Dental x-ray facility with four or less x-ray units -
$30.
(7) 	Podiatry x-ray facility - 	$30.
45-42 265
Complete printed PDF page 42
C
§11-45-24
(8) 	Veterinary x-ray facility - 	$30.
(9) 	Radiation therapy facility - 	$100.
(10) 	Medical radionuclide 	facility - 	$100.
(11) 	Industrial 	radiography - 	$50
(12) 	All other radiation facilities - 	$30.
(13) 	Radiation services - 	$30 .
(k) 	The license fee 	for an entity with more than one type
facility or services 	shall be the fee 	for the \Y9o·nP~ ti3-~Aw or
services with the highest dollar value. 	[Eff 	i\J '. V j_ l 	l~!j~
(Auth: 	HRS §§321-10, 	321-11, 	321-11.5, 	321-27, 	321-71) 	(Imp:
HRS §§321-1, 	321-11 (21), 	321-11. 5, 	321-27, 	321-71)
§11-45-23 	Issuance of license. 	(a) 	Upon determination
that an applicant meets the requirements of the rules, 	the
department shall issue a license.
(b) 	All persons with sources of radiation registered with
the department before the effective date of these 	rules shall be
issued a 	license after payment of the appropriate license fee
unless 	the use and/or the source of radiation is prohibited by
section 11-45-11. 	The department shall determine the duration and
expiration date of the initial license to establish the beginning
of the biennial period.
(c) 	All persons approved by the department as qualified
experts under chapter 11-40 before the effective date of these
rules 	shall be issued a 	license after payment of the appropriate
license 	fee. 	The department shall determine the duration and
expiration date of the initial license to establish the beginning
of the biennial period.
(d) 	The department may incorporate in the license at the
time of issuance or thereafter by appropriate notification, 	rule,
or order, 	such additional requirements and conditions as it deems
appropriate or necessary.
(e) 	No person shall conduct activities with sources of
radiation or provide radiation services specified in section 11-
45-22 without a 	license.
(f) 	Each license shall expire on the date specified by the
department biennially.
(g) 	The original license shall be posted in a 	location
which ti,-li"nyp.i' J:ri.~Mni ndividuals utilizing the facility.
[Eff 	!'JUV 	iG 	;~~:., 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-24 Renewal of li~ense. 	(a) 	All licensees 	shall
renew their license biennially before the expiration date by
submitting a 	renewal application on a 	form furnished by the
department and a 	renewal license fee 	according to the initial fee
schedule in section 11-45-22.
(b) 	Notices to renew licenses 	shall be mailed biennially
to licensees at the addresses 	recorded by the department. 	Failure
to 	receive the notice shall not be a valid reason for not renewing
licenses.
(c) 	A renewal application shall contain all 	the
45-43 2265
Complete printed PDF page 43
C
L
§11-45-24
information required by the form and accompanying instructions.
This includes:
(1) 	Licensee identification;
(2) 	Change in name, 	type, 	purpose, 	or location of facility
or service;
(3) 	Current listing of each radiation machine according to
the manufacturer's name, 	model number, 	serial number,
and date of manufacture of the control assembly;
(4) 	Current listing of radioactive materials and maximum
activity;
(5) 	Current curriculum vitae and copies of new board
certification for health physicists and medical
physicists; 	and
(6) 	Such other information deemed as 	the director may
request.
(d) 	The director shall not act upon or consider any
incomplete renewal application for a 	license. 	An application
shall be deemed complete only when all 	required and requested
information, 	including application form, 	have been submitted.
(e) 	Every renewal application shall be signed by the
licensee and shall constitute an acknowledgment and agreement that
the licensee shall comply with all the conditions of the license
and this chapter.
(f) 	The director may require the submission of additional
information after the renewal application has been submitted, 	and
may ensure that, 	if an application is incomplete or otherwise
deficient, 	processing of the application shall not be completed
until 	such time as the licensee has submitted all 	required
information or otherwise corrected the deficiency.
(g) 	The 	failure 	of the director to act on a completed
renewal application within sixty days of the receipt of such
application, 	shall be deemed a 	grant of such application; 	provided
that the licensee acts consistently wi t h the application process.
(h) 	Renewal 	fees 	received by mail 	shall be considered as
paid when due if the envelope bears the postmark of the expiration
date or earlier.
(i) 	Any license which is not renewed shall be declared
defunct.
(j) 	The licensee shall meet all requirements pursuant to
this chapter before renewal of a 	license. 	At the discretion of
the director, 	renewal of license may be allow ~q'4-f§ co w ective
actions are acceptable to the director. 	[Eff l~Uv 	1l 	1aa9
(Auth: 	HRS §§321-10, 	321-11, 	321-11.5, 	321-27, 	321-71) 	1~mp:
HRS §§321-1, 	321-11(21), 	321-11.5, 	321-27, 	321-71)
§11-45-25 	Report of changes. 	The licensee shall notify the
department in writing of any change which would render the
infor ~ N\C)Il f'tfl~nd in the license no longer accurate.
[Eff 	!' UV 	~ 	I~~~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
45-44
2 265
Complete printed PDF page 44
(
L
§11-45-26 	Approval not implied. 	No person, 	in any
advertisement, 	shall 	refer to the 	fact that a 	facility with
sources of radiation is licensed, 	and no person shall state or
imply that any activity under such license has been approved by
the department. 	[Eff 	t+.!n / 	i 9 I 9a 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 1 ~ 	§§3 '2'1- , ~ 21-11 (21), 	321-71).
§11-45-27
§11-45-27 	Assembly and transfer. 	(a ) 	No 	radiation
machines shall be assembled or installed in a 	facilit y which does
not ha v e a 	current license.
(b) 	No source of radiation shall be transferred to a
facility which does not have a current license.
(c) 	No person shall manufacture, 	sell, 	lease, 	transfer,
lend, 	assemble, 	or install one or more components of radiation
machines or the supplies used in connection with such machines
without having a 	current license. 	Radiation machines and supplies
when properly placed in operation and used shall meet the
requirements of this chapter.
(d) 	Any person who sells, 	leases, 	transfers, 	lends,
disposes, 	assembles, 	or installs one or more components of
radiation machines in this State shall notify the department
within fifteen days of:
(1) 	The name and address 	of persons who have 	received
(2)
(3)
(e)
specified
(1)
(2)
these machines;,
The manufacturer, 	model, 	and serial number of the
control assembly of each radiation machine
transferred; 	and
The date of transfer of one or more 	radiation machine
components.
The following items are e x empt from the 	requirements
in subsection 	(d):
Reloaded or replacement tube housing a ssemblies that
are reinstalled in or newly assembled into an existing
x -ray system;
Certified accessory components that have been
identified as such to the U.S. 	Department of Health
and Human Services;
(3) 	Repaired components, 	whether or not removed from the
system and reinstalled during the course of repair,
provided the original installation into the x-ray
system was 	reported; 	or
(4) 	Components installed temporarily in an x -ray system in
place of components removed temporarily for 	repair,
provided the temporarily installed component is
identified by a 	tag or label bearing the 	following
information:
"Temporarily Installed Component
This certified component has been assembled,
installed, 	adjusted, 	and tested by me according
to the instructions provided by the
manufacturer.
45-45
2 265
Complete printed PDF page 45
(
§11-45-27
Signature
Company Name
Street Address, 	P.O. 	Box
City, 	State, 	Zip Code
Date of Installation"
The replacement of the temporarily installed component
by a component other than the component originally
removed for 	repair shall be reported as 	specified by
subsection 	(d) .
(f) 	In the case of diagnostic x-ray systems which contain
one or more certified components, 	a copy of the assembler's 	report
prepared in compliance with requirements of the 	federal 	diagnostic
x-ray standard, 	21 C.F.R. 	§1020.30(d), 	shall be submitted to the
department within fifteen days 	following completion of the
assembly. 	Such report ~q~/ S,¥Wi.fng tr lieu of any other report
by the assembler. 	[Eff NU\I 	1 t{, 	I~ 	~ 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-28 	Reserved.
§11-45-29 	Reserved.
§11-45-30 	Out-of-state source of radiation. 	(a) 	Before
any source of radiation currently licensed by another state or the
U.S. 	Nuclear Regulatory Commission is brought into the State, 	for
any temporary use, 	the person proposing to bring such source into
the State shall give written notice to the department at least
seven working days before such source is brought into the State.
The notice shall include:
(1) 	The type of source of radiation;
(2) 	The nature, 	duration, 	and scope of use;
(3) 	The exact location(s) 	where the source of radiation is
to be used, 	stored, 	and secured; 	and
(4) 	Copy of a 	license issued by the U.S. 	Nuclear
Regulatory Commission, 	or copy of a 	license issued by
other states 	for the source of radiation.
(b) 	The person referred to in subsection 	(a) 	shall:
(1) 	Comply with all applicable rules; 	and
(2) 	Supply the department with such other information as
the d N artment may reasonably request.
[Eff 	ov 	12 1009 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-7 	(Imp: 	~~ §§321-1, 	321-11(21), 	321-71)
§11-45-31 	Prohibitions. 	No 	radiation facility licensee
shall allow any person in the business of providing radiation
machine installation or radiation services to furnish such
services to their sources of radiation or to a 	facility that does
not have a valid license issued by the department. 	This does not
45-46 2 265
Complete printed PDF page 46
(
§11-45-33
include persons who provide radiation machine
radia \{ oa ~'I'R- ces ~ or their own facility.
[Eff 	N JV 	'1 199~ 	] 	(Auth: 	HRS §§321 - 10,
(Imp: 	HRS 	321-1, 	321-11(21), 	321-71)
installation or
321-11, 	321-71)
SUBCHAPTER 3
SHIELDING EVALUATION AND AREA RADIATION SURVEY
§11-45-32 	Purpose and scope. 	This subchapter establishes
requirements 	for 	shielding of facilities with sources of
radiation. 	This 	subchapter applies to all licensees and persons
who i ntfA~ , t ~ 8 b ~AH'h sources of radiation.
[Eff 	NUV 	l (;, 	I~~~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-33 	Shielding evaluation. 	(a) 	Before construction,
the construction plans of all new facilities, 	or modifications of
existing facilities, 	utilizing sources of radiation shall be
submitted to the department 	for review and approval. 	These plans
shall show, 	as a minimum, 	the 	following:
(1) 	Normal location of the source of radiation;
(2) 	General direction(s) 	of the useful beam for x-ray
(3)
(4)
(5)
devices;
Locations of any windows and doors;
If the source of radiation is an x -ray device, 	the
location of the operator's booth or position and the
location of the control assembly;
Structural composition, 	dimensions, 	and thickness of
all walls, 	doors, 	partitions, 	floor, 	and ceiling of
room ( s) ;
(6) 	Dimensions and thickness of lead 	(in inches,
millimeters, 	or nominal weight in pounds 	for a one
square foot 	section) 	specified for walls, 	doors,
partitions, 	floors, 	or ceiling of room(s);
(7) 	Dimensions of the room(s);
(8) 	Type of occupancy of all adjacent rooms and areas
inclusive of space above and below the room(s) 	with
sources of radiation; 	and
(9) 	If there is an exterior wall, 	the distance to the
closest area(s) 	where it is likely that individuals
may be present.
(b) 	The licensee or person intending to establish a
facility or operation with sources of radiation shall utilize the
services of a 	qualified medical physicist to determine the
shielding requirements of the new facility or modification of an
existing facility. 	A written shielding evaluation report by the
qualified medical physicist shall be submitted with the
architectural and electrical plans to the department as 	specified
in subsection 	(a). 	The evaluation report shall 	include, 	but not
limited to:
(1) 	The anticipated workload or e x posure from the source
45-47 2265
Complete printed PDF page 47
C
§11-45-33
of radiation; 	and
(2) 	All basic assumptions used in the development of the
shielding specifications.
(3) 	All other relevant shielding specifications.
(c) 	The shielding specifications on the construction plan
shall be equal or greater than the shielding specifications
identified on the qualified medical physicist's 	shielding
evaluation report .
(d) 	T he approval of radiation shielding by the department
shall not preclude the requirement of additiona l 	modifications
should a 	subsequent analysis of operat i ng conditions indicate a
change in spatial relationship in the radiation facility or
conditions which may be contrary to the requirements in subchapter
4.
(e) 	A shielding evaluation before construction is optional
for the 	following:
( 1) 	Dental 	facilities;
(2) 	Podiatric 	facilities; 	or
( 3) 	Facili tiin sv di t n r !f.¾ned by the department.
[Eff 	1'1U 	l (,, 	1~9Q 	J 	(Auth: 	HRS §§321-10, 	321 - 11,
321-71) 	(Imp: 	HRS 1 §321-1, 	321-11 (21), 	321-71)
§11-45-34 	Area 	radiation survey. 	(a) 	An area radiation
surve y of a new facility or modification to an existing facility
shall be conducted by, 	or under the direction of, 	a 	qualified
medical physicist within si x months after the beginning of its
operation to determine compliance with subchapter 4.
(b) 	The licensee shall 	forward a 	report of the area
radiation sur v ey to the department within thirty days 	after the
completion of the survey.
(c) 	A subsequent area 	radiation survey of the facility
shall be conducted if the following occur:
(1) 	Changes in the number of sources of radiation; 	or
(2) 	Changes in the spatial relationship of sources of
(3)
(4)
radiation, 	relative to walls, 	partitions, 	doors,
floors, 	ceilings; 	or
Changes in workload or utilization of the source of
radiation; 	or
Condi ~'1'1ti' , w~io h
1
fff~ be contrary to subchapter 4 .
[Eff 	NUV 	l (,, 	::,::,::, 	J 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11 (21), 	321-71)
§11 - 45 - 35 	Reserved.
SUBCHAPTER 4
STANDARDS FOR PROTECTION AGAINST RADIATION
§11 - 45-36 	Purpose and scope. 	This 	subchapter establishes
standards 	for protection against ionizing radiation resulting from
activities conducted pursuant to licenses issued by the
45 - 48 2265
Complete printed PDF page 48
C
§11-45-40
department. 	These standards are designed to control the receipt,
possession, 	use, 	transfer, 	and disposal of sources of radiation by
any licensee so the total dose to an individual, 	including doses
resulting from all sources of radiation other than background
radiation, 	does not exceed the standards 	for protection against
radiation prescribed in this 	subchapter. 	However, 	nothing in this
subchapter shall be construed as limiting actions that may be
necessary to protect health and safety. 	The limits in this
subchapter do not apply to doses due to background radiation, 	to
exposure of patients to 	radiation for the purpose of medical
diagnosis or therapy, 	or t i 1 ~ oluntary 1 articipation in medical
research programs. 	[Eff 	NUV 	1 ? 1 go 	l 	(Auth: 	HRS §§321-
10, 	321-11, 	321-71) 	(Imp: 	HRS md'21 - , "'3 21-11 (21), 	321-71)
§11-45-37 	Exemptions. 	Sources of radiation licensed by the
U.S. 	Nuclear Regulatory Commission are exempt from the
requirements of this subchapter to the extent they are 	regulated
by th 	~) 	. r UJ: ~rf- egulatory Commission .
[Eff 	U 	' 	l;:J;:J~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	3 21-71)
(Imp: 	HRS §§3 2 1-1, 	321-11(21), 	321-71)
§11-45-38 	Implementation. 	(a) 	Any existing license
condition that is more 	restrictive than subchapter 4 remains in
force until 	there is an amendment or renewal of the license.
(b) 	If a 	license condition exempts a 	licensee from a
provision of subchapter 4 in effect on or before effective date of
these rules, 	it also exempts the lic i~1'~~ ft'E1 t,li; e corresponding
provision of this subchapter. 	[Eff 	NUV 	l 	1!::199 	] 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS§ 321-1, 	321-11(21),
321-71)
§11-45-39 	Radiation protection programs. 	(a) 	Each
licensee shall develop, 	document, 	and implement a 	radiation
protection program sufficient to ensure compliance with the
provisions 	of this subchapter and acceptable to the department.
(b) 	The licensee shall use, 	to the extent practicable,
procedures and engineering controls based upon sound radiation
protection principles to achie v e occupational doses and doses 	to
members of the public that are as low as 	is 	reasonably achievable
(ALARA).
(c) 	The department shall specify the minimum requirements
for the 	radiation protection program according to sources of
radiation and type of radiation activity and facility.
(d) 	The licensee shall, 	at least annually, 	review the
radiation 'l rotection program content and implementation.
[Eff 	Nn 	1 ? 1090 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	~ 	§§3 ~ 1- Y, ~ 21-11( 2 1), 	32 1-71)
§11-45-40 	Occupational dose limits 	for adults. 	(a) 	The
licensee shall control the occupational dose to indi v idual adults,
45-49 2 265
Complete printed PDF page 49
§11-45-40
except for planned special exposures, 	to the 	following dose
limits:
(1)
(2)
An annual limit, 	which is the more restrictive of:
(A) 	The total effective dose equivalent being equal
(B)
to five rems 	(0.05 sievert); 	or
The sum of the deep dose equivalent and the
committed dose equivalent to any individual
organ or tissue other than the lens of the eye
being equal to fifty rems 	(0.50 sievert).
The annual limits to the lens of the eye, 	to the skin,
and to the extremities which are:
(A) 	An eye dose equivalent of fifteen rems 	(0.15
(B)
sievert), 	and
A shallow dose equivalent of fifty rems 	(0.50
sievert) 	to the skin or to any extremity.
(b) 	Doses received in excess of the annual limits,
including doses 	received during accidents, 	emergencies, 	and
planned special exposures, 	shall be subtracted from the limits 	for
planned special exposures that the individual may receive during
the current year and during the individual's lifetime.
(c) 	The assigned deep dose equivalent and shallow dose
equivalent shall be for the portion of the body receiving the
highest exposure determined as follows:.
(1) 	The deep dose equivalent, 	eye dose equivalent and
shallow dose equivalent may be assessed from surveys
or other radiation measurements 	for the purpose of
demonstrating compliance with the occupational dose
limits, 	if the individual monitoring device was not in
the region of highest potential exposure, 	or the
results of individual monitoring are unavailable.
(2) 	Reserved.
(d) 	Derived air concentration 	(DAC) 	and annual limit on
intake 	(ALI) 	values are presented in Table I 	of Appendix Band may
be used to determine the individual's dose and to demonstrate
compliance with the occupational dose limits.
(e) 	Notwithstanding the annual dose limits, 	the licensee
shall limit the soluble uranium intake by an individual to ten
milligrams in a week in consideration of chemical toxicity. 	See
footnote 3 of Appendix B.
(f) 	The licensee shall reduce the dose that an individual
may be allowed to receive in the current year by the amount of
occupational dose received while employed by any other person.
(g) 	Appendix B to Part D of the Conference of Radiation
Control Program Directors, 	Inc. 	Suggested State Regulations 	for
Control of Radiation, 	Volume I, 	December 1995 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations for Release to Sanitary Sewerage",
is made a part of ~P,1,$,S~c~i~t\A-J,d is available from the
department. 	[Eff jiJLJ\! 	12 	1:;:;,~ 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
45-50
Complete printed PDF page 50
(
§11-45-42
§11-45-41 	Compliance with requirements 	for summation of
external and internal doses. 	(a) 	If the licensee is 	required to
monitor pursuant to section 11-45-54(a) 	and 	(b), 	the licensee
shall demonstrate compliance with the dose limits by summing
external and internal doses. 	If the licensee is required to
monitor only pursuant to section 11-45-54(a) 	or only pursuant to
section 11-45-54(b), 	then summation is not 	required to demonstrate
compliance with the dose limits. 	The licensee may demonstrate
compliance with the 	requirements 	for summation of external and
internal doses pursuant to subsections 	(b), 	(c), 	and 	(d). 	The
dose equivalents 	for the lens of the eye, 	the skin, 	and the
extremities are not included in the summation, 	but are subject to
separate limits.
(b) 	If the only intake of radionuclides is by inhalation,
the total effective dose equivalent limit is not e x ceeded if the
sum of the deep dose equivalent divided by the total effective
dose equivalent limit, 	and one of the following, 	does not exceed
unity:
(1) 	The sum of the fractions 	of the inhalation ALI 	for
each radionuclide, 	or
(2) 	The total number of derived air concentration-hours
(DAC-hours) 	for all 	radionuclides divided by two
. thousand, 	or
(3) 	The sum of the calculated committed effective dose
equivalents to all 	significantly irradiated organs or
tissues 	(T) 	calculated from bioassay data using
appropriate biological models and expressed as a
fraction of the annual limit. 	For purposes of this
requirement, 	an organ or tissue is deemed to be
significantly irradiated if, 	for that organ or tissue,
the product of the weighting 	factors, 	wT, 	and the
committed dose equivalent, 	HT ,so , 	per unit intake is
greater than ten percent of the maximum weighted value
of H50 , 	that is, 	wTH T, so , 	per unit intake 	for any organ
or _tissue.
(c) 	If the occupationally exposed individual also 	receives
an intake of radionuclides by oral ingestion greater than ten
percent of the applicable oral ALI, 	the licensee shall account 	for
this intake and include it in demonstrating compliance with the
limits.
(d) 	The licensee shall evaluate and, 	to the extent
practi ~1 bvaccount 	for intakes through wounds or skin absorption.
[Eff 	N 	12 1990 ] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321- ~, 	:Y21-11(21), 	321-71)
§11-45-42 	Determination of external dose from airborne
radioactive material. 	(a) 	Licensees 	shall, 	when determining the
dose 	from airborne radioactive material, 	include the contribution
to the deep dose equivalent, 	eye dose equivalent, 	and shallow dose
equivalent from external exposure to 	the 	radioactive cloud. 	See
Appendix B, 	footnotes 	1 and 2.
(b) 	Airborne 	radioactivity measurements and DAC values
shall not be used as 	the primary means 	to assess 	the deep dose
45-51 	2 265
Complete printed PDF page 51
C
§11-45-42
equivalent when the airborne 	radioactive material includes
radionuclides other than noble gases or if the cloud of airborne
radioactive material is not relatively uniform. 	The determination
of the deep dose equivalent to an individual 	shall be based upon
measurements using instruments or individual monitoring devices.
(c) 	Appendix B to Part D of the Conference of Radiation
Control 	Program Directors, 	Inc. 	Suggested State Regulations 	for
Control of Radiation, 	Volume I, 	December 1995 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations for Release to Sanitary Sewerage",
is made a part of t"i\-H , f ef t"ri offl WW is available from the
department. 	[ Eff 	NUV 	l 	I~~~ 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-43 	Determination of internal exposure. 	(a) 	For
purposes of assessing dose used to determine compliance with
occupational dose equivalent limits, 	the licensee shall, 	when
required under section 11-45-54, 	take 	suitable and timely
measurements of:
(1) 	Concentrations of radioactive materials in air in work
areas; 	or
(2) 	Quantities of radionuclides in the body; 	or
(3) 	Quantities of radionuclides excreted from the body; 	or
(4) 	Combinations of these measurements.
(b) 	Unless 	respiratory protective equipment is used, 	as
provided in section 11-45-60, 	or the assessment of intake is based
on bioassays, 	the licensee shall assume that an individual inhales
radioactive material at the airborne concentration in which the
individual is p r esent.
(c) 	When specific information on the physical and
biochemical properties of the radionuclides taken into the body or
the behavior or the material in an individual is known, 	the
licensee may:
(1) 	Use that information to calculate the committed
effective dose equivalent, 	and, 	if used, 	the licensee
shall document that information in the individual's
record; 	and
(2) 	Upon prior approval of the department, 	adjust the DAC
or ALI values to 	reflect the actual ph y sical and
chemical characteristics of airborne radioactive
material, 	for e x ample, 	aerosol size distribution or
density; 	and
( 3 ) 	Separately assess the contribution of fractional
intakes of Class D, 	W, 	or Y compounds of a 	given
radionuclide to the committed effective dose
equivalent. 	See Appendi x B.
(d) 	If the licensee chooses to assess intakes of Class Y
material using the measurements given in subsection 	(a) (2) 	or
(a) (3), 	the licensee may delay the recording and reporting of the
assessments 	for periods up to seven months, 	unless otherwise
required by sections 11-45- 8 9 and 11-45-90. 	This delay licenses
the licensee to make additional measurements basic to the
45-52 	.!...!..
2 2 6 5 ,, "
Complete printed PDF page 52
(
§11-45-43
assessments.
(e) 	If the identity and concentration of each radionuclide
in a mixture are known, 	the fraction 	of the DAC applicable to the
mixture for use in calculating DAC-hours 	shall be either:
(1) 	The sum of the ratios of the concentration to the
appropriate DAC value, 	that is, 	D, 	W, 	or Y, 	from
Appendix B for each radionuclide in the mixture; 	or
(2) 	The 	ratio of the total concentration for all
radionuclides in the mixture to the most restrictive
DAC value for any radionuclide in the mixture.
(f) 	If the identity of each radionuclide in a mixture is
known, 	but the concentration of one or more of the radionuclides
in the mixture is not known, 	the DAC for the mi x ture shall be the
most 	restrictive DAC of any radionuclide in the mixture.
(g) 	When a mixture of radionuclides in air e x ists, 	a
licensee may disregard certain radionuclides in the mixture if:
(1) 	The licensee uses the total activity of the mixture in
demonstrating compliance with the dose limits in
section 11-45-40 and in complying with the monitoring
requirements in section 11-45-54(b), 	and
(2 ) 	The concentration of any radionuclide disregarded is
less than ten percent of its DAC, 	and
(3) 	The sum of these percentages 	for all of the
radionuclides disregarded in the mixture does not
e x ceed thirty percent .
(h) 	When determining the committed effective dose
equivalent, 	the 	following information may be considered:
(1) 	In order to calculate the committed effective dose
equivalent, 	the licensee may assume that the
inhalation of one ALI, 	or an exposure of two thousand
DAC-hours, 	results in a 	committed effective dose
equivalent of five 	rems 	(0.05 sievert) 	for
radionuclides that have their ALis or DACs based on
the committed effective dose equivalent.
(2) 	For an ALI 	and the associated DAC determined by the
nonstochastic organ dose limit of fifty rems 	(0.50
sievert), 	the intake of radionuclides that would
result in a committed effective dose equivalent of
five 	rems 	(0.05 	sievert), 	that is, 	the stochastic ALI,
is listed in parentheses in Table I 	of Appendix B.
The licensee may, 	as a 	simplifying assumption, 	use the
stochastic ALis to determine committed effective dose
equivalent. 	However, 	if the licensee uses 	the
stochastic ALis, 	the licensee shall also demonstrate
that the limit in section 11-45-40 (a) (1) (B) 	is met.
(i) 	Appendi x B to 	Part D of the Conference of Radiation
Control 	Program Directors, 	Inc. 	Suggested State Regulations 	for
Control of Radiation, 	Volume I, 	December 19 95 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations for Release to Sanitary Sewerage",
is made a part of t p~fi\\ l e i t ~ on and is available from the
department. 	[Eff 	NUV 	! l 	1990 	] 	(Auth: 	HRS §§321 - 10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-I, 	~ 21-11 (21), 	321-71)
45-53
2265
Complete printed PDF page 53
C
L
§11-45-44
§11-45-44 	Determination of prior occupational dose. 	(a)
For each individual who may enter the licensee's restricted or
controlled area and is likely to 	receive, 	in a 	year, 	an
occupational dose requiring monitoring pursuant to section 11-45-
54, 	the licensee shall:
(1) 	Determine the occupational 	radiation dose 	received
during the current year; 	and
(2) 	Attempt to obtain the 	records of lifetime cumulative
occupational 	radiation dose.
(b) 	Before permitting an individual to participate in a
planned special exposure, 	the licensee shall determine:
(1) 	The internal and external doses 	from all previous
planned special exposures; 	and
(2) 	All doses in e x cess of the limits, 	including doses
received during accidents and emergencies, 	received
during the lifetime of the individual.
(c) 	In complying with the 	requirements of subsection 	(a),
a 	licensee may:
(1) 	Accept, 	as a 	record of the occupational dose that the
individual received during the current year, 	a written
signed statement from the individual, 	or from the
individual's most recent employer for work involving
radiation exposure, 	that discloses the nature and the
amount of any occupational dose that the individual
received during the current year; 	and
(2 ) 	Accept, 	as 	the record of lifetime cumulative radiation
dose, 	an up-to-date department 	form, 	signed by the
individual and countersigned by an appropriate
official of the most recent employer for work
involving radiation e x posure, 	or the individual's
current employer, 	if the individual is not employed by
the licensee; 	and
(3) 	Obtain reports of the individual's dose equivalent
from the most 	recent employer for work involving
radiation exposure, 	or the individual's current
employer, 	if the individual is not employed by the
licensee, 	by telephone, 	telegram, 	facsimile, 	or
letter. 	The licensee shall request a written
verification of the dose data if the authenticity of
the transmitted report cannot be established.
(d) 	The licensee shall record the exposure history, 	as
required by subsection 	(a), 	on a 	form provided by the department,
or other clear and legible record, 	of all the information required
on that form. 	The form or record shall show each period in which
the individual 	received occupational exposure to radiation or
radioactive material and shall be signed by the individual who
received the exposure. 	For each period for which the licensee
obtains 	reports, 	the licensee shall use the dose shown in the
report in preparing the department 	form. 	For any period in which
the licensee does not obtain a 	report, 	the licensee shall place a
notation on the department 	form indicating the periods of time for
which data are not a v ailable. 	Licensees are not required to
reevaluate the separate e x ternal dose equivalents and internal
45-54 2 265
Complete printed PDF page 54
§11-45-45
committed dose equivalents or intakes of radionuclides assessed
before the effective date of these 	rules. 	Further, 	occupational
exposure histories obtained and recorded on the department form
before the effective date of these rules, 	would not have included
effective dose equivalent, 	but may be used in the absence of
specific information on the intake of radionuclides by the
individual.
(e) 	If the licensee is unable to obtain a complete record
of an individual's current and previously accumulated occupational
dose, 	the licensee shall assume:
(1) 	In establishing administrative controls under section
11-45-40(f). 	for the current year, 	that the allowable
dose limit for the individual is reduced by 1.25 rems
(12.5 millisieverts) 	for each quarter for which
records were unavailable and the individual was
engaged in activities that could have resulted in
occupational radiation exposure; 	and
(2) 	That the individual is not available for planned
special exposures.
(f) 	The licensee shall retain records until the department
termiiates each.nertinent license requiring this 	record.
[Eff 	!~OV 12 1899 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-~, 	321-11(21), 	321-71)
§11-45-45 	Planned special exposures. 	A licensee may
authorize an adult worker to receive doses in addition to and
accounted for separately from the doses 	received under the limits
specified in section 11-45-40 provided that each of the following
conditions is satisfied:
(1) 	The licensee authorizes a planned special exposure
only in an exceptional situation when alternatives
that might avoid the higher exposure are unavailable
or impractical.
(2) 	The licensee, 	and employer if the employer is not the
licensee, 	specifically authorizes the planned special
exposure, 	in writing, 	before the exposure occurs.
(3) 	Before a planned special exposure, 	the licensee
ensures that each individual involved is:
(A) 	Informed of the purpose of the planned
operation; 	and
(B) 	Informed of the estimated doses and associated
potential risks and specific radiation levels or
other conditions that might be involved in
performing the task; 	and
(C) 	Instructed in the measures to be taken to keep
the dose ALARA considering other risks that may
be present.
(4) 	Before permitting an individual to participate in a
planned special exposure, 	the licensee ascertains
prior doses as required by section 11-45-44(b) 	during
the lifetime of the individual 	for each individual
involved.
45-55
Complete printed PDF page 55
§11-45-45
(5) 	subject to section 11-45-40(b), 	the licensee shall not
authorize a planned special exposure that would cause
an individual to receive a dose from all planned
special exposures and all doses in excess of the
limits to exceed:
(A) 	The numerical values of any of the dose limits
in section ll-45-40(a) 	in any year; 	and
(B) 	Five times the annual dose limits in section ll-
45-40(a) 	during the individual's lifetime.
(6) 	The licensee maintains 	records of the conduct of a
planned special exposure in accordance with section
11-45-82 and submits a 	written report in accordance
with section 11-45-91.
(7) 	The licensee records 	the best estimate of the dose
resulting from the planned special exposure in the
individual's record and informs the individual, 	in
writing, 	of the dose within thirty days 	from the date
of the planned special exposure. 	The dose from
planned special exposures shall not be considered in
controlling future occupational dose of the individual
pursuant to section 11-45-40(a), 	but shall be included
~e~hred by paragraphs 	(4) 	and 	(5).
I~!:;~ 	] 	(Auth: 	HRS §§321-10, 	321-11,
HRS §§321-1, 	321-11(21), 	321-71)
§11-45-46 	Occupational dose limits for minors. 	The annual
occupational dose limits 	for minors are ten percent of the annual
occupational dose livi-its specified for adult workers in section
11-45-40. 	[Eff 	NQ\/ 1 2 1990 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	32':r-11 (21), 	321-71)
§11-45-47 	Dose to an embryo/fetus. 	(a) 	The licensee shall
ensure that the dose to an embryo/fetus during the entire
pregnancy, 	due to occupational exposure of a declared pregnant
woman, 	does not exceed 0.5 rem 	(five millisieverts).
(b) 	The licensee shall make efforts to avoid substantial
variation above a uniform monthly exposure rate to a declared
pregnant woman so as to satisfy the limit in subsection 	(a).
(c) 	The dose to an embryo/fetus shall be taken as 	the sum
of:
(1) 	The deep dose equivalent to the declared pregnant
woman; 	and
(2) 	The dose to the embryo/fetus from radionuclides in the
embryo/fetus and radionuclides in the declared
pregnant woman.
(d) 	If by the time the woman declares pregnancy to the
licensee, 	the dose to the embryo/fetus has exceeded 0.45 rem 	(4.5
millisieverts), 	the licensee shall be deemed to be in compliance
with subsection 	(a). 	if the additional dose to the embryo/fetus
does not exceed 0.05 rem 	) during the remainder
of the pregnancy. 	[Eff 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	11(21), 	321-71)
45-56
Complete printed PDF page 56
(
§11-45-49
§11-45-48 	Dose limits 	for individual members of the public.
(a) 	Each licensee shall conduct operations so that:
(1) 	The total effective dose equivalent to individual
members of the public from the licensed operation does
not exceed 0.1 rem 	(one millisieverts) 	in a 	year,
exclusive of the dose contribution from the licensee's
disposal of radioactive material into sanitary
sewerage in accordance with section 11-45-72 	(Retrofit
shall not be required for locations within facilities
where only radiation machines existed before the
effective date of these 	rules and met the previous
requirements of 0.5 	rem 	(five millisieverts) 	in a
year), 	and
(2) 	The dose in any unrestricted area from e x ternal
sources does not exceed 0.002 	rem 	(0.02 millisieverts)
in any one hour.
(b) 	If the licensee allows members of the public to have
access to controlled areas, 	the limits 	for members of the public
continue to apply to those individuals.
(c) 	A licensee may apply for prior department
authorization to operate up to an annual dose limit 	for an
individual member of the public of 0.5 	rem 	(five millisieverts)
This application shall include the 	following information:
(1) 	Demonstration of the need for and the expected
duration of operations in excess of the limit in
subsection 	(a); 	and
(2) 	The licensee's program to assess and control dose
within the 	0.5 rem 	(five millisieverts) 	annual limit;
and
(3) 	The procedures to be 	followed to maintain the dose
ALARA.
(d) 	In addition to the requirements of subchapter 4, 	a
licensee subject to the provisions of the U.S. 	Environmental
Protection Agency's generally applicable environmental 	radiation
standards in 40 C.F.R. 	Part 190 shall comply with those standards.
(e) 	The department may impose additional 	restrictions on
radiation levels in unrestricted areas and on the total quantity
of radionuclides that a 	licensee may release in effluents in order
to 	restrict the collective dose . 	[Eff ~1(1\f 	1 9 1900 	] 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	hi'1t§ iz~ l, 	~~ -11(21),
321-71)
§11-45-49 	Compliance with dose limits 	for individual
members of the public. 	(a) 	The licensee shall make or cause to
be made surveys of radiation levels in unrestricted and controlled
areas and radioactive materials in effluents 	released to
unrestricted and controlled areas 	to demonstrate compliance with
the dose limits 	for individual members of the public in section
11-45-48.
(b) 	A licensee shall show compliance with the annual dose
limit in section 11-45-48 by:
45-57 	2 265
Complete printed PDF page 57
(
§11-45-49
(1) 	Demonstrating by measurement or calculation that the
tota l 	effective dose equivalent to the individual
likely to 	receive the highest dose from the licensed
operation does not exceed the annual dose limit; 	or
(2) 	Demonstrating that :
(A) 	The annual average concentrations of radioactive
material released in gaseous and liquid
effluents at the boundary of the unrestricted
area do not exceed the values specified in Table
II 	of Appendix B; 	and
(B) 	If an individual were continually present in an
unrestricted area, 	the dose 	from external
sources would not exceed 0.002 	rem 	(0 . 02
mi l l isieverts) 	in an hour and 0.05 	rem 	(0 . 50
millisieverts) 	in a 	year.
(c) 	Upon approval 	from the department, 	the licensee may
adjust the effluent concentration values in Appendi x B, 	Table II,
for members of the public, 	to take into account the actual
physical and chemical characteristics of the effluents, 	such as,
aerosol size distribution, 	solubility, 	density, 	radioactive decay
equilibrium, 	and chemical form.
(d) 	Appendix B to Part D of the Conference of Radiation
Control 	Program Directors, 	Inc . 	Suggested State Regulations 	for
Control of Radiation, 	Volume I, 	December 1995 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations 	for Release to Sanitary Sewerage",
is made a part of this 	section and is available from the
department. 	[Eff t, ny__ 	&) 1~99 	) 	(Auth: 	HRS §§321-10, 	321-
11, 	321 - 71) 	(Imp: 	~ RS 	3"21 ~ , 	3 2 1-11 (21), 	321 - 71)
§11-45-50 	Reser v ed.
§11 - 45-51 	Reserved.
§11-45-52 	Testing for leakage or contamination of sealed
sources. 	(a) 	Each sealed source with a half-life greater than
thirty days and in any form other than gas, 	shall be tested for
leakage or contamination before initial use and, 	unless otherwise
authorized by the department, 	at intervals not to e x ceed six
months, 	e x cept that each source designed for the purpose of
emitting alpha particles shall be tested at intervals not to
exceed three months. 	If, 	at any other time, 	there is reason to
suspect that a 	sealed source might have been damaged or might be
l eaking, 	it sha ll be tested for leakage before further use. 	In
the absence of a 	certificate from a 	transferor indicating that a
test for leakage has been made within six months before the
transfer, 	the sea l ed source shall not be put into use until 	tested
and the 	results 	rece i ved.
(1) 	Tests 	for leakage 	for all sealed sources, 	except those
manufactured to contain radium, 	shall be capable of
45-58
2265
Complete printed PDF page 58
C
§11-45-52
detecting the presence of 0.005 microcuries 	(one
hundred eighty-five becquerels) 	of radioactive
material on a test sample. 	Test samples shall be
taken from the sealed source or from the surfaces of
the container in which the sealed source is stored or
mounted on which one might expect contamination to
accumulate.
(2) 	The test for leakage 	for sources manufactured to
contain radium shall be capable of detecting an
absolute leakage rate of 0.001 microcuries 	(thirty-
seven becquerels) 	of radon-222 	in a 	twenty-four hour
period when the collection efficiency for radon-222
and its daughters has been determined with respect to
collection method, 	volume and time.
(3) 	Test samples shall also be taken from the interior
surfaces of the container in which sealed sources of
radium are stored. 	This test shall be capable of
detecting the presence of 0.005 microcuries 	(one
hundred eighty-five becquerels) 	of a 	radium daughter
which has a half-life greater than four days.
(4) 	Notwithstanding the periodic test for leakage
required, 	any sealed source is exempt from such tests
for leakage when the sealed source contains one
hundred microcuries 	(3.7 megabecquerels) 	or less of
beta or gamma emitting material or ten microcuries
(three hundred seventy becquerels) 	or less of alpha
emitting material.
(b) 	Tests 	for leakage or contamination shall be performed
by persons 	specifically authorized by the department to perform
such services .
(c) 	The following shall be considered evidence that the
sealed source is leaking:
(1) 	The presence of 0.005 microcuries 	(one hundred e i ghty-
five becquerels) 	or more of removable contamination on
any test sample. 	If the test of a 	sealed source,
other than radium, 	reveals the presence of 0.005
microcuries 	(one hundred eighty-five becquerels) 	or
more of removable contamination, 	the licensee shall
immediately withdraw the sealed source from use, 	take
action to prevent the spread of contamination, 	and
cause the sealed source to be decontaminated and
repaired or to be disposed of in accordance with
subchapter 4.
(2) 	Leakage of 0.001 microcuries 	(thirty-seven becquerels)
of radon-222 per twenty-four hours 	for 	sealed sources
manufactured to contain radium. 	If the test of a
sealed source manufactured to contain radium reveals
the presence of removable contamination resulting from
the decay of 0.005 microcuries 	(one hundred eighty-
five becquerels) 	or more of radium-226, 	the licensee
shall immediately withdraw the sealed source from use,
take action to prevent the spread of contamination,
cause the sealed source to be decontaminated and
repaired or to be disposed of in accordance with
45-59 2 265
Complete printed PDF page 59
(
§11-45-52
subchapter 4.
(d) 	Test results shall be kept in units of becquerel or
microcurie and maintained for 	inspection by the department 	for
three years.
(e) 	Reports of test results 	for leaking or contaminated
seale ~1 Qyr~ eN ~ hall be made pursuant to section 11-45-96.
[Eff !\luV 	1' 	,999 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-53 	Surveys and monitoring. 	(a) 	Radiation surveys
shall be conducted in all areas in the vicinity of sources of
radiation by, 	or under the direction of, 	a 	qualified medical
physicist.
(b) 	Instruments and equipment used for quantitative
radiation measurements shall be calibrated at least annually.
(c) 	All personnel dosimeters, 	except 	for direct and
indirect reading pocket ionization chambers and those dosimeters
used to measure the dose to any extremity, 	that require processing
to determine the 	radiation dose and that are used by licensees to
comply with section 11-45-40, 	with other applicable provisions of
this chapter, 	or with conditions specified in a license shall be
processed and evaluated by a dosimetry processor:
(1) 	Holding current personnel dosimetry accreditation from
the National Voluntary Laboratory Accreditation
Program 	(NVLAP) 	of the National 	Institute of Standards
and Technology; 	and
(2) 	Approved in this accreditation process 	for the type of
radiation or radiations included in the NVLAP program
that most closely approximates the type of radiation
or radiations 	for which the individual wearing the
dosimeter is monitored.
(d) 	The licensee shall ensure that adequate precautions
are taken to prevent a dece R tive exposure of an individual
monitoring device. 	[Eff Nu\/ 1 ? 1000 	l 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§32".l. ~ , '1~~-11(21), 	321-71)
§11-45-54 	Conditions 	requiring individual monitoring of
external and internal occupational dose. 	(a) 	Each licensee shall
monitor occupational exposure to 	radiation and shall supply and
require the use of individual monitoring devices by:
(1) 	Adults likely to 	receive, 	in one year from sources
external to the body, 	a dose in excess of ten percent
of the limits in section 11-45-40(a); 	and
(2) 	Minors and declared pregnant women likely to 	receive,
in one year from sources external to the body, 	a dose
in e x cess of ten percent of any of the applicable
limits in sections 11-45-46 and 11-45-47; 	and
( 3 ) 	Individuals entering a high or very high radiation
area.
(b) 	Each licensee shall monitor, 	to determine compliance
with section 11-45-43, 	the occupational intake of radioactive
material by and assess the committed effective dose equivalent to :
45-60
2 265
Complete printed PDF page 60
§11-45-55
(1) 	Adults likely to receive, 	in one year, 	an intake in
excess of ten percent of the applicable ALI in Table
I, 	Columns 1 and 2, 	of Appendix B; 	and
(2) 	Minors and declared pregnant women likely to receive,
in one year, 	a committed effective dose equivalent in
excess of 0.05 rem 	(0.50 millisievert).
(c) 	Appendix B to Part D of the Conference of Radiation
Control Program Directors, 	Inc. 	suggested state Regulations for
Control of Radiation, 	Volume I, 	December 1995 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations for Release to Sanitary Sewerage",
is made a part of t,hirt,section and is available from the
department. 	[Eff 	NOv 1 ,, 1090 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	'HRS §§3~1-~, ~21-11 (21), 	321-71)
§11-45-55 	Control of access to high radiation areas. 	(a)
The licensee shall ensure that each entrance or access point to a
high radiation area has one or more of the following features:
(1) 	A control device that, 	upon entry into the area,
causes the level of radiation to be reduced below that
level at which an individual might receive a deep dose
equivalent of 0.1 	rem 	(one millisievert) 	in one hour
at thirty centimeters from the source of radiation
from any surface that the radiation penetrates; 	or
(2) 	A control device that energizes a conspicuous visible
or audible alarm signal so that the individual
entering the high radiation area and the supervisor of
the activity are made aware of the entry; 	or
(3) 	Entryways that are locked, 	except during periods when
access to the areas 	is required, 	with positive control
over each individual entry.
(b) 	In place of the controls required by subsection 	(a)
for a 	high radiation area, 	the licensee may substitute continuous
direct or electronic surveillance that is capable of preventing
unauthorized entry.
(c) 	The licensee may apply to the department for approval
of alternative methods 	for controlling access to high radiation
areas.
(d) 	The licensee shall establish the controls 	required by
subsections 	(a) 	and 	(c) 	in a way that does not prevent individuals
from leaving a high radiation area.
(e) 	The licensee is not required to control each entrance
or access point to a 	room or other area that is a 	high radiation
area solely because of the presence of radioactive materials
prepared for transport and packaged and labeled in accordance with
the regulations of the U.S. 	Department of Transportation provided
that:
(1) 	The packages do not remain in the area longer than
three days; 	and
(2) 	The dose 	rate at one meter from the external surface
of any package does not exceed 0.01 	rem 	(0.1
millisievert) 	per hour.
45-61 2 6 5
Complete printed PDF page 61
§11-45-55
(f) 	The licensee is not required to control entrance or
access 	to rooms or"other areas in hospitals solely because of the
presence of patients containing radioactive material, 	provided
that there are personnel in attendance who are taking the
necessary precautions to prevent the exposure of individuals to
radiation or radioactive material in excess of the established
limits in this subchapter and to operate within the ALARA
provisions of the licensee's radiation protection program.
(g) 	The licensee is not required to control entrance or
access 	to rooms or other areas containing sources of radiation
capable of producing a high radiation area as described in this
section if the licensee has met all the specific requirements 	for
acces~I 	c.;;v~~·Jd./l'!!n 	in other applicable subchapters.
[Eff 	r\J 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-56 	Control of access 	to very high radiation areas.
(a) 	In addition to the requirements in section 11-45-55, 	the
licensee shall institute additional measures to ensure that an
individual is not able to gain unauthorized or inadvertent access
to areas in which radiation levels could be encountered at five
hundred rads 	(five grays) 	or more in one hour at one meter from a
source of radiation or any surface through which the 	radiation
penetrates. 	This requirement does not apply to rooms or areas in
which diagnostic x-ray systems are the only source of radiation,
or to non-self-shielded irradiators.
(b) 	The licensee is not required to control entrance or
access 	to rooms or other areas containing sources of radiation
capable of producing a very high radiation area as described in
subsection 	(a) 	if the licensee has met all the specific
requirements 	for acc~rAi1n~ 	specified in other applicable
subchapters. 	[Eff 	rHJ\ 	1 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§32 	(21), 	321-71)
§11-45-57 	Control of access to very high radiation areas --
irradiators. 	(a) 	This section applies to licensees with sources
of radiation in non-self-shielded irradiators, 	and does not apply
to sources of radiation that are used in teletherapy, 	in
industrial radiography, 	or in completely self-shielded irradiators
in which the source of radiation is both stored and operated
within the same shielding radiation barrier and, 	in the designed
configuration of the irradiator, 	is always physically inaccessible
to any individual and cannot create a 	radiation level of five
hundred rads 	(five grays) 	or more in one hour at one meter in an
area that is accessible to any individual.
(b) 	Each area in which there may exist radiation levels in
excess of five hundred rads 	(five grays) 	in one hour at one meter
from a 	source of radiation that is used to irradiate materials
shall meet the following requirements:
(1) 	Each entrance or access point shall be equipped with
entry control devices which:
(A) 	Function automatically to prevent any individual
45-62 6
Complete printed PDF page 62
from inadvertently entering a very high
radiation area; 	and
§11-45-57
(B) 	Permit deliberate entry into the area only after
a 	control device is actuated that causes the
radiation level within the area, 	from the source
of radiation, 	to be reduced below that at which
it would be possible for an individual to
receive a deep dose equivalent in excess of 0.1
rem 	(one millisievert) 	in one hour; 	and
(C) 	Prevent operation of the source of radiation if
it would produce radiation levels in the area
that could result in a deep dose equivalent to
an individual in excess of 0.1 rem 	(one
millisievert) 	in one hour.
(2) 	Additional control devices shall be provided so that,
upon failure of the entry control devices to 	function
as 	required by paragraph 	(1):
(A) 	The radiation level within the area, 	from the
source of radiation, 	is reduced below that at
which it would be possible for an individual to
receive a deep dose equivalent in excess of 0.1
rem 	(one millisievert) 	in one hour; 	and
(B) 	Conspicuous visible and audible alarm signals
are generated to make an individual attempting
to enter the area aware of the hazard and at
least one other authorized individual, 	who is
physically present, 	familiar with the activity,
and prepared to render or summon assistance,
aware of the failure of the entry control
devices.
(3) 	The licensee shall provide control devices 	so that,
upon failure 	or removal of physical radiation barriers
other than the source's shielded storage container:
(A) 	The radiation level from the source of radiation
is reduced below that at which it would be
possible 	for an individual to 	receive a deep
dose equivalent in excess of 0.1 	rem 	(one
millisievert) 	in one hour; 	and
(B) 	Conspicuous visible and audible alarm signals
are generated to make potentially affected
individuals aware of the hazard and the licensee
or at least one other individual, 	who is
familiar with the activity and prepared to
render or summon assistance, 	aware of the
failure or removal of the physical barrier.
(4) 	When the shield for stored sealed sources is a 	liquid,
the licensee shall provide means to monitor the
integrity of the shield and to signal, 	automatically,
loss of adequate shielding.
(5) 	Physical radiation barriers that comprise permanent
structural components, 	such as walls, 	that have no
credible probability of failure or removal in ordinary
circumstances need not meet the requirements of
paragraphs 	(3) 	and 	(4).
45-63
265
Complete printed PDF page 63
§11-45-57
(6) 	Each area shall be equipped with devices that shall
automatically generate conspicuous visible and audible
alarm signals to alert personnel in the area before
the source of radiation can be put into operation and
in time for any individual in the area to operate a
clearly identified control device, 	which shall be
installed in the area and which can prevent the source
of radiation from being put into operation.
(7) 	Each area shall be controlled by use of such
administrative procedures and such devices as are
necessary to ensure that the area is cleared of
personnel before each use of the source of radiation.
(8) 	Each area shall be checked by a 	radiation measurement
to ensure that, 	before the first individual's entry
into the area after any use of the source of
radiation, 	the radiation level from the source of
radiation in the area is below that at which it would
be possible for an individual to receive a deep dose
equivalent in excess of 0.1 rem 	(one millisievert) 	in
one hour.
(9) 	The entry control devices 	required in paragraph 	(1)
shall have been tested for proper functioning.
(A) 	Testing shall be conducted before initial
operation with the source of radiation on any
day, 	unless operations were continued
uninterrupted from the previous day; 	and
(B) 	Testing shall be conducted before resumption of
operation of the source of radiation after any
unintentional interruption; 	and
(C) 	The licensee shall submit and adhere to a
schedule for periodic tests of the entry control
and warning systems.
(10) 	The licensee shall not conduct operations, 	other those
necessary to place the source of radiation in safe
condition or to effect repairs on controls, 	unless
control devices are functioning properly.
(11) 	Entry and exit portals that are used in transporting
materials to and from the irradiation area, 	and that
are not intended for use by individuals, 	shall be
controlled by such devices and administrative
procedures as are necessary to physically protect and
warn against inadvertent entry by any individual
through these portals. 	Exit portals 	for irradiated
materials 	shall be equipped to detect and signal the
presence of any loose radioactive material that is
carried toward such an exit and automatically to
prevent loose radioactive material from being carried
out of the area.
(c) 	Licensees of sources of radiation within the purview
of subsection 	(b) 	which shall be used in a variety of positions or
in locations, 	such as open fields 	or forests, 	that make it
impracticable to comply with certain requirements of subsection
(b), 	such as 	those for the automatic control of radiation levels,
may apply to the department for approval of alternative safety
45-64 265,
Complete printed PDF page 64
§11-45-60
measures. 	Alternative safety measures shall provide personnel
protection at least equivalent to those specified in subsection
(b). 	At least one of the alternative measures shall include an
entry-preventing interlock control based on a measurement of the
radiation that ensures the absence of high radiation levels before
an individual can gain access to the area where such sources of
radiation are used.
(d) 	The entry control devices 	required by subsections 	(b)
and 	(c) 	shall be established in such a way that no individual
shall be prevented from leaving the area. 	[Effi\J('iv 	I?
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	H:kS-s§3Zl.!.'1.,
11 (21), 	321-71)
§11-45-58 	Use of process or other engineering controls.
The licensee shall use, 	to the extent practicable, 	process or
other engineering controls, 	such as, 	containment or ventilation,
to 	~h~ i~1rfiintrations of radioactive material in air.
[Eff 	1 l 	;:j::I~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-59 	Use of other controls. 	When it is not
practicable to apply process or other engineering controls to
control the concentrations of radioactive material in air to
values below those that define an airborne radioactivity area,
licensee shall, 	consistent with maintaining the total effective
dose equivalent ALARA, 	increase monitoring and limit intakes by
one or more of the following means:
(1) 	Control of access; 	or
; 	or
the
(2) 	Limitation of exposure times; 	or
(3) 	Use of respiratory pr<),1~fl~f
( 4) 	Other controls. 	[Eff 11
, UV 	l
§§321-10, 	321-11, 	321-71) 	(Imp:
11 (21), 	321-71)
] 	(Auth: 	HRS
§§321-1, 	321-
§11-45-60 	Use of individual respiratory protection
equipment. 	(a) 	If the licensee uses 	respiratory protection
equipment to limit intakes pursuant to section 11-45-59:
(1) 	Except as provided in paragraph 	(2), 	the licensee
shall use only respiratory protection equipment that
is tested and certified or had certification extended
by the National Institute 	for Occupational Safety and
Health and the Mine Safety and Health Administration.
(2) 	If the licensee wishes to use equipment that has not
been tested or certified by the National Institute for
Occupational Safety and Health and the Mine Safety and
Health Administration, 	and has not had certification
extended by the National Institute for Occupational
Safety and Health and the Mine Safety and Health
Administration, 	or for which there is no schedule for
testing or certification, 	the licensee shall submit an
application for authorized use of that equipment,
45-65 u
265 	I
Complete printed PDF page 65
§11-45-60
including a demonstration by testing, 	or a
demonstration on the basis of reliable test
information, 	that the material and performance
characteristics of the equipment are capable of
providing the proposed degree of protection under
anticipated conditions of use.
(3) 	The licensee shall implement and maintain a
respiratory protection program that includes:
(A) 	Air sampling sufficient to identify the
potential hazard, 	permit proper equipment
selection, 	and estimate exposures; 	and
(B) 	Surveys and bioassays, 	as appropriate, 	to
evaluate actual intakes; 	and
(C) 	Testing of respirators 	for operability
immediately before each use; 	and
(D) 	Written procedures regarding selection, 	fitting,
issuance, 	maintenance, 	and testing of
respirators, 	including testing for operability
immediately before each use; 	supervision and
training of personnel; 	monitoring, 	including air
sampling and bioassays; 	and recordkeeping; 	and
(E) 	Determination by a physician before initial
fitting of respirators, 	and at least every
twelve months thereafter, 	that the individual
user is physically able to use the respiratory
protection equipment.
(4) 	The licensee shall issue a written policy statement on
respirator usage covering:
(A) 	The use of process or other engineering
controls, 	instead of respirators; 	and
(B) 	The routine, 	nonroutine, 	and emergency use of
respirators; 	and
(C) 	The length of periods of respirator use and
relief from respirator use.
(5) 	The licensee shall advise each respirator user that
the user may leave the area at any time for 	relief
from respirator use in the event of equipment
malfunction, 	physical or psychological distress,
procedural or communication failure, 	significant
deterioration of operating conditions, 	or any other
conditions that might require such relief.
(6) 	The licensee shall use equipment within the equipment
manufacturer's expressed limitations for type and mode
of use and shall provide proper visual, 	communication,
and other special capabilities, 	such as adequate skin
protection, 	when needed.
(b) 	When estimating exposure of individuals to airborne
radioactive materials, 	the licensee may make allowance for
respiratory protection equipment used to limit intakes pursuant to
section 11-45-59, 	provided that the following conditions, 	in
addition to those in subsection 	(a), 	are satisfied:
(1) 	The licensee selects respiratory protection equipment
that provides a protection factor, 	specified in
Appendix A, 	greater than the multiple by which peak
45-66 Ll
26
Complete printed PDF page 66
§11-45-61
concentrations of airborne radioactive materials in
the working area are expected to exceed the values
specified in Appendix B, 	Table I, 	Column 3. 	However,
if the selection of respiratory protection equipment
with a protection factor greater than the peak
concentration is inconsistent with the goal specified
in section 11-45-59 of keeping the total effective
dose equivalent ALARA, 	the licensee may select
respiratory protection equipment with a lower
protection factor provided that such a 	selection would
result in a 	total effective dose equivalent that is
ALARA. 	The concentration of radioactive material in
the air that is inhaled when respirators are worn may
be initially estimated by dividing the average
concentration in air, 	during each period of
uninterrupted use, 	by the protection factor. 	If the
exposure is later found to be greater than initially
estimated, 	the corrected value shall be used; 	if the
exposure is later found to be less than initially
estimated, 	the corrected value may be used.
(2) 	The licensee shall obtain authorization from the
department before assigning respiratory protection
factors in excess of those specified in Appendix A.
The department may authorize a 	licensee to use higher
protection factors on receipt of an application that:
(A) 	Describes the situation for which a need exists
for higher protection factors, 	and
(B) 	Demonstrates that the respiratory protection
equipment provides these higher protection
factors under the proposed conditions of use.
(c) 	In an emergency, 	the licensee shall use as emergency
equipment only respiratory protection equipment that has been
specifically certified or had certification extended for emergency
use by the National Institute for Occupational Safety and Health
and the Mine Safety and Health Administration.
(d) 	The licensee shall notify the department in writing at
least thirty days before the date that respiratory protection
equipment is 	first used pursuant to either subsection 	(a) 	or 	(b).
(e) 	The exhibit at the end of chapter 11-45 entitled,
"Appendix A of Subchapter 4, 	Protection Factors for Respirators
(2/2/93)", 	is made a part of this section.
(f) 	Appendix B to Part D of the Conference of Radiation
Control 	Program Directors, 	Inc. 	Suggested State Regulations 	for
Control of Radiation, 	Volume I, 	December 1995 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations for Release to Sanitary Sewerage",
is made a part of this section and is available from the
department. 	[Eff 	NOV 1 2 1999 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, ~21-11(21), 	321-71)
§11-45-61 	Security of stored sources of radiation. 	The
licensee shall secure from unauthorized removal or access licensed
45-67
Complete printed PDF page 67
§11-45-61
sources of radfatto9, ~h~oMe stored in controlled or unrestricted
areas. 	[Eff f,,lQ\/ 	1 ;l 	18;j:::J 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-62 	Control of sources of radiation not in storage.
(a) 	The licensee shall control and maintain constant surveillance
of licensed radioactive material that is in a 	controlled or
unrestricted area and that is not in storage or in a patient.
(b) 	The licensee shall maintain control of licensed
radiation machine~ that are in a coi.,ff,q1.,l~.dt:;P~~estricted area
and that are not in storage. 	[Eff 	{\UV 	1.l 	!::,~~ 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-63 	Caution signs. 	(a) 	Unless otherwise authorized
by the department, 	the radiation symbol is the three-bladed design
specified in Appendix D.
(b) 	Notwithstanding the requirements of subsection 	(a),
licensees are authorized to label sources, 	source holders, 	or
device components containing sources of radiation that are
subjected to high temperatures, 	with conspicuously etched or
stamped radiation caution symbols and without a 	color requirement.
(c) 	In addition to the contents of signs and labels
prescribed in this subchapter, 	the licensee shall provide, 	on or
near the required signs and labels, 	additional information, 	as
appropriate, 	to make individuals aware of potential radiation
exposures and to minimize the exposures.
(d) 	The exhibit at the end of chapter
"Appendix D of Subchapter 4,
a part of this 	section. 	[Eff
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1,
11-45 entitled,
(8/19/98)", 	is made
] 	(Auth: 	HRS
321-11(21), 	321-71)
§11-45-64 	Posting requirements. 	(a) 	The licensee shall
post each radiation area with a conspicuous sign or signs bearing
the radiation symbol and the words "CAUTION, 	RADIATION AREA."
(b) 	The licensee shall post each high radiation area with
a conspicuous sign or signs bearing the radiation symbol and the
words 	"CAUTION, 	HIGH RADIATION AREA" or "DANGER, 	HIGH RADIATION
AREA."
(c) 	The licensee shall post each very high radiation area
with a conspicuous sign or signs bearing the radiation symbol and
words 	"GRAVE DANGER, 	VERY HIGH RADIATION AREA."
(d) 	The licensee shall post each airborne radioactivity
area with a conspicuous sign or signs bearing the radiation symbol
and the words 	"CAUTION, 	AIRBORNE RADIOACTIVITY AREA" or "DANGER,
AIRBORNE RADIOACTIVITY AREA."
(e) 	The licensee shall post each area or room in which
there is used or stored an amount of licensed material exceeding
ten times the quantity of such material specified in Appendix C
with a conspicuous sign or signs bearing the radiation symbol and
the words 	"CAUTION, 	RADIOACTIVE MATERIAL(S)" 	or "DANGER,
RADIOACTIVE MATERIAL(S) ."
45-68
5
Complete printed PDF page 68
§11-45-66
(f) 	The exhibit at the end of chapter 11-45 entitled,
"Appendix C of Subchapter 4, 	Quantities of Material Requiring
Labeli3t<jr 	(~/2/93)", 	is made a part of this section.
[Eff 	NO\/ 	1 2 1990 ] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	t21-11(21), 	321-71)
§11-45-65 	Exceptions to posting requirements. 	(a) 	A
licensee is not 	required to post caution signs in areas or rooms
containing sources of radiation for periods of less than eight
hours, 	if each of the following conditions is met:
(1) 	The sources of radiation are constantly attended
during these periods by an individual who takes the
precautions necessary to prevent the exposure of
individuals to sources of radiation in excess of the
limits established in this subchapter; 	and
(2) 	The area or room is subject to the licensee's control.
(b) 	Rooms or other areas in hospitals that are occupied by
patients are not required to be posted with caution signs pursuant
to this section provided that the patient could be released from
confinement pursuant to subchapter 7.
(c) 	A room or area is not required to be posted with a
caution sign because of the presence of a 	sealed source provided
the radiation level at thirty centimeters from the surface of the
sealed source container or housing does not exceed 0.005 	rem 	(0.05
millisieverts) 	per hour.
(d) 	A room or area is not 	required to be posted with a
caution sign because of the presence of a diaip~~ti~ h-~~Aystem
used solely for healing arts purposes. 	[Eff PH.r✓ 	l l 	l!'.::t~~ 	]
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-66 	Labeling containers and radiation machines. 	(a)
The licensee shall ensure that each container of licensed
material bears a durable, 	clearly visible label bearing the
radiation symbol and the words 	"CAUTION, 	RADIOACTIVE MATERIAL" or
"DANGER, 	RADIOACTIVE MATERIAL." The label shall also provide
information, 	such as the radionuclides present, 	an estimate of the
quantity of radioactivity, 	the date for which the activity is
estimated, 	radiation levels, 	kinds of materials, 	and mass
enrichment, 	to license individuals handling or using the
containers, 	or working in the vicinity of the containers, 	to take
precautions to avoid or minimize exposures.
(b) 	Each licensee shall, 	before removal or disposal of
empty uncontaminated containers to unrestricted areas, 	remove or
deface the radioactive material label or otherwise clearly
indicate that the container no longer contains radioactive
materials.
(c) 	Each licensee shall ensure that each radiation machine
is labeled in a 	conspicuous manner which cautions individuals that
radia~'r'rlf-S,s 	uM..;i..,':A-' when it is energized.
[Eff 	hu '( 	l 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§32 	1-11(21), 	321-71)
45-69
2
Complete printed PDF page 69
§11-45-67
§11-45-67 	Exemptions to labeling requirements. 	(a) 	A
licensee is not required to label:
(1) 	Containers holding licensed material in quantities
less than the quantities listed in Appendix C; 	or
(2) 	Containers holding licensed material in concentrations
less than those specified in Table III of Appendix B;
or
(3) 	Containers attended by an individual who takes the
precautions necessary to prevent the exposure of
individuals in excess of the limits established by
this subchapter; 	or
(4) 	Containers when they are in transport and packaged and
labeled in accordance with the regulations of the U.S.
Department of Transportation, 	or
(5) 	Containers that are accessible only to individuals
authorized to handle or use them, 	or to work in the
vicinity of the containers, 	if the contents are
identified to these individuals by a 	readily available
written record. 	Examples of containers of this type
are containers in locations such as water-filled
canals, 	storage vaults, 	or hot cells. 	The record shall
be retained as long as the containers are in use for
the purpose indicated on the record; 	or
(6) 	Installed manufacturing or process equipment, 	such as
chemical process equipment, 	piping, 	and tanks.
(b) 	Appendix B to Part D of the Conference of Radiation
Control Program Directors, 	Inc. 	Suggested State Regulations 	for
Control of Radiation, 	Volume I, 	December 1995 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations for Release to Sanitary Sewerage
(2/2/93)", 	is made a part of this section and is available from
the department.
(c) 	The exhibit at the end of chapter 11-45 entitled,
"Appendix C of Subchapter 4, 	Quantities of Material Requiring
Labeli~W)(i/~ -~~-•- 	made a part of this section.
[Eff 	1\1U V 	1. 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	1-11 (21), 	321-71)
§11-45-68 	Reserved.
§11-45-69 	Reserved.
§11-45-70 	Waste disposal. 	(a) 	A licensee shall dispose of
licensed radioactive material only:
(1) 	By transfer to an authorized recipient as provided in
section 11-45-75; 	or
(2) 	By decay in storage; 	or
(3) 	By release in effluents within the limits in section
45-70
Complete printed PDF page 70
11-45-48; 	or
(4) 	As authorized in this subchapter.
(b) 	A person shall be specifically licensed to receive
waste containing licensed material from other persons 	for:
(1) 	Treatment before disposal; 	or
(2) 	Treatment or disposal by incineration; 	or
(3) 	Decay in storage; 	or
(4) 	Storage until transferred to a disposal 	facility
authorized to receive the waste.
§11-45-72
(c) 	Disposal of licensed radioactive material other than
that 5Mnffif4,i~QWsections 	(a) 	and 	(b) 	shall be prohibited.
[Eff 	l~U \l 	_ ;(, 	1::1:::1::t 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-71 	Method for obtaining approval of proposed
disposal procedures. 	A licensee may apply to the department for
approval of proposed procedures, 	not otherwise authorized in these
rules, 	to dispose of licensed material generated in the licensee's
operations. 	Each application shall include:
(1) 	A description of the waste containing licensed
material to be disposed of, 	including the physical and
chemical properties that have an impact on risk
evaluation, 	and the proposed manner and conditions of
waste disposal; 	and
(2) 	An analysis and evaluation of pertinent information on
the nature of the environment; 	and
(3) 	The nature and location of other potentially affected
facilities; 	and
(4) 	Analyses and procedures to ensure that doses are
maintained ALARA at:?n.\:~t~ip. the dose limits in this
subchapter. 	[ E ff 	l"HJ \/ 	1 ;i 1999 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-72 	Disposal by release into sanitary sewerage. 	(a)
A licensee may discharge licensed material into sanitary sewerage
if each of the following conditions is satisfied:
(1) 	The material is readily soluble, 	or is readily
dispersible biological material, 	in water; 	and
(2) 	The quantity of licensed radioactive material that the
licensee releases into the sewer in one month divided
by the average monthly volume of water released into
the sewer by the licensee does not exceed the
concentration listed in Table III of Appendix B; 	and
(3) 	If more than one radionuclide is released, 	the
following conditions shall also be satisfied:
(A) 	The licensee shall determine the fraction of the
limit in Table III of Appendix B represented by
discharges into sanitary sewerage by dividing
the actual monthly average concentration of each
radionuclide released by the licensee into the
sewer by the concentration of that 	radionuclide
45-71
Complete printed PDF page 71
§11-45-72
listed in Table III 	of Appendix B; 	and
(B) 	The sum of the fractions 	for each radionuclide
required by subparagraph 	(A) 	does not exceed
unity; 	and
(4) 	The total quantity of licensed radioactive material
that the licensee releases into the sanitary sewerage
system in a 	year does not exceed one curie 	(thirty
seven gigabecquerels) 	of all other radioactive
materials combined.
(b) 	Excreta from individuals undergoing medical diagnosis
or therapy with radioactive material are not subject to the
limitations contained in subsection 	(a).
(c) 	Appendix B to Part D of the Conference of Radiation
Control 	Program Directors, 	Inc. 	Suggested State Regulations 	for
Control of Radiation, 	Volume I, 	December 1995 Edition, 	entitled,
"Annual Limits on Intake 	(ALI) 	and Derived Air Concentrations
(DAC) 	of Radionuclides 	for Occupational Exposure; 	Effluent
Concentrations; 	Concentrations 	for Release to Sanitary Sewerage",
is made a part of this section and is available from the
department. 	[Eff 	Mf)\/ 1 2 1090 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: u 0
HRS §§3"'21-Y, ~21-11 (21), 	321-71)
§11-45-73 	Treatment or disposal by incineration. 	A
licensee may treat or dispose of licensed material by incineration
only as 	specifically approved b_y the ~7flartment pursuant to
section 11-45-71. 	[Eff 	NO\! 	12 19~~ 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-74 	Reserved.
§11-45-75 	Transfer for disposal and manifests. 	The
transfer of low-level radioactive waste intended for disposal at a
low-level 	radioactive waste disposal 	facility, 	shipment manifests,
and certification of the waste generator, 	shall meet the
requirements of the Northwest Interstate Compact on Low-Level
Radioactive Waste Management. 	[Eff 	Nn\/ 1 ? 1090 	l 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	H~i' §§3~1-t', ~21-11 (21),
321-71)
§11-45-76 	Compliance with environmental and health
protection regulations. 	Nothing in subchapter 4 relieves 	the
licensee from complying with other applicable federal, 	state, 	and
local 	regulations governing any other toxic or hazardous
properties of materiiJ-~\Jh<Jlrfl~.R1' disposed of pursuant to
subchapter 4. 	[Eff NUv 	l l 	1::,H::, 	l 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11 (21), 	321-71)
§11-45-77 	Units and quantities on records. 	(a) 	Each
licensee shall use the SI units becquerel, 	gray, 	sievert and
45-72
Complete printed PDF page 72
(
§11-45-80
coulomb per kilogram, 	or the special units: 	curie, 	rad, 	rem, 	and
roentgen, 	including multiples and subdivisions, 	and shall clearly
indicate the units of all quantities on records 	required by this
subchapter.
(b) 	The licensee shall make a 	clear distinction among the
quantities entered on the records 	required by this subchapter such
as, 	total effective dose equivalent, 	shallow dose equivalent, 	eye
dose equivalent, 	deep dose equivalent, 	or committed effective dose
equivalent. 	[Eff NOV 1 ? ;agg 	l 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	FiRS §§:r21!...~, 	321-11 (21), 	321-71)
§11-45-78 	Records of radiation protection programs. 	(a)
Each licensee shall maintain records of the radiation protection
program, 	including:
(1) 	The provisions of the program; 	and
(2) 	Audits and other reviews of program content and
implementation.
(b) 	The licensee shall retain the records 	required by
paragraphs 	(1) 	and 	(2) 	for inspection by the department for three
years. 	[Eff 	NQV 1 2 j 	9 ] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-, 	321-11(21), 	321-71)
§11-45-79 	Records of surveys. 	(a) 	Each licensee shall
maintain records showing the results of surveys and calibrations
required by section 11-45-53. 	The licensee shall retain these
records 	for inspection by the department for three years.
(b) 	The licensee shall retain each of the following
records 	for inspection by the department for three years:
(1) 	Records of the results of surveys to determine the
dose from external sources of radiation used, 	in the
absence of or in combination with individual monitor-
ing data, 	in the assessment of individual dose
equivalents; 	and
(2) 	Records of the results of measurements and
calculations used to determine individual intakes of
radioactive material and used in the assessment of
internal dose; 	and
(3) 	Records showing the results of air sampling, 	surveys,
and bioassays required pursuant to section 11-45-
60 (a) (3) (A) 	and 	(B); 	and
(4) 	Records of the results of measurements and
calculations used to evaluate the release of
radioactive effluents to the environment.
[Eff NOV 12 1900 	l 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	H~s"'§§321-l, 	321-11(21), 	321-71)
§11-45-80 	Records of tests 	for leakage or contamination of
sealed sources. 	Records of tests 	for leakage or contamination of
sealed sources shall be kept in units of becquerel or curies and
maintained for inspection by the department for three years.
[Eff NQV 12 1999 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
45-73
2 2 6 5
Complete printed PDF page 73
§11-45-80
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-81 	Records of prior occupational dose. 	(a) 	The
licensee shall retain the records of prior occupational dose and
exposure history as specified in section 11-45-44 for inspection
by the department for three years.
(b) 	Upon termination of the license, 	the licensee shall
permanently store records. 	[Eff f\l flV 	1 9 1990 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 1 i'I~s §~.tl-L 	~1-11 (21), 	321-71)
§11-45-82 	Records of planned special exposures. 	(a) 	For
each use of the provisions of section 11-45-45 	for planned special
exposures, 	the licensee shall maintain records that describe:
(1) 	The exceptional circumstances requiring the use of a
planned special exposure; 	and
(2) 	The name of the management official who authorized the
planned special exposure and a copy of the signed
authorization; 	and
(3) 	What actions were necessary; 	and
(4) 	Why the actions were necessary; 	and
(5) 	What precautions were taken to assure that doses were
maintained ALARA; 	and·
(6) 	What individual and collective doses were expected to
result; 	and
(7) 	The doses actually received in the planned special
exposure.
(b) 	The licensee shall retain records 	for inspection by
the department for three years.
(c) 	Upon termination of tll,Ei~fe.Jl~, AAn licensee shall
permanently store records. 	[Eff 	l~UV 	1 l, 1::J:::J~ 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-83 	Records of individual monitoring results. 	(a)
Each licensee shall maintain records of doses 	received by all
individuals 	for whom monitoring was 	required pursuant to section
11-45-54, 	and records of doses 	received during planned special
exposures, 	accidents, 	and emergency conditions. 	Assessments of
dose equivalent and records made using units in effect before the
effective date of this chapter need not be changed. 	These records
shall include, 	when applicable:
(1) 	The deep dose equivalent to the whole body, 	eye dose
equivalent, 	shallow dose equivalent to the skin, 	and
shallow dose equivalent to the extremities; 	and
(2) 	The estimated intake or body burden of radionuclides;
and
(3) 	The committed effective dose equivalent assigned to
the intake or body burden of radionuclides; 	and
(4) 	The specific information used to calculate the
committed effective dose equivalent pursuant to
section 11-45-43(c) .; 	and
(5) 	The total effective dose equivalent when required by
45-74
65
Complete printed PDF page 74
section 11-45-41; 	and
(6) 	The total of the deep dose equivalent and the
committed dose to the organ receiving the highest
total dose.
(b) 	The licensee shall make entries of the records
specified in subsection 	(a) 	at least annually.
§11-45-87
(c) 	The licensee shall maintain the 	records specified in
subsection 	(a) .
(d) 	The licensee shall maintain the records of dose to an
embryo/fetus with the records of dose to the declared pregnant
woman. 	The declaration of pregnancy, 	including the estimated date
of conception, 	shall also be kept on file, 	but may be maintained
separately from the dose records.
(e) 	The licensee shall retain each required form or record
for 	inspection by the department for three years.
(f) 	Upon termination of tiyn1,~c~ne, 	t~~ licensee shall
permanently store records. 	[Eff 	l'HJV 	1 ,t; 19~ 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	3 1-11(21), 	321-71)
§11-45-84 	Records of dose to individual members of the
public. 	(a) 	Each licensee shall maintain records sufficient to
demonstrate compliance with the dose limit for individual members
of the public.
(b) 	The licensee shall retain the records 	required by
subsef(A~/ 	ia~ i~r Anspection by the department for three years.
[Eff r~U ,, 	1 '1 -,~g~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-85 	Records of waste disposal. 	(a) 	Each licensee
shall maintain records of the disposal of licensed materials made
pursuant to sections 11-45-71, 	11-45-72, 	and 11-45-73.
(b) 	The licensee shall retain the records 	required by this
sectio~dff jnspect~on by the department for three years.
[Eff 	. 	V l 2 199ij 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §$321-1, 	321-11(21), 	321-71)
§11-45-86 	Records of testing entry control devices for very
high radiation areas. 	(a) 	Each licensee shall maintain records
of tests made pursuant to section 11-45-57(b) (9). 	on entry control
devices 	for very high radiation areas. 	These records shall
include the date, 	time, 	and results of each such test of function.
(b) 	The license,.Ej rifflJ- 	retain the records 	required by
subsection 	(a). 	[Eff 	l'iU V 1 ? 1090 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §;321-t, ~21-11(21), 	321-71)
§11-45-87 	Form of records. 	Each record required by this
subchapter shall be legible. 	The record shall be the original or
a 	reproduced copy or a microform, 	provided that the copy or
microform is authenticated by authorized personnel and that the
microform is capable of producing a 	clear copy or the record may
45-75
Complete printed PDF page 75
§11-45-87
also be stored in electronic media with the capability for
producing legible, 	accurate, 	and complete records. 	Records, 	such
as 	letters, 	drawings, 	and specifications, 	shall include all
pertinent information, 	such as stamps, 	initials, 	and signatures.
The licensee shall maintain adequate safeguards against tampering
with and loss of records. 	[Eff Nfl\J 	1 ') 1090 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	h"Rs §§3'21-'f, ~21-11 (21), 	321-71)
§11-45-88 	Reports of stolen, 	or lost, 	or missing sources of
radiation. 	(a) 	Each licensee shall report to the department by
telephone as 	follows:
(1) 	Immediately after its occurrence becomes known to the
licensee, 	stolen, 	lost, 	or missing radioactive
material in an aggregate quantity equal to or greater
than one thousand times the quantity specified in
Appendix C under such circumstances that it appears to
the licensee that an exposure could result to
individuals in unrestricted areas; 	or
(2) 	Within thirty days after its occurrence becomes known
to the licensee, 	lost, 	stolen, 	or missing radioactive
material in an aggregate quantity greater than ten
times the quantity specified in Appendix C that is
still missing.
(3) 	Immediately after its occurrence becomes known to the
licensee, 	a 	stolen, 	lost, 	or missing radiation
machine.
(b) 	Each licensee required to make a 	report pursuant to
subsection 	(a) 	shall, 	within thirty days after making the
telephone report, 	make a written report to the department setting
forth the following information:
(1) 	A description of the source of radiation involved,
including, 	for radioactive material, 	the kind,
quantity, 	and chemical and physical 	form; 	and, 	for
radiation machines, 	the manufacturer, 	model and serial
number, 	type and maximum energy of radiation emitted;
(2) 	A description of the circumstances under which the
loss or theft occurred; 	and
(3) 	A statement of disposition, 	or probable disposition,
of the source of radiation involved; 	and
(4) 	Exposures of individuals to radiation, 	circumstances
under which the exposures occurred, 	and the possible
total effective dose equivalent to persons in
unrestricted areas; 	and
(5) 	Actions that have been taken, 	or shall be taken, 	to
recover the source of radiation; 	and
(6) 	Procedures or measures that have been, 	or shall be,
adopted to ensure against a 	recurrence of the loss or
theft of sources of radiation.
(c) 	Subsequent to filing the written report, 	the licensee
shall also report additional substantive information
on the loss or theft within thirty days 	after the
licensee learns of such information.
(d) 	The licensee shall prepare any report filed with the
45-76
6
Complete printed PDF page 76
§11-45-89
department pursuant to this section so that names of individuals
who may have 	received exposure to radiation are stated in a
separate and detachable portion of the report.
(e) 	The exhibit at the end of this chapter entitled,
"Appendix C of Subchapter 4, 	Quantities of Materials Requiring
Label~1,~\tlJ/~3
1
"gg:tr made a part of this section.
[Eff 	NUv 	l t, 	. 	1:::, 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-89 	Notification of incidents. 	(a) 	Notwithstanding
other requirements 	for notification, 	each licensee shall
immediately report each event involving a 	source of radiation
possessed by the licensee that may have caused or threatens to
cause any of the following conditions:
(1) 	An individual to receive:
(A) 	A total effective dose equivalent of twenty-five
rems 	(0.25 sievert) 	or more; 	or
(B) 	An eye dose equivalent of seventy-five rems
(0.75 sievert) 	or more; 	or
(C) 	A shallow dose equivalent to the skin or
extremities of two hundred fifty rads 	(2.5
grays) 	or more; 	or
(2) 	The release of radioactive material, 	inside or outside
of a 	restricted area, 	so that, 	had an individual been
present for twenty-four hours, 	the individual could
have received an intake five times the occupational
ALI.
(b) 	Each licensee shall, 	within twenty-four hours of
discovery of the event, 	report each event involving loss of
control of licensed source of radiation possessed by the licensee
that may have caused, 	or threatens to cause, 	any of the following
conditions:
(1) 	An individual to 	receive, 	in a period of twenty-four
hours:
(A) 	A total effective dose equivalent exceeding five
rems 	(0.05 	sievert); 	or
(B) 	An eye dose equivalent exceeding fifteen rems
(0.15 	sievert); 	or
(C) 	A shallow dose equivalent to the skin or
extremities exceeding fifty rems 	(0.5 sievert);
or
(2) 	The release of radioactive material, 	inside or outside
of a 	restricted area, 	so that, 	had an individual been
present for twenty-four hours, 	the individual could
have received an intake in excess of one occupational
ALI.
(c) 	The licensee shall prepare each report filed with the
department pursuant to this section so that names of individuals
who have received exposure to sources of radiation are stated in a
separate and detachable portion of the report.
(d) 	Licensees shall make the reports 	required by
subsections 	(a) 	and 	(b) 	by telephone, 	telegram, 	mailgram, 	or
facsimile to the department.
45-77
Complete printed PDF page 77
§11-45-89
(e) 	The provisions of this section do not apply to doses
that 	result from planned special exposures, 	provided such doses
are within the limits 	for planned special exp9it5,'tf~? f1W-
reported pursuant to section 11-45-91. 	[Eff 	l~U\f 	1 l,
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321- ,
11 (21), 	321-71)
§11-45-90 	Reports of exposures, 	radiation levels, 	and
concentrations of radioactive material exceeding the limits. 	(a)
In addition to the notification required by section 11-45-89,
each licensee shall submit a written report within thirty days
after learning of any of the following occurrences:
(1) 	Incidents 	for which notification is required by
section 11-45-89; 	or
(2) 	Doses in excess of any of the following:
(A) 	The occupational dose limits 	for adults in
section 11-45-40; 	or
(B) 	The occupational dose limits 	for a minor in
section 11-45-46; 	or
(C) 	The limits 	for an embryo/fetus of a declared
pregnant woman in section 11-45-47; 	or
(D) 	The limits 	for an individual member of the
public in section 11-45-48; 	or
(E) 	Any applicable limit in the license; 	or
(3) 	Levels of radiation or concentrations of radioactive
material in:
(A) 	A restricted area in excess of applicable limits
in the license; 	or
(B) 	An unrestricted area in excess of ten times the
applicable limit set forth in this subchapter or
in the license, 	whether or not involving
exposure of any individual in excess of the
limits in section 11-45-48; 	or
(4) 	For licensees subject to the provisions of U.S.
Environmental Protection Agency's generally applicable
environmental radiation standards in 40 C.F.R. 	Part
190, 	levels of radiation or releases of radioactive
material in excess of those standards, 	or of license
conditions related to those standards.
(b) 	Contents of reports.
(1) 	Each report required by subsection 	(a) 	shall describe
the extent of exposure of individuals to radiation and
radioactive material, 	including, 	as appropriate:
(A) 	Estimates of each individual's dose; 	and
(B) 	The levels of radiation and concentrations of
radioactive material involved; 	and
(C) 	The cause of the elevated exposures, 	dose rates,
or concentrations; 	and
(D) 	Corrective steps taken or planned to ensure
against a 	recurrence, 	including the schedule for
achieving conformance with applicable limits,
generally applicable environmental standards,
and associated license conditions.
45-78
Complete printed PDF page 78
§11-45-96
(2) 	Each report filed pursuant to subsection 	(a) 	shall
include for each individual exposed: 	the name, 	social
security account number, 	and date of birth. 	With
respect to the limit for the embryo/fetus in section
11-45-47, 	the identifiers should be those of the
declared pregnant woman. 	The report shall be prepared
so that this information is stated in a 	separate and
detachable portion of the report.
(c) 	All licensees who make reports pursuant to subsection
(a) 	shall submit the report in writing to the department.
[Eff 	Nnv 1 2 1090 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	Bis §t3~1-t, ~21-11(21), 	321-71)
§11-45-91 	Reports of planned special exposures. 	The
licensee shall submit a written report to the department within
thirty days 	following any planned special exposure conducted in
accordance with section 11-45-45, 	informing the department that a
planned special exposure was conducted and indicating the date the
planned special exposuret.l~PLF~ffi#'\rfhe information required by
section 11-45-82. 	[Eff Pl!UV 	l 	(., 	1::,::,::, 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-92 	Reserved.
§11-45-93 	Reserved.
§11-45-94 	Reserved.
§11-45-95 	Notifications and reports to individuals. 	(a)
Requirements 	for notification and reports to individuals of
exposure to radiation or radioactive material are specified in
subchapter 10.
(b) 	When a 	licensee is required pursuant to section 11-45-
90 	to 	report to the department any exposure of an individual to
radiation or radioactive material, 	the licensee shall also notify
the individual. 	Such notice shall be transmitted at a time not
later than the transmittal to the department, 1 ang shall comply
with the provisions of subchapter 10. 	[Eff p"QV 	1 ? 1090 	]
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§~1-'i, ~21-
11 (21), 	321-71)
§11-45-96 	Reports of leaking or contaminated sealed
sources. 	If the test for leakage or contamination indicates a
sealed source is leaking or contaminated, 	a 	report of the test
shall be filed within five days with the department describing the
equipment inv~J~~1, \the test results and the corrective action
taken. 	[Eff 	NU\/ 	? 1909 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS 	§32~-Y, 	321-11 (21), 	321-71)
45-79
Complete printed PDF page 79
§11-45-97
§11-45-97 	Vacating premises. 	Each licensee shall, 	no less
than thirty days before vacating or relinquishing possession or
control of premises which may have been contaminated with
radioactive material as a 	result of his activities, 	notify the
department in writing of intent to vacate. 	When deemed necessary
by the department, 	the licensee shall decontaminate the premises
in SUCW} rli pia~rwr A-A the department may specify.
[Eff 	1\!IJV 	.t l 	1::t::t9 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
SUBCHAPTER 5
INDUSTRIAL RADIOGRAPHIC OPERATIONS
§11-45-98 	Purpose and scope. 	This subchapter establishes
requirements 	for the use of sources of radiation for industrial
radiography. 	The requirements of this subchapter are in addition
to, 	and not in subst~t}\\io~ io5
9oAher applicable requirements in
this chapter. 	[Eff rJuv' 	l' 	I 9:; 	l 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-99 	Exemptions. 	(a) 	Sources of radiation licensed
by the U.S. 	Nuclear Regulatory Commission are exempt from the
requirementq of this subchapter to the extent the sources are
regulated by the U.S. 	Nuclear Regulatory Commission.
(b) 	Industrial uses of portable J:ight-intensified imaging
deviceNs ~;r.e exem!it f.rom the requirements in this part.
[Eff O
,ILJV 1 2 ,ago 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §t321-t, ~21-11(21), 	321-71)
§11-45-100 	Storage precautions. 	X-ray systems shall be
physically secured to frevent tampering or removal by unauthorized
personnel. 	[Eff 	Nn\ 	1 ? 1909 	l 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS s§32r-f, 	~_:11 (21), 	321-71)
§11-45-101 	Radiation survey instruments. 	(a) 	The licensee
shall maintain sufficient calibrated and operable radiation survey
instruments to make physical radiation surveys as 	required by this
subchapter and subchapter 4. 	Instrumentation required by this
section shall have a 	range such that two milliroentgens 	(5.16 x
10- 7 coulombs per kilogram) 	per hour through one roentgen 	(2.58 x
10- 4 coulombs per kilogram) 	per hour can be measured.
(b) 	Each radiation survey instrument shall be calibrated:
(1) 	At energies appropriate for use and at intervals not
to exceed three months and after each instrument
servicing;
(2) 	Such that accuracy within plus or minus twenty percent
can be demonstrated; 	and
45-80
Complete printed PDF page 80
§11-45-104
(3) 	At two points located approximately one-third and two-
thirds of full-scale on each scale for linear scale
instruments; 	at midrange of each decade, 	and at two
points of at least one decade for logarithmic scale
instruments; 	and at appropriate points for digital
instruments.
(c) 	Records of these calibrations shall be maintained after
the calibration date for inspection by the department for three
years.
(d) 	Each radiation survey instrument shall be checked with
a 	radiation source at the beginning of each day of use and at the
beginn\r,-}"\\'}f -·-·,,,,,.,.. 	shift to ensure it is operating properly.
[Eff 	l,UV 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-102 	Utilization logs. 	Each licensee shall maintain
current logs which shall be kept available for inspection by the
department for three years, 	showing for each x-ray system the
following information:
(1) 	A unique identification, 	such as a serial number, 	of
(2)
(3)
(4)
each x-ray system;
The identity of the radiographer to whom assigned;
Locations where used and dates of use; 	and
The date(s) 	each x-ray system 	~n5torage
and returned to storage. 	[Eff 	I~~~ 	]
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS
§§321-1, 	321-11(21), 	321-71)
§11-45-103 	Inspection and maintenance. 	(a) 	Each licensee
shall ensure that checks are made for obvious defects in x-ray
systems before use each day.
(b) 	Each licensee shall conduct a program of at least
quarterly inspection and maintenance of x-ray systems to assure
proper functioning of components important to safety. 	All
appropriate parts shall be maintained in accordance with
manufacturer's specifications. 	Records of inspection and
maintenance shall be maintained for inspection by the department
for three years.
(c) 	If any inspection reveals damage to components
critical to 	radiation safety, 	the x-ray system shall be removed
from service and labeled as defective until 	repairs have been
made.
(d) 	Inspections of x-ray systems utilized for non-
destructive testing shall be conducted by, 	or under the
superv~Jffi~ ~fh 	--:i:·~--fied medical physicist.
[Eff 	i\U V 1 it, 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321- , 	1-11 (21), 	321-71)
§11-45-104 	Permanent radiographic installations. 	Permanent
radiographic installations having high radiation area entrance
controls of the type described in subchapter 4 shall also meet the
45-81
Complete printed PDF page 81
§11-45-104
following
(1)
(2)
requirements:
Each entrance that is used for personnel access 	to the
high radiation area shall have both visible and
audible warning signals to warn of the presence of
radiation. 	The visible signal shall be activated by
radiation. 	The audible signal shall be activated when
an attempt is made to enter the installation during
exposure.
The control device or alarm system shall be tested for
proper operation at the beginning of each day of
equipment use. 	If a control device or alarm system is
operating improperly, 	it shall be immediately labeled
as defective and repaired before industrial
radiographic operations are resumed. 	Records of these
tests shall be maintained for inspection by the
department for three years. 	[Eff M 	1 ? 1aoa
(Auth: 	HRS §§321-10, 	321-11, 	321-11 	(ntrp'': 	~fl!
§§321-1, 	321-11(21), 	321-71)
§11-45-105 	Reserved.
§11-45-106 	Training and testing of radiographic personnel.
(a) 	No licensee shall permit any individual to act as a
radiographer trainee unless 	such individual has 	received copies
of, 	instructions in, 	and has demonstrated an understanding of:
(1) 	The subjects outlined in Appendix A of this
subchapter;
(2) 	This subchapter and subchapter 4;
(3) 	The appropriate license; 	and
(4) 	The licensee's operating and emergency procedures.
(b) 	No licensee shall permit any individual to act as a
radiographer, 	as defined in this part, 	until such individual:
(1) 	Has met the requirements of subsection 	(a);
(2) 	Has provided the department with documentation showing
completion of at least thirty days of on-the-job
training by a 	radiographer instructor as a
radiographer trainee following completion of the
requirements of subsection 	(a};
(3) 	Has demonstrated competence in the use of x-ray
systems, 	related handling tools, 	and radiation survey
instruments which may be employed in industrial
radiographic assignments; 	and
(4) 	Has demonstrated an understanding of the instructions
in subsection 	(a} 	by successful completion of a
written test and a 	field examination on the subjects
covered.
(c) 	Records of the above training, 	including copies of
written tests and dates of oral tests and field examinations,
shall be maintained by the licensee for inspection by the
department for three years 	following termination of employment.
(d) 	Each licensee shall conduct an internal audit program
to ensure that the department's license conditions or licensee's
45-82
Complete printed PDF page 82
§11-45-108
operating and emergency procedures are followed by each
radiographer. 	These internal audits shall be performed at least
quarterly, 	and each radiographer shall be audited at least
quarterly. 	Records of internal audits shall be maintained for
inspection by the department for three years.
(e) 	The exhibit at the end of chapter 11-45 entitled,
"Appendix A of Subchapter 5, 	Subjects for Instruction of
RadiogrfM:P.fr.tl T"'at·aww 	(2/2/93) ", 	is made a part of this section.
[Eff 	!~UV 	l '1 	~:,;::J 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-107 	Operating and emergency procedures. 	The
licensee's operating and emergency procedures shall include
instructions in at least the following:
(1) 	Handling and use of sources of radiation to be
employed such that no individual is likely to be
exposed to 	radiation doses in excess of the limits
established in subchapter 4;
(2) 	Methods and occasions 	for conducting radiation
(3)
(4)
(5)
(6)
(7)
(8)
surveys;
Methods 	for controlling access to radiographic areas;
Methods and occasions 	for locking and securing x-ray
systems;
Personnel monitoring and the use of personnel
monitoring equipment, 	including steps that shall be
taken immediately by radiography personnel in the
event a pocket dosimeter is found to be off-scale;
Notification procedures in the event of an accident;
Maintenance of records; 	and
The iniPA~tion
2
a~~~;intenance of x-ray systems.
[Eff 	NUV 	1 	l~:;:t: 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS 	§321-1, 	321-11(21), 	321-71)
§11-45-108 	Personnel monitoring control. 	(a) 	The licensee
shall not permit any individual to act as a 	radiographer or as a
radiographer trainee unless, 	at all times during radiographic
operations, 	the individual wears a direct reading pocket
dosimeter, 	an alarm ratemeter, 	and either a 	film badge or a
thermoluminescent dosimeter 	(TLD). 	Pocket dosimeters shall have a
range 	from zero to two hundred milliroentgens 	(5.16 x 10-5
coulombs per kilogram) 	and shall be recharged daily or at the
start of each shift. 	Each film badge or TLD shall be assigned to
and worn by only one individual.
(b) 	Pocket dosimeters shall be read and exposures 	recorded
at least once daily.
(c) 	Pocket dosimeters shall be checked for correct response
to 	radiation at periods not to exceed one year. 	Acceptable
dosimeters shall read within plus or minus thirty percent of the
true radiation exposure. 	Records of this check shall be
maintained for inspection by the department for three years.
(d) 	Each alarm ratemeter shall:
(1) 	Be checked to ensure that the alarm functions properly
45-83
Complete printed PDF page 83
§11-45-108
(sounds) 	before use at the start of each shift;
(2) 	Be set to give an alarm signal at a preset dose rate
of five hundred milliroentgens per hour;
(3) 	Require special means to change the preset alarm
function; 	and
(4) 	Be calibrated at periods not to exceed one year for
correct response to radiation. 	Acceptable ratemeters
shall alarm within plus or minus twenty percent of the
true radiation dose rate. 	Records of these
calibrations shall be maintained for inspection by the
department for three years.
(e) 	If an individual's pocket dosimeter is inoperable or
discharged beyond its range, 	industrial radiographic operations by
that individual shall cease and the individual's 	film badge or TLD
shall be processed immediately. 	The individual shall not return
to work with x-ray systems until a determination of the radiation
exposure has been made.
(f) 	Reports 	received from the film badge or TLD processor
and records of daily pocket dosimeter readings 	shall be kept for
inspection by the department for three years.
(g) 	If a 	film badge or TLD is lost or damaged, 	the worker
shall cease work immediately until a 	replacement film badge or TLD
is provided and the exposure is calculated for the time period
from i~~fpc~ ,to~A~l'f or damage of the film badge or TLD.
[Eff 	r,u V 1 l 	l!:f!:j:,j 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-109 	Security. 	During each radiographic operation,
the radiographer, 	radiographer instructor or radiographer trainee
shall maintain a direct surveillance of the operation to protect
against unauthorized entry into a high radiation area, 	as defined
in subchapter 1, 	except:
(1) 	Where the high radiation area is equipped with a
control device or alarm system as described in
subchapter 4, 	or
(2) 	Where the high radiation area is locked to protect
againsNt unaut.horized or accidental entry.
[Eff 	IQV 	12 109Q 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	H~S~§321-1, 	321-11 (21), 	321-71)
§11-45-110 	Posting. 	Areas in which radiography is being
performed l\iW\V f ?':I c.i'WA~icuously posted as 	required by subchapter
4. 	[Eff 	l~UV 	.. , 	l~~tj 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-111 	Radiation surveys and survey records. 	(a) 	No
radiographic operation shall be conducted unless calibrated and
operable radiation survey instrumentation is available and used at
each site where radiographic exposures are made.
(b) 	A physical radiation survey shall be made after each
radiographic exposure to determine that the machine is 	"off".
45-84
Complete printed PDF page 84
§11-45-113
(c) 	A survey with a 	radiation survey instrument shall be
made after each radiographic exposure to determine that x-ray
system is not operational. 	The area at and surrounding the
collimator of the x-ray system shall be surveyed.
(d) 	Records shall be kept for the required surveys. 	such
records shall be mairA~tn~d2f£~ i~spection by the department for
three years. 	[Eff f'iUV 	1 ., 	l::.t9 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	21-11 (21), 	321-71)
§11-45-112 	Documents and records 	required at temporary
jobsites. 	Each licensee conducting industrial radiography at a
temporary jobsite shall have the following records available at
that site for inspection by the department for three years:
(1) 	Copy of license;
(2) 	Operating and emergency procedures;
(3) 	Chapter 11-45;
(4) 	Survey records pursuant to subchapter 5 and area
survey records pursuant to subchapter 4 for the period
of operation at the site;
(5) 	Daily pocket dosimeter records 	for the period of
operation at the site; 	and
(6) 	The latest instrument calibration records 	for survey
instruments in use at the site. 	Acceptable records
include tags or labels 	affixed to the device
or survey meter. 	[Eff 	1009 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-7 	'M<.~ §§321-1, 	321-
11 (21), 	321-71)
§11-45-113 	Specific requirements 	for radiographic personnel
performing industrial radiography. 	(a) 	At a 	jobsite, 	the
following shall be supplied by the licensee:
(1) 	At least one operable, 	calibrated survey instrument;
(2) 	A current whole body personnel monitor 	(TLD or film
badge) 	for each individual;
(3) 	An operable, 	calibrated pocket dosimeter with a 	range
of zero to two hundred milliroentgens 	(5.16 x 10-5
coulombs per kilogram) 	for each worker; 	and
(4) 	The appropriate barrier ropes and signs.
(b) 	Industrial radiographic operations shall not be
performed if any of the items in subsection 	(a) 	are not available
at the jobsite or are inoperable.
(c) 	Each licensee shall provide a minimum of two
radiographic personnel when x-ray systems are used at temporary
jobsites. 	If one of the personnel is a 	radiographer trainee, 	the
other shall be a 	radiographer instructor.
(d) 	No individual other than a 	radiographer or a
radiographer trainee who is under the personal supervision of a
radiographer instructor shall manipulate controls or operate
equipment used in industrial 	radiographic operations.
(e) 	No individual shall act as a 	radiographer instructor
unless such individual:
(1) 	Has met the requirements of subchapter 5;
45-85
6
Complete printed PDF page 85
§11-45-113
(2) 	Has one year of documented experience as a
radiographer; 	and
(3) 	Has been named as a 	radiographer instructor on the
license issued by the department.
(f) 	During an inspection by the department, 	the department
inspector may terminate an operation if any of the items in
subsection 	(a) 	are not available and operable or if the required
number of radiographic personnel are not present. 	Operations
shall n2~,b~ ~e~~ until such conditions are met.
[Eff NUv 	1' 	l~!:j~ 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-114 	Special requirements and exemptions 	for cabinet
radiography. 	(a) 	Systems for cabinet radiography designed to
allow admittance of individuals shall:
(1) 	Comply with all applicable requirements of this
subchapter and subchapter 4. 	If such a 	system is a
certified cabinet x-ray system, 	it shall comply with
all applicable requirements of this subchapter and 21
C.F.R. 	§1020.40.
(2) 	Be evaluated at intervals not to exceed one year to
assure compliance with the applicable requirements of
this subchapter. 	Records of these evaluations shall
be maintained for inspection by the department for
three years.
(b) 	Certified cabinet x-ray systems designed to exclude
individuals 	from the interior of the cabinet are exempt from the
requirements of this subchapter except that:
(1) 	Operating personnel shall be provided with either a
film badge or a 	thermoluminescent dosimeter, 	and
reports of the results shall be maintained for
inspection by the department for three years.
(2) 	No licensee shall permit any individual to operate a
cabinet x-ray system until 	such individual has
received a 	copy of and instruction in the operating
procedures 	for the unit and has demonstrated compe-
tence in its use. 	Records which demonstrate
compliance with paragraph 	(1) 	shall be maintained for
inspection by the department for three years.
(3) 	Tests for proper operation of high radiation area
control devices or alarm systems, 	where applicable,
shall be conducted, 	recorded, 	and maintained in
accordance with section 11-45-104.
(4) 	The licensee shall perform an evaluation, 	at intervals
not to exceed one year, 	to determine conformance with
subchapter 4. 	If such a system is a 	certified cabinet
x-ray system, 	it shall be evaluated at intervals not
to exceed one year to determine conformance with 21
C.F.R. 	§1020.40. 	Records of these evaluations shall
be maintained for inspection by the department for
three years.
(c) 	Certified cabinet x-ray systems shall be maintained in
compliance with 21 C.F.R. 	§1020.40 unless prior approval has been
45-86
Complete printed PDF page 86
§11-45-116
granti1~~ ~r)~lf~ntment pursuant to subchapter 1.
[Eff 	i'tUV 	l 	~ 	l~:j~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
SUBCHAPTER 6
X-RAYS IN THE HEALING ARTS
§11-45-115 	Purpose and scope. 	This subchapter establishes
requirements 	for use of x-ray equipment in the healing arts. 	The
requirements of this subchapter are in addition to, 	and not in
subst~j}k~~OI}, f,pr~ ~~her applicable requirements of this chapter.
[Eff 	r1
~dV 	l, 	19~~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-116 	General requirements. 	(a) 	The licensee shall
be responsible for directing the operation of the x-ray system(s)
under the licensee's administrative control. 	The licensee or the
licensee's agent shall assure that the 	requirements of this
subchapter are met in the operation of the x-ray system(s).
(1) 	An x-ray system which does not meet the provisions of
this subchapter shall not be operated for diagnostic
purposes if so directed by the department.
(2) 	Individuals who will be operating x-ray systems shall
be adequately instructed on the operating procedures
and be competent in the safe use of the equipment.
(3) 	Individuals practicing radiologic technology shall be
licensed under chapter 11-44.
(4) 	A chart shall be provided in the vicinity of the
diagnostic x-ray system's control panel which
specifies, 	for all examinations performed with that
system, 	the following information:
(A) 	Patient's body part and anatomical 	size, 	or body
part thickness, 	or age 	(for pediatrics), 	versus
technique factors 	to be utilized;
(Bl 	Type and size of the film or screen-film
combination to be used;
(C) 	Type and focal distance of the grid to be used,
if any;
(D) 	Source-image receptor distance to be used
(except for dental intraoral radiography);
(El 	Type and location of placement of patient
shielding 	(e.g., 	gonad, 	etc.) 	to be used; 	and
(Fl 	For mammography, 	indication of kilovoltage
peak/target/filter combination.
(5) 	The licensee of a 	facility shall establish and make
available to x-ray operators written safety
procedures, 	including patient holding and any
restrictions of the operating technique required for
the safe operation of the particular x-ray system.
The operator shall be able to demonstrate familiarity
with these procedures.
45-87
Complete printed PDF page 87
§11-45-116
(6) 	Except for human patients who cannot be moved out of
the room, 	only the staff, 	ancillary personnel, 	or
other persons required for the medical procedure or
training shall be in the room during the radiographic
exposure. 	Other than the patient being examined:
(A) 	All individuals shall be positioned such that no
part of the body will be struck by the useful
beam unless protected by not less than 0.5
millimeter lead equivalent material or otherwise
authorized by a 	qualified medical physicist.
(B) 	The x-ray operator, 	other staff, 	ancillary
personnel and other persons required for the
medical procedure shall be protected from the
direct scatter radiation by protective aprons or
whole body protective barriers of not less than
0.25 millimeter lead equivalent material, 	unless
otherwise authorized by a qualified medical
physicist.
(C) 	Human patients who cannot be removed from the
room shall be protected from the direct scatter
radiation by whole body protective barriers of
not less than 0.25 millimeter lead equivalent
material, 	or shall be so positioned that the
nearest portion of the body is at least two
meters 	from both the tube head and the nearest
edge of the image receptor, 	unless otherwise
authorized by a 	qualified medical physicist.
(7) 	Gonad shielding of not less than 0.5 millimeter lead
equivalent material shall be used for human patients,
who have not passed the reproductive age, 	during
radiographic procedures in which the gonads are in the
useful beam, 	except for cases in which this would
interfere with the diagnostic procedure.
(8) 	Individuals shall not be exposed to the useful beam
except for healing arts purposes and unless such
exposure has been authorized by a 	licensed
practitioner of the healing arts. 	This provision
specifically prohibits deliberate exposure for the
following purposes:
(A) 	Exposure of an individual 	for training,
demonstration, 	or other non-healing arts
purposes; 	and
(B) 	Reserved.
(9) 	When a patient or film requires auxiliary support
during a 	radiation exposure:
(A) 	Mechanical holding devices 	shall be used. 	The
written safety procedures, 	required by paragraph
(5), 	shall list individual projections where
holding devices cannot be utilized;
(B) 	Written safety procedures, 	as required by
paragraph 	(5), 	shall indicate the 	requirements
for 	selecting a holder and the procedure the
holder shall follow;
(C) 	The human holder shall be instructed in personal
45-88
5
Complete printed PDF page 88
§11-45-116
radiation safety and protected as 	required by
paragraph 	(6);
(D) 	No individual shall be used routinely to hold
film or human patients;
(E) 	In those cases where an individual must hold the
film 	(except during intraoral examinations), 	any
portion of the body other than the area of
clinical interest struck by the useful beam
shall be protected by not less than 0.5
millimeter lead equivalent material; 	or
(F) 	Each facility shall have leaded aprons and
gloves available in sufficient numbers to
provide protection for all personnel who are
involved with x-ray operations and who are
otherwise not shielded.
(10) 	Procedures and auxiliary equipment designed to
minimize patient and personnel exposure commensurate
with the needed diagnostic information shall be
utilized.
(A) 	The speed of the screen and film combinations
used shall be the fastest speed consistent with
the diagnostic objective of the examinations.
Film cassettes without intensifying screens
shall not be used for any routine diagnostic
radiological imaging, 	with the exception of
standard film packets 	for intraoral use in
dental radiography.
(B) 	The radiation exposure to the patient shall be
the minimum exposure required to produce images
of good diagnostic quality.
(C) 	Portable or mobile x-ray equipment shall be used
only for examinations where it is impractical to
transfer the patient(s) 	to a 	stationary x-ray
installation.
(D) 	If grids are used between the patient and the
image receptor to decrease scatter to the film
and improve contrast the grid shall:
(i) 	Be positioned properly, 	i.e., 	tube side
facing the right direction, 	and grid
centered to the central ray; 	and
(ii) 	If of the focused type, 	be of the proper
focal distance for the SID's being used.
(11) 	All individuals who are associated with the operation
of an x-ray system are subject to the requirements of
subchapter 4.
(12) 	Reserved.
(13) 	The licensee shall maintain the following information
for each x-ray system for inspection by the
department:
(A) 	Model and serial numbers of all major
components, 	and user's manuals 	for those
components;
(B) 	Tube rating charts and cooling curves;
(C) 	Records of surveys, 	calibrations, 	maintenance,
45-89
265
Complete printed PDF page 89
§11-45-116
and modifications performed on the x-ray
system(s);
(D) 	Copy of all correspondence with this department
regarding that x-ray system.
(14) 	Each facility shall maintain either an x-ray log
containing the name of the patient, 	the type of x-ray
examination, 	the date of the x-ray examination, 	the
name of the referring practitioner, 	and the name of
the individual who performed the x-ray examination
(facilities subject to chapter 11-44), 	or a patient
record specifying the type of x-ray examination and
the date of the x-ray examination. 	When the patient
or film must be provided with human auxiliary support,
the name of the human holder shall be recorded.
(b) 	Each installation using a 	radiographic x-ray system
and using analog image receptors 	(radiographic film) 	shall have
available suitable equipment for handling and processing
radiographic film in accordance with the following provisions:
(1) 	Manually developed film:
(A) 	Processing tanks shall be constructed of
mechanically rigid, 	corrosion resistant
material; 	and
(B) 	The temperature of solutions in the tanks shall
be maintained within the range of sixty degrees
Fahrenheit to eighty degrees Fahrenheit 	(sixteen
degrees Celsius to twenty-seven degrees
Celsius). 	Film shall be developed in accordance
with the time-temperature relationships
recommended by the film manufacturer, 	or, 	in the
absence of such recommendations, 	with Appendix A
of this 	subchapter.
(C) 	Devices shall be utilized which will indicate
the actual temperature of the developer and
signal the passage of a preset time appropriate
to the developing time required.
(2) 	Automatic processors and other closed processing
systems:
(A) 	Films shall be developed in accordance with the
time-temperature relationships 	recommended by
the film manufacturer; 	in the absence of such
recommendations, 	the film shall be developed
according to Appendix B of this subchapter.
(B) 	The specified developer temperature and
immersion time shall be posted in the darkroom
or on the automatic processor.
(3) 	Processing deviations 	from the requirements of
paragraph 	(1) 	shall be documented by the licensee in
such manner that the requirements are shown to be met
or exceeded 	(e.g., 	extended processing, 	and special
rapid chemistry).
(c) 	Other requirements:
(1) 	Pass boxes, 	if provided, 	shall be so constructed as to
exclude light from the darkroom when cassettes are
placed in or removed from the boxes, 	and shall
45-90 	6
Complete printed PDF page 90
§11-45-117
incorporate adequate shielding from stray radiation to
prevent exposure of undeveloped film.
(2) 	The darkroom shall be light tight and use proper
safelighting such that any film type in use exposed in
a 	cassette to x-radiation sufficient to produce an
optical density from 1.0 to 2.0 when processed shall
not have a 	fog density greater than 0.1 	(0.05 	for
mammography) 	when exposed in the darkroom for two
minutes with all safelights on. 	If used, 	daylight
film handling boxes shall preclude fogging of the
film.
(3) 	Darkrooms typically used by more than one individual
shall be provided a method to prevent accidental entry
while undeveloped films 	are being handled or
processed.
(4) 	Film shall be stored in a 	cool, 	dry place and shall be
protected from exposure to stray radiation. 	Film in
open packages shall be stored in a 	light tight
container.
(5) 	Film cassettes and intensifying screens shall be
inspected periodically and shall be cleaned and
replaced as necessary to best assure radiographs of
good diagnostic quality.
(6) 	Outdated x-ray film shall not be used for human
diagnostic radiographs, 	unless the film has been
stored in accordance with the manufacturer's
recommendations and a sample of the film passes a
sensitometric test for normal ranges of base plus fog
and speed.
(7) 	Film developing solutions shall be prepared in
accordance with the directions given by the
manufacturer, 	and shall be maintained in strength by
replenishment or renewal so that full development is
accomplished within the time specified by the
manufacturer.
(d) 	The exhibits at the end of chapter 11-45 entitled,
"Appendix A of Subchapter 6, 	Time-Temperature Chart 	(2/2/93)" and
"Appendix B of Subchapter 6, 	Temperature-Immersion Chart
(2/2/93i •:_.,. ,are made a part of this section.
[Eff 	NUV 12 1990 ] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-~, 	t21-11(21), 	321-71)
§11-45-117 	All diagnostic x-ray systems. 	(a) 	The control
panel containing the main power switch shall bear the warning
statement, 	legible and accessible to view: 	"WARNING: 	This x-ray
unit may be dangerous 	to patient and operator unless safe exposure
factors 	and operating instructions are observed."
(b) 	On battery-powered x-ray generators, 	visual means
shall be provided on the control panel to indicate whether the
battery is in a 	state of charge adequate for proper operation.
(c) 	The leakage radiation from the diagnostic source
assembly measured at a 	distance of one meter in any direction from
the source shall not exceed one hundred milliroentgens 	(25.8
45-91 	2 	5
Complete printed PDF page 91
§11-45-117
microcoulombs per kilogram) 	in one hour when the x-ray tube is
operated at its leakage technique factors. 	Compliance shall be
determined by measurements averaged over an area of one hundred
square centimeters with no linear dimension greater than
twenty centimeters.
(d) 	The radiation emitted by a component other than the
diagnostic source assembly shall not exceed two milliroentgens
(0.516 microcoulombs per kilogram) 	in one hour at five centimeters
from any accessible surface of the component when it is 	operated
in an assembled x-ray system under any conditions 	for which it was
designed. 	Compliance shall be determined by measurements averaged
over an area of one hundred square centimeters with no 	linear
dimension greater than twenty centimeters.
(e) 	Beam quality:
(1) 	The half-value layer of the useful beam for a 	given x-
ray tube potential shall not be less than the
appropriate value specified in Appendix C of this
subchapter under "specified dental systems," 	for any
dental system designed for use with intraoral image
receptors and manufactured after December 1, 	1980; 	and
under "other x-ray systems," for all other x-ray
systems subject to this section. 	If it is necessary
to determine such half-value layer at an x-ray tube
potential which is not listed in Appendix C, 	linear
interpolation or extrapolation may be made. 	Positive
means shall be provided to insure that at least the
minimum filtration needed to achieve the above beam
quality requirements is in the useful beam during each
exposure.
(2) 	For capacitor energy storage equipment, 	compliance
with the requirements of paragraph 	(1) 	shall be
determined with the maximum quantity of charge per
exposure. 	This shall be deemed to have been met if an
milliamperes-second of ten has been used.
(3) 	The required minimal half-value layer of the useful
beam shall include the 	filtration contributed by all
materials which are permanently between the 	source and
the patient.
(4) 	For x-ray systems which have variable kilovoltage peak
and variable filtration 	for the useful beam, 	a device
shall link the kilovolts peak selector with the
filter(s) 	and shall prevent an exposure unless the
minimum amount of filtration necessary to produce the
half-value layer required by paragraph 	(1) 	is in the
useful beam for the given kilovolts peak which has
been selected.
(f) 	Where two or more radiographic tubes are controlled by
one exposure switch, 	the tube or tubes which have been selected
shall be clearly indicated before initiation of the exposure.
This indication shall be both on the x-ray control panel and at or
near the tube housing assembly which has been selected.
(g) 	The tube housing assembly supports 	shall be adjusted
such that the tube housing assembly shall remain stable during an
exposure unless tube housing movement is a designed function of
45-92 6
Complete printed PDF page 92
§11-45-118
the x-ray
(h)
(1)
system.
Technique indicators:
The technique factors to be used during an exposure
shall be indicated before the exposure begins. 	If
automatic exposure controls are used, 	the technique
factors which are set before the exposure shall be
indicated.
(2) 	The requirement of paragraph 	(1) 	may be met by
permanent markings on equipment having fixed technique
factors. 	Indication of technique factors 	shall be
visible from the operator's position except in the
case of spot films made by the fluoroscopist.
(i) 	Diagnostic x-ray systems and their associated
components used on humans and certified pursuant to the 	federal x-
ray equipment performance standard, 	21 C.F.R. 	Part 1020, 	shall be
maintained in compliance with applicable requirements of that
standard and this subchapter.
(j) 	All position locking, 	holding, 	and centering devices
on x-ray system components and systems shall function as intended.
(k) 	The exhibit at the end of chapter 11-45 entitled,
"Appendix C of Subchapter 6, 	kyVa\un !Mnn (7 /25/96) ", 	is made
a part of this 	section. 	[Eff 	U 	1 t 	I~~~ 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-118 	Fluoroscopic x-ray systems. 	(a) 	All
fluoroscopic x-ray systems shall be image intensified or used in
conjunction with a 	digital imaging system.
(b) 	The useful beam shall be limited as 	follows:
(1) 	Primary barrier:
(A) 	The fluoroscopic 	imaging assembly shall be
provided with a primary protective barrier which
intercepts the entire cross section of the
useful beam at any source-image receptor
distance; 	and
(B) 	The x-ray tube used for 	fluoroscopy shall not
produce x-rays unless the barrier is in position
to intercept the entire useful beam.
(2) 	Fluoroscopic beam limitation:
(A) 	For certified fluoroscopic systems with or
without a 	spot film device, 	neither the length
nor the width of the x-ray field in the plane of
the image receptor shall exceed that of the
visible area of the image receptor by more than
three percent of the source-image receptor
distance. 	The sum of the excess length and the
excess width shall be no greater than four
percent of the source-image receptor distance.
(B) 	For uncertified fluoroscopic systems with a spot
film device, 	the x-ray beam with the shutters
fully opened 	(during fluoroscopy or spot
filming) 	shall be no larger than the largest
spot film size 	for which the device is designed.
Measurements shall be made at the minimum
45-93
6
Complete printed PDF page 93
§11-45-118
source-image receptor distance available but at
no less than twenty centimeters table top to the
film plane distance.
(C) 	For uncertified fluoroscopic systems without a
spot film device, 	the requirements of
subparagraph 	(A) 	apply.
(D) 	Other requirements 	for fluoroscopic beam
limitation:
(i) 	Means shall be provided to permit further
limitation of the field. 	Beam-limiting
devices manufactured after May 22, 	1979,
and incorporated in equipment with a
variable source-image receptor distance
and/or a visible area of greater than
three hundred square centimeters shall be
provided with means 	for stepless
adjustment of the x-ray field;
(ii) 	All equipment with a 	fixed source-image
receptor distance and a visible area of
three hundred square centimeters or less
shall be provided with either stepless
adjustment of the x-ray field or with
means to 	further limit the x-ray field
size at the plane of the image receptor to
one hundred twenty-five square centimeters
or less;
(iii) 	If provided, 	stepless adjustment shall, 	at
the greatest source-image receptor
distance, 	provide continuous 	field sizes
from the maximum attainable to a 	field
size of five centimeters by five
centimeters or less;
(iv) 	For equipment manufactured after February
25, 	1978, 	when the angle between the image
receptor and beam axis 	is variable, 	means
shall be provided to indicate when the
axis of the x-ray beam is perpendicular to
the plane of the image receptor; 	and
(v) 	For rectangular x-ray fields used with
circular image reception, 	the error in
alignment shall be determined along the
length and width dimensions of the x-ray
field which pass through the center of the
visible area of the image receptor.
(E) 	Spot-film devices which are certified components
shall meet the following additional
requirements:
(i) 	Means shall be provided between the source
and the patient for adjustment of the x-
ray field size in the plane of the film to
the size of that portion of the film which
has been selected on the spot film
selector. 	Such adjustment shall be
automatically accomplished except when the
45-94
265
Complete printed PDF page 94
§11-45-118
x-ray field size in the plane of the film
is smaller than that of the selected
portion of the film. 	For spot 	film
devices manufactured after June 21, 	1979,
if the x-ray field size is less than the
size of the selected portion of the film,
the means 	for adjustment of the field size
shall be only at the operator's 	option;
(ii) 	It shall be possible to adjust the x-ray
field size in the plane of the 	film to a
size smaller than the selected portion of
the film. 	The minimum field size at the
greatest source-image receptor distance
shall be equal to, 	or less than, 	five
centimeters by five centimeters;
(iii) 	The center of the x-ray field in the plane
of the film shall be aligned with the
center of the selected portion of the film
to within two percent of the source-image
receptor distance; 	and
(iv) 	on spot-film devices manufactured after
February 25, 	1978, 	if the angle between
the plane of the image receptor and beam
axis is variable, 	means shall be provided
to indicate when the axis of the x-ray
beam is perpendicular to the plane of the
image receptor, 	and compliance shall be
determined with the beam axis indicated to
be perpendicular to the plane of the image
receptor.
(F) 	If a means exists to override any of the
automatic x-ray field size adjustments 	required
in subparagraph 	(E), 	that means:
(i) 	Shall be designed for use only in the
event of system failure;
(ii) 	Shall incorporate a signal visible at the
fluoroscopist's position which shall
indicate whenever the automatic field size
(iii)
adjustment is overridden; 	and
Shall be clearly and durably labeled as
follows:
FOR X-RAY FIELD
LIMITATION SYSTEM FAILURE
(c) 	X-ray production in the fluoroscopic mode shall be
controlled by a 	device which requires continuous pressure by the
fluoroscopist 	for the entire time of any exposure. 	When recording
serial 	fluoroscopic images, 	the fluoroscopist 	shall be able to
terminate the x-ray exposure(s) 	at any time, 	but means may be
provided to permit completion of any single exposure of the series
in process.
(d)
(1)
Entrance exposure rate allowable limits:
Fluoroscopic equipment which is provided with
45-95
Complete printed PDF page 95
§11-45-118
automatic exposure rate control shall not be operable
at any combination of tube potential and current which
shall result in an exposure rate in excess of ten
roentgens 	(2.58 millicoulombs per kilogram) 	per minute
at the point where the center of the useful beam
enters the patient, 	except:
(A) 	During recording of fluoroscopic images, 	or
(B) 	When an optional high level control is provided.
When so provided, 	the equipment shall not be
operable at any combination of tube potential
and current which shall 	result in an exposure
rate in excess of five 	roentgens 	(1.29
millicoulombs per kilogram) 	per minute at the
point where the center of the useful beam enters
the patient unless the high level control is
activated. 	Special means of activation of high
level controls shall be required. 	The high
level control shall only be operable when
continuous manual activation is provided by the
operator. 	A continuous signal audible to the
fluoroscopist 	shall indicate that high level
control is being employed.
(2) 	Fluoroscopic equipment which is not provided with
automatic exposure rate control 	(manual mode) 	shall
not be operable at any combination of tube potential
and current which shall result in an exposure rate in
excess of five 	roentgens 	(1.29 millicoulombs per
kilogram) 	per minute at the point where the center of
the useful beam enters the patient, 	except:
(A) 	During recording of fluoroscopic images, 	or
(B) 	When an optional high level control is
activated. 	Special means of activation of high
level controls shall be required. 	The high
level control shall only be operable when
continuous manual activation is provided by the
operator. 	A continuous signal audible to the
fluoroscopist 	shall indicate that the high level
control is being employed.
(3) 	Fluoroscopic equipment which is provided with both
automatic exposure rate control mode and a manual mode
shall not be operable at any combination of tube
potential and current which shall 	result in an
exposure rate in excess of ten roentgens 	(2.58
millicoulombs per kilogram) 	per minute in either mode
at the point where the center of the useful beam
enters the patient, 	except:
(A) 	During recording of fluoroscopic 	images; 	or
(B) 	When the mode or modes have an optional high
level control, 	in which case that mode or modes
shall not be operable at any combination of tube
potential and current which shall result in an
exposure rate in excess of five 	roentgens 	(1.29
millicoulombs per kilogram) 	per minute at the
point where the center of the useful beam enters
45-96
Complete printed PDF page 96
§11-45-118
the patient, 	unless the high level control is
activated. 	Special means of activation of high
level controls shall be required. 	The high
level control shall only be operable when
continuous manual activation is provided by the
operator. 	A continuous signal audible to the
fluoroscopist shall indicate that the high level
control is being employed.
(4) 	Any fluoroscopic equipment manufactured after May 19,
1995 which can exceed five 	roentgens 	(1.29
millicoulombs per kilogram) 	per minute shall be
equipped with an automatic exposure rate control. 	All
entrance exposure rate limits shall be ten roentgens
(2.58 millicoulombs per kilogram) 	per minute with an
upper limit of twenty roentgens 	(5.16 millicoulombs
per kilogram) 	per minute when high level 	control is
activated.
(5) 	Compliance with the requirements of this 	subsection
shall be determined as 	follows:
(A) 	If the source is below the x-ray table, 	the
exposure rate shall be measured one centimeter
above the tabletop or cradle.
(B) 	If the source is above the x-ray table, 	the
exposure rate shall be measured at thirty
centimeters above the tabletop with the end of
the beam-limiting device or spacer positioned as
closely as possible to the point of measurement.
(C) 	For a c-arm type of fluoroscope, 	the exposure
rate shall be measured thirty centimeters from
the input surface of the fluoroscopic imaging
assembly, 	with the source positioned at any
available SID, 	provided that the end of the
beam-limiting device or spacer is no closer than
thirty centimeters from the input surface of the
fluoroscopic imaging assembly.
(D) 	For a 	lateral type fluoroscope, 	the 	exposure
rate shall be measured at a point fifteen
centimeters from the centerline of the x-ray
table and in the direction of the x-ray source
with the end of the beam-limiting device or
spacer positioned as closely as possible to the
point of measurement. 	If the tabletop is
movable, 	it shall be positioned as closely as
possible to the lateral x-ray source, 	with the
end of the beam-limiting device or spacer no
closer than fifteen centimeters to the
centerline of the x-ray table.
(6) 	Periodic measurement of entrance exposure rate shall
be performed as 	follows:
(A) 	Such measurements shall be made by, 	or under the
supervision of, 	a 	qualified medical physicist at
least annually or after any maintenance of the
system which might affect the exposure rate.
(B) 	Results of these measurements shall be posted
45-97
Complete printed PDF page 97
§11-45-118
where any fluoroscopist may have ready access to
such results while using the fluoroscope and in
the record required in section 11-45-116
(a) (13) (C). 	The measurement results 	shall be
stated in roentgens 	(coulombs per kilogram) 	per
minute and include the technique factors used in
determining such results. 	The name of the
individual performing the measurements and the
date the measurements were performed shall be
included in the results.
(C) 	Conditions of periodic measurement of maximum
entrance exposure rate are as 	follows:
(i) 	The measurement shall be made under the
conditions that satisfy the 	requirements
of this subsection;
(ii) 	The kilovolts peak and milliamperes shall
be typical of clinical use of the x-ray
system;
(iii) 	The x-ray system(s) 	that incorporates
automatic exposure rate control shall have
an appropriate phantom placed in the
useful beam to produce a milliamperage
and/or kilovoltage typical of the use of
the x-ray system; 	and
(e) 	Barrier transmitted radiation rate limits:
(1) 	The exposure rate due to transmission through the
primary protective barrier with the attenuation block
in the useful beam, 	combined with radiation from the
image intensifier, 	if provided, 	shall not exceed two
milliroentgens 	(0.516 microcoulombs per kilogram) 	per
hour at ten centimeters from any accessible surface of
the fluoroscopic imaging assembly beyond the plane of
the image receptor for each roentgen per minute of
entrance exposure rate.
(2) 	Measuring compliance of barrier transmission.
(A) 	The exposure rate due to transmission through
the primary protective barrier combined with
radiation from the image intensifier shall be
determined by measurements averaged over an area
of one hundred square centimeters with no linear
dimension greater than twenty centimeters.
(B) 	If the source is below the tabletop, 	the
measurement shall be made with the input surface
of the fluoroscopic imaging assembly positioned
thirty centimeters above the tabletop.
(C) 	If the source is above the tabletop and the
source-image receptor distance is variable, 	the
measurement shall be made with the end of the
beam-limiting device or spacer as close to the
tabletop as it can be placed, 	provided that it
shall not be closer than thirty centimeters.
(D) 	Movable grids and compression devices shall be
removed from the useful beam during the
measurement.
45-98
Complete printed PDF page 98
(f) 	During fluoroscopy and cinefluorography the tube
potential and tube current shall be continuously indicated.
§11-45-118
(g) 	The source-to-skin distance shall not be less than:
(1) 	Thirty-eight centimeters on stationary fluoroscopic
systems manufactured on or after August 1, 	1974,;
(2) 	A distance of 35.5 centimeters on stationary
fluoroscopic systems manufactured before August 1,
1974;
(3) 	Thirty centimeters on all mobile and portable
fluoroscopes; 	and
(4) 	Twenty centimeters for all image-intensified
fluoroscopes used for specific surgical application.
(h) 	Fluoroscopic timer:
(1) 	Means shall be provided to preset the cumulative on-
time of the fluoroscopic x-ray tube. 	The maximum
cumulative time of the timing device shall not exceed
five minutes without resetting.
(2) 	A signal audible to the fluoroscopist 	shall indicate
the completion of any preset cumulative on-time. 	Such
signal shall continue to sound while x-rays are
produced until the timing device is reset.
(i) 	Control of scattered radiation:
(1) 	Fluoroscopic table designs when combined with
procedures utilized shall be such that no unprotected
part of any staff or ancillary individual's body shall
be exposed to unattenuated scattered radiation which
originates from under the table. 	The attenuation
required shall be not less than 0.25 millimeter lead
equivalent.
(2) 	Equipment configuration when combined with procedures
shall be such that no portion of any staff or
ancillary individual's body, 	except the extremities,
shall be exposed to the unattenuated scattered
radiation emanating from above the tabletop unless
that individual:
(A) 	Is at least one hundred twenty centimeters from
the center of the useful beam; 	or
(B) 	The radiation has passed through not less than
0.25 millimeter lead equivalent material
including, 	but not limited to, 	drapes, 	bucky-
slot cover panel, 	or self-supporting curtains,
in addition to any lead equivalency provided by
the protective apron.
(j) 	Fluoroscopic systems equipped with spot film
(radiographic) 	mode shall meet the exposure reproducibility
requirements of section 11-45-119 when operating in the spot film
mode.
(k) 	Radiation therapy simulation systems shall be exempt
from all the requirements of subsections 	(b), 	(d), 	(e), 	and 	(h)
provided that:
(1) 	Such systems are designed and used in such a manner
that no individual other than the patient is in the x-
ray room during periods of time when the system is
producing x-rays; 	and
45-99
65
Complete printed PDF page 99
§11-45-118
(2) 	Systems which do not meet the requirements of
subsection 	(h) 	are provided with a means of indicating
the cumulative time that an individual patient has
been exposed to x-rays. 	Procedures shall require in
such cases that the timer be reset between
examinations.
(l) 	All newly-assembled, 	installed, 	or reassembled
fluoroscopic x-ray systems shall be evaluated by, 	or under the
supervision of, 	a 	qualified medical physicist before use on human
patien»rt\I}l~St, <:iHwnwise authorized by the director.
[Eff 	UV 	1 {., 	I~~::, J 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-119 	Radiographic systems other than fluoroscopic,
dental intraoral, 	or computed tomography systems. 	(a) 	The useful
beam shall be limited to the area of clinical interest. 	This
shall be deemed to have been met if a positive beam limiting
device has been properly used or if evidence of collimation is
shown on at least three sides or three corners of the film, 	(e.g.,
projections from the shutters of the collimator, 	cone cutting at
the corners, 	or borders at the film's 	edge).
(1) 	General purpose stationary, 	mobile, 	and portable x-ray
systems, 	including veterinary systems installed after
the effective date of this chapter:
(A) 	Means shall be provided for independent stepless
adjustment of the size of the x-ray field. 	Each
dimension of the minimum field size at an
source-image receptor distance of one hundred
centimeters shall be equal to or less than five
centimeters.
(B) 	A method shall be provided for visually defining
the perimeter of the x-ray field. 	The total
misalignment of the edges of the visually
defined field with the respective edges of the
x-ray field along either the length or width of
the visually defined field shall not exceed two
percent of the distance from the source to the
center of the visually defined field when the
surface upon which it appears 	is perpendicular
to the axis of the x-ray beam.
(2) 	In addition to the requirements of paragraph 	(1), 	all
stationary general purpose x-ray systems shall meet
the 	following requirements:
(A) 	A method shall be provided to indicate when the
axis of the x-ray beam is perpendicular to the
plane of the image receptor, 	to align the center
of the x-ray field with respect to the center of
the image receptor to within two percent of the
source-image receptor distance, 	and to indicate
the source-image receptor distance to within two
percent;
(B) 	The beam-limiting device shall indicate
numerically the field size in the plane of the
45-100
Complete printed PDF page 100
§11-45-119
image receptor to which it is adjusted; 	and
(C) 	Indication of field size dimensions and source-
image receptor distances shall be specified in
inches and/or centimeters, 	and shall be such
that aperture adjustments 	result in x-ray field
dimensions in the plane of the image receptor
which correspond to those indicated by the
beam-limiting device to within two percent of
the source-image receptor distances when the
beam axis is indicated to be perpendicular to
the plane of the image receptor.
(3) 	Radiographic equipment designed for only one image
receptor size at a 	fixed source-image receptor
distances shall be provided with means to limit the
field at the plane of the image receptor to dimensions
no greater than those of the image receptor, 	and to
align the center of the x-ray field with the center of
the image receptor to within two percent of the
source-image receptor distance, 	or shall be provided
with means to both size and align the x-ray field such
that the x-ray field at the plane of the image
receptor does not extend beyond any edge of the image
receptor.
(4) 	Radiographic systems other than those designated in
paragraphs 	(1), 	(2), 	and 	(3), 	and veterinary systems
installed before the effective date of.this chapter:
(A) 	Means shall be provided to limit the x-ray field
in the plane of the image receptor so that such
field does not exceed each dimension of the
image receptor by more than two percent of the
source-image receptor distance when the axis of
the x-ray beam is perpendicular to the plane of
the image receptor.
(B) 	Means shall be provided to align the center of
the x-ray field with the center of the image
receptor to within two percent of the source-
image receptor distance, 	or means shall be
provided to both size and align the x-ray field
such that the x-ray field at the plane of the
image receptor does not extend beyond any edge
of the image receptor. 	Compliance shall be
determined with the axis of the x-ray beam
perpendicular to the plane of the image
receptor.
(C) 	The requirements of subparagraphs 	(A) 	and 	(B)
may be met with a 	system that meets the
requirements 	for a 	general purpose x-ray system
as specified in paragraph 	(1) 	or, 	when alignment
means are also provided, 	may be met with either:
(i) 	An assortment of removable,
fixed-aperture, 	beam-limiting devices
sufficient to meet the 	requirement for
each combination of image receptor size
and source-image receptor distance for
45-101
2 	5
Complete printed PDF page 101
§11-45-119
which the unit is designed with each such
device having clear and permanent markings
to indicate the image receptor size and
source-image receptor distance for which
it is designed; 	or
(ii) 	A beam-limiting device having multiple
fixed apertures sufficient to meet the
requirement for each combination of image
receptor size and source-image receptor
distance for which the unit is designed.
Permanent, 	clearly legible markings shall
indicate the image receptor size and
source-image receptor distance for which
each aperture is designed and shall
indicate which aperture is in position for
use.
(b) 	Radiation exposure control:
(1) 	Means shall be provided to initiate the 	radiation
exposure by a deliberate action on the part of the
operator, 	such as the depression of a 	switch.
Radiation exposure shall not be initiated without such
an action. 	In addition, 	it shall not be possible to
initiate an exposure when the timer is 	set to a 	"zero"
or "off" position if either position is provided.
(2) 	Means shall be provided to terminate the exposure at a
preset time interval, 	preset product of current and
time, 	a preset number of pulses, 	or a preset radiation
exposure to the image receptor. 	Termination of an
exposure shall cause automatic resetting of the timer
to its initial setting or to 	"zero."
(A) 	When manual exposure control is provided, 	an x-
ray exposure control shall be incorporated into
each x-ray system such that an exposure can be
terminated by the operator at any time
("deadman" switch) 	except for:
(i) 	Exposure of one-half second or less; 	and
(ii) 	During serial radiography when means shall
be provided to permit completion of any
single exposure of the series in progress.
(B) 	When an automatic exposure control is provided:
(i) 	Indication shall be made on the control
panel when this mode of operation is
selected;
(ii) 	If the x-ray tube potential is equal to or
greater than fifty-one 	kilovolts peak, 	the
minimum exposure time for 	field emission
equipment rated for pulsed operation shall
be equal to or less than a 	time interval
equivalent to two pulses;
(iii) 	The minimum exposure time for all
equipment other than that specified in
clause 	(ii) 	shall be equal to or less than
one-sixtieth second or a 	time interval
required to deliver five milliamperes-
45-102
26
Complete printed PDF page 102
seconds, 	whichever is greater;
(iv) 	Either the product of peak x-ray tube
potential, 	current, 	and exposure time
§11-45-119
shall be limited to not more than sixty
kilowatt-seconds per exposure, 	or the
product of x-ray tube current and exposure
time shall be limited to not more than six
hundred milliamperes-seconds per exposure,
except that when the x-ray tube potential
is less than fifty-one 	kilovolts peak, 	the
product of x-ray tube current and exposure
time shall be limited to not more than two
thousand milliamperes-seconds per
exposure; 	and
(v) 	A visible signal shall indicate when an
exposure has been terminated at the limits
required by clause 	(iv), 	and manual
resetting shall be required before 	further
automatically timed exposures can be made.
(3) 	Means shall be provided for visual indication of x-ray
production observable at or from the operator's
protected position whenever x-rays are produced. 	In
addition, 	a 	signal audible to the operator shall
indicate that the exposure has terminated.
(4) 	With a 	timer setting of 0.5 seconds or less, 	the
difference between the maximum exposure time 	(Tmax)
and the minimum exposure time 	(Tmin) 	shall be less
than or equal to ten percent of the average exposure
time 	(T), 	when four timer tests are performed:
(Tmax - 	Tmin) 	~ 0.lT
(5) 	The x-ray exposure control shall be so placed that the
operator can view the patient while making any
exposure.
(6) 	Operator protection, 	except veterinary systems:
(A) 	Stationary x-ray systems shall be required to
have the x-ray exposure control permanently
mounted in a protected area so that the operator
is required to remain in that protected area
during the entire exposure.
(B) 	Mobile and portable x-ray systems which are:
(i) 	Used continuously for greater than one
week in the same location, 	i.e., 	a 	room or
suite shall be considered as stationary
systems under subparagraph 	(A).
(ii) 	Used for less than one week in the same
location shall be provided with either a
protective barrier at least 6.5 	feet 	(two
meters) 	high for operator protection
during exposures, 	or means shall be
provided to allow the operator to be at
least twelve feet 	(3.7 meters) 	from the
tube housing assembly during the exposure.
45-103
Complete printed PDF page 103
§11-45-119
(7) 	Operators of all stationary, 	mobile or portable x-ray
systems used for veterinary work shall be provided
with a 	6.5 	foot 	(two meters) 	high protective barrier
for operator protection during exposures, 	or a means
to allow the operator to be at least twelve feet 	(3.7
meters) 	from the tube housing assembly during
exposures, 	or a protective apron and protective
gloves.
(c) 	All mobile or portable radiographic systems shall be
provided with means to limit the source-to-skin distance to equal
to or greater than thirty centimeters, 	except for veterinary
systems.
(d) 	When all technique factors 	are held constant,
including control panel selections associated with automatic
exposure control systems, 	the coefficient of variation of exposure
for both manual and automatic exposure control systems shall not
exceed 0.05. 	This requirement shall be deemed to have been met
if, 	when four exposures are made at identical technique factors,
the difference between the maximum exposure 	(Emax) 	and the minimum
exposure 	(Emin) 	shall be less than or equal to ten percent of the
average exposure 	(E):
(Emax - 	Emin) 	< 0.lE
(e) 	Radiation from capacitor energy storage equipment in
standby status: 	Radiation emitted from the x-ray tube when the
exposure switch or timer is not activated shall not exceed a 	rate
of two milliroentgens 	(0.516 microcoulombs per kilogram) 	per hour
at 	five 	centimeters from any accessible surface of the diagnostic
source assembly, 	with the beam-limiting device fully open.
(f) 	Deviation of technique factors 	from indicated values
shall not exceed the limits specified for that system by its
manufacturer. 	In the absence of manufacturer's specifications,
the deviation shall not exceed ten percent of the indicated value.
(g) 	Linearity, 	uncertified x-ray systems only: 	The
following 	requirements apply when the equipment is operated on a
power supply as specified by the manufacturer for any fixed x-ray
tube potential within the range of forty percent to one hundred
percent of the maximum rated.
(1) 	Equipment having independent selection of x-ray tube
current: 	The average ratios of exposure to the
indicated milliampere seconds product 	(mR/mAs 	(or
C/kg/mAs)) 	obtained at any two tube current settings
shall not differ by more than 0.10 times their sum:
[Xl - 	X2] 	~ 	0.l0(Xl + X2)
where Xl and X2 are the average mR/mAs 	(or C/kg/mAs)
values obtained at any two tube current settings.
(2) 	Equipment having a combined x-ray tube current-
exposure time product 	(mAs) 	selector, 	but not a
separate tube current 	(mA) 	selector: 	The average
ratios of exposure to the indicated milliampere
seconds product 	(mR/mAs 	(or C/kg/mAs)) 	obtained at any
45-104
6
Complete printed PDF page 104
two rnAs selector settings shall not differ by more
than 0.10 times their sum:
[Xl - 	X2] 	< 	0.l0(Xl + X2)
§11-45-119
where Xl and X2 are the average mR/rnAs 	(or C/kg/rnAs)
values obtained at any two rnAs selector settings.
(3) 	Determination of compliance shall be based on four
exposures, 	of no less than 0.05 seconds each, 	taken
within a 	time period of one hour, 	at each of the two
settings. 	These two settings may include any two
focal 	spot sizes provided that neither focal 	spot size
is equal to or less than 0.45 millimeter, 	in which
case the two settings shall be restricted to the same
focal 	spot size. 	For purposes of this 	requirement,
focal 	spot size is the nominal focal 	spot size
specified by the x-ray tube manufacturer.
(h) 	Additional requirements applicable to certified
systems only: 	diagnostic x-ray systems incorporating one or more
certified component(s) 	shall be required to comply with the
following additional requirement(s) 	which relate to that certified
component(s).
(1) 	Linearity: 	When the equipment is operated on a power
supply as specified by the manufacturer in accordance
with the requirements of applicable federal 	standards,
for any fixed x-ray tube potential within the range of
forty to one hundred percent of the maximum rated.
(A) 	For equipment having independent selection of x-
ray tube current, 	the average ratios of exposure
to the indicated milliampere seconds product
obtained at any two consecutive tube current
settings shall not differ by more than 0.10
times their sum:
[Xl - 	X2] 	< 	0.l0(Xl + X2)
where Xl and X2 	are the average mR/rnAs
(C/kg/rnAs) 	values obtained at each of two
consecutive tube current settings or at two
settings differing by no more than a 	factor of
two where the tube current selection is
continuous.
(B) 	For equipment having selection of x-ray tube
current-exposure time product manufactured after
May 3, 	1994, 	the average ratios of exposure to
the indicated milliampere seconds product
obtained at any two consecutive milliampere
seconds selector settings shall not differ by
more than 0.10 times their sum:
[Xl - 	X2] 	< 	0.l0(Xl + X2)
where Xl and X2 	are the average mR/rnAs
(C/kg/rnAs) 	values obtained at each of two
45-105
Complete printed PDF page 105
§11-45-119
consecutive milliampere seconds selector
settings or at two settings differing by no more
than a 	factor of two where the milliampere
seconds selector provides continuous 	selection.
(2) 	Beam limitation for stationary and mobile general
purpose x-ray systems:
(A) 	There shall be provided a means of stepless
adjustment of the size of the x-ray field. 	The
minimum field size at an source-image receptor
distance of one hundred centimeters shall be
equal to or less than five centimeters by five
centimeters.
(B) 	When a 	light localizer is used to define the x-
ray field, 	it shall provide an average
illumination of not less than one hundred sixty
lux or fifteen 	footcandles 	at one hundred
centimeters or at the maximum source-image
receptor distance, 	whichever is less. 	The
average illumination shall be based upon
measurements made in the approximate center of
each quadrant of the light field. 	Radiation
therapy simulation systems manufactured on and
after May 27, 	1980, 	are exempt from this
requirement.
(C) 	The edge of the light field at one hundred
centimeters or at the maximum source-image
receptor distance, 	whichever is less, 	shall have
a 	contrast ratio, 	corrected for ambient
lighting, 	of not less than four in the c~se of
beam-limiting devices designed for use on
stationary equipment, 	and a 	contrast ratio of
not less than three in the case of beam-limiting
devices designed for use on mobile equipment.
The contrast ratio is defined as I 1 /I 2 where I 1
is the illumination three millimeters from the
edge of the light field toward the center of the
field; 	and I 2 is the illumination three
millimeters from the edge of the light field
away from the center of the field. 	Compliance
shall be determined with a measuring instrument
aperture of one millimeter in diameter.
(3) 	Beam limitation for portable x-ray systems shall meet
the beam limitation requirements of paragraphs 	(1) 	and
(2).
(4) 	stationary, 	general purpose x-ray systems which
contain a 	tube housing assembly, 	an x-ray control,
and, 	for those systems so equipped, 	a 	table, 	all
certified in accordance with 21 C.F.R. 	§1020.30(c)
shall have manual collimation for 	field limitation and
alignment.
(5) 	If positive beam limitation is installed on
stationary, 	general purpose x-ray systems after the
effective date of these rules, 	the following
requirements shall be met:
45-106
Complete printed PDF page 106
§11-45-119
(A) 	Positive beam limitation shall prevent the
production of x-rays when:
(i) 	Either the length or width of the x-ray
field in the plane of the image receptor
differs, 	except as permitted by
subparagraph 	(E), 	from the corresponding
image receptor dimensions by more than
three percent of the source-image receptor
distance; 	or
(ii) 	The sum of the length and width
differences as stated in clause 	(i)
without regard to sign exceeds 	four
percent of the source-image receptor
distance.
(iii) 	The beam limiting device is at an source-
image receptor distance for which positive
beam limitation is not designed for
sizing.
(B) 	If a means of overriding the positive beam
limitation system exists, 	that means 	shall be
designed for use only in the event of positive
beam limitation system failure or if the system
is being serviced.
(C) 	If in a position that the operator would
consider it part of the operational controls or
if it is referenced in the operator's manual or
in other materials intended for the operator,
(i) 	Shall 	require that a 	key be utilized to
defeat the positive beam limitation;
(ii) 	Shall require that the key remain in place
during the entire time the positive beam
limitation system is overridden; 	and
(iii) 	Shall 	require that the key or key switch
be clearly and durably labeled as 	follows:
FOR X-RAY FIELD LIMITATION SYSTEM FAILURE
(D) 	Compliance with subparagraph 	(A) 	shall be
determined when the equipment indicates that the
beam axis is perpendicular to the plane of the
image receptor. 	Compliance shall be determined
no sooner than five seconds after insertion of
the image receptor.
(E) 	The positive beam limitation system shall be
capable of operation, 	at the discretion of the
operator, 	such that the size of the 	field may be
made smaller than the size of the image receptor
through stepless adjustment of the 	field size.
The minimum field size at an source-image
receptor distance of one hundred centimeters
shall be equal to or less than five centimeters
by five centimeters.
(F) 	Positive beam limitation shall be designed such
that if a 	change in image receptor does not
45-107
2 	5
Complete printed PDF page 107
§11-45-119
cause an automatic return to positive beam
limitation function as described in subparagraph
(B), 	then any change of image receptor size or
source-image receptor distance shall cause the
automatic return.
(6) 	Except for dental panoramic systems, 	termination of
exposure shall cause automatic resetting of the timer
to its initial setting or to 	"zero."
(i) 	A tube stand or other mechanical support shall be used
for portable x-ray systems so that the x-ray tube housiAJ-g aA~embly
need not be hand-held during exposures. 	[Eff 	NQ\[ 	1 ;:;; 19:::, 	l
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	32 -
11(21), 	321-71)
§11-45-120 	Intraoral dental 	radiographic systems. 	(a)
X-ray systems designed for use with an intraoral image receptor
shall be provided with means to limit source-to-skin distance to
not less than eighteen centimeters.
(b) 	Radiographic systems designed for use with an
intraoral image receptor shall be provided with means to limit the
x-ray field such that the x-ray field at the minimum source-to-
skin distance shall be containable in a circle having a diameter
of no more than seven centimeters.
(c) 	Radiation exposure control for certified and non-
certified systems:
(1) 	Exposure initiation:
(A) 	Means shall be provided to initiate the
radiation exposure by a deliberate action on the
part of the operator, 	such as the depression of
a 	switch. 	Radiation exposure shall not be
initiated without such an action; 	and
(B) 	It shall not be possible to make an exposure
when the timer is set to a 	"zero" 	or "off"
position if either position is provided.
(2) 	Exposure termination:
(A) 	Means shall be provided to terminate the
exposure at a preset time interval, 	preset
product of current and time, 	a preset number of
pulses, 	or a preset radiation exposure to the
image receptor.
(B) 	An x-ray exposure control shall be incorporated
into each x-ray system such that an exposure can
be terminated by the operator at any time,
except for exposures of one-half second or less.
(C) 	Termination of an exposure shall cause automatic
resetting of the timer to its initial setting or
to 	"zero".
(3) 	Means shall be provided for visual indication
observable at or from the operator's protected
position whenever x-rays are produced. 	In addition, 	a
signal audible to the operator shall indicate that the
exposure has terminated.
(4) 	With a 	timer setting of 0.5 seconds or less, 	the
45-108
Complete printed PDF page 108
§11-45-120
difference between the maximum exposure time 	(Tmax)
and the minimum exposure time 	(Tmin) 	shall be less
than or equal to ten percent of the average exposure
time 	(T), 	when four timing tests are performed:
(Tmax - 	Tmin) 	~ 0.10T
(5) 	Exposure control location and operator protection:
(A) 	Stationary x-ray systems shall be required to
have the x-ray exposure control permanently
mounted in a protected area, 	so that the
operator is 	required to remain in that protected
area during the entire exposure; 	and
(B) 	Mobile and portable x-ray systems which are:
(i) 	Used for greater than one week in the same
location, 	i.e., 	a 	room or suite, 	shall
meet the requirements of subparagraph 	(A);
(ii) 	Used for less than one week in the same
location shall be provided with either a
protective barrier at least 6.5 	feet 	(two
meters) 	high for operator protection, 	or
means to allow the operator to be at least
twelve feet 	(3.7 meters) 	from the tube
housing assembly while making exposures.
(d) 	The coefficient of variation for exposure
reproducibility shall not exceed 0.05 when all technique factors
are held constant. 	This 	requirement shall be deemed to have been
met if, 	when four exposures are made within a period of one hour
at identical technique factors, 	the difference between the maximum
exposure value 	(Emax) 	and the minimum exposure value 	(Emin) 	shall
be less than or equal to ten percent of the average exposure:
(Emax - 	Emin) 	~ 0.l0E
(e) 	When the equipment allows a choice of x-ray tube
current settings and is operated on a power supply as specified by
the manufacturer for any fixed x-ray tube potential within the
range of forty to one hundred percent of the maximum rating, 	the
average ratios of exposure to the indicated milliampere seconds
product obtained at any two consecutive tube current settings
shall not differ by more than 0.10 times their sum:
(Xl - 	X2) 	~ 0.l0(Xl + X2)
where Xl and X2 are the average wR/mAs 	(C/kg/mAs) 	values obtained
at each of two consecutive tube current settings.
(f) 	Deviation of technique factors 	from indicated values
shall not exceed the limits specified for that system by its
manufacturer. 	In the absence of manufacturer's specifications the
deviation shall not exceed ten percent of the indicated value.
(g) 	Dental x-ray machines with a nominal 	fixed kilovolts
peak of less than fifty kilovolts peak shall not be used to make
diagnostic dental 	radiographs of humans.
(h) 	Administrative controls:
45-109
265
Complete printed PDF page 109
§11-45-120
(1)
(2)
(3)
(4)
Patient and film holding devices shall be used when
appropriate.
The tube housing and the position indicating device
shall not be hand-held during an exposure.
The x-ray system shall be operated in such a manner
that the useful beam at the patient's skin does not
exceed the requirements of subsection 	(b).
Dental fluoroscopy w~1~~~t
1
~a~gnifrtensification shall
not be used. 	[Eff 	l~U \J 	iG 	I 	;:;J;:;J 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-121 	Computed tomography x-ray systems. 	(a)
General requirements:
(1) 	Means shall be provided to terminate the x-ray
exposure automatically by either de-energizing the x-
ray source or shuttering the x-ray beam in the event
of equipment failure affecting data collection. 	Such
termination shall occur within an interval that limits
the total scan time to no more than one hundred ten
percent of its preset value through the use of either
a backup timer or devices which monitor equipment
function.
(2) 	A visible signal shall indicate when the x-ray
exposure has been terminated through the means
required by paragraph 	(1).
(3) 	The operator shall be able to terminate the x-ray
exposure at any time during a 	scan, 	or series of scans
under computed tomography x-ray system control, 	of
greater than one-half second duration.
(4) 	For any single tomogram system, 	means shall be
provided to permit visual determination of the
tomographic plane or a 	reference plane offset from the
tomographic plane.
(5) 	For any multiple tomogram system, 	means shall be
provided to permit visual determination of the
location of a 	reference plane. 	This 	reference plane
can be offset from the location of the tomographic
planes.
(6) 	If a device using a light source is used to satisfy
paragraphs 	(4) 	and 	(5), 	the light source shall provide
illumination levels sufficient to permit visual
determination of the location of the tomographic plane
or reference plane under ambient light conditions of
up to five hundred lux.
(7) 	The computed tomography x-ray control and gantry shall
provide visual indication whenever x-rays are produced
and, 	if applicable, 	whether the shutter is open or
closed.
(8) 	Each emergency button or switch shall be clearly
labeled as to its 	function.
(9) 	The computed tomography x-ray system shall be designed
such that the computed tomography conditions of
45-110 l.!.
265
Complete printed PDF page 110
§11-45-121
operation to be used during a 	scan or a 	scan sequence
shall be indicated before the initiation of a 	scan or
a 	scan sequence. 	On equipment having all or some of
these conditions of operation at fixed values, 	this
requirement may be met by permanent markings.
Indication of computed tomography conditions of
operation shall be visible from any position from
which scan initiation is possible.
(10) 	When data are not being collected for image
production, 	the radiation adjacent to the tube port
shall not exceed that permitted by section 11-45-117
(c).
(11) 	The angular position where the maximum surface
computed tomography dose index occurs shall be
identified to allow for reproducible positioning of a
computed tomography dosimetry phantom.
(12) 	Additional requirements applicable to computed
tomography x-ray systems containing a 	gantry
manufactured after September 3, 	1985:
(A) 	The total error in the indicated location of the
tomographic plane or reference plane shall not
exceed five millimeters.
(B) 	If the x-ray production period is less than one-
half second, 	the indication of x-ray production
shall be actuated for at least one-half second.
Indicators at or near the gantry shall be
discernible from any point external to the
patient opening where insertion of any part of
the human body into the primary beam is
possible.
(C) 	The deviation of indicated scan increment versus
actual increment shall not exceed plus or minus
one millimeter with any mass 	from zero to one
hundred kilograms 	resting on the support device.
The patient support device shall be incremented
from a typical starting position to the maximum
incremented distance or thirty centimeters,
whichever is less, 	and then returned to the
starting position. 	Measurement of actual versus
indicated scan increment may be taken anywhere
along this travel.
(D) 	Premature termination of the x-ray exposure by
the operator shall necessitate resetting of the
computerized tomography conditions of operation
before the initiation of another scan.
(b) 	Provision shall be made for two-way aural
communication between the patient and the operator at the control
panel.
(c) 	Viewing systems:
(1) 	Windows, 	mirrors, 	closed-circuit television, 	or an
equivalent shall be provided to permit continuous
observation of the patient during irradiation and
shall be so located that the operator can observe the
patient from the control panel.
45-111
Complete printed PDF page 111
§11-45-121
(2) 	When the primary viewing system is by electronic
means, 	an alternate viewing system 	(which may be
electronic) 	shall be available for use in the event of
failure of the primary viewing system.
(d) 	Surveys:
(1) 	All computed tomography x-ray systems installed after
the effective date of these rules and those systems
not previously surveyed shall have a 	survey made by,
or under the direction of, 	a 	qualified medical
physicist before use on human patients unless
otherwise authorized by the director. 	In addition,
such surveys shall be done after any change in the
facility or equipment which might cause a 	significant
increase in radiation hazard.
(2) 	A written report of the survey shall be available for
inspection by the department for three years.
(e) 	Radiation calibrations:
(1) 	Calibration of the radiation output of the computed
tomography x-ray system shall be performed by, 	or
under the direction of, 	a 	qualified medical physicist
who is physically present at the facility during such
calibration.
(2) 	Calibration of a 	computed tomography x-ray system
shall be performed at intervals specified by a
qualified medical physicist and after any change or
replacement of components which, 	in the opinion of the
qualified medical physicist, 	could cause a 	change in
the radiation output.
(3) 	Calibration of the radiation output of a computed
tomography x-ray system shall be performed with a
calibrated dosimetry system. 	The calibration of such
system shall be traceable to a national standard. 	The
dosimetry system shall have been calibrated within the
preceding two years.
(4) 	Computed tomography dosimetry phantom(s) 	shall be used
in determining the radiation output of a computed
tomography x-ray system. 	Such phantom(s) 	shall meet
the following specifications and conditions of use:
(A) 	Computed tomography dosimetry phantom(s) 	shall
be right circular cylinders of polymethyl
methacrylate of density 1.19 plus or minus 	0.01
grams per cubic centimeter. 	The phantom(s)
shall be at least fourteen centimeters in length
and shall have diameters of 32.0 centimeters for
testing computed tomography x-ray systems
designed to image any section of the body and
16.0 centimeters for systems designed to image
the head or for whole body scanners operated in
the head scanning mode.
(B) 	Computed tomography dosimetry phantom(s) 	shall
provide means 	for the placement of a
dosimeter(s) 	along the axis of rotation and
along a 	line parallel to the axis of rotation
1.0 centimeter from the outer surface and within
45-112
265
Complete printed PDF page 112
the phantom. 	Means 	for the placement of
dosimeters or alignment devices at other
locations may be provided.
§11-45-121
(C) 	Any effects on the doses measured due to the
removal of phantom material to accommodate
dosimeters shall be accounted for through
appropriate corrections to the 	reported data or
included in the statement of maximum deviation
for the values obtained using the phantom.
(D) 	All dose measurements shall be performed with
the computed tomography dosimetry phantom placed
on the patient couch or support device without
additional attenuation materials present.
(5) 	Calibration shall be required for each type of head,
body, 	or whole-body scan performed at the facility.
(6) 	Calibration shall include the center multiple scan
average dose and the surface multiple scan average
dose.
(7) 	Calibration procedures shall be in writing. 	Records
of calibrations performed shall be maintained for
inspection by the department for three years.
(fl 	Spot checks shall be conducted:
(1) 	Procedures shall be in writing and shall have been
developed by a 	qualified medical physicist.
(2) 	Procedures shall incorporate the use of a computed
tomography dosimetry phantom which has a 	capability of
providing an indication of contrast scale, 	noise,
nominal tomographic section thickness, 	the resolution
capability of the system for low and high contrast
objects, 	and measuring the mean computed tomography
number for water or other reference material.
(3) 	Spot checks shall be included in the calibration
required by subsection 	(e) 	and at time intervals and
under system conditions specified by a 	qualified
medical physicist.
(4) 	Spot checks shall include acquisition of images
obtained with the computed tomography dosimetry
phantom(s) 	using the same processing mode and computed
tomography conditions of operation as are used to
perform calibrations required by subsection 	(e). 	The
images shall be retained, 	until a 	new calibration is
performed, 	in two forms as 	follows:
(A) 	Photographic copies of the images obtained from
the image display device; 	and
(B) 	Images 	stored in digital 	form on a 	storage
medium compatible with the computed tomography
x-ray system.
(5) 	Written records of the spot checks 	shall be
maintained for inspection by the department for three
years.
(g) 	Operating procedures:
(1) 	The computed tomography x-ray system shall be operated
by an individual who has been specifically trained in
its operation and licensed under chapter 11-44.
45-113
Complete printed PDF page 113
§11-45-121
(2) 	Information shall be available at the control panel
regarding the operation and calibration of the system.
Such information shall include the 	following:
(A) 	Dates of the latest calibration and spot checks
and the location within the facility where the
results of those tests may be obtained;
(B) 	Instructions on the use of the computed
tomography dosimetry phantom(s) 	including a
schedule of spot checks appropriate for the
system, 	allowable variations 	for the indicated
parameters, 	and the results of at least the most
recent spot checks conducted on the system;
(C) 	The distance in millimeters between the
tomographic plane and the reference plane if a
reference plane is utilized; 	and
(D) 	A current technique chart available at the
control panel which specifies for each routine
examination the computed tomography conditions
of operation and the number of scans per
examination.
(3) 	If the calibration or spot check of the computed
tomography x-ray system identifies that a 	system
operating parameter has exceeded a 	tolerance
established by the qualified medical physicist, 	use of
the computed tomography x-ray system on patients shall
be limited to those uses permitted by established
written instructirM\9f1tR.e 	ualified medical
physicist. 	[Eff l~U V .1. l 	1 	9 	J 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	( mp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-122 	Bone densitometry systems. 	(a) 	Only systems
certified by the U.S. 	Department of Health and Human Services
shall be utilized for healing arts purposes.
(b) 	System performance shall be maintained in accordance
with all performance standards specified by the U.S. 	Department of
Health and Human Services.
(c) 	Operators of bone
licensed under chapter 11-44.
HRS §§321-10, 	321-11, 	321-71)
321-71)
§11-45-123 	Reserved.
§11-45-124 	Reserved.
densit~f~\Jy 1s¥,3~
9shall be
[Eff 	i'IU, 	., 	1::1:::J 	] 	(Auth:
(Imp: 	HRS §§321-1, 	321-11(21),
§11-45-125 	Mammography x-ray systems and facilities. 	(a)
Only x-ray systems, 	pursuant to the Mammography Quality Standards
Act of 1992, 	Public Law 102-539, 	42 U.S.C. 	263b, 	and 21 C.F.R.
Part 900, 	shall be used for screening and diagnostic mammography.
45-114
265
Complete printed PDF page 114
§11-45-127
(b) 	A facility performing mammography shall have a valid
certificate issued by the U.S. 	Department of Health and Human
Services, 	pursuant to the Mammography Quality Standards Act of
1992, 	Public Law 102-539, 	42 U.S.C. 	263b, 	and 21 C.F.R. 	Part 900.
(c) 	A facility performing mammography shall ensure that
the additional mammography activities of processing the x-ray
film, 	interpreting the image, 	and maintaining viewing conditions,
wherever performed, 	meet all quality standards pursuant to the
Mammography Quality Standards Act of 1992, 	Public Law 102-539, 	42
U.S.C. 	263b, 	and 21 C.F.R. 	Part 900.
(d) 	A facility without a valid certificate issued by the
U.S. 	Department of Health and Human Services, 	pursuant to the
Mammography Quality Standards Act of 1992, 	Public Law 102-539, 	42
U.S.C. 	263b, 	and 21 C.F.R. 	Part 900 shall electronically
disconnect diagnostic mammography x-ray systems 	from the power
source.
(e) 	Each mammography facility shall maintain accurate
records of mammography services performed at the facility with at
least the following information:
(1) 	Patient's name;
(2) 	Referring physician's name;
(3) 	Mammography examination date;
(4) 	Type of mammography examination; 	and
(5) 	Name of radiologic technologist who performed the
mammography examination.
(f) 	No individual or facility shall obstruct, 	prevent, 	or
deny a 	holder of a valid certificate issued by the U.S. 	Department
of Health and Human Services, 	pursuant to the Mammography Quality
Standards Act of 1992, 	Public Law 102-539, 	42 U.S.C. 	263b, 	and 21
C.F.R. 	Part 900 from acquiring data, 	informat~~?i\Jo~ ~taw~~ics on
outcome required by Public Law 102-539. 	[Eff t~ LJ ii 	1 ~ 1::t~!:'J 	l
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
SUBCHAPTER 7
USE OF RADIONUCLIDES IN THE HEALING ARTS
§11-45-126 	Purpose and scope. 	This subchapter establishes
requirements 	for the receipt, 	possession, 	production, 	preparation,
compounding, 	use, 	and transfer of radionuclides in the healing
arts. 	The requirements of this subchapter are in addition to, 	and
not in substitution for, 	other requirements in this chapter.
[Eff 	1 ') 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	~~ 	1-11(21), 	321-71)
§11-45-127 	Exemptions. 	Radionuclides licensed by the U.S.
Nuclear Regulatory Commission are exempt from the requirements of
this subchapter to the extent they arr 1 ;;;t9ula~e~g~gthe U.S.
Nuclear Regulatory Commission. 	[Eff NU\} 	l 'l, 	I I~; 	] 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21),
321-71)
45-115
2
Complete printed PDF page 115
§11-45-128
§11-45-128 	Notifications. 	A licensee shall notify the
department in writing within thirty days when an authorized user,
radiation safety officer, 	or qualified medical physicist,
perma1l..in.tJ.~ P-\fc.fr?~H1ues performance of duties.
[Eff 	NU\J 	12 	1::J~~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-129 	ALARA program. 	(a) 	Each licensee shall develop
and implement a written program to maintain radiation doses and
releases of radioactive material in effluents to unrestricted
areas 	as low as 	reasonably achievable.
(b) 	To satisfy the requirement of subsection 	(a):
(1) 	The management, 	radiation safety officer, 	and all
authorized users shall participate in the
establishment, 	implementation, 	and operation of the
program as 	required by this subchapter or the
radiation safety committee; 	or
(2) 	For licensees that are not medical institutions,
management and all authorized users shall participate
in the program as required by the radiation safety
officer.
(c) 	The ALARA program shall include an annual review by
the radiation safety committee for licensees that are medical
institutions, 	or management and the radiation safety officer for
licensees that are not medical institutions, 	of summaries of the
types and amounts of radioactive material used, 	occupational dose
reports, 	and continuing education and training for all personnel
who work with or in the vicinity of radioactive material. 	The
purpose of the review is to ensure that individuals make every
reasonable effort to maintain occupational doses, 	doses 	to the
general public, 	and releases of radioactive material as low as
reasonably achievable, 	taking into account the state of
technology, 	and the cost of improvements in relation to benefits.
(d) 	The licensee shall retain a 	current written
description of the ALARA program which shall include:
(1) 	A commitment by management to keep occupational doses
as low as 	reasonably achievable;
(2) 	A requirement that the radiation safety officer brief
management once each year on the 	radiation safety
program;
(3) 	Personnel exposure investigational levels as
established in accordance with section ll-45-13l(c) (8)
that, 	when exceeded, 	shall initiate an investigation
by the radiation safety officer of the cause of the
exposure; 	and
(4) 	Personnel exposure investigational levels that, 	when
exceeded, 	shall initiate a prompt investigation by the
radiation safety officer of the cause of the exposure
and a 	consideration of actions that might be taken to
redu'i
1 of recurrence.
[Eff l'J 	(Auth: 	HRS §§321-10, 	321-11,
45-116 	2 	5
Complete printed PDF page 116
§11-45-131
321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-130 	Radiation safety officer. 	(a) 	A licensee shall
appoint a 	radiation safety officer responsible for implementing
the radiation safety program. 	The licensee, 	through the radiation
safety officer, 	shall ensure that radiation safety activities are
being performed in accordance with approved procedures and
regulatory requirements in the daily operation of the licensee's
radioactive material program.
(b) 	The radiation safety officer shall:
(1) 	Investigate overexposures, 	accidents, 	spills, 	losses,
thefts, 	unauthorized receipts, 	uses, 	transfers, 	and
disposals, 	and other deviations 	from approved
radiation safety practice and implement corrective
actions as necessary;
(2) 	Implement written policy and procedures 	for:
(A) 	Authorizing the purchase of radioactive
material;
(B) 	Receiving and opening packages of radioactive
material;
(C) 	Storing radioactive material;
(D) 	Keeping an inventory record of radioactive
material;
(E) 	Using radioactive material 	safely;
(F) 	Taking emergency action if control of
radioactive material is lost;
(G) 	Performing periodic radiation surveys;
(H) 	Performing checks and calibrations of survey
instruments and other safety equipment;
(I) 	Disposing of radioactive material;
(J) 	Training personnel who work in or frequent 	areas
where radioactive material is used or stored;
and
(K) 	Keeping a copy of all records and reports
required by this chapter; 	and
(3) 	For medical use not sited at a medical institution,
approve or disapprove radiation safety program changes
with the advice and consent of management before
submittal to the department for action; 	or
(4) 	For medical use sited at a medical institution, 	assist
the radiation safef~nf9IIUJU~tee i~ the performance of
its duties. 	[Eff rfLJV 	1 t:. 199::J 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-131 	Radiation safety committee. 	(a) 	Each medical
institution licensee shall establish a 	radiation safety committee
to oversee the use of radioactive material.
(b) 	The committee shall meet the following administrative
requirements:
(1) 	Membership shall consist of at least three individuals
and shall include an authorized user of each type of
45-117
265
Complete printed PDF page 117
§11-45-131
shall:
use permitted by the license, 	the radiation safety
officer, 	a 	representative of the nursing service, 	and
a 	representative of management who is neither an
authorized user nor a 	radiation safety officer. 	Other
members may be included as the licensee deems
appropriate.
(2) 	The committee shall meet at least once each calendar
quarter.
(3) 	To establish a quorum and to conduct business, 	one-
half of the committee's membership shall be present,
including the radiation safety officer and the
management's representative.
(4) 	The minutes of each radiation safety committee meeting
shall include:
(A) 	Date of the meeting;
(B) 	Members present;
(C) 	Members absent;
(D) 	Summary of deliberations and discussions;
(E) 	Recommended actions and the numerical 	results of
all ballots; 	and
(F) 	Document any reviews required in section ll-45-
129(c) 	and subsection 	(c).
(5) 	The committee shall provide each member with a 	copy of
the meeting minutes, 	and retain one copy.
(c) 	To oversee the use of licensed material, 	the committee
(1) 	Be responsible for monitoring the institutional
program to maintain occupational doses as low as
reasonably achievable;
(2) 	Review, 	on the basis of safety and with regard to the
training and experience standards of this part, 	and
approve or disapprove any individual who is to be
listed as an authorized user, 	the radiation safety
officer, 	or qualified medical physicist before
submitting an application for a 	license or request for
amendment or renewal;
(3) 	Review on the basis of safety and approve or
disapprove each proposed method of use of radioactive
material;
(4) 	Review on the basis of safety, 	and approve with the
advice and consent of the radiation safety officer and
the management representative, 	or disapprove
procedures and radiation safety program changes before
submittal to the department for license action;
(5) 	Review quarterly, 	with the assistance of the radiation
safety officer, 	occupational 	radiation exposure
records of all personnel working with radioactive
material;
(6) 	Review quarterly, 	with the assistance of the radiation
safety officer, 	all incidents involving radioactive
material with respect to cause and subsequent actions
taken;
(7) 	Review annually, 	with the assistance of the radiation
safety officer, 	the radioactive material program; 	and
45-118
5
Complete printed PDF page 118
§11-45-133
(8) 	Establish a 	table of investigational levels for
occupational dose that, 	when exceeded, 	shall initiate
investigations and consideration\ 1
w; 1
a~t.tp~A>~gthe
radiation safety officer. 	[Eff 	l~U\ 	l t 	I~~ 	l
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS
§§321-1, 	321-11(21), 	321-71)
§11-45-132 	Statement of authorities and responsibilities.
(a) 	A licensee shall provide sufficient authority and
organizational freedom to the radiation safety officer and the
radiation safety committee to:
(1) 	Identify radiation safety problems;
(2) 	Initiate, 	recommend, 	or provide solutions; 	and
(3) 	Verify implementation of corrective actions.
(b) 	A licensee shall establish in writing the authorities,
duties, 	responsibilities, 	and radiation safety activities of the
radiat~M\tafe,t,y
1
R™cer and the radiation safety committee.
[Eff 	r~aJv 	l l 	::,::,::, 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-133 	Supervision. 	(a) 	A licensee who allows 	the
receipt, 	possession, 	production, 	preparation, 	compounding, 	use, 	or
transfer of radioactive material by an individual under the
supervision of an authorized user shall:
(1) 	Instruct the supervised individual in the principles
of radiation safety appropriate to that individual's
use of radioactive material;
(2) 	Review the supervised individual's use of radioactive
material, 	provide reinstruction as needed and review
records 	kept to reflect this use;
(3) 	Require an authorized user to be immediately available
to communicate with the supervised individual; 	and
(4) 	Require that only those individuals specifically
trained, 	and designated by the authorized user, 	shall
be permitted to administer radionuclides or radiation
to patients.
(b) 	The supervised individual receiving, 	possessing,
producing, 	preparing, 	compounding, 	using or transferring
radioactive material shall:
(1) 	Follow the instructions of the supervising authorized
user;
(2) 	Follow the written radiation safety procedures
established by the licensee;
(3) 	Follow the procedures established by the radiation
safety officer; 	and
(4) 	Comply with this chapter and the license conditions
with re~ect to the use of radioactive material.
[Eff 	Nu\/ 1 ? 1qa9. 	l 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp:'"" 	Ui?! s§321-1, 	321-11 (21), 	321-71)
45-119
265
Complete printed PDF page 119
§11-45-134
§11-45-134 	Visiting authorized user. 	(a) 	A licensee may
permit any visiting authorized user to use radioactive materials
for medical use for sixty days each year if:
(1) 	The visiting authorized user has the prior written
permission of the licensee's management and, 	if the
use occurs on behalf of an institution, 	the
institution's radiation safety committee; 	and
(2) 	Only those procedures 	for which the visiting
authorized user is specifically authorized are
performed by that individual.
(b) 	A licensee shall retain~n9~ie1 Rf~hl;l{nrecords
specified in subsection 	(a). 	[Eff l~UV 	, i(, 	I~~::, 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-135 	Mobile nuclear medicine service administrative
requirements. 	(a) 	The mobile nuclear medicine service shall be
licensed if the service receives, 	uses, 	or possesses radioactive
material. 	The client of the mobile nuclear medicine service shall
be licensed if the client receives or possesses radioactive
material to be used by a mobile nuclear medicine service.
(bl 	Mobile nuclear medicine service licensees shall 	retain
for the duration of service a 	letter signed by the management of
each location where services are rendered that authorizes use of
radioactive material. 	If the client is licensed, 	the letter shall
document procedures 	for notification, 	receipt, 	storage, 	and
documentation of transfer of radioactive material delivered to the
client's location for use by the mobile nuclear medicine service.
(c) 	A mobile nuclear medicine service shall not have
radioactive material delivered directly from the manufacturer or
the distributor to the client's address of use, 	unless the client
has a 	license. 	Radioactive material delivered to the client's
address of use shall be received in conformance with the client's
license.
(d) 	A mobile nuclear medicine service shall inform a
responsible individual, 	such as a 	representative of management or
nursing staff in charge of the patient or the nursing unit, 	who is
on site at the time trn~,r~d1~Raceuticals are being
administered. 	[Eff NlJV 	l k, 	J:,,,;;!j 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-136 	Procedures 	for human administrations. 	(a) 	Each
licensee shall establish and maintain a 	written program to provide
assurance that radioactive material or radiation therefrom is
administered to humans as directed by the authorized user. 	The
program shall include procedures 	for:
(1) 	Preparing written directives 	for the administration of
radiation, 	however, 	a written directive is not
required when an authorized user personally assays and
administers a dosage provided the pertinent 	facts 	are
documented as otherwise required;
(2) 	Verifying by more than one method the identity of the
individual to be administered radiation or radioactive
45-120 65
Complete printed PDF page 120
§11-45-137
material;
(3) 	Updating the diagnostic clinical procedures manual;
(4) 	Verifying that final plans of treatment and related
calculations for brachytherapy, 	teletherapy, 	and gamma
stereotactic radiosurgery are in accordance with
respective written directives;
(5) 	Assuring that administration of radiation is carried
out as specified in the written directive or the
diagnostic clinical procedures manual;
(6) 	Identifying and evaluating unintended deviations 	from
the written directive or diagnostic clinical
procedures manual including taking appropriate action
for recordable events and misadministrations;
(b) 	Each licensee shall evaluate and respond to
misadministrations in accordance with section 11-45-137.
(c) 	Each licensee shall evaluate and respond to recordable
events within thirty days after discovery by assembling the
relevant facts, 	identifying the cause of the recordable event, 	and
taking appropriate action to prevent recurrence.
(d) 	Each licensee shall conduct an annual evaluation of
the human administration program including any recommendations for
changes to be made as well as any modifications made since the
last evaluation and, 	if required, 	revise procedures to assure that
radioactive material and the radiation therefrom is administered
as directed by the authorized user. 	Modifications made to the
program shall not decrease the effectiveness of the program.
(e) 	Each licensee shall retain, 	in auditable form:
(1) 	Each written directive;
(2) 	A record of each administered radiation dose or
radiopharmaceutical dosage where a written directive
is 	required;
(3) 	A record of each annual review of the program
including the evaluations and findings 	of the review;
(4) 	A record of each recordable event, 	the 	relevant facts,
and any corrective actions taken.
(f) 	Individuals practicing nuclear medicine technology
shall be licensed under chapter 11-44. 	[Eff ~{QV 	1_2 ·mga, 	l
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§3~1.!'11; ~21-
11 (21), 	321-71)
§11-45-137 	Records, 	notifications, 	and reports of
misadministrations. 	(a) 	The following actions shall be taken for
a misadministration by the licensee:
(1) 	The licensee shall notify the department by telephone
no later than twenty-four hours after discovery of the
misadministration.
(2) 	The licensee shall submit a written report to the
department within fifteen days after discovery of the
misadministration. 	The written report shall include
the licensee's name; 	the prescribing physician's name;
a brief description of the event; 	why the event
occurred; 	the effect on the patient; 	what improvements
are needed to prevent recurrence; 	actions taken to
45-121 265
Complete printed PDF page 121
§11-45-137
prevent recurrence; 	whether the licensee notified the
patient; 	or the patient's responsible relative or
guardian, 	and if not, 	why not, 	and if the patient was
notified, 	what information was provided to the
patient. 	The report shall not include the patient's
name or other information that could lead to
identification of the patient.
(3) 	The licensee shall notify the referring physician and
also notify the patient of the misadroinistration not
later than twenty-four hours after its discovery,
unless the referring physician personally informs the
licensee either that he or she shall inform the
patient or that, 	based on medical judgment, 	telling
the patient would be harmful. 	The licensee is not
required to notify the patient without first
consulting the referring physician. 	If the referring
physician or patient cannot be reached within twenty-
four hours, 	the licensee shall notify the patient as
soon as possible thereafter. 	The licensee may not
delay any appropriate medical care for the patient,
including any necessary remedial care as a 	result of
the misadroinistration, 	because of any delay in
notification.
(4) 	If the patient was notified, 	the licensee shall also
furnish, 	within fifteen days after discovery of the
misadroinistration, 	a written report to the patient by
sending either:
(A) 	A copy of the report that was submitted to the
department, 	or
(B) 	A brief description of both the event and the
consequences, 	as they may affect the patient,
provided a 	statement is included that the report
submitted to the department can be obtained from
the licensee.
(b) 	Each licensee shall retain a 	record of each
misadroinistration. 	The record shall contain the names of all
individuals involved including the prescribing physician, 	allied
health personnel, 	the patient, 	and the patient's 	referring
physician, 	the patient's social security number or identification
number if one has been assigned, 	a brief description of the
misadroinistration, 	why it occurred, 	the effect on the patient,
what improvements are needed to prevent recurrence, 	and the action
taken to prevent recurrence.
(c) 	Aside from the notification requirement, 	nothing in
subsections 	(a) 	and 	(b) 	shall affect any rights or duties of
licensees, 	and physicians in relation to each other, 	patients, 	or
patiel\"ti',.$\feflWnxARi'\e relatives or guardians.
[Eff 	r"U I 	_" 	i::J!:t!:t 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-138 	Suppliers. 	A licensee shall use for medical use
only:
(1) 	Radioactive material manufactured, 	produced, 	labeled,
45-122 65
Complete printed PDF page 122
§11-45-140
prepared, 	compounded, 	packaged, 	and distributed
pursuant to these rules or the equivalent regulations
of an agreement state or the U.S. 	Nuclear Regulatory
Commission; 	and
(2) 	Reagent kits that have been manufactured, 	labeled,
packaged, 	and distributed in accordance with an
approval issued by the U.S. 	Department of Health and
Human Services.
(3) 	Radiopharmaceuticals compounded from a prescription in
accordance with chapter 16-95.
(4) 	Teletherapy and brachytherapy sources manufactured and
distributed in accordance with a 	license issued
pursuant to these rules or the equivalent regulations
of an agreement s\
1te 1 pr 	Nuclear Regulatory
Commission. 	[Eff r.. Ov 	1 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-139 	Quality control of diagnostic equipment. 	Each
licensee shall establish written quality control procedures 	for
all diagnostic equipment used for radionuclide studies. 	As a
minimum, 	quality control procedures and frequencies 	shall be those
recommended by equipment manufacturers and procedures which have
been approved by the department. 	The licensee shall conduct
qualii¥n·~9n~r~1~wRedures in accordance with written procedures.
[Eff NLV 	l l 	l~~:;j 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-140 	Possession, 	use, 	calibration, 	and check of dose
calibrators. 	(a) 	A medical use licensee authorized to administer
radiopharmaceuticals shall possess a 	dose calibrator and use it to
measure the amount of activity administered to each patient. 	In
the case where the ionization type dose calibrator cannot be used
effectively to verify administered activity, 	the licensee shall
use an alternative method. 	Any alternative method for the use of
a dose calibrator shall be approved by the department. 	Any
alternative method shall provide for acceptable verification of
constancy, 	accuracy, 	linearity, 	and geometry dependence as
applicable.
(b) 	Each licensee shall establish written quality control
procedures 	for all dose calibrators used for measuring the amount
of activity administered to a patient. 	As a minimum, 	quality
control procedures and frequencies 	shall be those recommended by
the American National Standards Institute:
(1) 	Check each dose calibrator for constancy with a
dedicated check source at the beginning of each day of
use. 	To satisfy the requirement of this section, 	the
check shall be done on a 	frequently used setting with
a 	sealed source of not less than fifty microcuries
(1.85 megabecquerels) 	with energies representative of
the radionuclides in clinical use at the facility;
(2) 	Test each dose calibrator for accuracy upon
45-123
2 	5
Complete printed PDF page 123
§11-45-140
installation and at intervals not to exceed twelve
months thereafter by assaying at least two sealed
sources containing different radionuclides, 	the
activity of which the manufacturer has determined
within five percent of the stated activity, 	with
minimum activity of fifty microcuries 	(1.85
megabecquerels) 	and energies representative of the
radionuclides in clinical use at the facility;
(3) 	Test each dose calibrator for linearity upon
installation and at intervals not to exceed three
months thereafter over the range of use between ten
microcuries 	(three hundred seventy kilobecquerels) 	and
the highest dosage that shall be assayed; 	and
(4) 	Test each dose calibrator for geometry dependence upon
installation over the range of volumes and volume
configurations for which it shall be used. 	The
licensee shall keep a 	record of this test for the
duration of the use of the dose calibrator.
(c) 	A licensee shall mathematically correct dosage
readings 	for any geometry or linearity error that exceeds ten
percent if the dosage is greater than ten microcuries 	(three
hundred seventy kilobecquerels) 	and shall repair or replace the
dose calibrator if the accuracy or constancy error exceeds ten
percent.
(d) 	A licensee shall also perform checks and tests
required by subsection 	(b) 	following adjustment or repair of the
dose calibrator.
(e) 	A licensee shall retain a 	record of each check and
test required by this section. 	The records 	required by subsection
(b) 	shall include:
(1) 	For subsection 	(b) (1), 	the model and serial number of
the dose calibrator, 	the identity and calibrated
activity of the radionuclide contained in the check
source, 	the date of the check, 	the activity measured,
the instrument settings, 	and the initials of the
individual who performed the check;
(2) 	For subsection 	(b) (2), 	the model and serial number of
the dose calibrator, 	the model and serial number of
each source used and the identity of the 	radionuclide
contained in the source and its activity, 	the date of
the test, 	the results of the test, 	the instrument
settings, 	and the signature of the individual who
performed the test;
(3) 	For subsection 	(b) (3), 	the model and serial number of
the dose calibrator, 	the calculated activities, 	the
measured activities, 	the date of the test, 	and the
signature of the radiation safety officer; 	and
(4) 	For subsection 	(b) (4), 	the model and serial number of
the dose calibrator, 	the configuration and calibrated
activity of the source measured, 	the activity of the
source, 	the activity measured and the instrument
setting for each volume measured, 	the date of the
test, 	and the signature of the individual who
performed the test.
45-124
265
Complete printed PDF page 124
§11-45-142
(f) 	A licensee shall use dose calibrator reference and
calibration sources traceable to the National Institute of
Standards and Technology or other standards recognized as being
equivtlffl~ qy2t~n~9tional Institute of Standards and Technology.
[Eff 	pJJ V 	1 ,, 	1:::t~: 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-141 	Calibration and check of survey instruments.
(a) 	A licensee shall ensure that the survey instruments used to
show compliance with this part have been calibrated before first
use, 	annually, 	and following repair.
(b) 	To satisfy the requirements of subsection 	(a), 	the
licensee shall:
(1) 	Calibrate all required scale readings up to one
thousand millirems 	(ten millisieverts) 	per hour with a
radiation source;
(2) 	For each scale that shall be calibrated, 	calibrate two
readings separated by at least fifty percent of scale
rating; 	and
(3) 	Conspicuously note on the instrument the apparent dose
rate from a dedicated check source as determined at
the time of calibration, 	and the date of calibration.
(c) 	To satisfy the requirements of subsection 	(b), 	the
licensee shall consider a point as calibrated if the indicated
exposure rate differs from the calculated exposure rate by not
more than twenty percent, 	and shall conspicuously attach a
correction chart or graph to the instrument.
(d) 	A licensee shall check each survey instrument for
proper operation with the dedicated check source before each use.
The licensee is not required to keep records of these checks.
(e) 	The licensee shall retain a 	record of each calibration
required in subsection(a). 	The record shall include:
(1) 	A description of the calibration procedure; 	and
(2) 	A description of the source used and the certified
dose 	rates 	from the source, 	and the rates indicated by
the instrument being calibrated, 	the correction
factors deduced from the calibration data, 	the
signature of the individual who performed the
calibration, 	and the date of calibration.
(f) 	To meet the requirements of subsections 	(a), 	(b), 	and
(c), 	the licensee may obtain the services of individuals licensed
to perform calibrations of survey instruments. 	Records of
calibrations which contain information requ~red by subsection 	(e)
shall be maintained by the licensee. 	[Eff NQV 	l ~ j009 	J
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS.!C§f3zf'-q, 	321-
11 (21), 	321-71)
§11-45-142 	Assay of radiopharmaceutical dosages. 	(a) 	An
assay shall be conducted, 	before medical use, 	of the activity of
each radiopharmaceutical dosage that contains more than ten
microcuries 	(three hundred seventy kilobecquerels) 	of a photon-
emitting radionuclide;
45-125 2 	5
Complete printed PDF page 125
§11-45-142
(b) 	Radiopharmaceuticals emitting alpha and/or beta
radiations as the radiation of principal interest shall be
obtained:
(1)
(2)
In unit dose form, 	calibrated by the supplier for
individual patients; 	and
From a 	supplier which participates in a measurement
quality assurance program with the National Institute
of Standards and Technology, 	and which is designed to
ensure that unit doses have a 	calibration traceable to
a national standard.
(c) 	The licensee shall retain a 	record of the assays or
calibrations required by subsections 	(a) 	and 	(b). 	To satisfy this
requirement, 	the record shall contain the:
(1) 	Radiopharmaceutical, 	or the radionuclide administered;
(2) 	Patient's name, 	and identification number if one has
been assigned;
(3) 	Prescribed dosage and measured activity of the dosage
at the time of assay, 	or a notation that the total
activity was determined by a 	calibration traceable to
a 	national standard;
(4) 	Date and time of the assay or calibration and the date
and time of the administration; 	and
(5) 	Initials of the individual who performed the assay or
documentation of the supplier's participation in the
meaSUf~~rvtA 	assurance program.
[Eff i'HJV 	.! 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	§§321-1, 	321-11(21), 	321-71)
§11-45-143 	Authorization for calibration and reference
sources. 	A licensee for medical use of radioactive material may
receive, 	possess, 	and use the following radioactive material for
check, 	calibration and reference use:
(1) 	Sealed sources manufactured and distributed by persons
specifically licensed pursuant to these rules or
provisions of the U.S. 	Nuclear Regulatory Commission,
and that do not exceed fifteen millicuries 	(five
hundred fifty-five megabecquerels) 	each;
(2) 	Any radioactive material with a half-life of one
hundred days or less in individual amounts not to
exceed fifteen millicuries 	(five hundred fifty-
five megabecquerels); 	and
(3) 	Any radioactive material with a half-life greater than
one hundred days in individual amounts not to exceed
two 	·~·~-~-~;1t}f~'lf~~ies 	(7.4 megabecquerels) 	each.
[Eff 	1 (, 	,~~~ 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-144 	Possession of sealed sources and brachytherapy
sources. 	(a) 	A licensee in possession of any sealed source or
brachytherapy source shall follow the radiation safety and
handling instructions supplied by the manufacturer or equivalent
instructions approved by the department and shall maintain the
45-126
5
Complete printed PDF page 126
§11-45-144
instructions 	for the duration of source use in a 	legible form
convenient to users.
(b) 	A licensee
assure that:
in possession of a 	sealed source shall
(b),
(1)
(2)
The source is tested before its 	first use unless the
licensee has a certificate from the supplier
indicating that the source was tested within six
months before transfer to the licensee; 	and
The source is tested for leakage at intervals not to
exceed six months or at intervals approved by the
department, 	another agreement state, 	a 	licensing
state, 	or the U.S. 	Nuclear Regulatory Commission.
(c) 	To satisfy the leak test requirements of subsection
the licensee shall assure that:
(1) 	Leak tests are capable of detecting the presence of
0.005 microcuries 	(one hundred eighty-five becquerels)
of radioactive material on the test sample, 	or in the
case of radium, 	the escape of radon at the rate of
0.001 microcuries 	(thirty-seven becquerels) 	per
(2)
twenty-four hours;
Test samples are taken from the source or from the
surfaces of the device in which the source is mounted
or stored on which radioactive contamination might be
expected to accumulate; 	and
(3) 	Test samples are taken when the source is in the "off"
position.
(d) 	A licensee shall retain leak test records. 	The
records 	shall contain the model number and serial number, 	if
assigned, 	of each source tested, 	the identity of each source
radionuclide and its estimated activity, 	the measured activity of
each test sample expressed in microcuries 	(becquerels), 	a
description of the method used to measure each test sample, 	the
date of the test, 	and the signature of the individual who
performed the test.
(e) 	If the leak test reveals the presence of 0.005
microcuries 	(one hundred eighty-five becquerels) 	or more of
removable contamination, 	the licensee shall:
(1) 	Immediately withdraw the sealed source from use and
store, 	repair, 	or dispose of it in accordance with the
requirements of subchapter 4; 	and
(2) 	File a 	report with the department within five days of
receiving the leak test results describing the
equipment involved, 	the test results, 	and the action
taken.
(f)
following
(1)
(2)
(3)
A licensee need not perform a 	leak test on the
sources:
Sources containing only radioactive material with a
half-life of less than thirty days;
Sources containing only radioactive material as a 	gas;
Sources containing one hundred microcuries 	(3.7
megabecquerels) 	or less of beta- or photon-emitting
material or ten microcuries 	(three hundred seventy
kilobecquerels) 	or less of alpha-emitting material;
and
45-127
265
Complete printed PDF page 127
§11-45-144
(4) 	Sources stored and not being used. 	The licensee
shall, 	however, 	test each such source for leakage
before any use or transfer unless it has been tested
for leakage within six months before the date of use
or transfer.
(g) 	A licensee in possession of a 	sealed source or
brachytherapy source shall conduct a physical inventory of all
such sources at intervals not to exceed three months. 	The
licensee shall retain each inventory record. 	The inventory
records 	shall contain the model number of each source, 	and serial
number if one has been assigned, 	the identity of each source
radionuclide and its estimated activity, 	the location of each
source, 	date of the inventory, 	and the signature of the individual
who performed the inventory.
(h) 	A licensee in possession of a 	sealed source or
brachytherapy source shall survey with a 	radiation survey
instrument at intervals not to exceed three months all areas where
such sources are stored. 	This does not apply to teletherapy units
or sealed sources in diagnostic devices.
(i) 	A licensee shall retain a 	record of each survey
required in subsection 	(h). 	The record shall include the date of
the 	survey, 	a 	sketch of each area that was surveyed, 	the measured
dose rate at several points in each area expressed in millirem
(millisieverts) 	per hour, 	the model number and serial number of
the 	~ur~e~ intrument used to make the survey, 	a49,
0
k~e.f¼$nature of
the individual who performed the survey. 	[Eff 	I~ 	v 1 , 	1aa9 	i
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	:t~-
11 (21), 	321-71)
§11-45-145 	syringe shields. 	(a) 	A licensee shall keep
syringes that contain radioactive material to be administered in a
radiation shield.
(b) 	A licensee shall require each individual who prepares
or administers radiopharmaceuticals to use a 	syringe radiation
shield unless 	o~ the shield is contraindicated for that
patient. 	[Eff 	1 l 	199~ 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	§§321-1, 	:'r21-ll (21), 	321-71)
§11-45-146 	syringe labels. 	Unless utilized immediately, 	a
licensee shall conspicuously label each syringe, 	or syringe
radiation shield as to contents or intended patient.
[Eff 	Nnv 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	Ms 	1-11(21), 	321-71)
§11-45-147 	Vial shields. 	A licensee shall require each
individual preparing or handling a vial that contains a
radio~~;n-m~ceutical to keep the vial in a vial 	radiation shield.
[Eff 	14UV 	12 1qgo 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-T, ~21-11(21), 	321-71)
45-128 	2
Complete printed PDF page 128
§11-45-148 	Vial shield labels. 	A licensee shall
conspicuously label each vial radiation shield that contains a
§11-45-150
~ial 	of a 	~ad~opharmaceu\~ffl-; w_;- 1~~the radiopharmaceutical name or
its 	abbreviation. 	[Eff 	f\JU \r 	j_ 'i(, 	,gao 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321:-1, 	:fiJ.-11 (21), 	321-71)
§11-45-149 	surveys 	for contamination and ambient radiation
dose 	rate. 	(a) 	A licensee shall survey with a 	radiation
detection survey instrument at the end of each day of use all
areas where radiopharmaceuticals are routinely prepared for use or
administered.
(b) 	A licensee shall survey with a 	radiation detection
survey instrument at least once each week all areas where
radiopharmaceuticals or radioactive wastes are stored.
(c) 	A licensee shall conduct the surveys required by
subsections 	(a) 	and 	(b) 	so as to be able to measure dose rates as
low as 	0.1 millirem 	(one microsievert) 	per hour.
(d) 	A licensee shall establish dose rate action levels 	for
the 	surveys 	required by subsections 	(a) 	and 	(b) 	and shall require
that the individual performing the survey immediately notify the
radiation safety officer if a dose 	rate exceeds an action level.
(e) 	A licensee shall survey for 	removable contamination
each day of use all areas where 	radiopharmaceuticals are routinely
prepared for use or administered and each week where radioactive
materials are stored.
(f) 	A licensee shall conduct the surveys required by
subsection 	(e) 	so as to be able to detect contamination on each
wipe sample of two thousand disintegrations per minute 	(33.3
becquerels) .
(g) 	A licensee shall establish removable contamination
action levels 	for the surveys required by subsection 	(e) 	and shall
require that the individual performing the survey immediately
notify the radiation safety officer if contamination exceeds
action levels.
(h) 	A licensee shall retain a 	record of each survey
required by subsections 	(a), 	(b), 	and 	(e). 	The record shall
include the date of the survey, 	a 	sketch of each area surveyed,
action levels established for each area, 	the measured dose rate at
several points in each area expressed in millirems 	(microsieverts)
per hour or the removable contamination in each area expressed in
disintegrations per minute 	(becquerels) 	per one hundred square
centimeters, 	the serial number and the model number of the
instrument used to make the survey or analyze the samples, 	and the
initials of the individual who performed the survey.
[Eff 	1 ') 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	§t3~ 	1-11 (21), 	321-71)
§11-45-150 	Release of patients containing
radiopharmaceuticals or permanent implants. 	A licensee shall not
authorize 	release from confinement for medical care any patient
administered a 	radiopharmaceutical or administered a permanent
implant until the dose rate from the patient at the time of
45-129 265
Complete printed PDF page 129
§11-45-150
release shall not result in a 	total effective dose equivalent
exceeding five hundred millirems 	(five millisieverts) 	to any other
indiv½?~~~ qurinftcomplete decay of the radionuclide.
[Eff 	NUV 	'i 2 1 go 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321- 	,V:321-11(21), 	321-71)
§11-45-151 	Mobile nuclear medicine service technical
requirements. 	A licensee providing mobile nuclear medicine
service shall:
(1) 	Transport to each address of use only syringes or
vials containing prepared radiopharmaceuticals or
radiopharmaceuticals that are intended for
reconstitution of radiopharmaceutical kits;
(2) 	Bring into each area of use all 	radioactive material
to be used and, 	before leaving, 	remove all unused
radioactive material and associated radioactive waste;
(3) 	Secure or keep under constant surveillance and
immediate control all 	radioactive material when in
transit or at an area of use;
(4) 	In addition to complying with sections 11-45-140 and
11-45-141, 	check survey instruments and dose
calibrators for constancy and response, 	and check all
other transported equipment for proper function before
medical use at each area of use;
(5) 	Carry a 	survey meter calibrated in accordance with
section 11-45-141 in each vehicle that is being used
to transport radioactive material, 	and, 	before leaving
a 	client area of use, 	survey all areas of
radiopharmaceutical use with a 	radiation detection
survey instrument to ensure that all
radiopharmaceuticals and all associated radioactive
waste have been removed;
(6) 	Retain a 	record of each survey. 	The record shall
include the date of the survey, 	a plan of each area
that was surveyed, 	the measured dose rate at several
points in each area of use expressed in millirem
(millisieverts) 	per hour, 	the model and serial number
of the instrument used to make the survey, 	and the
initials of the individual who performed the survey;
(7) 	Use radioactive gases only in areas of use and under
conditions which have been evaluated and approved by
the department for compliance with airborne release
standards; 	and
(8) 	Remove all radioactive material from the mobile
vehicle and monitor the vehicle fq~.f'lqoptamination at
the end of each day of use. 	[Eff r~ IJ V 1 2 1900 	]
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	~M,
§§321-1, 	321-11(21), 	321-71)
§11-45-152 	Storage of volatiles and gases. 	(a) 	A licensee
shall store volatile radiopharmaceuticals and radioactive gases in
the shippers' 	radiation shield and container.
45-130
265
Complete printed PDF page 130
(b) 	A licensee shall store and use
in a 	properly functioning fume hood. 	[Eff
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp:
11 (21), 	321-71)
§11-45-155
§11-45-153 	Decay-in-storage. 	(a) 	A licensee shall hold
radioactive material 	for decay-in-storage before disposal in
ordinary trash and is exempt from the waste disposal 	requirements
of subchapter 4 if the licensee:
(1) 	Holds radioactive material 	for decay a minimum of ten
half-lives;
(2) 	Monitors radioactive material at the container surface
before disposal as ordinary trash and determines that
its radioactivity cannot be distinguished from the
background radiation level with a 	radiation detection
survey instrument set on its most sensitive scale and
with no interposed shielding;
(3) 	Removes or obliterates all 	radiation labels; 	and
(4) 	Separates and monitors each generator column
individually with all 	radiation shielding removed to
ensure that its contents have decayed to background
radiation level before disposal.
(b) 	For radioactive material disposed in accordance with
subsection 	(a), 	the licensee shall retain a 	record of each
disposal. 	The record shall include the date of the disposal, 	the
date on which the radioactive material was placed in storage, 	the
radionuclides disposed, 	the model and serial number of the survey
instrument used, 	the background dose rate, 	the radiation dose rate
measured at the surface of each waste container, 	and the name of
the individual who performed the disposal.
[Eff NQ\/ 1-2 1099 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HR's §§321~'7 	321-11 (21), 	321-71)
§11-45-154 	Reserved.
§11-45-155 	Specific requirements 	for the use of
radiopharmaceuticals 	for uptake, 	dilution, 	or excretion studies.
(a) 	A licensee may use any radioactive material in a
radiopharmaceutical 	for a 	diagnostic use involving measurements of
uptake, 	dilution, 	or excretion;
(1) 	Which has been granted acceptance or approval by the
U.S. 	Department of Health and Human Services; 	or
(2) 	Which is prepared and compounded in accordance with
chapter 16-95.
(b) 	A licensee authorized to use radioactive material 	for
uptake, 	dilution, 	excretion, 	imaging, 	and localization studies
shall possess a portable radiation detection survey instrument
capable of detecting dose rates over the range 0.1 millirem 	(1.0
microsievert) 	per hour to fifty millirems 	(five hundred
microsieverts) 	per hour. 	The instrument shall be operable and
calibrated in accordance with section 11-45-140.
45-131 .L
265
Complete printed PDF page 131
§11-45-155
[Eff 	2 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
1-11(21), 	321-71)	(Imp: 	HRS §§321-1,
§11-45-156 	Specific requirements 	for the use of
radiopharmaceuticals, 	generators, 	and reagent kits for imaging and
localization studies. 	(a) 	A licensee may use any radioactive
material in a non-gaseous or non-aerosol diagnostic
radiopharmaceutical or any generator or reagent kit for
preparation and diagnostic use of a non-gaseous or non-aerosol
radiopharmaceutical containing radioactive material:
(1) 	Which has been granted acceptance or approval by the
U.S. 	Department of Health and Human Services; 	or
(2) 	Which has been prepared and compounded in accordance
with chapter 16-95.
(3) 	A licensee shall elute generators.
(b) 	Radionuclide contaminants.
(1) 	A licensee shall not administer a 	radiopharmaceutical
containing:
(A) 	Reserved.
(B) 	More than 0.02 microcurie 	(0.02 kilobecquerel)
of strontium-82 per millicurie 	(megabecquerel)
of rubidium-82 chloride injection;
(C) 	More than 0.2 microcurie 	(0.2 	kilobecquerel) 	of
strontium-85 per millicurie 	(megabecquerel) 	of
rubidium-82 chloride injection.
(2) 	A licensee preparing radiopharmaceuticals from
radionuclide generators shall measure the
concentration of radionuclide contaminant in each
eluate or extract, 	as appropriate for the generator
system, 	to determine compliance with the limits
specified in paragraph 	(1).
(3) 	A licensee who shall measure radionuclide contaminant
concentration shall retain a 	record of each
measurement. 	The record shall include, 	for each
elution or extraction tested, 	the measured activity of
the radiopharmaceutical expressed in millicuries
(megabecquerels), 	the measured activity of contaminant
expressed in microcuries 	(kilobecquerels), 	the 	ratio
of the measures expressed in microcuries
(kilobecquerels) 	of contaminant per millicurie
(megabecquerel), 	the date of the test, 	and the
initials of the individual who performed the test.
(4) 	A licensee shall report immediately to the department
each occurrence of radionuclide contaminant
concentration exceeding the limits specified in
paragraph 	( 1) .
(c) 	Control of aerosols and gases.
(1) 	A licensee who administers radioactive aerosols or
gases shall do 	so with a 	system that shall keep
airborne concentrations within the limits prescribed
in subchapter 4.
(2) 	The system shall provide collection and decay or
disposal of the aerosol or gas in a 	shielded
45-132 	!...
2 6 5
Complete printed PDF page 132
§11-45-157
container.
(3) 	A licensee shall only administer radioactive gases in
rooms that are at negative pressure compared to
surrounding rooms.
(4) 	Before receiving, 	using, 	or storing a 	radioactive gas,
the licensee shall calculate the amount of time needed
after a 	release to 	reduce the concentration in the
area of use to the occupational limit in subchapter 4.
The calculation shall be based on the highest
activity of gas handled in a 	single container and the
measured available air exhaust rate.
(5) 	A licensee shall post the time calculated in paragraph
(4) 	at the area of use and require that, 	in case of a
gas spill, 	individuals evacuate the room until the
posted time has elapsed.
(6) 	A licensee shall check the operation of collection
systems monthly and measure the ventilation rates in
areas of use at intervals not to exceed six months.
Records of these checks and measurements shall be
maintained.
(7) 	A copy of the calculations required in paragraph 	(4)
shall be recorded and retained for the duration of the
license.
(d) 	A licensee authorized to use radioactive material 	for
imaging and localization studies shall possess a portable
radiation detection survey instrument capable of detecting dose
rates over the range of 0.1 millirem 	(one microsievert) 	per hour
to 	fifty millirems 	(five hundred microsieverts) 	per hour, 	and a
portable radiation measurement survey instrument capable of
measuring dose 	rates over the range one millirem 	(ten
microsieverts) 	per hour to one thousand millirems 	(ten
millisieverts) 	per hour. 	The instruments shall be operable and
calibrated in accordance with section 11-45-141.
[Eff 	1 2 1099 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	§'§3"21-"1_, 	321-11 (21), 	321-71)
§11-45-157 	Specific requirements for the use of
radiopharmaceuticals 	for therapy. 	(a) 	A licensee may use any
radioactive material in a 	radiopharmaceutical and for a
therapeutic use:
(1) 	Which has been granted acceptance or approval by the
U.S. 	Department of Health and Human Services; 	or
(2) 	Which has been prepared and compounded in accordance
with chapter 16-95.
(b) 	Safety instruction.
(1) 	A licensee shall provide oral and written radiation
safety instruction for all personnel caring for
patients undergoing radiopharmaceutical therapy.
Refresher training shall be provided at intervals not
to exceed one year.
(2) 	To satisfy paragraph 	(1), 	the instruction shall
describe the licensee's procedures 	for:
(A) 	Patient control;
45-133
265
Complete printed PDF page 133
§11-45-157
(B) 	Visitor control;
(C) 	Contamination control;
(D) 	Waste control;
(E) 	Notification of the radiation safety officer or
authorized user in case of the patient's death
or medical emergency; 	and
(F) 	Training for workers.
(3) 	A licensee shall keep a 	record of individuals
receiving instruction required by paragraph 	(1), 	a
description of the instruction, 	the date of
instruction, 	and the name of the individual who gave
the instruction. 	Such record shall be maintained for
inspection by the department for three years.
(c) 	Safety precautions.
(1) 	For each patient receiving radiopharmaceutical therapy
and hospitalized for compliance with section 11-45-
150, 	a 	licensee shall:
(A) 	Provide a private room with a private sanitary
facility;
(B) 	Post the patient's door with a 	"Caution:
Radioactive Material" sign and note on the door
or on the patient's chart where and how long
visitors may stay in the patient's room;
(C) 	Authorize visits by individuals under eighteen
years of age only on a case-by-case basis with
the approval of the authorized user after
consultation with the radiation safety officer;
(D) 	Promptly after administration of the dosage,
measure the dose rates in contiguous restricted
and unrestricted areas with a 	radiation
measurement survey instrument to demonstrate
compliance with the requirements of subchapter 4
and retain a 	record of each survey that includes
the time and date of the survey, 	a plan of the
area or list of points surveyed, 	the measured
dose rate at several points expressed in
millirems 	(microsieverts) 	per hour, 	the
instrument used to make the survey, 	and the
initials of the individual who made the survey;
(E) 	Either monitor material and items removed from
the patient's room to determine that any
contamination cannot be distinguished from the
natural background radiation level with a
radiation detection survey instrument set on its
most sensitive scale and with no interposed
shielding, 	or handle these materials and items
as 	radioactive waste;
(F) 	Instruct the patient and, 	if practical, 	the
patient's 	family, 	orally and in writing
concerning radiation safety precautions that
shall help to keep 	radiation dose to household
members and the public as low as 	reasonably
achievable before authorizing release of the
patient;
45-134 265
Complete printed PDF page 134
§11-45-159
(G) 	Survey the patient's room and private sanitary
facility for 	removable contamination with a
radiation detection survey instrument before
assigning another patient to the room. 	The room
shall not be reassigned until removable
contamination is less than two hundred
disintegrations per minute 	(3.33 becquerels) 	per
one hundred square centimeters.
(2) 	For each non-hospitalized patient receiving
radiopharmaceutical therapy, 	the licensee shall
instruct the patient and, 	if practical, 	the patient's
family, 	orally and in writing concerning radiation
safety precautions that shall help to keep radiation
doses 	to the household members and the public as low
as 	reasonably achievable.
(3) 	A licensee shall notify the radiation safety officer
or the authorized user immediately if the patient
expires or has a medical emergency.
(d) 	A licensee authorized to use radioactive material 	for
radiopharmaceutical therapy shall possess a portable radiation
detection survey instrument capable of detecting dose rates over
the 	range 0.1 millirem 	(one microsievert) 	per hour to fifty
millirems 	(five hundred microsieverts) 	per hour, 	and a portable
radiation measurement survey instrument capable of measuring dose
rates over the range one millirem 	(ten microsieverts) 	per hour to
one thousand millirems 	(ten millisieverts) 	per hour. 	The
instruments shall be operable and calibrated in accordance with
section 11-45-141. 	[Eff NQ:J 	1 2 1900 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	!Ms §§3Zl-l, 	~'f-11 (21), 	321-71)
§11-45-158 	Specific requirements for the use of sealed
sources 	for diagnosis. 	A licensee authorized to use radioactive
material as a 	sealed source for diagnostic purposes shall have
available for use a portable radiation detection survey instrument
capable of detecting dose rates over the range 0.1 millirem 	(one
microsievert) 	per hour to fifty millirems 	(five hundred
microsieverts) 	per hour or a portable radiation measurement survey
instrument capable of measuring dose rates over the range one
millirem 	(ten microsieverts) 	per hour to one thousand millirems
(ten millisieverts) 	per hour. 	The instrument shall be operable
and citffiJatec;l in accordance with section 11-45-141.
[Eff 	t\ld\r 	1 l 	1qgg 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS ~321-'1, 	321-11(21), 	321-71)
§11-45-159 	Specific requirements for the use of sources 	for
brachytherapy. 	(a) 	A licensee shall use radioactive sources in
accordance with the manufacturer's radiation safety and handling
instructions.
(b) 	Safety instruction.
(1) 	The licensee shall provide oral and written radiation
safety instruction to all personnel caring 	for a
patient receiving implant therapy. 	Refresher training
45-135 26
Complete printed PDF page 135
§11-45-159
(2)
(3)
(c)
(1)
shall be provided at intervals not to exceed one year.
To satisfy paragraph 	(1), 	the instruction shall
describe:
(A) 	Size and appearance of the brachytherapy
sources;
(B) 	Safe handling and shielding instructions in case
of a dislodged source;
(C) 	Procedures 	for patient control;
(D) 	Procedures 	for visitor control;
(E) 	Procedures 	for notification of the radiation
safety officer or authorized user if the patient
expires or has a medical emergency; 	and
(F) 	Training for workers.
A licensee shall maintain a 	record of individuals
receiving instruction required by paragraph 	(1), 	a
description of the instruction, 	the date of
instruction, 	and the name of the individual who gave
the instruction.
Safety precautions.
For each patient receiving
shall:
implant therapy a 	licensee
(A) 	Not place the patient in the same room with a
patient who is not receiving radiation therapy
unless the licensee can demonstrate compliance
with the radiation dose limits 	for individual
members of the public as 	specified in subchapter
4 at a distance of one meter from the implant;
(B) 	Post the patient's door with a 11 Caution:
Radioactive Materials 11 sign and note on the door
or the patient's chart where and how long
visitors may stay in the patient's room;
(C) 	Authorize visits by individuals under eighteen
years of age only on a case-by-case basis with
the approval of the authorized user after
consultation with the radiation safety officer;
(D) 	Promptly after implanting the sources, 	survey
the dose 	rates in contiguous restricted and
unrestricted areas with a 	radiation measurement
survey instrument to demonstrate compliance with
subchapter 4 and retain a 	record of each survey
that includes the time and date of the survey, 	a
sketch of the area or list of points surveyed,
the measured dose rate at several points
expressed in millirems 	(microsieverts) 	per hour,
the instrument used to make the survey, 	and the
initials of the individual who made the survey;
and
(E) 	Provide the patient with radiation safety
guidance that shall help keep the radiation dose
to household members and the public as low as
reasonably achievable before releasing the
patient if the patient was administered a
permanent implant.
(2) 	A licensee shall notify the radiation safety officer
45-136 	2 6 5
Complete printed PDF page 136
§11-45-159
or authorized user immediately if the patient expires
or has a medical emergency.
(d) 	Brachytherapy sources inventory.
(1) 	Each time brachytherapy sources are returned to an
area of storage from an area of use, 	the licensee
shall immediately count or otherwise verify the number
returned to ensure that all sources taken from the
storage area have been returned.
(2) 	A licensee shall make a 	record of brachytherapy source
utilization which includes:
(A) 	The names of the individuals permitted to handle
the sources;
(B) 	The number and activity of sources removed from
storage, 	the room number of use and patient's
name, 	the time and date they were removed from
storage, 	the number and activity of sources in
storage after the removal, 	and the initials of
the individual who removed the sources from
storage; 	and
(C) 	The number and activity of sources returned to
storage, 	the room number of use and patient's
name, 	the time and date they were returned to
storage, 	the number and activity of sources in
storage after the return, 	and the initials of
the individual who returned the sources to
storage.
(3) 	Immediately after implanting sources in a patient and
immediately after removal of sources from a patient,
the licensee shall make a 	radiation survey of the
patient and the area of use to confirm that no sources
have been misplaced. 	The licensee shall make a 	record
of each survey.
(4) 	A licensee shall maintain records 	required in
paragraphs 	(2) 	and 	(3).
(e) 	Release of patients treated with temporary implants.
(1) 	Immediately after removing the last temporary implant
source from a patient, 	the licensee shall perform a
radiation survey of the patient with a 	radiation
detection survey instrument to confirm that all
sources have been removed. 	The licensee shall not
release from confinement for medical care a patient
treated by temporary implant until all sources have
been removed.
(2) 	A licensee shall maintain a 	record of patient surveys
which demonstrate compliance with paragraph 	(1). 	Each
record shall include the date of the survey, 	the name
of the patient, 	the dose rate from the patient
expressed in millirems 	(microsieverts) 	per hour and
measured within one meter from the patient, 	and the
initials of the individual who made the survey.
(f) 	A licensee authorized to use radioactive material 	for
implant therapy shall possess a portable radiation detection
survey instrument capable of detecting dose rates over a 	range 0.1
millirem 	(one microsievert) 	per hour to fifty millirems 	(five
45-137 	26
Complete printed PDF page 137
§11-45-159
hundred microsieverts) 	per hour, 	and a portable radiation
measurement survey instrument capable of measuring dose rates over
the 	range one millirem 	(ten microsieverts) 	per hour to one
thousand millirems 	(ten millisieverts) 	per hour. 	The instruments
shall be op¥~ir t:t'\4 ~g~rated in accordance with section 11-45-
141. 	[Eff 	l~UV 	l l 	I ~~ 	J 	(Auth: 	HRS §§321-10, 	321-11, 	321-
71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-160 	Reserved.
§11-45-161 	Reserved.
§11-45-162 	Reserved.
§11-45-163 	Reserved.
§11-45-164 	Reserved.
§11-45-165 	Training for 	radiation safety officer. 	An
individual 	fulfilling the 	responsibilities of the 	radiation safety
officer shall:
(1) 	Be certified by the:
(A) 	American Board of Health Physics in
Comprehensive Health Physics; 	or
(Bl 	American Board of Radiology in Radiological
Physics, 	Therapeutic Radiological Physics, 	or
Medical Nuclear Physics; 	or
(C) 	American Board of Nuclear Medicine; 	or
(D) 	American Board of Science in Nuclear Medicine;
or
(El 	Board of Pharmaceutical Specialties in Nuclear
Pharmacy or Science; 	or
(F) 	American Board of Medical Physics in Radiation
Oncology Physics; 	or
(Gl 	Royal College of Physicians and Surgeons of
Canada in Nuclear Medicine; 	or
(2l 	Have had two hundred hours of classroom and laboratory
training covering:
(A} 	Radiation physics and instrumentation;
(B) 	Radiation protection;
(C) 	Mathematics pertaining to the use and
measurement of radioactivity;
(D) 	Radiation biology; 	and
(E} 	Radiopharmaceutical chemistry; 	and
(F) 	Have had one year of full-time 	experience in
radiation safety at a medical institution under
the supervision of the individual identified as
the radiation safety officer on a department
45-138
Complete printed PDF page 138
§11-45-168
license, 	agreement state, 	licensing state, 	or a
U.S. 	Nuclear Regulatory Commission license that
authorizes the medical use of radioactive
material; 	or
(3) 	Be an authorized user for those radioactive material
uses that come within t~~n~?i~tAo~ s~ety officer's
responsibilities. 	[Eff NuV 	l' 	199~ 	l 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1,
321-11(21), 	321-71)
§11-45-166 	Training for experienced radiation safety
officer. 	An individual identified as a 	radiation safety officer
on a 	license, 	or a U.S. 	Nuclear Regulatory Commission license on
the effective date of this chapter who oversees only the use of
radioactive material for which the licensee was authorized on that
date need not comply with the training requirements of section 11-
45-165. 	[Eff NQV 1 2 1999 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11 (21), 	321-71)
§11-45-167 	Training for uptake, 	dilution, 	or excretion
studies. 	Except as provided in sections 11-45-169 and 11-45-170,
the licensee shall require the authorized user of a
radiopharmaceutical listed in section 11-45-155 to be a physician
who:
(1) 	Is certified in:
(A) 	Nuclear medicine by the American Board of
Nuclear Medicine; 	or
(B) 	Diagnostic radiology by the American Board of
Radiology; 	or
(C) 	Diagnostic radiology or radiology by the
American Osteopathic Board of Radiology; 	or
(D) 	Nuclear medicine by the American Osteopathic
Board of Nuclear Medicine; 	or
(E) 	Nuclear medicine by the Royal College of
Physicians and Surgeons of Canada; 	or
(2) 	Has successfully completed a 	six-month training
program in nuclear medicine as part of a training
program that has been approved by the Accreditation
Council 	for Graduate Medical Education and that
included classroom and laboratory training, 	work
experi1
e~ce, Al aIJ.d°"AW'ffrvised clinical experience.
[Eff NuV 	1 2 t:,:;::, 	l 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-168 	Training for imaging and localization studies.
Except as provided in sections 11-45-169 and 11-45-170, 	the
licensee shall require the authorized user of a
radiopharmaceutical, 	generator, 	or reagent 	kit specified in
section 11-45-156 to be a physician who:
(1) 	Is certified in:
(A) 	Nuclear medicine by the American Board of
45-139 2
Complete printed PDF page 139
§11-45-168
Nuclear Medicine; 	or
(B) 	Diagnostic radiology by the American Board of
Radiology; 	or
(C) 	Diagnostic radiology or radiology by the
American osteopathic Board of Radiology; 	or
(D) 	Nuclear medicine by the American Osteopathic
Board of Nuclear Medicine; 	or
(E) 	Nuclear medicine by the Royal College of
Physicians and surgeons of Canada; 	or
(2) 	Has successfully completed a 	six-month training
program in nuclear medicine as part of a training
program that has been approved by the Accreditation
Council for Graduate Medical Education and that
included classroom and laboratory training, 	work
experiff~1.'.' 1n~ s~e~vised clinical experience.
[Eff 	NU if 	l l 	1~9 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS 	§321-1, 	321-11 (21), 	321-71)
§11-45-169 	Training for experienced authorized users.
Practitioners of the healing arts identified as authorized users
for the human use of radioactive material on an agreement state
license, 	or U.S. 	Nuclear Regulatory Commission license on the
effective date of this chapter who perform only those methods of
use 	for which they were authorized on that date need not comply
with the training reguirements of sections 11-45-165 and 11-45-
171. 	[Eff 	NOV 	l 2 1999 	J 	(Auth: 	HRS §§321-10, 	321-11, 	321-
71) 	(Imp: 	HRS §§321-1, 	321-11 (21), 	321-71)
§11-45-170 	Physician training in a three-month program. 	A
physician who, 	before July 1, 	1984, 	began a 	three-month nuclear
medicine training program approved by the Accreditation Council
for Graduate Medical Education and has 	successfully completed the
program, 	is exempted fro~ tpe re~.u~ements of sections 11-45-167
or 11-45-168. 	[Eff 	NQ\J 	12 i~9~ 	J 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-171 	Recentness of training. 	The training and
experience specified in sections 11-45-165 through 11-45-168 shall
have been obtained within the five years preceding the date of
application or the individual shall have had continuing applicable
experience since thh~r1
eqp:W.re~~;i;aining an? experience was
completed. 	[Eff 	Nu ! 	l ~ ,~~::1 	J 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS § 321-1, 	321-11 (21), 	321-71)
SUBCHAPTER 8
ANALYTICAL X-RAY EQUIPMENT
§11-45-172 	Purpose and scope. 	This subchapter provides
requirements 	for analytical x-ray equipment 	(x-ray diffraction and
45-140
Complete printed PDF page 140
§11-45-174
fluorescence analysis). 	The requirements of this subchapter are
in addition to, 	and not in substitutipA.tor, 	other applicable
requirements in this chapter. 	[Eff r~uv 	1 2 1090 	] 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp:• HRS §'§321.!."'.f'., ~21-11 (21),
321-71)
§11-45-173 	Exemptions. 	Equipment utilizing sources of
radiation licensed by the U.S. 	Nuclear Regulatory Commission are
exempt from the requirements of this subchapter to the extent they
are 	re~u}at~.d2bXa™ U.S. 	Nuclear Regulatory Commission.
[Eff ~iUV 	1. ls! I::,~~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-174 	Equipment requirements. 	(a) 	A safety device
which prevents the entry of any portion of an individual's body
into the primary x-ray beam path or which causes the beam to be
shut off upon entry into its path shall be provided on all open-
beam configurations. 	A licensee may apply to the department for
an exemption from the 	requirement of a 	safety device. 	such
application shall include:
(1) 	A description of the various safety devices 	that have
been evaluated;
(2) 	The reason each of these devices cannot be used; 	and
(3) 	A description of the alternative methods that shall be
employed to minimize the possibility of an accidental
exposure, 	including procedures to assure that
operators and others in the area shall be informed of
the absence of safety devices.
(b) 	Open-beam configurations shall be provided with a
readily discernible indication of:
(1) 	X-ray tube "on-off" status located near the 	radiation
source housing, 	if the primary beam is controlled in
this manner; 	and/or
(2) 	Shutter "open-closed" status located near each port on
the radiation source housing, 	if the primary beam is
controlled in this manner.
(c) 	An easily visible warning light labeled with the words
"X-RAY ON", 	or words having a similar intent, 	shall be located:
(1) 	Near any switch that energizes an x-ray tube and shall
be illuminated only when the tube is energized; 	or
(2) 	In the case of a 	radioactive source, 	near any switch
that opens a housing shutter and shall be illuminated
only when the shutter is open.
(d) 	Warning devices shall be labeled so that their purpose
is easily identified. 	On equipment installed after the effective
date of these rules, 	warning devices shall have fail-safe
characteristics.
(e) 	Unused ports on radiation source housings shall be
secured in the closed position in a manner which shall prevent
casual opening.
(f) 	All analytical x-ray equipment shall be labeled with a
readily discernible sign or signs bearing the radiation symbol and
45-141
6
Complete printed PDF page 141
§11-45-174
the words:
(1)
(2)
(3)
"CAUTION - 	HIGH INTENSITY X-RAY BEAM", 	or words having
a 	similar intent, 	on the x-ray source housing; 	and
"CAUTION RADIATION - 	THIS EQUIPMENT PRODUCES RADIATION
WHEN ENERGIZED", 	or words having a 	similar intent,
near any switch that energizes an x-ray tube if the
radiation source is an x-ray tube; 	or
"CAUTION - 	RADIOACTIVE MATERIAL", 	or words having a
similar intent, 	on the source housing in accordance
with subchapter 4 if the radiation source is a
radionuclide.
(g) 	on open-beam configurations installed after the
effective date of these rules, 	each port on the radiation source
housing shall be equipped with a 	shutter that cannot be opened
unless a 	collimator or a 	coupling has been connected to the port.
(h) 	Each radiation source housing shall be subject to the
following requirements:
(1) 	Each x-ray tube housing shall be equipped with an
interlock that shuts off the tube if it is removed
from the radiation source housing or if the housing is
disassembled; 	and
(2) 	Each radioactive source housing or port cover or each
x-ray tube housing shall be so constructed that, 	with
all shutters closed, 	the radiation measured at a
distance of five centimeters 	from its surface is not
capable of producing a dose in excess of 2.5 millirems
(0.025 millisievert) 	in one hour. 	For systems
utilizing x-ray tubes, 	this limit shall be met at any
specified tube rating.
(i) 	Each x-ray generator shall be supplied with a
protective cabinet which limits leakage radiation measured at a
distance of five centimeters from its surface such that it is not
capable of producing a dose in excess of 0 25 	• lirem 	(2.5
microsieverts) 	in one hour. 	[Eff NQV 	12 19 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	3 1-11(21), 	321-71)
§11-45-175 	Area requirements. 	(a) 	The local components of
an analytical x-ray system shall be located and arranged and shall
include sufficient shielding or access control such that no
radiation levels exist in any area surrounding the local component
group which could result in a dose to an individual present
therein in excess of the dose limits given in subchapter 4. 	For
systems utilizing x-ray tubes, 	these levels shall be met at any
specified tube rating.
(b) 	Radiation surveys, 	as 	required by subchapter 4, 	of all
analytical x-ray systems sufficient to show compliance with
subsection 	(a) 	shall be performed:
(1) 	Upon installation of the equipment, 	and at least once
every twelve months thereafter;
(2) 	Following any change in the initial arrangement,
number, 	or type of local components in the system;
(3) 	Following any maintenance requiring the disassembly or
removal of a 	local component in the system;
45-142 265
Complete printed PDF page 142
§11-45-177
(4) 	During the performance of maintenance and alignment
procedures if the procedures require the presence of a
primary x-ray beam when any local component in the
system is disassembled or removed;
(5) 	Any time a visual inspection of the local components
in the system reveals an abnormal condition; 	and
(6) 	Whenever personnel monitoring devices show a
significant increase over the previous monitoring
period or the readings are approaching the limits
specified in subchapter 4.
(c) 	Radiation survey measurements shall not be required if
a 	licensee can demonstrate compliance with subchapter 4 to the
satisfaction of the department.
(d) 	Each area or room containing analytical x-ray
equipment shall be conspicuously posted with a 	sign or signs
bearing the radiation symbol and the words 	"CAUTION - 	X-RAY
EQUIPMENT" 	or words :t1fJ4,~1Jg,,.,ar;;s~iF intent in accordance with
subchapter 4. 	[Eff 	hl lJ V 	J. l 	l:::f:::f~ 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-176 	Operating requirements. 	(a) 	Normal operating
procedures shall be written and available to all analytical x-ray
equipment workers. 	No individual shall be permitted to operate
analytical x-ray equipment in any manner other than that specified
in the procedures unless 	such individual has obtained written
approval of the radiation safety officer.
(b) 	No individual shall bypass a safety device or
interlock unless such individual has obtained the approval of the
radiation safety officer. 	such approval shall be for a 	specified
period of time. 	When a 	safety device or interlock has been
bypassed, 	a 	readily discernible sign bearing the words 	"SAFETY
DEVICE NOT WORKING", 	or words having a similar intent, 	shall be
placed on the radiation source housing.
(c) 	Except as specified in subsection 	(b), 	no operation
involving removal of covers, 	shielding materials or tube housings
or modifications to shutters, 	collimators, 	or beam stops 	shall be
performed without ascertaining that the tube is off and shall
remain off until safe conditions have been restored. 	The main
switch, 	rather than interlocks, 	shall be used for 	routine shutdown
in preparation for 	repairs.
(d) 	Radioactive source housings shall be opened for 	source
replacement, 	leak testing, 	or other maintenance or repair
procedures only by individuals authorized to specifically conduct
such procedures licensed by the U.S. 	Nuclear Regulatory
Commission, 	an agreement state, 	or a 	licensing state.
[Eff 	1 2 1909 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp 	-'§§32l-t, 	321-11(21), 	321-71)
§11-45-177 	Personnel requirements. 	(a) 	No individual
shall be permitted to operate or maintain analytical x-ray
equipment unless such individual has 	received instruction in and
demonstrated competence as 	to:
45-143
Complete printed PDF page 143
§11-45-177
(1) 	Identification of radiation hazards associated with
the use of the equipment;
(2) 	Significance of the various 	radiation warning, 	safety
devices, 	and interlocks incorporated into the
equipment, 	or the reasons they have not been installed
on certain pieces of equipment and the extra
precautions required in such cases;
(3) 	Proper operating procedures for the equipment;
(4) 	Recognition of symptoms of an acute localized
exposure; 	and
(5) 	Proper procedures 	for 	reporting an actual or suspected
exposure.
(b) 	Finger or wrist dosimetric devices shall be provided
to and shall be used by:
(1) 	Analytical x-ray equipment workers using systems
having an open-beam configuration and not equipped
with a safety device; 	and
(2) 	Personnel maintaining analytical x-ray equipment if
the maintenance procedures require the presence of a
primary x-ray beam when any local component in the
analytical x-ray system is disassembled or removed.
(c) 	Reported dose values shall not be used for the purpose
of determining compliance with subchapter 4 ~r~ fv~luated by a
qualified health or medical physicist. 	[Eff NU 'I/ 	_ I,, 1990 	](Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	t21-
ll (21), 	321-71)
SUBCHAPTER 9
PARTICLE ACCELERATORS
§11-45-178 	Purpose and scope. 	(a) 	This subchapter
establishes 	requirements for the non-healing arts use of particle
accelerators. 	The requirements of this subchapter are in addition
to, 	and not in subs~~R\tfo,p 10.«'ig~ther applicable requirements in
this chapter. 	[Effl~Uv 	1 i"'1 I~ 	::.t 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-179 	General 	requirements. 	(a) 	The licensee shall
be qualified by reason of training and experience to use the
accelerator in such a manner as to minimize danger to public
health and safety or property.
(b) 	The licensee's equipment, 	facilities, 	and operating and
emergency procedures shall be adequate to protect health and
minimize danger to public health and safety or property.
(c) 	The licensee shall appoint a 	radiation safety officer.
(d) 	The licensee and the licensee's staff shall have
experience in the use of particle accelerators and training for
application to its intended uses.
(e) 	The licensee shall establish a 	radiation safety
committee to approve, 	in advance, 	proposals 	for uses of particle
accelerators, 	whenever deemed necessary by the department.
45-144 2
Complete printed PDF page 144
§11-45-183
(f) 	The licensee shall have a traini~~~~fo~rfim
1~'rm
operators of particle accelerators. 	[Eff 	!~UV 	1 i(,, 	1:j;;J~ 	]
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-180 	Reserved.
§11-45-181 	Limitations. 	(a) 	No licensee shall permit an
individual to act as an operator of a particle accelerator until
such individual:
(1) 	Has been instructed in radiation safety and has
demonstrated an understanding thereof;
(2) 	Has received copies of and instruction in this
subchapter and the applicable requirements of this
chapter, 	pertinent license conditions and the
licensee's operating and emergency procedures, 	and
shall have demonstrated understanding thereof; 	and
(3) 	Has demonstrated competence to use the particle
accelerator, 	related equipment, 	and survey instruments
which shall be employed.
(b) 	The radiation safety committee or the radiation safety
officer shall have the authority to terminate the operations at a
particle accelerator facility if such action is deemed necessary
to 	e 1
dwi 	ho public health and safety or property.
[Eff 	_;_ 2 	::J 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-182 	Shielding and safety design requirements. 	(a)
A qualified health or medical physicist shall be consulted in the
design of a particle accelerator installation and called upon to
perform a 	radiation survey when the accelerator is first capable
of producing radiation.
(b) 	Each particle accelerator installation shall be
provided with such primary and secondary barriers as are necessary
to assure compliance with subchapter 4. 	[Eff NQ\/ 	I 2 1900 	]
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS ~§3,1-1, 	~~-
11 (21), 	321-71)
§11-45-183 	Particle accelerator controls and interlock
systems. 	(a) 	Instrumentation, 	readouts, 	and controls on the
particle accelerator control console shall be clearly identified
and easily discernible.
(b) 	Each entrance into a 	target room or other high
radiation area shall be provided with a 	safety interlock that
shuts down the accelerator under conditions of barrier
penetration.
(c) 	Each safety interlock shall be on a circuit which
shall allow it to operate independently of all other safety
interlocks.
(d) 	All 	safety interlocks shall be designed so that any
45-145
265
Complete printed PDF page 145
§11-45-183
defect or component failure in the safety interlock system
prevents operation of the accelerator.
{e) 	When a 	safety interlock system has been tripped, 	it
shall only be possible to resume operation of the accelerator by
manually resetting controls at the position where the safety
interlock has been tripped and, 	lastly, 	at the main control
console.
{f) 	A scram button or other emergency power cutoff switch
shall be located and easily identifiable in all high radiation
areas. 	Such a 	cutoff switch shall include a manual 	reset so that
the accelerator cannot be restarted from the accelerator control
console ~~t~o'\:lt ?'let\ntting the cutoff switch.
[Eff 	r~Ov 	1, 1::;::,~ 	] 	{Auth: 	HRS §§321-10, 	321-11, 	321-71)
{Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-184 	Warning devices. 	{a) 	Each location designated
as high radiation area, 	and each entrance to such location, 	shall
be equipped with easily observable warning lights that operate
when, 	and only when, 	radiation is being produced.
(b) 	Except in facilities designed for human exposure, 	each
high radiation area shall have an audible warning device which
shall be activated for 	fifteen seconds before the possible
creation of such high radiation area. 	Such warning device shall
be clearly discernible in all high radiation areas.
{c) 	Barriers, 	temporary or otherwise, 	and pathways leading
to high radiation areas shall be ~.osted in accordance with
subchapter 4. 	[Eff MIQ\/ 1 ? jO 	~ 	] 	{Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	{Imp: 	1:J:~ 	1§§Tit'-1,.:I 	91-11 (21), 	321-71)
§11-45-185 	Operating procedures. 	{a) 	Particle
accelerators, 	when not in operation, 	shall be secured to prevent
unauthorized use.
{b) 	The safety interlock system shall not be used to turn
off the accelerator beam except in an emergency.
(c) 	All 	safety and warning devices, 	including interlocks,
shall be checked for proper operation at intervals not to exceed
three months. 	Results of such tests 	shall be maintained at the
accelerator facility for inspection by the department for three
years.
{d) 	Electrical circuit diagrams of the accelerator and the
associated safety interlock systems shall be kept current and
maintained for inspection by the department and shall be available
to the operator at each accelerator facility.
{e) 	If, 	for any reason, 	it is necessary to intentionally
bypass a 	safety interlock or interlocks, 	such action shall be:
(1) 	Authorized by the radiation safety committee or
radiation safety officer;
(2) 	Recorded in a permanent log and a notice posted at the
accelerator control console; 	and
(3) 	Terminated as soon as possible.
{f) 	A copy of the current operating and the emergency
procedures shall be maintained at the accelerator control panel.
45-146 26
Complete printed PDF page 146
[Eff
(Imp:
1..L 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
HRS §§321-1, 	321-11(21), 	321-71)
§11-45-187
§11-45-186 	Radiation monitoring requirements. 	(a) 	There
shall be available at each particle accelerator facility
appropriate portable monitoring equipment which is operable and
has been appropriately calibrated for the radiation being produced
at the facility. 	Such equipment shall be tested for proper
operation daily and calibrated at intervals not to exceed one year
and after each servicing and repair.
(b) 	A radiation protection survey shall be performed and
documented by a 	qualified health or medical physicist when changes
have been made in shielding, 	operation, 	equipment, 	or occupancy of
adjacent areas.
(c) 	Radiation levels in all high radiation areas 	shall be
continuously monitored. 	The monitoring devices 	shall be
electrically independent of the accelerator control and safety
interlock systems and capable of providing a 	readout at the
control panel.
(d) 	All area monitors shall be calibrated at intervals not
to exceed one year and after each servicing and repair.
(e) 	Whenever applicable, 	periodic surveys shall be made to
determine the amount of airborne particulate radioactivity
present.
(f) 	Whenever applicable, 	periodic smear surveys shall be
made to determine the degree of contamination.
(g) 	All surveys shall be made in accordance with the
written procedures established by a 	qualified health or medical
physicist or the radiation safety officer.
(h) 	Records of all radiation protection surveys,
calibrations, 	and instrumentation tests shall be maintained at the
accelerator f~Al}t¥ ~oig-Hrrpection by the department for three
years. 	[Eff ,~uv 	1 l 	I ~::, 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-187 	Ventilation systems. 	(a) 	Ventilation systems
shall be provided to ensure that personnel entering any area where
airborne radioactivity may be produced shall not be exposed to
airborne radioactive material in excess of those limits specified
in subchapter 4.
(b) 	A licensee shall not vent, 	release, 	or otherwise
discharge airborne 	radioactive material to an unrestricted area,
except as authorized by subchapter 4. 	Concentrations may be
averaged over a period not greater than one year. 	Every effort
should be made to maintain releases of radioactive material to
unrei~flffef tj-r~A~s low as is reasonably achievable.
[Eff !'3lJ\f 	!. (, 	l;j;j::, 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
45-147 65
Complete printed PDF page 147
§11-45-188
Subchapter 10
NOTICES, 	INSTRUCTIONS, 	REPORTS, 	AND INSPECTIONS
§11-45-188 	Purpose and scope. 	This subchapter establishes
requirements for notices, 	instructions, 	and reports by licensees
to individuals engaged in activities under a license and options
available to such individuals in connection with department
inspections of licensees to ascertain compliance with this
chapter. 	This subchapter applies to all persons who receive,
possess, 	use, 	own, 	or transfer s~i-½~ 7 s 1
o~ 5~ 	tion licensed
pursuant to subchapter 2. 	[Eff 1'4 UV 	Ji. 2 l!:f 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-189 	Posting of notices to workers. 	(a) 	Each
licensee shall post current copies of the following documents:
(1) 	This subchapter and subchapter 4;
(2) 	License and documents incorporated into the license by
reference and amendments 	thereto;
(3) 	The operating procedures applicable to activities
under the license; 	and
(4) 	Any violation involving radiological working
conditions, 	or order issued pursuant to this chapter,
and any response from the licensee.
(b) 	If posting of a document specified in subsection 	(a)
is not practicable, 	the licensee may post a notice which describes
the document and states where it may be examined.
(c) 	Department form "Notice to Employees" 	shall be posted
by each licensee as required by this 	subchapter.
(d} 	Department documents posted pursuant to subsection 	(a)
shall be posted within five working days after receipt of the
documents 	from the department; 	the licensee's 	response, 	if any,
shall be posted within five working days after dispatch from the
licensee. 	Such documents shall remain posted for a minimum of
five working days or until action correcting the violation has
been completed, 	whichever is later.
(e) 	Documents, 	notices, 	or forms posted pursuant to this
section shall appear in a 	sufficient number of places to permit
individuals engaged in work under the license to observe them on
the way to or from any particular work location to which the
document applies, 	shall be 	and shall be replaced if
defaced or altered. 	[Eff 	] 	(Auth: 	HRS §§321-
10, 	321-11, 	321-71) 	(Imp: 	1-11(21), 	321-71)
§11-45-190 	Instructions to workers. 	(a) 	All individuals
likely to 	receive an occupational dose:
(1) 	Shall be kept informed of the storage, 	transfer, 	or
use of sources of radiation in such portions of the
restricted area;
(2) 	Shall be instructed in the health protection problems
associated with exposure to radiation or radioactive
material to the individual and potential offspring, 	in
45-148
6
Complete printed PDF page 148
§11-45-191
precautions or procedures to minimize exposure, 	and in
the purposes and functions 	of protective devices
employed;
(3) 	Shall be instructed in, 	and instructed to observe, 	to
the extent within the worker's control, 	the applicable
provisions of this chapter for the protection of
personnel from exposures to radiation or radioactive
material occurring in such areas;
(4) 	Shall be instructed of their responsibility to report
promptly to the licensee any condition which may
constitute, 	lead to, 	or cause a violation of this
chapter, 	or cause unnecessary exposure to radiation or
radioactive material;
(5) 	Shall be instructed in the appropriate response to
warnings made in the event of any unusual occurrence
or malfunction that may involve exposure to radiation
or radioactive material; 	and
(6) 	Shall be advised as to the radiation exposure reports
which workers shall be furnished pursuant to section
11-45-191.
(b) 	The extent of these instructions 	shall be commensurate
with potential radiologifP\\ih~aht~ti(\gtection problems in the
restricted area. 	[Eff l~UV 	1 l 	I~~ 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-191 	Notifications and reports to individuals. 	(a)
Radiation exposure data for an individual and the 	results of any
measurements, 	analyses, 	and calculations of radioactive material
deposited or retained in the body of an individual shall be
reported to the individual as specified in this section. 	The
information reported shall include data and results obtained
pursuant to this chapter, 	orders, 	or license conditions, 	as shown
in records maintained by the licensee pursuant to section 11-45-
83. 	Each notification and report shall:
(1) 	Be in writing;
(2) 	Include appropriate identifying data such as the name
of the licensee, 	the name of the individual, 	and the
individual's identification number, 	preferably social
security number;
(3) 	Include the individual's exposure information; 	and
(4) 	Contain the following statement:
"This report is 	furnished to you under the provisions
of chapter 45. 	You should preserve this report for
further reference."
(b) 	Each licensee shall advise each worker annually of the
worker's dose as shown in records maintained by the licensee
pursuant to section 11-45-83.
(c) 	Each licensee shall furnish a 	report of the worker's
exposure to sources of radiation at the request of a worker
formerly engaged in activities controlled by the licensee. 	The
report shall include the dose record for each year the worker was
required to be monitored pursuant to section 11-45-54. 	Such
report 	shall be furnished within thirty days 	from date the 	request
45-149 65
Complete printed PDF page 149
§11-45-191
was made, 	or within thirty days after the dose of the individual
has been determined by the licensee, 	whichever is later. 	The
report shall cover the period of time that the worker's activities
involved exposure to sources of radiation and shall include the
dates 	and locations of work under the license in which the worker
participated during this period.
(d) 	When a licensee is required pursuant to section 11-45-
90 to 	report to the department any exposure of an individual to
sources of radiation, 	the licensee shall also provide the
individual a 	report on the exposure data included therein. 	Such
reports 	shall be transmitted at a 	time not later than the
transmittal to the department.
(e) 	At the request of a worker who is terminating
employment with the licensee in work involving exposure to
radiation or radioactive material, 	during the current year, 	each
licensee shall provide at termination to each such worker, 	or to
the worker's designee, 	a written report regarding the radiation
dose 	received by that worker from operations of the licensee
during the current year or fraction thereof. 	If the most recent
individual monitoring results are not available at that time, 	a
written estimate of the dose shall be provided together with a
clear ~MtraiiRn ~l)..at this is an estimate.
[Eff 	NUV 	• {,, 1!:::f!:19 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-192 	Presence of representatives of licensees and
workers during inspection. 	(a) 	Each licensee shall afford to the
department at all reasonable times opportunity to inspect
materials, 	machines, 	activities, 	facilities, 	premises, 	and records
pursuant to this chapter.
(b) 	During an inspection, 	department inspectors may
consult privately with workers as specified in section 11-45-193.
The licensee may accompany department inspectors during other
phases of an inspection.
(c) 	If, 	at the time of inspection, 	an individual has been
authorized by the workers to represent them during department
inspections, 	the licensee shall notify the inspectors of such
authorization and shall give the workers' 	representative an
opportunity to accompany the inspectors during the inspection of
physical working conditions.
(d) 	Each workers' 	representative shall be routinely
engaged in work under control of the licensee and shall have
received instructions as specified in section 11-45-190.
(e) 	Different representatives of licensees and workers may
accompany the inspectors during different phases of an inspection
if there is no resulting interference with the conduct of the
inspection. 	However, 	only one workers' 	representative at a time
may accompany the inspectors.
(f) 	With the approval of the licensee and the workers'
representative, 	an individual who is not routinely engaged in work
under control of the licensee, 	for example, 	a 	consultant to the
licensee or to the workers' 	representative, 	shall be afforded the
opportunity to accompany department inspectors during the
45-150
65
Complete printed PDF page 150
§11-45-194
inspection of physical working conditions.
(g) 	Notwithstanding the other provisions of this section,
department inspectors are authorized to refuse to permit
accompaniment by any individual who deliberately interferes with a
fair 	and orderly inspection. 	With regard to any area containing
proprietary information, 	the workers' 	representative for that area
shall be an individual ije.vious:j,y authorized by the licensee to
enter that area. 	[Eff 	I nv 	1 "/ 199a 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	~s §§:32"1.-t 	3"21-11 (21), 	321-71)
§11-45-193 	Consultation with workers during inspections.
(a) 	Department inspectors may consult privately with workers
concerning matters of occupational radiation protection and other
matters 	related to applicable provisions of this chapter to the
extent the inspectors deem necessary for the conduct of an
effective and thorough inspection.
(b) 	During the course of an inspection, 	any worker may
bring privately to the attention of the inspectors, 	either orally
or in writing, 	any past or present condition which the worker has
reason to believe may have contributed to or caused any violation
of this chapter, 	or license condition, 	or any unnecessary exposure
of an individual to sources of radiation under the licensee's
control. 	Any such notice in writing shall comply with the
requirements of section 11-45-194.
(c) 	The provisions of subsection
inte 7preted as authorizatif~,to
section 11-45-190. 	[Eff l'I IJ ,/
321-11, 	321-71) 	(Imp: 	HRS §§321-1,
(b) 	shall not be
instructions pursuant to
] 	(Auth: 	HRS §§321-10,
11 (21), 	321-71)
§11-45-194 	Requests by workers 	for inspections. 	(a) 	Any
worker or representative of workers believing that a violation of
this 	chapter, 	or license conditions exists or has occurred in work
under a 	license with regard to radiological working conditions in
which the worker is engaged may request an inspection by giving
the alleged violation to the department. 	Any such notice shall be
in writing, 	shall set forth the ~pecific grounds 	for the notice,
and shall be signed by the worker or representative of the
workers. 	A copy shall be provided to the licensee by the
department no later than at the time of inspection except that,
upon the request of the worker giving such notice, 	such worker's
name and the name of individuals referred to therein shall not
appear in such copy or on any record published, 	released, 	or made
available by the department.
(b) 	If, 	upon receipt of such notice, 	the department
determines that the complaint meets the requirements set forth in
subsection 	(a), 	and that there are 	reasonable grounds to believe
that the alleged violation exists or has occurred, 	an inspection
shall be made as soon as practicable to determine if such alleged
violation exists or has occurred. 	Inspections pursuant to this
section need not be limited to matters 	referred to in the
complaint.
(c) 	No licensee, 	or contractor or subcontractor of a
45-151
65
Complete printed PDF page 151
§11-45-194
licensee shall discharge or in any manner discriminate against any
worker because such worker has 	filed any complaint or instituted
or caused to be instituted any proceeding under this chapter or
has 	testified or is about to testify in any such proceeding or
because of the exercise by such worker on behalf of such worker or
others of any option afforded by this subchapter.
[Eff 	Nnv 1 ? 1aaa 	l 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	•~s §§321-~7~21-ll (21), 	321-71)
§11-45-195 	Inspections not warranted; 	informal review. 	(a)
If the department determines, 	with respect to a complaint under
section 11-45-194, 	that an inspection is not warranted because
there are no 	reasonable grounds to believe that a violation exists
or has occurred, 	the department shall notify the complainant in
writing of such determination. 	The complainant may obtain review
of such determination by submitting a written statement of
position with the department. 	The department shall provide the
licensee with a 	copy of such statement by certified mail,
excluding, 	at the request of the complainant, 	the name of the
complainant. 	The licensee may submit an opposing written
statement of position with the department. 	The department shall
provide the complainant with a copy of such statement by certified
mail.
(b) 	Upon the request of the complainant, 	the department
may hold an informal conference in which the complainant and the
licensee may orally present their views. 	An informal conference
may also be held at the request of the licensee, 	but disclosure of
the 	identity of the complainant shall be made only following
receipt of written authorization from the complainant. 	After
considering all written and oral views presented, 	the department
shall affirm, 	modify, 	or reverse the determination and furnish the
complainant and the licensee a written notification of the
decision and the reason thereof.
(c) 	If the department determines that an inspection is not
warranted because the requirements of section ll-45-194(a) 	have
not been met, 	the complainant shall be notified in writing of such
determination. 	Such determination shall be without prejudice to
the 	filing of a new fltoA1~ia1·n~ ~~7'Wng the requirements of section
11-45-194 (a). 	[Eff nu\/ 	· l, 	1::1~~ 	] 	(Auth: 	HRS §§321-10, 	321-
11, 	321-71) 	(Imp: 	HRS '§§321-1, 	321-11 (21), 	321-71)
§11-45-196 	Reserved.
§11-45-197 	Reserved.
§11-44-198 	Reserved.
§11-44-199 	Reserved.
45-152 	65
Complete printed PDF page 152
§11-45-201
SUBCHAPTER 11
HEALTH PHYSICS SERVICES AND MEDICAL PHYSICS SERVICES
§11-45-200 	Purpose and scope. 	This 	subchapter establishes
requirements 	for health physics services and medical physics
services and the use of these services for 	radiation facilities.
The requirements of this subchapter apply to all persons who
provide health physics services or medical physics services and to
all licensees 	requiring these services. 	[Eff NQV 	·J 	l
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§TI 	,
11 (21), 	321-71)
§11-45-201 	General requirements. 	(a) 	All persons
licensed for health physics services shall have at least one
qualified health physicist certified by the American Board of
Health Physics in the appropriate fields 	or specialties in which
services are provided.
(b) 	All persons licensed for medical physics services shall
have at least one qualified medical physicist with qualifications
in the appropriate fields 	or specialties in which services are
provided.
(c)
(1)
(2)
Qualified medical physicists shall be:
Medical physicists certified in an appropriate field
or specialty by The American Board of Radiology; 	or
The American Board of Medical Physics; 	or The American
Board of Science in Nuclear Medicine; 	or The Canadian
College of Physicists in Medicine; 	or
Within six months after the effective date of these
rules, 	medical physicists who are not certified in an
appropriate specialty by one of the boards specified
in paragraph 	(1) 	but have the following 	formal
education, 	training, 	and experience:
(A) 	Master or doctorate degree in a physical science
or engineering with not less than thirty
semester hours or equivalent in college level
physics and/or radiation science with at least
one year of full 	time work experience in an
appropriate field under the supervision of a
qualified medical physicist who meets the
requirements specified in paragraph 	(1); 	and one
year of training in an appropriate field in
medical physics; 	or
(B) 	Bachelor degree in a physical science or
engineering with at least five years of full
time work experience in an appropriate field
under the supervision of a 	qualified medical
physicist who meets the 	requirements specified
in paragraph 	(1); 	and one year of training in an
appropriate field in medical physics.
(d) 	Qualified medical physicists 	for mammography shall be
medical physicists who are certified by the American Board of
45-153
265
Complete printed PDF page 153
§11-45-201
Radiology or the American Board of Medical 	Physics, 	and who meet
the 	requirements specified by the Mammography Quality Standards
Act of 1992, 	Public Law 102-539, 	42 U.S.C. 	263b, 	and 21 C.F.R.
Part 900.
(e) 	All surveys, 	audits, 	reports, 	and other work performed
by a 	health physics service or medical physics service for
licensed facilities 	shall be reviewed and signed by a qualified
health physicist or a qualified medical physicist, 	respectively.
[Eff 	_1 :) jQQn 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	~~1 ~P~21-11(21), 	321-71)
§11-44-202 	Records. 	(a) 	All health physics services or
medical physics services shall have records of services provided
to licensed facilities with sources of radiation.
(b) 	Records shall be maintained for inspection by the
department for three years. 	[Efff\l(lV 	1 ') ◄!gQO 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	1HR'S §§-3!!- 	, ~1-11 (21), 	321-71)
§11-44-203 	Use of medical physics services. 	(a) 	In
addition to specifications requiring the use of a qualified
medical physicist pursuant to this chapter, 	medical physics
services shall be utilized by licensees who provide the following
services:
(1)
(2)
Mammography;
Diagnostic and therapeutic use of radioactive
materials;
(3) 	Therapeutic use of sources of radiation; 	or
(4) 	Fluoroscopy.
(b) 	To assure radiation protection and quality control
pursuant to this chapter, 	licensees providing services specified
in subsection 	(a) 	shall have surveys physically conducted at least
annually by, 	or under the direction of, 	a 	qualified medical
physicist.
(c) 	Records of surveys shall be maintained by the licensee
for i41~~ctio1hb;; the department for three years.
[Eff 	NU\/ 	l ~ 	1P99 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-~, 	321-11(21), 	321-71)
§11-44-204 	Reserved.
SUBCHAPTER 12
TRANSPORTATION OF RADIOACTIVE MATERIAL
§11-45-205 	Purpose and scope. 	This subchapter establishes
requirements 	for transportation of radioactive material and
applies to any person who transports 	radioactive material or
deliVJfn\~a'f<nactiv~ material to a carrier for transport.
[Eff 	1\l U ,J 	l 	l 	199::1 ] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
45-154 265
Complete printed PDF page 154
§11-45-209
§11-45-206 	General requirements. 	(a) 	Each licensee who
transports 	radioactive material outside of the confines of the
licensee's premises or other place of use, 	or who delivers
radioactive material to a 	carrier for transport, 	shall:
(1) 	Comply with the regulations of the U.S. 	Department of
Transportation; 	and
(2) 	Assure that any special instructions needed to safely
open the package are sent to or have been made
available to the consignee.
(b) 	If, 	for any reason, 	the regulations of the U.S.
Department of Transportation are not applicable to a 	shipment of
licensed radioactive material, 	the licensee shall conform to the
standards and requirements of those regulations to the same extent
as if~lffilsl'Jti.P{U~HflS subject to the regulations.
[Eff 	i'eU V l 7, 	1~::,~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
SUBCHAPTER 13
WIRELINE SERVICE OPERATIONS AND SUBSURFACE TRACER STUDIES
§11-45-207 	Purpose and scope. 	This subchapter establishes
requirements 	for using sources of radiation for wireline service
operations including mineral-logging, 	radioactive markers, 	and
subsurface tracer studies. 	The requirements of this subchapter
are in addition to, 	and not in substifi'\\~01 tor
9other applicable
requirements of this chapter. 	[Eff NUV 	tG 1 991 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-208 	Exemptions. 	Sources of radiation licensed by
U.S. 	Nuclear Regulatory Commission are exempt from the
requirements of this subchapter to the extent they are regulated
by th\ 1
U.S. pu.clear Regulatory Commission.
[Eff 	rJOV 	-, 2 1gaa 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1'.~~21-11 (21), 	321-71)
§11-45-209 	Prohibition. 	No licensee shall perform wireline
service operations with a 	sealed source(s) 	unless, 	before
commencement of the operation, 	the licensee has a written
agreement with the well-operator, 	well-owner, 	drilling contractor,
or land owner that:
(1) 	In the event a sealed source is lodged downhole, 	a
reasonable effort at recovery shall be made; 	and
(2) 	In the event a decision is made to abandon the sealed
source downhole, 	the requir~~i,\~s 	section 11-45-
231(c) 	shall be met. 	[Eff 	NIJI/ 	199Q 	] 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS "'§§321-1,
321-11(21), 	321-71)
45-155
Complete printed PDF page 155
§11-45-210
§11-45-210 	Limits on levels of radiation. 	Sources of
radiation shall be used, 	stored, 	and transported in such a manner
that the transportation requirements of subchapter 12 and the dose
limit¾t;¼~ rfC:Wi~n1Alfts of subchapter 4 are met.
[Eff 	NUV 	i{, 	1::J~:j 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-211 	Storage precautions. 	(a) 	Each source of
radiation, 	except accelerators, 	shall be provided with a 	storage
or transport container. 	The container shall be provided with a
lock, 	or tamper seal for calibration sources, 	to prevent
unauthorized removal of, 	or exposure to, 	the source of radiation.
(b) 	Sources of radiation shall be stored in a manner which
shall i{f~~mi~e dan~er from explosion or fire.
[Eff 	r'luv 	12 10:::tO 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-Y, ~21-11(21), 	321-71)
§11-45-212 	Transport precautions. 	Transport containers
shall be physically secured to the transporting vehicle to prevent
accidental loss, 	taA1-pering, 	or unauthorized removal.
[Eff NQV 	1 2 199~ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp:'"HR~ §§321-1, 	321-11(21), 	321-71)
§11-45-213 	Radiation survey instruments. 	(a) 	The licensee
shall maintain sufficient calibrated and operable radiation survey
instruments at each field station to make physical radiation
surveys as 	required by this subchapter and subchapter 4. 	Instru-
mentation shall be capable of measuring 0.1 milliroentgen 	(25.8
nanocoulombs per kilogram) 	per hour through at least
fifty milliroentgens 	(12.9 microcoulombs per kilogram) 	per hour.
Survey instruments acquired before the effective date of this
chapter and capable of measuring 0.1 milliroentgen
(25.8 nanocoulombs per kilogram) 	per hour through at least
twenty milliroentgens 	(5.16 microcoulombs per kilogram) 	per hour
also satisfies this 	requirement 	five years after the effective
date of this chapter.
(b) 	Each radiation survey instrument shall be calibrated:
(1) 	At intervals not to exceed six months and after each
instrument servicing;
(2) 	For linear scale instruments, 	at two points located
approximately one-third and two-thirds of full-scale
on each scale; 	for logarithmic scale instruments, 	at
midrange of each decade, 	and at two points of at least
one decade; 	and for digital instruments, 	at
appropriate points; 	and
(3) 	So that accuracy within twenty percent of the true
radiation level can be demonstrated on each scale.
(c) 	Calibration records 	s~~ii be maintained at the
facility for three years. 	[Eff 	NOV 	121990] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRs'§§!21-, 	3'21-11(21), 	321-71)
45-156 	26
Complete printed PDF page 156
§11-45-215
§11-45-214 	Leak testing of sealed sources. 	(a) 	Each
licensee using sealed sources of radioactive material shall have
the sources tested for leakage. 	Records of leak test results
shall be kept in units of microcuries or becquerels and maintained
for inspection by the department for three years.
(b) 	Tests for leakage shall be performed only by persons
specifically authorized to perform such tests by the department.
The test sample shall be taken from the surface of the source,
source holder, 	or from the surface of the device in which the
source is stored or mounted and on which one might expect
contamination to accumulate. 	The test sample shall be analyzed
for 	radioactive contamination, 	and the analysis shall be capable
of detecting the presence of 0.005 microcurie 	(one hundred eighty-
five becquerels) 	of radioactive material on the test sample.
(c) 	Each sealed source of radioactive material shall be
tested at intervals not to exceed six months. 	In the absence of a
certificate from a 	transferor indicating that a test has been made
before the transfer, 	the sealed source shall not be put into use
until 	tested. 	If, 	for any reason, 	it is suspected that a 	sealed
source may be leaking, 	it shall be removed from service
immediately and tested for leakage as soon as practical.
(d) 	If the test reveals the presence of 0.005 microcurie
(one hundred eighty-five becquerels) 	or more of leakage or
contamination, 	the licensee shall immediately withdraw the source
from use and shall cause it to be decontaminated, 	repaired, 	or
disposed of in accordance with this chapter. 	A report describing
the equipment involved, 	the test results, 	and the corrective
action taken shall be filed with the department within five days
of receiving the test results.
(e) 	The following sources are exempt from the periodic
leak test requirements of subsections 	(a) 	through 	(d):
(1) 	Sources of radioactive material with a half-life of
thirty days or less;
(2) 	Sealed sources of radioactive material in gaseous
form;
(3) 	Sources of beta- or gamma-emitting radioactive
material with an activity of one hundred microcuries
(3.7 megabecquerels) 	or less; 	and
(4) 	Sources of alpha-emitting radioactive material with an
activity of ten microcuries 	(0.370 megabecquerels) 	or
less. 	[Eff Nnv 1 2 1900 	l 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(fmp: 	~ §§321-1, 	321-11(21), 	321-
71)
§11-45-215 	Quarterly inventory. 	Each licensee shall
conduct a 	quarterly physical inventory to account for all sources
of radiation. 	Records of inventories shall be maintained for
inspection by the department for three years and shall include the
quantities and kinds of sources of radiation, 	the location where
sources of radiation are assigned, 	the date of the inventory, 	and
the 	fW ~~vidual conducting the inventory.
[Eff 	.x. Z Jl8~:::J 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
45-157 265
Complete printed PDF page 157
§11-45-216
§11-45-216 	Utilization records. 	Each licensee shall
maintain current records, 	which shall be kept available for
inspection by the department 	for three years, 	showing the
following information for each source of radiation:
(1) 	Make, 	model number, 	and a 	serial number, 	or a
description of each source of radiation used;
(2) 	The identity of the well-logging supervisor or field
unit to whom assigned;
(3) 	Locations where used and dates of use; 	and
(4) 	In the case of tracer materials and radioactive
markers, 	the utilization record shall indicate the
radiopp~l~de and activity used in a particular well.
[Eff !1'lUV 	1? 1000 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp":' 	rf~§§321-1, 	321-11 (21), 	321-71)
§11-45-217 	Design, 	performance, 	and certification criteria
for sealed sources used in downhole operations. 	(a) 	Each sealed
source, 	except those containing radioactive material in gaseous
form, 	used in downhole operations and manufactured one year after
the effective date of this chapter shall be certified by the
manufacturer, 	or other testing organization acceptable to the
department, 	to meet the following minimum criteria:
(1) 	Be of doubly encapsulated construction;
(2) 	Contain radioactive material whose chemical and
physical forms 	are as insoluble and non-dispersible as
practical; 	and
(3) 	Has been individually pressure tested to at least
24,656 pounds per square inch absolute 	(one hundred
seventy MN/m2 ) 	without failure.
(b) 	For sealed sources, 	except those containing
radioactive material in gaseous 	form, 	acquired one year after the
effective date of this chapter, 	in the absence of a 	certificate
from a 	transferor certifying that an individual sealed source
meets the requirements of subsection 	(a), 	the sealed source shall
not be put into use until such determinations and testing have
been performed.
(c) 	Each sealed source, 	except those containing
radioactive material in gaseous form, 	used in downhole operations
two years after the effective date of this chapter shall be
certified by the manufacturer, 	or other testing organization
acceptable to the department, 	as meeting the sealed source
performance requirements for oil well-logging as contained in the
American National Standard N43.6, 	"Classification of Sealed
Radioactive Sources," 	formerly N542, 	ANSI/NBS 126.
(d) 	Certification documents shall be maintained for
inspection by the department for three years. 	If the source is
abandoned downhole, 	the certification documents shall be
mainti~~~ 	department authorizes disposition.
[Eff IH,n/ 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
45-158
Complete printed PDF page 158
§11-45-221
§11-45-218 	Labeling. 	(a) 	Each source, 	source holder, 	or
logging tool containing radioactive material shall bear a durable,
legible, 	and clearly visible marking or label, 	which has, 	as a
minimum, 	the standard radiation caution symbol, 	without the
conventional color requirement, 	and the 	following wording:
DANGER 	(OR CAUTION)
RADIOACTIVE
This labeling shall be on the smallest component transported as a
separate piece of equipment.
(b) 	Each transport container shall have permanently
attached to it a durable, 	legible, 	and clearly visible label which
has, 	as a minimum, 	the standard radiation caution symbol and the
following wording:
[Eff
(Imp:
DANGER 	(OR CAUTION)
RADIOACTIVE
NOTIFY CIVIL AUTHORITIES 	[OR NAME OF COMPANY]
1 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
HRS §§321-1, 	321-11(21), 	321-71)
§11-45-219 	Inspection and maintenance. 	(a) 	Each licensee
shall conduct, 	at intervals not to exceed six months, 	a program of
inspection and maintenance of source holders, 	logging tools,
source handling tools, 	storage containers, 	transport containers,
and injection tools 	to assure proper labeling and physical
condition. 	Records of inspection and maintenance shall be
maintained for inspection by the department for three years.
(b) 	If any inspection conducted pursuant to subsection 	(a)
reveals damage to labeling or components critical to 	radiation
safety, 	the device shall be removed from service until 	repairs
have been made.
(c) 	If a 	sealed source is stuck in the source holder, 	the
licensee shall not perform any operation, 	such as drilling,
cutting, 	or chiseling, 	on the source holder unless the licensee is
specifically approved by the department.
(d) 	The repair, 	opening, 	or modification of any sealed
source shall be performed only by [rsons 	fically authorized
to do so by the department. 	[Eff 	Q\I 	1 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	~§3~ 	11(21), 	321-71)
§11-45-220 	Reserved.
§11-45-221 	Training requirements. 	(a) 	No licensee shall
permit any individual to act as a logging supervisor as defined in
this chapter until such individual has:
(1) 	Received, 	in a 	course recognized by the department,
45-159
Complete printed PDF page 159
§11-45-221
instruction in the subjects outlined in Appendix A of
this subchapter and demonstrated an understanding
thereof;
(2) 	Read and received instruction in the requirements
contained in this subchapter and the applicable
requirements of this chapter, 	license conditions, 	and
the licensee's operating and emergency procedures, 	and
demonstrated an understanding thereof; 	and
(3) 	Demonstrated competence to use sources of radiation,
related handling tools, 	and radiation survey
instruments which shall be used on the 	job.
(b) 	No licensee shall permit any individual to assist in
the handling of sources of radiation until such individual has:
(1) 	Read or received instruction in the licensee's
operating and emergency procedures and demonstrated an
understanding thereof; 	and
(2) 	Demonstrated competence to use, 	under the personal
supervision of the logging supervisor, 	the sources of
radiation, 	related handling tools, 	and radiation
survey instruments which shall be used on the 	job.
(c) 	The licensee shall maintain employee training records
for inspection by the department for three years.
(d) 	The exhibit at the end of chapter 11-45 entitled,
"Appendix A of Subchapter 13, 	Subjects to be Included in Training
Courses 	for Logg~\ ~u~e,.i;-vj.A~H 	(2/2/93) ", 	is made a part of this
section. 	[Eff 	l"iU i/ 	l i::, 	I~~:=, 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-222 	Operating and emergency procedures. 	The
licensee's operating and emergency procedures shall include
instructions in at least the following:
(1) 	Handling and use of sources of radiation to be
employed so that no individual is likely to be exposed
to radiation doses in excess of the standards
established in subchapter 4;
(2) 	Methods and occasions 	for conducting radiation
surveys;
(3) 	Methods and occasions 	for locking and securing sources
of radiation;
(4) 	Personnel monitoring and the use of personnel
monitoring equipment;
(5) 	Transportation to temporary jobsites and field
stations, 	including the packaging and placing of
sources of radiation in vehicles, 	placarding of
vehicles, 	and securing sources of radiation during
transportation;
(6) 	Minimizing exposure of individuals in the event of an
accident;
(7) 	Procedure for notifying proper personnel in the event
of an accident;
(8) 	Maintenance of records;
(9) 	Use, 	inspection and maintenance of source holders,
logging tools, 	source handling tools, 	storage
45-160 	2
Complete printed PDF page 160
§11-45-226
containers, 	transport containers, 	and injection tools;
(10) 	Procedure to be followed in the event a 	sealed source
is lodged downhole;
(11) 	Procedures to be used for picking up, 	receiving, 	and
opening packages containing radioactive material;
(12) 	For the use of tracers, 	decontamination of the
environment, 	equipment, 	and personnel;
(13) 	Maintenance of records generated by logging personnel
at temporary jobsites;
(14) 	Notifying proper persons in the event of an accident;
and
(15) 	Actions to be taken if a 	sealed source is ruptured,
including actions to prevent the spread of
contamination and minimize inhalation and ingestion of
radioactive material and actions to obtain suitable
radiation survey lll-~IllPW:5.c~~ required by section
11-45-213. 	[Eff 	I'! U \f 	l 	(, 	1::1!:f9 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-
11 (21), 	321-71)
§11-45-223 	Personnel monitoring. 	(a) 	No licensee 	shall
permit any individual to act as a logging supervisor or to assist
in the handling of sources of radiation unless each such
individual wears either a 	film badge or a 	thermoluminescent
dosimeter 	(TLD). 	Each film badge or TLD shall be assigned to and
worn by only one individual. 	Film badges shall be replaced at
least monthly and TLDs replaced at least quarterly. 	After
replacement, 	each film badge or TLD shall be promptly processed.
(b) 	Personnel monitoring records shall be maintained for
inspectt:1)~1?/ bv :}h..fhf¾Wartment for three years.
[Eff 	I\JU • 	l ,v 	f:;:J:;:J::J 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-224 	Security. 	During each logging or tracer
application, 	the logging supervisor or other designated employee
shall maintain direct surveillance of the operation to protect
against unauthorized or unnecessar~. entry into a 	restricted area,
as defined in this chapter. 	[Eff 	I\Jn\J 1 2 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HR'.s4§§321•.:!1 	(21), 	321-71)
§11-45-225 	Handling tools. 	The licensee shall provide and
require the use of tools that shall assure remote handling of
~ov~~~'ll-nthan low-activity calibration sources.
] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
HRS §§321-1, 	321-11(21), 	321-71)
§11-45-226 	Subsurface tracer studies. 	(a) 	Protective
gloves and other appropriate protective clothing and equipment
shall be used by all personnel handling radioactive tracer
material. 	Precautions shall be taken to avoid ingestion or
45-161 2
Complete printed PDF page 161
§11-45-226
inhalation of radioactive material.
(b) 	No licensee shall cause tl\E;. d.pp eftJolh~~ radioactive
material into potable aquifers. 	[Eff N 	V 	l Z I~~~ 	] 	(Auth:
HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21),
321-71)
§11-45-227 	Particle accelerators. 	No licensee shall permit
above-ground testing of particle accelerators, 	designed for use in
well-logging, 	which results in the production of radiation, 	except
in areas or facilities 	so controlled or shielded that the
requirements of subchapter 4, 	as applicable, 	are met.
[Eff 	1 9 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	§§3"2 	1-11(21), 	321-71)
§11-45-228 	Radiation surveys. 	(a) 	Radiation surveys or
calculations shall be made and recorded for each area where
radioactive materials are used and stored.
(b) 	Radiation surveys shall be made and recorded for the
radiation levels in occupied positions and on the exterior of each
vehicle used to transport radioactive material. 	Such surveys
shall include each source of radiation or combination of sources
to be transported in the vehicle.
(c) 	If the sealed source assembly is removed from the
logging tool before departing the 	jobsite, 	the logging tool
detector shall be energized, 	or a 	survey meter used, 	to assure
that the logging tool is 	free of contamination.
(d) 	Radiation surveys shall be made and recorded at the
jobsite or well-head for each tracer operation. 	These surveys
shall include measurements of radiation levels before and after
the operation.
(e) 	Records 	required pursuant to subsections 	(a) 	through
(d) 	shall include the dates, 	the identification of individual(s)
making the survey, 	the identification of survey instrument(s)
used, 	and an exact description of the location of the survey.
Records of these surveys shall be ~~iQ~ained for inspection by the
department for three years. 	[Eff NOV 	1 2 1909 	] 	(Auth: 	HRS
§§321-10, 	321-11, 	321-71) 	(Imp: 	HRS ~§3~1-1, 	t21-11(21), 	321-71)
§11-45-229 	Documents and records 	required at field
stations. 	Each licensee shall maintain, 	for inspection by the
department, 	the following documents and records 	for the specific
devices and sources used at the field station:
(1) 	License;
(2) 	Operating and emergency procedures;
(3) 	Chapter 11-45;
(4) 	Records of the latest survey instrument calibrations
pursuant to section 11-45-213;
(5) 	Records of the latest leak test results pursuant to
section 11-45-214;
(6) 	Records of quarterly inventories required pursuant to
section 11-45-215;
45-162
5
Complete printed PDF page 162
(7)
(8)
(9)
(10)
§11-45-231
Utilization records 	required pursuant to section 11-
45-216;
Records of inspection and maintenance required
pursuant to section 11-45-219;
Survey records 	required pursuant to section 11-45-228;
and
Training rR~r~s,
1
r~q~~hEW pursuant to section 11-45-
221. 	[Eff1'<UV 	it, 	l:.:f::1::J 	) 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-
71)
§11-45-230 	Documents and records required at temporary
jobsites. 	Each licensee conducting operations at a temporary
jobsite shall have the following documents and records available
at that site for inspection by the department:
(1) 	Operating and emergency procedures;
(2) 	Survey records 	required pursuant to section 11-45-228
for the period of operation at the site;
(3) 	Evidence of current calibration for the radiation
(4)
(5)
survey instruments in use at the site;
Copy of license; 	and
Shipping papers for the.trans~ortation of radioactive
material. 	[Eff Mnv 1 2 19a 	] 	(Auth: 	HRS §§321-
10, 	321-11, 	321..:7:r') 	r.tmp: 	'rt s §§321-1, 	321-11 (21),
321-71)
§11-45-231 	Notification of incidents, 	abandonment, 	and lost
sources. 	(a) 	Notification of incidents and sources lost in other
than downhole logging operations shall be made in accordance with
appropriate provisions of subchapter 4.
(b) 	Whenever a 	sealed source or device containing
radioactive material is lodged downhole, 	the licensee shall:
(1) 	Monitor at the surface for the presence of radioactive
contamination with a 	radiation survey instrument or
logging tool during logging tool 	recovery operations;
and
(2) 	Notify the department immediately by telephone and
subsequently, 	within thirty days, 	by confirmatory
letter if the licensee knows or has 	reason to believe
that a 	sealed source has been ruptured. 	This letter
shall identify the well or other location, 	describe
the magnitude and extent of the escape of 	radioactive
material, 	assess the consequences of the rupture, 	and
explain efforts planned or being taken to mitigate
these consequences.
(c) 	When it becomes apparent that efforts to 	recover the
radioactive source shall not be successful, 	the licensee shall
(1) 	Advise the well-operator of an appropriate method of
abandonment, 	which shall include:
(A) 	The immobilization and sealing in place of the
radioactive source with a cement plug,
(B) 	The setting of a whipstock or other deflection
45-163
Complete printed PDF page 163
§11-45-231
device, 	and
(C) 	The mounting of a permanent identification
plaque at the surface of the well, 	containing
the appropriate information required by
subsection 	{d);
(2) 	Notify the department by telephone, 	giving the
circumstances of the loss, 	and request approval of the
proposed abandonment procedures; 	and
(3) 	File a written report with the department within
thirty days of the abandonment. 	The licensee shall
send a copy of the report to the agency who issued
permits or otherwise approved of the drilling
operation. 	The report shall contain the following
information:
(A) 	Date of occurrence;
(B) 	A description of the well logging source
involved, 	including the radionuclide and its
quantity, 	chemical, 	and physical form;
(C) 	Surface location and identification of the well;
(D) 	Results of efforts 	to immobilize and seal the
source in place;
(E) 	A brief description of the attempted recovery
effort;
(F) 	Depth of the source;
(G) 	Depth of the top of the cement plug;
(H) 	Depth of the well;
(I) 	Any other information, 	such as a warning
statement, 	contained on the permanent
identification plaque; 	and
(J) 	The names of state agencies 	receiving a 	copy of
this 	report.
(d) 	Whenever a 	sealed source containing radioactive
material is abandoned downhole, 	the licensee shall provide a
permanent plaque for posting the well or well-bore. 	This plaque
shall:
(1)
(2)
Be constructed of long-lasting material, 	such as
stainless steel or monel; 	and
Contain the following information engraved on its
face:
(A)
(B)
(C)
(D)
(E)
(F)
The word "CAUTION";
The radiation symbol without the conventional
color requirement;
The date of abandonment;
The name of the well-operator or well-owner;
The well name and well identification number(s)
or other designation;
The sealed source(s) 	by radionuclide and
activity;
(G) 	The source depth and the depth to the top of the
plug; 	and
(H) 	An appropriate warning, 	depending on the
specific circumstances of each abandonment.
(e) 	The licensee shall immediately notify the department
by telephone and subsequently by confirming letter if the licensee
45-164
265
Complete printed PDF page 164
§11-45-233
knows or has 	reason to believe that 	radioactive material has been
lost in or to an underground potable aquifer. 	such notice shall
designate the well location and shall describe the magnitude and
extent of loss of radioactive material, 	assess the consequences of
such loss, 	and explain ef¾~¾fgP}a;;i.ni~ W being taken to mitigate
these consequences. 	[Eff f\luv 	l' 	1~9~ 	] 	(Auth: 	HRS §§321-
10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
SUBCHAPTER 14
THERAPEUTIC RADIATION MACHINES
§11-45-232 	Purpose and scope. 	This subchapter establishes
requirements 	for the use of therapeutic radiation machines. 	The
requirements of this subchapter are in addition to, 	and not in
subst~t
0
utfop forn 	other applicable requirements of this chapter.
[Eff N ,v 	I 2 i::,09 	] 	(Auth: 	HRS §§321-10, 	321-11, 	321-71)
(Imp: 	HRS s§321-t', 	321-11(21), 	321-71)
§11-45-233 	General administrative requirements 	for
facilities using therapeutic radiation machines. 	(a) 	The
licensee shall be responsible for directing the operation of the
therapeutic radiation machine. 	The licensee or the licensee's
agent shall ensure that the requirements of this subchapter are
met in the operation of the therapeutic radiation machine(s).
(b) 	A therapeutic radiation machine which does not meet
the provisions of this subchapter shall not be used for
irradiation of patients.
(c) 	The licensee for any therapeutic radiation machine
subject to section 11-45-236 or section 11-45-238 shall 	require
the authorized user to be a physician who:
(1) 	Is certified in:
(A) 	Radiology or therapeutic radiology by the
American Board of Radiology; 	or
(B) 	Radiation oncology by the American Osteopathic
Board of Radiology; 	or
(C) 	Radiology, 	with specialization in radiotherapy,
as a British "Fellow of the Faculty of
Radiology" 	or "Fellow of the Royal College of
Radiology"; 	or
(D) 	Therapeutic radiology by the Canadian Royal
College of Physicians and Surgeons; 	or
(2) 	Is in the active practice of therapeutic radiology,
and has completed two hundred hours of instruction in
basic radiation techniques applicable to the use of an
external beam radiation therapy unit, 	five hundred
hours of supervised work experience, 	and a minimum of
three years of supervised clinical experience.
(A) 	To satisfy the 	requirement for instruction, 	the
classroom and laboratory training shall include:
(i) 	Radiation physics and instrumentation;
(ii) 	Radiation protection;
45-165
2 	5
Complete printed PDF page 165
§11-45-233
(iii) 	Mathematics pertaining to the use and
measurement of radioactivity; 	and
(iv) 	Radiation biology.
(B) 	To satisfy the requirement for supervised work
experience, 	training shall be under the
supervision of an authorized user and shall
include:
(i) 	Review of the full 	calibration
measurements and periodic quality
assurance checks;
(ii) 	Preparing treatment plans and calculating
treatment times;
(iii) 	Using administrative controls to prevent
misadministrations;
(iv) 	Implementing emergency procedures to be
followed in the event of the abnormal
operation of a external beam radiation
therapy unit or console; 	and
(v) 	Checking and using radiation survey
meters.
(C) 	To satisfy the requirement for a period of
supervised clinical experience, 	training shall
include one year in a 	formal 	training program
approved by the Residency Review Committee for
Radiology of the Accreditation Council 	for
Graduate Medical Education or the Committee on
Postdoctoral Training of the American
Osteopathic Association and an additional two
years of clinical experience in therapeutic
radiology under the supervision of an authorized
user. 	The supervised clinical experience shall
include:
(i) 	Examining individuals and reviewing their
case histories to determine their
suitability for external beam radiation
therapy treatment, 	and any
limitations/contraindications;
(ii) 	Selecting proper dose and how it is to be
administered;
(iii) 	Calculating the external beam radiation
therapy doses and collaborating with the
authorized user in the review of patients'
progress and consideration of the need to
modify originally prescribed doses as
warranted by patients' 	reaction to
radiation; 	and
(iv) 	Post-administration follow-up 	and review
of case histories.
(3) 	Notwithstanding the requirements of subparagraphs 	(A)
and 	(B), 	the licensee for any therapeutic 	radiation
machine less than five hundred kilovolts may also
submit the training of the prospective authorized user
physician for department review on a case-by-case
basis.
45-166
Complete printed PDF page 166
§11-45-233
(4) 	A physician shall not act as an authorized user for
any therapeutic radiation machine until such time as
said physician's training has been reviewed and
approved by the department.
{d) 	The licensee for any photon therapy systems 	(five
hundred kilovolts and above) 	and electron therapy systems 	(five
hundred kilo electron volts and above) 	shall utilize the services
of a 	qualified medical physicist in radiation therapy.
(e) 	Individuals operating a 	therapeutic radiation machine
and practicing radiation therapy technology shall be licensed
under chapter 11-44.
(f) 	Written safety procedures shall be developed by a
qualified medical physicist and shall be available at the control
area of a therapeutic radiation machine, 	including any
restrictions required for the safe operation of the particular
therapeutic radiation machine. 	The operator shall be able to
demonstrate 	familiarity with these rules.
{g) 	Individuals shall not be exposed to the useful beam
except for medical therapy purposes and unless such exposure has
been ordered in writing by a 	licensed practitioner of the healing
arts 	facility. 	This provision specifically prohibits deliberate
exposure of an individual for training, 	demonstration or other
non-healing arts purposes.
{h) 	Notwithstanding the provisions of subsection 	(c), 	a
licensee may permit any physician to act as a visiting authorized
user for up to sixty days per calendar year under the following
conditions:
(1)
(2)
(i)
therapeutic
comply with
program.
The visiting authorized user has the prior written
permission of the licensee's management and, 	if the
use occurs on behalf of an institution, 	the
institution's radiation safety committee; 	and
The visiting authorized user meets the requirements
established for authorized user{s) 	in subsection 	(c).
All individuals associated with the operation of a
radiation machine shall be instructed in and shall
the provisions of the licensee's quality management
{j) 	The licensee shall maintain the following information
in a separate file or package for each therapeutic radiation
machine:
(1)
(2)
(3)
(4)
Report of acceptance testing;
Records of all surveys, 	calibrations, 	and periodic
quality assurance checks of the therapeutic radiation
machine required by this subchapter, 	as well as the
name{s) 	of person{s) 	who performed such activities;
Records of major maintenance and/or modifications
performed on the therapeutic radiation machine after
the effective date of this chapter, 	as well as the
name(s) 	of person(s) 	who performed such services;
Signature of person authorizing the return of
therapeutic radiation machine to cli-ffiffl
1 use after
service, 	repair, 	or upgrade. 	[Eff NU\I 	..1. 2 1aa9(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	f:r'~
§§321-1, 	321-11(21), 	321-71)
45-167 2 6 5
Complete printed PDF page 167
§11-45-234
§11-45-234 	General technical 	requirements 	for 	facilities
using therapeutic radiation machines. 	(a) 	Protection surveys.
(1) 	The licensee shall ensure that radiation protection
surveys of all new facilities, 	and existing facilities
not previously surveyed are performed with an operable
radiation measurement survey instrument calibrated in
accordance with this subchapter. 	The radiation
protection survey shall be performed by, 	or under the
direction of, 	a qualified medical physicist and shall
verify that, 	with the therapeutic radiation machine in
a 	"BEAM-ON" condition, 	with the largest clinically
available treatment field and with a 	scattering
phantom in the useful beam of radiation:
(A) 	Radiation levels in restricted areas are not
likely to cause personnel exposures in excess of
the limits specified in subchapter 4; 	and
(B) 	Radiation levels in unrestricted areas do not
exceed the limits specified in subchapter 4.
(2) 	In addition to the requirements of paragraph 	(1), 	a
radiation protection survey shall also be performed
before any subsequent medical use and:
(A) 	After making any change in the treatment room
shielding;
(B) 	After making any change in the location of the
therapeutic radiation machine within the
treatment room;
(C) 	After relocating the therapeutic 	radiation
machine; 	or
(D) 	Before using the therapeutic radiation machine
in a manner that could result in increased
radiation levels in areas outside the external
beam radiation therapy treatment room.
(3) 	The survey record shall indicate all instances where
the facility, 	in the opinion of the qualified medical
physicist, 	is in violation of applicable chapters.
The survey record shall also include the date of the
measurements, 	the reason the survey is required, 	the
manufacturer's name, 	model number and serial number of
the therapeutic radiation machine, 	the instrument(s)
used to measure radiation levels, 	a plan of the areas
surrounding the treatment room that were surveyed, 	the
measured dose rate at several points in each area
expressed in millirems 	(microsieverts) 	per hour, 	the
calculated maximum level of radiation over a period of
one week for each restricted and unrestricted area,
and the signature of the individual responsible for
conducting the survey;
(4) 	If the results of the surveys required by subsection
(a) 	indicate any radiation levels in excess of the
respective limit specified in paragraph 	(1), 	the
licensee shall lock the control in the "OFF" position
and not use the unit;
(A) 	Except as may be necessary to 	repair, 	replace,
or test the therapeutic radiation machine, 	the
45-168
Complete printed PDF page 168
§11-45-234
therapeutic radiation machine shielding, 	or the
treatment room shielding; 	or
(B) 	Until the licensee has 	received a 	specific
exemption from the department.
(b) 	If the survey required by subsection 	(a) 	indicates
that an individual in an unrestricted area may be exposed to
levels of radiation greater than those permitted by subchapter 4,
before beginning the treatment program the licensee shall:
(1) 	Either equip the unit with beam direction interlocks
or add additional radiation shielding to ensure
compliance with subchapter 4;
(2) 	Perform the survey required by subsection 	(a); 	and
(3) 	Include in the report 	required by subsection 	(d), 	the
results of the initial survey, 	a description of the
modification made to comply with paragraph 	(1), 	and
the results of the second survey.
(c) 	Dosimetry equipment.
(1) 	The licensee shall have a calibrated dosimetry system
available for use. 	The system shall have been
calibrated for Cobalt-60 by the National Institute of
Standards and Technology or by an American Association
of Physicists in Medicine Accredited Dosimetry
Calibration Laboratory. 	The calibration shall have
been performed within the previous twenty-four months
and after any servicing that may have affected system
calibration;
(2) 	The licensee shall have available for use a dosimetry
system for quality assurance check measurements. 	To
meet this 	requirement, 	the system may be compared with
a 	system that has been calibrated in accordance with
paragraph 	(1). 	This comparison shall have been
performed within the previous twelve months 	(six
months if the dosimetry system is an ionization
chamber) 	and after each servicing that may have
affected system calibration. 	The quality assurance
check system may be the same system used to meet the
requirement in paragraph 	(1);
(3) 	The licensee shall maintain a 	record of each dosimetry
system calibration, 	intercomparison, 	and comparison
for the duration of the license. 	For each
calibration, 	intercomparison, 	or comparison, 	the
record shall include the date, 	the model numbers and
serial numbers of the instruments that were
calibrated, 	intercompared, 	or compared as 	required by
paragraphs 	(1) 	and 	(2), 	the correction factors 	that
were determined, 	the names of the individuals who
performed the calibration, 	intercomparison, 	or
comparison, 	and evidence that the intercomparison was
performed by, 	or under the direct supervision of, 	a
qualified medical physicist.
(d) 	The licensee for any therapeutic radiation machine
subject to section 11-45-236 or section 11-45-238 shall furnish a
copy of the records 	required in subsections 	(a) 	and 	(b) 	to the
department within thirty days 	following completion of the action
45-169
6
Complete printed PDF page 169
§11-45-234
that initiated the record requirement.
(Auth: 	HRS §§321-10, 	321-11, 	321-71)
11 (21), 	321-71)
[Eff
(Imp:
]
HRS §§321-1, 	321-
§11-45-235 	Quality management program. 	(a) 	Each applicant
or licensee subject to section 11-45-236 or section 11-45-238
shall establish and maintain a written quality management program
to provide high confidence that radiation shall be administered as
directed by the authorized user. 	The quality management program
shall include written policies and procedures to meet the
following specific objectives:
(1) 	Before administration, 	a written directive is prepared
for any external beam radiation therapy dose;
(A) 	Notwithstanding this paragraph, 	a written
revision to an existing written directive may be
made provided that the revision is dated and
signed by an authorized user before the
administration of the external beam radiation
therapy dose or the next external beam radiation
therapy fractional dose;
(B) 	Notwithstanding this paragraph, 	if, 	because of
the patient's condition, 	a delay in order to
provide a written revision to an existing
written directive would jeopardize the patient's
health, 	an oral revision to an existing written
directive shall be acceptable, 	provided that the
oral revision is documented immediately in the
patient's record and a 	revised written directive
is signed by an authorized user within forty-
eight hours of the oral revision;
(C) 	Notwithstanding this paragraph, 	if, 	because of
the emergent nature of the patient's condition,
a delay in order to provide a written directive
would jeopardize the patient's health, 	an oral
directive shall be acceptable, 	provided that the
information contained in the oral directive is
documented immediately in the patient's record
and a written directive is prepared and signed
by an authorized user within twenty-four hours
of the oral directive.
(2) 	Before the administration of each course of radiation
treatments, 	the patient's identity is verified, 	by
more than one method, 	as the individual named in the
written directive;
(3) 	External beam radiation therapy final 	plans of
treatment and related calculations are in accordance
with the respective written directives;
(4) 	Each administration is in accordance with the written
directive; 	and
(5) 	Any unintended deviation from the written directive is
identified and evaluated, 	and appropriate action is
taken.
(c) 	Development of quality management program.
45-170
Complete printed PDF page 170
§11-45-235
(1) 	Each application for a 	license subject to section 11-
45-236 or section 11-45-238 shall develop a 	quality
management program as part of the application required
by subchapter 2. 	The licensee shall implement the
program upon issuance of a license by the department;
(2) 	Each existing licensee subject to sections 11-45-236
and 11-45-238 shall, 	within thirty days of the
effective date of this chapter, 	submit to the
department a written certification that a 	quality
management program has been implemented.
(d) 	As a part of the quality management program, 	the
licensee shall:
(1) 	Develop procedures 	for, 	and conduct a 	review of, 	the
quality management program including, 	since the last
review, 	an evaluation of a 	representative sample of
patient administrations, 	all recordable events, 	and
all misadministrations to verify compliance with all
aspects of the quality management program;
(2) 	Conducted these reviews at intervals not to exceed
(3)
(4)
(e)
days after
event by:
(1)
(2)
(3)
twelve months;
Evaluate each of these reviews to determine the
effectiveness of the quality management program and,
if required, 	make modifications to meet the
requirements of subsection 	(b); 	and
Maintain records of each review, 	including the
evaluations and findings 	of the review.
The licensee shall evaluate and respond, 	within thirty
discovery of the recordable event, 	to each recordable
Assembling the relevant facts including the cause;
Identifying what, 	if any, 	corrective action is
required to prevent recurrence; 	and
Retaining a 	record, 	in an auditable form, 	of the
relevant facts 	and what corrective action, 	if any, 	was
taken.
(f) 	The licensee shall retain:
(1) 	Each written directive; 	and
(2) 	A record of each administered radiation dose, 	in an
auditable form.
(g) 	The licensee may make modifications to the quality
management program to increase the program's efficiency provided
the program's effectiveness is not decreased.
(h) 	The licensee shall evaluate each misadministration and
shall take the following actions in response to a
misadministration:
(1) 	Notify the department by telephone no later than the
next calendar day after discovery of the
misadministration;
(2) 	Submit a written report to the department within
fifteen days after discovery of the misadministration.
The written report shall include: 	the licensee's
name; 	the prescribing physician's name; 	a brief
description of the event; 	why the event occurred; 	the
effect on the patient; 	what improvements are needed to
45-171 2265
Complete printed PDF page 171
§11-45-235
prevent recurrence; 	actions taken to prevent
recurrence; 	whether the licensee notified the patient
or the patient's responsible relative or guardian
(this person shall subsequently be referred to as 	"the
patient"), 	and if not, 	why not, 	and if the patient was
notified, 	what information was provided to the
patient. 	The report shall not include the patient's
name or other information that could lead to
identification of the patient;
(3) 	Notify the referring physician and also notify the
patient of the misadministration no later than twenty-
four hours after its discovery, 	unless 	the referring
physician personally informs the licensee either that
he/she shall inform the patient or that, 	based on
medical 	judgment, 	telling the patient would be
harmful. 	The licensee is not required to notify the
patient without first 	consulting the referring
physician. 	If the referring physician or patient
cannot be reached within twenty-four hours, 	the
licensee shall notify the patient as soon as possible
thereafter. 	The licensee shall not delay any
appropriate medical care for the patient, 	including
any necessary remedial care as a 	result of the
misadministration, 	because of any delay in
notification;
(4) 	Retain a 	record of each misadministration. 	The record
shall contain the names of all individuals involved
(including the prescribing physician, 	allied health
personnel, 	the patient, 	and the patient's 	referring
physician), 	the patient's social security number or
identification number if one has been assigned, 	a
brief description of the event, 	why it occurred, 	the
effect on the patient, 	what improvements are needed to
prevent recurrence, 	and the actions taken to prevent
recurrence; 	and
(5) 	If the patient was notified, 	furnish, 	within fifteen
days after discovery of the misadministration, 	a
written report to the patient by sending either a copy
of the 	report that was submitted to the department, 	or
a brief description of both the event and the
consequences as they may effect the patient, 	provided
a 	statement is included that the report submitted to
the department can be obtained from the licensee;
(i) 	Aside from the notification requirement, 	nothing in
subsection 	(h) 	affects any rights or duties of licensees and
physicians in relation to each other, 	patients, 	or the patient's
responsible relatives or guardians. 	[Eff NQ\/ 	1 9 10Q9 	]
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§3Wv, 	321-
11 (21), 	321-71)
§11-45-236 	Therapeutic radiation machines of less than five
hundred kilovolts.
(a) 	When the x-ray tube is operated at its maximum rated
45-172
265
Complete printed PDF page 172
§11-45-236
tube current for the maximum kilovolts, 	the leakage air kerma rate
shall not exceed the value specified at the distance specified for
that classification of therapeutic radiation machine:
(1) 	Five to fifty kilovolt systems: 	The leakage air kerma
rate measured at any position five centimeters from
(2)
the tube housing assembly shall not exceed one hundred
miilirad 	(one milligray) 	in any one hour.
Greater than fifty and less than five hundred kilovolt
systems: 	The leakage air kerma rate measured at a
distance of one meter from the source in any direction
shall not exceed one rad 	(one centigray) 	in any one
hour. 	This air kerma rate measurement may be averaged
over areas no larger than one hundred square
centimeters. 	In addition, 	the air kerma 	rate at a
distance of five centimeters from the surface of the
tube housing assembly shall not exceed thirty rad
(thirty centigray) 	per hour.
(b) 	Permanent beam limiting devices: 	Permanent diaphragms
or cones used for limiting the useful beam shall provide at least
the same degree of attenuation as 	required for the tube housing
assembly.
(c)
(1)
(2)
(d)
(1)
(2)
(3)
(4)
Adjustable or removable beam limiting devices:
All adjustable or removable beam limiting devices,
diaphragms, 	cones or blocks shall not transmit more
than five percent of the useful beam for the most
penetrating beam used;
When adjustable beam limiting devices are used, 	the
position and shape of the radiation field shall be
indicated by a 	light beam.
The filter system shall be so designed that:
Filters can not be accidentally displaced at any
possible tube orientation;
For equipment installed after the effective date of
this chapter, 	an interlock system prevents irradiation
if the proper filter is not in place;
The air kerma rate escaping from the 	filter slot shall
not exceed one rad 	(one centigray) 	per hour at one
meter under any operating conditions; 	and
Each filter shall be marked as to its material of
construction and its thickness.
(e) 	Tube immobilization:
(1) 	The x-ray tube shall be so mounted that it can not
accidentally turn or slide with respect to the housing
aperture; 	and
(2) 	The tube housing assembly shall be capable of being
immobilized for 	stationary portal treatments.
(f) 	The tube housing assembly shall be so marked that it
is possible to determine the location of the source to within five
millimeters, 	and such marking shall be readily accessible 	for use
during calibration procedures.
(g) 	Contact therapy tube housing assemblies 	shall have a
removable shield of material, 	equivalent in attenuation to 0.5
millimeters of lead at one hundred kilovolts, 	which can be
positioned over the entire useful beam exit port during periods
45-173
Complete printed PDF page 173
§11-45-236
when the beam is not in use.
(h) 	A suitable irradiation control device shall be
provided to terminate the irradiation after a pre-set time
interval.
(1)
(2)
A timer which has a display shall be provided at the
treatment control panel. 	The timer shall have a
pre-set time selector and an elapsed time indicator;
The timer shall be a cumulative timer which activates
with an indication of "BEAM-ON" and retains its
reading after irradiation is interrupted or
terminated. 	After irradiation is terminated and
before irradiation can be reinitiated, 	it shall be
necessary to reset the elapsed time indicator;
(3) 	The timer shall terminate irradiation when a
pre-selected time has elapsed, 	if any dose monitoring
system present has not previously terminated
irradiation;
(4) 	The timer shall permit accurate pre-setting and
determination of exposure times as short as one
second;
(5) 	The timer shall not permit an exposure if set at zero;
(6) 	The timer shall not activate until the shutter is
opened when irradiation is controlled by a shutter
mechanism unless calibration includes a timer error
(7)
(i)
correction to compensate 	for mechanical lag; 	and
Timer shall be accurate to within one percent of the
selected value or one second, 	whichever is greater.
The control panel, 	in addition to the displays
required by
(1)
other provisions in this 	section, 	shall have:
An indication of whether electrical power is available
at the control panel and if activation of the x-ray
tube is possible;
tube:
(2)
(3)
(4)
(5)
(6)
(j)
An indication of whether x-rays are being produced;
Means for indicating x-ray tube potential and current;
The means for terminating an exposure at any time;
A locking device which shall prevent unauthorized use
of the therapeutic radiation machine; 	and
For therapeutic radiation machines manufactured after
the effective date of this chapter, 	a positive display
of specific filter(s) 	in the beam.
When a 	control panel may energize more than one x-ray
(1) 	It shall be possible to activate only one x-ray tube
at any time;
(2) 	There shall be an indication at the control panel
identifying which x-ray tube is activated; 	and
(3) 	There shall be an indication at the tube housing
assembly when that tube is energized.
(k) 	Target-to-skin distance 	(TSD): 	There shall be a means
of determining the central axis TSD to within one centimeter and
of reproducing this measurement to within two millimeters
thereafter.
(1) 	Unless it is possible to bring the x-ray output to the
prescribed exposure parameters within five 	seconds after the x-ray
45-174 	265
Complete printed PDF page 174
§11-45-236
"ON" switch is energized, 	the beam shall be attenuated by a
shutter having a 	lead equivalency not less than that of the tube
housing assembly. 	In addition, 	after the unit is at operating
parameters, 	the shutter shall be controlled electrically by the
operator from the control panel. 	An indication of shutter
position shall appear at the control panel.
(m) 	Each therapeutic radiation machine equipped with a
beryllium or other low-filtration window shall be clearly labeled
as such upon the tube housing assembly and shall be provided with
a permanent warning device on the control panel that is activated
when no additional 	filtration is present, 	to indicate that the
dose rate is very high.
(n) 	Facility design requirements 	for therapeutic radiation
machines capable of operating in the range fifty kilovolts to five
hundred kilovolts: 	In addition to shielding adequate to meet
requirements of section 11-45-241, 	the treatment room shall meet
the 	following design requirements:
(1) 	Provision shall be made for continuous two-way aural
communication between the patient and the operator at
the control panel;
(2) 	Provision shall be made to permit continuous
observation of the patient during irradiation and the
viewing system shall be so located that the operator
can observe the patient from the control panel. 	The
therapeutic radiation machine shall not be used for
patient irradiation unless at least one viewing system
is operational.
(o) 	Treatment rooms which contain a 	therapeutic radiation
machine capable of operating above one hundred fifty kilovolts
shall meet the following additional requirements:
(1) 	All protective barriers shall be fixed except for
entrance doors or beam interceptors;
(2) 	The control panel shall be located outside the
treatment room or in a totally enclosed booth, 	which
has a 	ceiling, 	inside the room;
(3) 	Interlocks shall be provided such that all entrance
doors, 	including doors to any interior booths, 	shall
be closed before treatment can be initiated or
continued. 	If the 	radiation beam is interrupted by
any door opening, 	it shall not be possible to 	restore
the machine to operation without closing the door and
reinitiating irradiation by manual action at the
control panel; 	and
(4) 	When any door referred to in paragraph 	(3) 	is opened
while the x-ray tube is activated, 	the air kerma rate
at a distance of one meter from the source shall be
reduced to less than one hundred millirad 	(one
milligray) 	per hour.
(p) 	Full calibration measurements:
(1) 	Full calibration of a therapeutic radiation machine
subject to section 11-45-236 shall be performed by, 	or
under the direct supervision of, 	a 	qualified medical
physicist:
(A) 	Before the first medical use following
45-175
265
Complete printed PDF page 175
§11-45-236
(2)
(3)
installation or reinstallation of the
therapeutic radiation machine;
(B) 	At intervals not exceeding one year; 	and
(C) 	Before medical use under the following
conditions:
(D)
(i) 	Whenever quality assurance check
measurements indicate that the radiation
output differs by more than five percent
from the value obtained at the last full
calibration and the difference cannot be
(ii)
easily discerned; 	and
Following any component replacement, 	major
repair, 	or modification of components that
could significantly affect the
characteristics of the radiation beam.
Notwithstanding the requirements of subparagraph
(C):
(i)
(ii)
Full calibration of therapeutic radiation
machines with multi-energy capabilities is
required only for those modes and/or
energies that are not within their
acceptable range; 	and
If the repair, 	replacement or modification
does not affect all energies, 	full
calibration shall be performed on the
affected energy that is in most frequent
clinical use at the facility. 	The
remaining energies may be validated with
quality assurance check procedures against
the criteria in clause 	(i).
To satisfy the requirement of paragraph 	(1), 	full
calibration shall include all measurements recommended
for annual calibration by NCRP Report 69, 	"Dosimetry
of X-Ray and Gamma Ray Beams for Radiation Therapy in
the Energy Range 10 keV to 50 MeV" 	(1981).
The licensee shall maintain a 	record of each
calibration for the duration of the license. 	The
record shall include the date of the calibration, 	the
manufacturer's name, 	model number, 	and serial number
for both the therapeutic radiation machine and the
x-ray tube, 	the model numbers and serial numbers of
the instruments used to calibrate the therapeutic
radiation machine, 	and the signature of the qualified
medical physicist responsible for performing the
calibration.
(q) 	Periodic quality assurance checks:
(1) 	Periodic quality assurance checks shall be performed
on therapeutic radiation machines subject to this
section, 	which are capable of operation at greater
than fifty kilovolts.
(2) 	To satisfy the requirement of paragraph 	(1), 	quality
assurance checks shall meet the following
requirements:
(A) 	The licensee shall perform quality assurance
45-176 265
Complete printed PDF page 176
§11-45-236
checks in accordance with written procedures
established by the qualified medical physicist;
and
(B) 	The quality assurance check procedures shall
specify the frequency at which tests or
measurements are to be performed. 	The quality
assurance check procedures shall specify that
the quality assurance check shall be performed
during the calibration specified in subsection
(p). 	The acceptable tolerance for each
parameter measured in the quality assurance
check, 	when compared to the value for that
parameter determined in the calibration
specified in subsection 	(p), 	shall be stated.
(3) 	The cause for a parameter exceeding a 	tolerance set by
the qualified medical physicist shall be investigated
and corrected before the system is used for patient
irradiation;
(4) 	Whenever a 	quality assurance check indicates a
significant change in the operating characteristics of
a 	system, 	as specified in the qualified medical
physicist's quality assurance check procedures, 	the
system shall be recalibrated as 	required in subsection
{p);
(5) 	The licensee shall use the dosimetry system described
in section 11-45-234(c) (2) 	to make the quality
assurance check required in paragraph 	(2);
(6) 	The licensee shall have the qualified medical
physicist review and sign the results of each
radiation output quality assurance check at intervals
not to exceed one month;
(7) 	Therapeutic radiation machines subject to this 	section
shall have safety quality assurance checks of each
external beam radiation therapy facility performed at
intervals not to exceed one month;
(8) 	To satisfy the requirement of paragraph 	(7), 	safety
quality assurance checks shall ensure proper operation
of:
(A) 	Electrical interlocks at each external beam
radiation therapy room entrance;
(B) 	Proper operation of the "BEAM-ON" and
termination switches;
(C) 	Beam condition indicator lights on the access
door(s), 	control console, 	and in the radiation
therapy room;
(D) 	Viewing systems;
(E) 	Electrically operated treatment room doors from
inside and outside the treatment room;
(9) 	The licensee shall promptly repair any system
identified in paragraph 	(8) 	that is not operating
properly.
(10) 	The licensee shall maintain a 	record of each quality
assurance check required by paragraphs 	(1) 	and 	(7) 	for
three years. 	The record shall include the date of the
45-177
2265
Complete printed PDF page 177
§11-45-236
quality assurance check, 	the manufacturer's name,
model number, 	and serial number for the therapeutic
radiation machine, 	the manufacturer's name, 	model
number and serial number of the instrument(s) 	used to
measure the radiation output of the therapeutic
radiation machine, 	and the signature of the individual
who performed the periodic quality assurance check.
(r) 	Operating procedures:
(1) 	The therapeutic radiation machine shall not be used
for irradiation of patients unless the requirements of
subsections 	(p) 	and 	(q) 	have been met;
(2) 	Therapeutic radiation machines shall not be left
unattended unless it is secured pursuant to subsection
(i);
(3) 	When a patient shall be held in position for 	radiation
therapy, 	mechanical supporting or restraining devices
shall be used;
(4) 	The tube housing assembly shall not be held by an
individual during operation unless 	the assembly is
designed to require such holding and the peak tube
potential of the system does not exceed fifty
kilovolts. 	In such cases, 	the holder shall wear
protective gloves and apron of not less than 0.5
millimeters lead equivalency at one hundred kilovolts;
(5) 	A copy of the current operating and emergency
procedures shall be maintained at the therapeutic
radiation machine control console; 	and
(6) 	No individual other than the patient shall be in the
treatment room during exposures 	from therapeutic
radiation machines operating above one hundred fifty
kilovolts. 	At energies less than or equal to one
hundred fifty 	kilovolts, 	any individual, 	other than
the patient, 	in the treatment room shall be protected
by a barrier sufficient to meet the requirements of
subchapter 4.
(s) 	Each facility location authorized to use a 	therapeutic
radiation machine in accordance with this section shall possess
appropriately calibrated portable monitoring equipment. 	As a
minimum, 	such equipment shall include a portable radiation
measurement survey instrument capable of measuring dose rates over
the 	range one millirem 	(ten microsieverts) 	per hour to one
thousand millirems 	(ten millisieverts) 	per hour. 	The survey
instrument(s) 	shall be Of\~-H,Ple~~d
1
~~¾brated in accordance with
section 11-45-240. 	[Eff l'JUV 	l (, 	!:j~~ 	] 	(Auth: 	HRS §§321-10,
321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	321-11(21), 	321-71)
§11-45-237 	Reserved.
§11-45-238 	Therapeutic radiation machines - 	photon therapy
systems 	(five hundred kilovolts and above) 	and electron therapy
45-178 	2 6 5,
Complete printed PDF page 178
§11-45-238
systems 	(five hundred kilo electron volts and above). 	(a) 	Each
facility location authorized to use a 	therapeutic radiation
machine in accordance with this section shall possess
appropriately calibrated portable monitoring equipment. 	As a
minimum, 	such equipment shall include a portable radiation
measurement survey instrument capable of measuring dose rates over
the 	range one millirem 	(ten microsieverts) 	per hour to one
thousand millirems 	(ten millisieverts) 	per hour. 	The survey
instrument(s) 	shall be operable and calibrated in accordance with
section 11-45-240.
(b) 	Leakage radiation outside the maximum useful beam in
photon and electron modes:
(1) 	The absorbed dose rate due to leakage radiation
(excluding neutrons) 	at any point outside the maximum
sized useful beam, 	but within a 	circular plane of
radius two meters which is perpendicular to and
centered on the central axis of the useful beam at the
nominal treatment distance 	(i.e. 	patient plane), 	shall
not exceed a maximum of 0.2 percent and an average of
0.1 percent of the absorbed dose rate on the central
axis of the beam at the nominal treatment distance.
Measurements shall be averaged over an area not
exceeding one hundred square centimeters at a minimum
of sixteen points uniformly distributed in the plane;
(2) 	Except for the area defined in paragraph 	(1), 	the
absorbed dose rate in water 	(excluding that 	from
neutrons) 	at one meter from the electron path between
the source and the target or electron window shall not
exceed 0.5 percent of the absorbed dose rate in water
on the central axis of the beam at the nominal
treatment distance. 	Measurements shall be averaged
over an area not exceeding one hundred square
centimeters at a minimum of sixteen points uniformly
distributed in the plane;
(3) 	The neutron absorbed dose rate outside the useful beam
shall be kept as low as practicable. 	Measurements of
the portion of the leakage radiation dose contributed
by neutrons shall be averaged over an area not
exceeding eight hundred square centimeters; 	and
(4) 	For each therapeutic radiation machine, 	the licensee
shall determine, 	or obtain from the manufacturer, 	the
leakage radiation existing at the positions specified
in paragraphs 	(1) 	through 	(3) 	for the specified
operating conditions. 	Records on leakage radiation
measurements shall be maintained at the installation
for inspection.
(c) 	Leakage radiation through beam limiting devices:
(1) 	Photon radiation: 	All adjustable or interchangeable
beam limiting devices 	shall attenuate the useful beam
such that at the nominal treatment distance, 	the
maximum absorbed dose anywhere in the area shielded by
the beam limiting device(s) 	shall not exceed two
percent of the maximum absorbed dose on the central
axis of the useful beam measured in a 	ten centimeters
45-179
265
Complete printed PDF page 179
§11-45-238
by ten centimeters radiation field;
(2) 	Electron radiation: 	All adjustable or interchangeable
electron applicators 	shall attenuate the radiation,
including but not limited to photon radiation
generated by electrons incident on the beam limiting
device and electron applicator and other parts of the
radiation head, 	such that the absorbed dose in a plane
perpendicular to the central axis of the useful beam
at the nominal treatment distance shall not exceed:
(A) 	A maximum of two percent of the absorbed dose on
the central axis of the useful beam at the
nominal treatment distance. 	This limit shall
apply beyond a 	line seven centimeters outside
the periphery of the useful beam; 	and
(B) 	A maximum of ten percent of the absorbed dose on
the central axis of the useful beam at the
nominal treatment distance. 	This limit shall
apply beyond a 	line two centimeters outside the
periphery of the useful beam.
(3) 	Measurement of leakage radiation:
(A) 	Photon radiation: 	Measurements of leakage
radiation through the beam limiting devices
shall be made with the beam limiting devices
closed and any residual aperture blocked by at
least two tenth value layers of suitable
absorbing material. 	In the case of overlapping
beam limiting devices, 	the leakage radiation
through each set shall be measured independently
at the depth of maximum dose. 	Measurements
shall be made using a 	radiation detector of area
not exceeding ten square centimeters;
(B) 	Electron radiation: 	Measurements of leakage
radiation through the electron applicators 	shall
be made with the electron beam directed into the
air and using a 	radiation detector of area up to
but not exceeding one square centimeter suitably
protected against radiation which has been
scattered from material beyond the radiation
detector. 	Measurements shall be made using one
centimeter of water equivalent build up
material.
(4) 	When adjustable beam limiting devices are used, 	the
position and shape of the radiation field shall be
indicated by a 	light field.
(d) 	Filters/wedges:
(1) 	Each filter and/or wedge which is removable from the
system shall be clearly marked with an identification
number. 	For removable wedge filters, 	the nominal
wedge angle shall appear on the wedge or wedge tray
(if permanently mounted to the tray). 	If the wedge or
wedge tray is damaged, 	the wedge transmission factor
shall be redetermined;
(2) 	If the absorbed dose 	rate information required by
subsection 	(a) 	relates exclusively to operation with a
45-180
265
Complete printed PDF page 180
§11-45-238
field flattening or beam scattering filter in place,
such filter shall be removable only by the use of
tools;
(3) 	For equipment manufactured after the effective date of
this chapter which utilize a 	system of wedge filters,
interchangeable field 	flattening filters, 	or
interchangeable beam scattering foils:
(A) 	Irradiation shall not be possible until a
selection of a 	filter or a positive selection to
use "no filter" 	has been made at the treatment
control panel, 	either manually or automatically;
(B) 	An interlock system shall be provided to prevent
irradiation if the filter selected is not in the
correct position;
(C) 	A display shall be provided at the treatment
control panel showing the wedge filter(s),
interchangeable 	field flattening 	filter(s),
and/or interchangeable beam scattering foil(s)
in use; 	and
(D) 	An interlock shall be provided to prevent
irradiation if any filter and/or beam scattering
foil 	selection operation carried out in the
treatment room does not agree with the 	filter
and/or beam scattering foil 	selection operation
carried out at the treatment control panel.
(e) 	For equipment manufactured after the effective date of
this chapter, 	the licensee shall determine during acceptance
testing, 	or obtain from the manufacturer, 	data sufficient to
ensure that x-ray stray radiation in the useful electron beam,
absorbed dose at the surface during x-ray irradiation and stray
neutron radiation in the useful x-ray beam are in compliance with
International Electrotechnical Commission 	(IEC) 	Document 601-2-1.
(f) 	All therapeutic radiation machines subject to this
section shall be provided with beam monitoring devices. 	The
sensors 	for this device shall be fixed in the useful beam during
treatment to indicate the air kerma rate or dose rate.
(1) 	Equipment manufactured after the effective date of
this chapter shall be provided with at least two
independently powered integrating dose meters.
Alternatively, 	a common power supply may be used if
the production of radiation is terminated upon failure
of any common element.
(2) 	Equipment manufactured on or before the effective date
of this chapter shall be provided with at least one
radiation detector. 	This detector shall be
incorporated into a 	useful beam monitoring system;
(3) 	The detector and the system into which that detector
is incorporated shall meet the following 	requirements:
(A) 	Each detector shall be removable only with tools
and, 	if movable, 	shall be interlocked to prevent
incorrect positioning;
(B) 	Each detector shall form part of a beam
monitoring system from which readings in dose
monitor units the absorbed dose at a 	reference
45-181 265
Complete printed PDF page 181
§11-45-238
point in the treatment volume can be calculated;
(C) 	Each beam monitoring system shall be capable of
independently monitoring, 	interrupting, 	and
terminating irradiation; 	and
(D) 	For equipment manufactured after the effective
date of this chapter, 	the design of the beam
monitoring systems shall ensure that the:
(i) 	Malfunctioning of one system shall not
affect the correct functioning of the
secondary system; 	and
(ii) 	Failure of any element common to both
systems which could affect the correct
function of both systems shall terminate
irradiation or prevent the initiation of
radiation.
(E) 	Each beam monitoring system shall have a 	legible
display at the treatment control panel. 	For
equipment manufactured after the effective date
of this chapter, 	each display shall:
(i) 	Maintain a 	reading until intentionally
reset;
(ii) 	Have only one scale and no electrical or
mechanical scale multiplying factors;
(iii) 	Utilize a design such that increasing dose
is displayed by increasing numbers; 	and
(iv) 	In the event of power failure, 	the beam
monitoring information required in clause
(iii) 	displayed at the control panel at
the time of failure shall be retrievable
in at least one system for a 	twenty minute
period of time.
(g) 	Beam symmetry:
(1) 	Bent-beam linear accelerators subject to this section
shall be provided with auxiliary device(s) 	to monitor
beam symmetry;
(2) 	The device(s) 	referenced in paragraph 	(1) 	shall be
able to detect field asymmetry greater than ten
percent; 	and
(3) 	The device(s) 	referenced in paragraph 	(1) 	shall be
configured to terminate irradiation if the
specifications in paragraph 	(2) 	cannot be maintained.
(h) 	Selection and display of dose monitor units:
(1) 	Irradiation shall not be possible until a 	selection of
a number of dose monitor units has been made at the
treatment control panel;
(2) 	The pre-selected number of dose monitor units shall be
displayed at the treatment control panel until 	reset
manually for the next irradiation;
(3) 	After termination of irradiation, 	it shall be
necessary to reset the dosimeter display before
subsequent treatment can be initiated; 	and
(4) 	For equipment manufactured after the effective date of
this chapter, 	after termination of irradiation, 	it
shall be necessary for the operator to 	reset the
45-182
265
Complete printed PDF page 182
§11-45-238
pre-selected dose monitor units before irradiation can
be initiated.
(i) 	For equipment manufactured after the effective date of
this chapter, 	a 	system shall be provided from which readings the
air kerma rate or absorbed dose rate at a 	reference point in the
treatment volume can be calculated. 	The radiation detectors
specified in subsection 	(f) 	may form part of this system. 	In
addition:
(1) 	The dose monitor unit dose rate shall be displayed at
the treatment control panel;
(2) 	If the equipment can deliver under any conditions an
air kerma rate or absorbed dose rate at the nominal
treatment distance more than twice the maximum value
specified by the manufacturer, 	a device shall be
provided which terminates irradiation when the air
kerma rate or absorbed dose rate exceeds a value twice
the specified maximum. 	The dose 	rate at which the
irradiation shall be terminated shall be a 	record
maintained by the licensee;
(3) 	For equipment manufactured after the effective date of
this chapter, 	if the equipment can deliver under any
conditions an air kerma rate or absorbed dose rate at
the nominal treatment distance more than ten times the
maximum value specified by the manufacturer, 	a device
shall be provided to prevent the air kerma rate or
absorbed dose rate anywhere in the radiation field
(4)
( j)
system or
(1)
(2)
(3)
(k)
equipment
from exceeding twice the specified maximum value and
to terminate irradiation if the excess absorbed dose
at the nominal treatment distance exceeds 	four hundred
rads 	(four grays); 	and
For each therapeutic radiation machine, 	the licensee
shall determine, 	or obtain from the manufacturer, 	the
maximum value(s) 	specified in paragraphs 	(2) 	and 	(3)
for the specified operating conditions. 	Records of
these maximum value(s) 	shall be maintained at the
installation for inspection.
Termination of irradiation by the beam monitoring
systems during stationary beam radiation therapy:
Each primary system shall terminate irradiation when
the pre-selected number of dose monitor units has been
detected by the system;
If the original design of the equipment included a
secondary dose monitoring system, 	that system shall be
capable of terminating irradiation when not more than
fifteen percent or forty dose monitor units above the
pre-selected number of dose monitor units set at the
control panel has been detected by the secondary dose
monitoring system; 	and
For equipment manufactured after the effective date of
this chapter, 	an indicator on the control panel shall
show which monitoring system has terminated
irradiation.
It shall be possible to terminate irradiation and
movement or go from an interruption condition to
45-183
Complete printed PDF page 183
§11-45-238
termination condition at any time from the operator's position at
the treatment control panel.
(1) 	If a 	therapeutic radiation machine has an interrupt
mode, 	it shall be possible to interrupt irradiation and equipment
movements at any time from the treatment control panel. 	Following
an interruption it shall be possible to restart irradiation by
operator action without any reselection of operating conditions.
If any change is made of a pre-selected value during an
interruption, 	irradiation and equipment movements shall be
automatically terminated.
(m) 	A suitable irradiation control device shall be
provided to terminate the irradiation after a pre-set time
interval.
(1)
(2)
A timer shall be provided which has a display at the
treatment control panel. 	The timer shall have a
pre-set time selector and an elapsed time indicator;
The timer shall be a cumulative timer which activates
with an indication of "BEAM-ON" and retains 	its
reading after irradiation is interrupted or
terminated. 	After irradiation is terminated and
before irradiation can be reinitiated, 	it shall be
necessary to 	reset the elapsed time indicator;
(3) 	For equipment manufactured after the effective date of
this chapter, 	after termination of irradiation and
before irradiation can be reinitiated, 	it shall be
necessary for the operator to reset the pre-set time
selector;
(4) 	The timer shall terminate irradiation when a
pre-selected time has elapsed, 	if the dose monitoring
systems have not previously terminated irradiation.
(n) 	Equipment capable of both x-ray therapy and electron
therapy shall meet the 	following additional requirements:
(1) 	Irradiation shall not be possible until a 	selection of
radiation type 	(x-rays or electrons) 	has been made at
the treatment control panel;
(2) 	The radiation type selected shall be displayed at the
treatment control panel before and during irradiation;
(3) 	An interlock system shall be provided to ensure that
the equipment can principally emit only the radiation
type which has been selected;
(4) 	An interlock system shall be provided to prevent
irradiation with x-rays, 	except to obtain a
verification film, 	when electron applicators are
fitted;
(5) 	An interlock system shall be provided to prevent
irradiation with electrons when accessories specific
for x-ray therapy are fitted; 	and
(6) 	An interlock system shall be provided to prevent
irradiation if any selected operations carried out in
the treatment room do not agree with the selected
operations carried out at the treatment control panel.
(o) 	Equipment capable of generating radiation beams of
different energies shall meet the following requirements:
(1) 	Irradiation shall not be possible until a 	selection of
45-184 26
Complete printed PDF page 184
§11-45-238
energy has been made at the treatment control panel;
(2) 	The measured energy value selected shall be displayed
(megavolts 	for photons and mega electron volts 	for
electrons) 	at the treatment control panel before and
during irradiation; 	and
(3) 	Irradiation shall not be possible until 	the
appropriate 	flattening 	filter or scattering foil 	for
the selected energy is in its proper location.
(p) 	Therapeutic radiation machines capable of both
stationary beam radiation therapy and rotational arc radiation
therapy shall meet the 	following requirements:
(1) 	Irradiation shall not be possible until a 	selection of
stationary beam radiation therapy or rotational arc
radiation therapy has been made at the treatment
control panel;
(2) 	The mode of operation shall be displayed at the
treatment control panel;
(3) 	An interlock system shall be provided to ensure that
the equipment can operate only in the mode which has
been selected;
(4) 	An interlock system shall be provided to prevent
irradiation if any selected parameter in the treatment
room does not agree with the selected parameter at the
treatment control panel;
(5) 	Rotational arc radiation therapy shall be controlled
to obtain the selected relationships between
incremental dose monitor units and incremental angle
of movement for equipment manufactured after the
effective date of this chapter:
(A) 	An interlock system shall be provided to
terminate irradiation if the number of dose
monitor units delivered in any fifteen degrees
of arc differs by more than twenty percent from
the selected value;
(B) 	Where gantry angle terminates the irradiation in
rotational arc radiation therapy, 	the dose
monitor units shall differ by less than five
percent from the value calculated from the
absorbed dose per unit angle and total angle
relationship;
(C) 	An interlock shall be provided to prevent the
gantry moving more than five degrees beyond the
selected angular limits during rotational arc
radiation therapy;
(D) 	An interlock shall be provided to 	require that a
selection of direction be made at the treatment
control panel in all units which are capable of
both clockwise and counter-clockwise rotational
arc radiation therapy.
(6) 	Where the beam monitor system terminates the
irradiation in rotational arc radiation therapy, 	the
termination of irradiation shall be as 	required by
subsection 	(j); 	and
(7) 	For equipment manufactured after the effective date of
45-185
2
Complete printed PDF page 185
§11-45-238
this chapter, 	an interlock system shall be provided to
terminate irradiation if movement of the gantry:
(A) 	Occurs during stationary beam radiation therapy;
or
(B) 	Stops during rotational arc radiation therapy
unless such stoppage is a pre-planned function.
(q) 	Facility design requirements 	for therapeutic radiation
machines operating above five hundred kilovolts: 	In addition to
shielding adequate to meet requirements of section 11-45-241, 	the
following design requirements are made:
(1) 	All protective barriers shall be fixed, 	except for
access doors to the treatment room or movable beam
interceptors;
(2) 	In addition to other requirements specified in this
subchapter, 	the control panel shall also:
(A) 	Be located outside the treatment room;
(B) 	Provide an indication of whether electrical
power is available at the control panel and if
activation of the radiation is possible;
(C) 	Provide an indication of whether radiation is
being produced; 	and
(D) 	Include an access control 	(locking) 	device which
shall prevent unauthorized use of the
therapeutic radiation machine;
(3) 	Windows, 	mirrors, 	closed-circuit television or an
equivalent viewing system shall be provided to permit
continuous observation of the patient following
positioning and during irradiation and shall be so
located that the operator may observe the patient from
the treatment control panel. 	The therapeutic
radiation machine shall not be used for patient
irradiation unless at least one viewing system is
operational;
(4) 	Provision shall be made for continuous two-way aural
communication between the patient and the operator at
the control panel. 	The therapeutic 	radiation machine
shall not be used for irradiation of patients unless
continuous two-way aural communication is possible;
(5) 	Treatment room entrances shall be provided with
warning lights in a 	readily observable position near
the outside of all access doors, 	which shall indicate
when the useful beam is 	"ON" and when it is 	"OFF";
(6) 	Interlocks shall be provided such that all entrance
doors shall be closed before treatment can be
initiated or continued. 	If the radiation beam is
interrupted by any door opening, 	it shall not be
possible to restore the machine to operation without
closing the door and reinitiating irradiation by
manual action at the control panel;
(7) 	If the shielding material in any protective barrier
requires the presence of a beam interceptor to ensure
compliance with subchapter 4, 	interlocks shall be
provided to prevent the production of radiation,
unless the beam interceptor is in place, 	whenever the
45-186
26
Complete printed PDF page 186
§11-45-238
useful beam is directed at the designated barrier(s);
(8) 	At least one "scram button" 	or other emergency power
cutoff switch shall be located in the radiation
therapy room and shall terminate all equipment
electrical power including radiation and mechanical
motion. 	This switch is in addition to the termination
switch required by subsection 	(k). 	All emergency power
cutoff switches shall include a manual 	reset so that
the therapeutic radiation machine cannot be restarted
from the unit's control console without resetting the
emergency cutoff switch;
(9) 	All safety interlocks shall be designed so that any
defect or component failure in the safety interlock
system prevents or terminates operation of the
therapeutic radiation machine; 	and
(10) 	Surveys for residual activity shall be conducted on
all therapeutic radiation machines capable of
generating photon and electron energies above ten
megavolts before machining, 	removing, 	or working on
therapeutic radiation machine components which may
have become activated due to photo-neutron production.
(r) 	Qualified medical physicist support:
(1) 	The services of a 	qualified medical physicist shall be
utilized in facilities having therapeutic radiation
machines with energies of five hundred kilovolts and
above. 	The qualified medical physicist shall be
responsible for:
(A) 	Full calibration(s) 	required by subsection 	(t)
and protection surveys required by section 11-
45-234(a);
(B) 	Supervision and review of dosimetry;
(C) 	Beam data acquisition and transfer for
computerized dosimetry, 	and supervision of its
use;
(D) 	Quality assurance, 	including quality assurance
check review required by subsections 	(u) 	and
(v);
(E) 	Consultation with the authorized user in
treatment planning, 	as needed; 	and
(F) 	Perform calculations/assessments regarding
misadministrations.
(2) 	If the qualified medical physicist is not a 	full-time
employee of the licensee, 	the operating procedures
required by subsection 	(s) 	shall also specifically
address how the qualified medical physicist is to be
contacted for problems or emergencies, 	as well as the
specific actions, 	if any, 	to be taken until the
qualified medical physicist can be contacted.
(s) 	Operating procedures:
(1) 	No individual, 	other than the patient, 	shall be in the
treatment room during treatment or during any
irradiation for testing or calibration purposes;
(2) 	Therapeutic radiation machines shall not be made
available 	for medical use unless the requirements of
45-187
26
Complete printed PDF page 187
§11-45-238
section 11-45-234(a), 	subsections 	(t) 	and 	{u) 	have
been met;
(3) 	Therapeutic radiation machines, 	when not in operation,
shall be secured to prevent unauthorized use;
(4) 	If a patient must be held in position during
treatment, 	mechanical supporting or restraining
devices 	shall be used; 	and
(5) 	A copy of the current operating and emergency
procedures shall be maintained at the therapeutic
radiation machine control console.
(t) 	Full calibration measurements:
(1) 	Full calibration of a 	therapeutic radiation machine
subject to this section shall be performed by, 	or
under the direct supervision of, 	a 	qualified medical
physicist:
(A) 	Before the first medical use following
installation or reinstallation of the
therapeutic radiation machine;
(B) 	At intervals not exceeding one year; 	and
(C) 	Before medical use under the following
conditions:
(D)
(i) 	Whenever quality assurance check
measurements indicate that the radiation
output differs by more than five percent
from the value obtained at the last full
calibration and the difference cannot be
easily discerned; 	and
(ii) 	Following any component replacement, 	major
repair, 	or modification of components that
could significantly affect the
characteristics of the radiation beam.
Notwithstanding the 	requirements of subparagraph
(C):
(i)
(ii)
Full calibration of therapeutic radiation
machines with multi-energy and/or
multi-mode capabilities is 	required only
for those modes and/or energies that are
not within their acceptable range; 	and
If the repair, 	replacement or modification
does not affect all modes and/or energies,
full 	calibration shall be performed on the
effected mode/energy that is in most
frequent clinical use at the facility.
The remaining energies/modes may be
validated with quality assurance check
procedures against the criteria in clause
(i).
(2) 	To satisfy the requirement of paragraph 	(1), 	full
calibration shall include all measurements required
for annual calibration by American Association of
Physicists in Medicine 	(AAPM) 	Report 13, 	"Physical
Aspects of Quality Assurance in Radiation Therapy";
(3) 	The licensee shall use the dosimetry system described
in section 11-45-234(c) 	to measure the radiation
45-188 6
Complete printed PDF page 188
§11-45-238
output 	for one set of exposure conditions. 	The
remaining radiation measurements required in paragraph
(2) 	may then be made using a dosimetry system that
indicates relative dose rates; 	and
(4) 	The licensee shall maintain a 	record of each
calibration for the duration of the license. 	The
record shall include the date of the calibration, 	the
manufacturer's name, 	model number, 	and serial number
for the therapeutic radiation machine, 	the model
numbers and serial numbers of the instruments used to
calibrate the therapeutic radiation machine, 	and the
signature of the qualified medical physicist
responsible for performing the calibration.
(u) 	Periodic quality assurance checks:
(1) 	Periodic quality assurance checks 	shall be performed
on all therapeutic radiation machines subject to this
section at intervals not to exceed one week;
(2) 	To satisfy the requirement of paragraph 	(1), 	quality
assurance checks shall include determination of all
parameters for periodic quality assurance checks
contained in American Association of Physicists in
Medicine 	(AAPM) 	Report 13, 	"Physical Aspects of
Quality Assurance in Radiation Therapy";
(3) 	The licensee shall use a dosimetry system which has
been inter-compared within the previous six months
with the dosimetry system described in section ll-45-
234(c) (1) 	to make the periodic quality assurance
checks 	required in paragraph 	(2);
(4) 	The licensee shall perform periodic quality assurance
checks 	required by paragraph 	(1) 	in accordance with
procedures established by the qualified medical
physicist;
(5) 	The licensee shall review the results of each periodic
radiation output check according to the following
procedures:
(A) 	The authorized user and qualified medical
physicist shall be immediately notified if any
parameter is not within its acceptable range.
The therapeutic 	radiation machine shall not be
made available for subsequent medical use until
the qualified medical physicist has determined
that all parameters are within their acceptable
range;
(B) 	If all quality assurance check parameters appear
to be within their acceptable range, 	the quality
assurance check shall be reviewed and signed by
either the authorized user or qualified medical
physicist within three treatment days; 	and
(C) 	The qualified medical physicist shall review and
sign the results of each radiation output
quality assurance check at intervals not to
exceed one month.
(6) 	Therapeutic radiation machines subject to this section
shall have safety quality assurance checks performed
45-189
6
Complete printed PDF page 189
§11-45-238
(7)
( 8)
(9)
at intervals not to exceed one week;
To satisfy the requirement of paragraph 	(6), 	safety
quality assurance checks 	shall ensure proper operation
of:
(A)
(B)
(C)
(D)
(E)
(F)
Electrical interlocks at each external beam
radiation therapy room entrance;
Proper operation of the "BEAM-ON", 	interrupt and
termination switches;
Beam condition indicator lights on the access
doors, 	control console, 	and in the radiation
therapy room;
Viewing systems;
Electrically operated treatment room door(s)
from inside and outside the treatment room;
At least one emergency power cutoff switch. 	If
more than one emergency power cutoff switch is
installed and not all switches are tested at
once, 	each switch shall be tested on a 	rotating
basis. 	Safety quality assurance checks of the
emergency power cutoff switches may be conducted
at the end of the treatment day in order to
minimize possible stability problems with the
therapeutic radiation machine.
The licensee shall promptly repair any system
identified in paragraph 	(7) 	that is not operating
properly; 	and
The licensee shall maintain a 	record of each quality
assurance check required by paragraphs 	(1) 	and 	(7) 	for
three years. 	The record shall include the date of the
quality assurance check, 	the manufacturer's name,
model number, 	and serial number for the therapeutic
radiation machine, 	the manufacturer's name, 	model
number and serial number of the instrument(s) 	used to
measure the radiation output of the therapeutic
radiation machine, 	and the signature of the individual
who pe~f~~~ed the,r.eriodic quality assurance check.
[Eff 	NUV 	12 1~90 ] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS ~§321-1, 	321-11 (21), 	321-71)
§11-45-239 	Reserved.
§11-45-240 	Calibration and check of survey instruments.
(a) 	The licensee shall ensure that the survey instruments used to
show compliance with this subchapter have been calibrated before
first use, 	at intervals not to exceed twelve months, 	and following
repair.
(b) 	To satisfy the requirements of subsection 	(a), 	the
licensee shall:
(1) 	Calibrate all 	required scale readings up to one
thousand millirems 	(ten millisieverts) 	per hour with
an appropriate radiation source;
(2) 	Calibrate at least two points on each scale to be
45-190
2 6 5
Complete printed PDF page 190
§11-45-241
calibrated. 	These points should be at approximately
one-third and two-thirds of scale rating.
(c) 	To satisfy the requirements of subsection 	(b), 	the
licensee shall:
(1) 	Consider a point as calibrated if the indicated dose
rate differs 	from the calculated dose rate by not more
(2)
(d)
required in
(1)
(2)
than ten percent; 	and
Consider a point as calibrated if the indicated dose
rate differs 	from the calculated dose rate by not more
than twenty percent if a 	correction factor or graph is
conspicuously attached to the instrument.
The licensee shall retain a 	record of each calibration
subsection 	(a). 	The record shall include:
A description of the calibration procedure; 	and
A description of the source used and the certified
dose rates 	from the source, 	and the rates indicated by
the instrument being calibrated, 	the correction
factors deduced from the calibration data, 	the
signature of the individual who performed the
calibration, 	and the date of calibration.
(e) 	The licensee may obtain the services of individuals
licensed to perform calibrations of survey instruments. 	Records
of calibrations which contain information required by subsection
(d) 	shall be maintained by the licensee. 	[Eff NO\/ 	12 19aa 	1
(Auth: 	HRS §§321-10, 	321-11, 	321-71) 	(Imp: 	HRS §§321-1, 	3~:t'-
11 (21), 	321-71)
§11-45-241 	Shielding and safety design requirements. 	(a)
Each therapeutic 	radiation machine subject to sections 11-45-236
and 11-45-238 shall be provided with such primary and/or secondary
barriers as are necessary to ensure compliance with subchapter 4.
(b) 	Facility design information for all new installations
of a 	therapeutic radiation machine or installations of a
therapeutic radiation machine of higher energy into a 	room not
previously approved for that energy shall be submitted for
department approval before actual installation of the therapeutic
radiation machine. 	The minimum facility design information that
shall be submitted is contained in Appendix A to subchapter 14.
(c) 	The exhibit at the end of chapter 45 entitled
"Appendix A of Subchapter 14, 	Information on Radiation Shielding
Requ~red for 	Pla.n ~fYixW.f 	93) ", 	is made a part of this
section. 	[Eff 	Nuv 	l l, 	] 	(Auth: 	HRS §§321-10, 	321-11,
321-71) 	(Imp: 	HRS §§321- , 	11 (21), 	321-71)
45-191 	26
Complete printed PDF page 191
Appendix A of Subchapter 1
QUALITY FACTORS AND ABSORBED DOSE EQUIVALENCIES 	(2/2/93)
Quality Factor
TYPE OF RADIATION
X, 	gamma, 	or beta radiation and
high-speed electrons
Alpha particles, 	multiple-charged
particles, 	fission 	fragments 	and
heavy particles of unknown charge
Neutrons of unknown energy
High-energy protons
Absorbed Dose
(Q)
1
20
10
10
Equal to
a Unit Dose
Equivalene
1
0.05
0.1
0.1
aAbsorbed dose 	in 	rad 	equal 	to 	one 	rem or 	the 	absorbed dose 	in 	gray equal 	to
one Sv.
l-A-1
22 	5
Complete printed PDF page 192
Appendix B of Subchapter 1
MEAN QUALITY FACTORS, 	Q, 	AND FLUENCE PER UNIT DOSE
EQUIVALENT FOR MONOENERGETIC NEUTRONS 	(2/2/93)
Neutron 	Quality 	Fluence per Unit 	Fluence per Unit
Energy 	Factora 	Dose Equi valentb 	Dose Equivalentb
(MeV) 	(Q) 	(neutrons
cm- 2 rem- 1 )
(neutrons
cm- 2 Sv- 1 )
(thermal) 	2.5 X 	10-8 	2 	980 X 10 6 	980 X 108
1 	X 	10-7 	2 	980 X 10 6 	980 X 108
1 	X 	10- 6 	2 	810 X 10 6 	810 X 108
1 	X 	10-5 	2 	810 X 10 6 	810 X 108
1 	X 	10-4 	2 	840 X 10 6 	840 X 108
1 	X 	10-3 	2 	980 X 10 6 	980 X 108
1 	X 	10- 2 	2.5 	1010 X 10 6 	1010 X 108
1 	X 	10-l 	7.5 	170 X 	10 6 	170 X 108
5 	X 	10- 1 	11 	39 X 10 6 	39 X 108
1 	11 	27 X 10 6 27 X 108
2.5 	9 	29 X 10 6 	29 X 108
5 	8 	23 X 10 6 	23 X 108
7 	7 	24 X 10 6 	24 X 108
10 	6.5 	24 X 10 6 	24 X 108
14 	7.5 	17 X 10 6 	17 X 108
20 	8 	16 X 10 6 	16 X 108
40 	7 	14 X 10 6 	14 X 108
60 	5.5 	16 X 10 6 	16 X 108
1 	X 	10 2 	4 	20 X 10 6 	20 X 108
2 	X 	10 2 	3.5 	19 X 10 6 	19 X 108
3 	X 	10 2 	3.5 	16 X 10 6 	16 X 108
4 	X 	10 2 	3.5 	14 X 10 6 	14 X 108
aValue of quality factor 	(Q) 	at the point where the dose equivalent is maximum
in a 	30-cm diameter cylinder tissue-equivalent phantom.
bMonoenergetic neutrons 	incident normally on a 	30-cm diameter 	cylinder tissue-
equivalent phantom.
1-B-1
2265
Complete printed PDF page 193
Appendix A of Subchapter 2
EXEMPT CONCENTRATIONS 	(2/2/93)
Column
II
Column 	Liquid
I 	and solid
Gas 	con- 	concen-
Element 	(atomic 	centration 	tration
number) 	Radionuclide 	μCi/ml 1/ 	μCi/ml 	2/
Antimony 	(51) 	Sb-122 	3Xl0-4
Sb-124 	2X10- 4
Sb-125 	lXl0-3
Argon 	(18) 	Ar-37 	lXl0-3
Ar-41 	4X10-7
Arsenic 	(33) 	As-73 	5X10-3
As-74 	5X10-4
As-76 	2X10- 4
As-77 	8X10- 4
Barium 	(56) 	Ba-131 	2X10- 3
Ba-140 	3X10-4
Beryllium 	(4) 	Be-7 	2x10-2
Bismuth 	(83) 	Bi-206 	4X10-4
Bromine 	(35) 	Br-82 	4X10- 7 	3X10-3
Cadmium 	(48) 	Cd-109 	2x10-3
Cd-115m 	3X10-4
Cd-115 	3X10-4
Calcium 	(20) 	Ca-45 	9X10-5
Ca-47 	5X10-4
Carbon 	(6) 	C-14 	1x10-6 	8x10-3
Cerium 	(58) 	Ce-141 	9X10- 4
Ce-143 	4X10-4
Ce-144 	lXl0-4
Cesium 	(55) 	Cs-131 	2x10-2
Cs-134m 	6X10-2
Cs-134 	9X10- 5
Chlorine 	(17) 	Cl-38 	9X10- 7 	4X10-3
Chromium 	(24) 	Cr-51 	2x10-2
Cobalt 	(27) 	Co-57 	5X10-3
Co-58 	lXl0-3
Co-60 	5X10- 4
1/ Values 	are given in Column I 	only for those materials normally
used as gases.
2/ μCi/g 	for 	solids.
2-A-1
il
2 6 5 .,. ·rr:
Complete printed PDF page 194
Column
II
Column 	Liquid
I 	and solid
Gas 	con- 	concen-
Element 	(atomic 	centration 	tration
number) 	Radionuclide 	μCi/ml _:!:./ 	μCi/ml '?:_/
Copper 	(29) 	Cu-64 	3X10-3
Dysprosium 	(66) 	Dy-165 	4X10-3
Dy-166 	4X10- 4
Erbium 	(68) 	Er-169 	9X10-4
Er-171 	lXl0-3
Europium 	(63) 	Eu-152(9.2 h) 	6Xl0-4
Eu-155 	2X10-3
Fluorine 	(9) 	F-18 	2x10- 6 	8x10-3
Gadolinium 	(64) 	Gd-153 	2x10-3
Gd-159 	8x10-4
Gallium 	(31) 	Ga-72 	4X10-4
Germanium 	(32) 	Ge-71 	2x10-2
Gold 	(79) 	Au-196 	2x10-3
Au-198 	5x10-4
Au-199 	2X10-3
Hafnium 	(72) 	Hf-181 	7X10- 4
Hydrogen 	(1) 	H-3 	sx10- 6 	3X10-2
Indium 	(49) 	In-113m 	lXl0-2
In-114m 	2X10- 4
Iodine 	(53) 	I-126 	3X10- 9 	2x10-5
I-131 	3X10- 9 	2X10- 5
I-132 	8x10-0 	6X10- 4
I-133 	lXl 0- 8 	7X10-5
I-134 	2x10- 7 	lXl0-3
Iridium 	(77) 	Ir-190 	2x10-3
Ir-192 	4X10- 4
Ir-194 	3X10-4
Iron 	(26) 	Fe-55 	8x10-3
Fe-59 	6X10- 4
Krypton 	(36) 	Kr-85m 	1x10-6
Kr-85 	3X10-6
Lanthanum 	(57) 	La-140 	2X10- 4
Lead 	(82) 	Pb-203 	4X10-3
Lutetium 	(71) 	Lu-177 	lXl0-3
1/ Values 	are given in Column I 	only for those materials normally
used as 	gases.
2/ μCi/g 	for 	solids.
2-A-2
2265
Complete printed PDF page 195
Column
II
Column 	Liquid
I 	and solid
Gas 	con- 	concen-
Element 	(atomic 	centration 	tration
number) 	Radionuclide 	μCi/ml _;I/ 	μCi/ml 	2/
Manganese 	(25) 	Mn-52 	3X10-4
Mn-54 	lXl0-3
Mn-56 	lXl0-3
Mercury 	(80) 	Hg-197m 	2X10- 3
Hg-197 	3X10-3
Hg-203 	2X10- 4
Molybdenum 	(42) 	Mo-99 	2Xl0-3
Neodymium 	(60) 	Nd-147 	6X10-4
Nd-149 	3X10-3
Nickel 	(28) 	Ni-65 	lXl0-3
Niobium 	(Columbium) 	(41) 	Nb-95 	lXl0-3
Nb-97 	9X10-3
Osmium 	(76) 	Os-185 	7Xl0-4
Os-191m 	3X10- 2
Os-191 	2X10-3
Os-193 	6X10- 4
Palladium 	(46) 	Pd-103 	3X10-3
Pd-109 	9Xl0- 4
Phosphorus 	(15) 	P-32 	2x10-4
Platinum 	(78) 	Pt-191 	lXl0-3
Pt-193m 	1x10- 2
Pt-197m 	1x10- 2
Pt-197 	lXl0-3
Potassium 	(19) 	K-42 	3X10-3
Praseodymium 	(59) 	Pr-142 	3X10-4
Pr-143 	5X10-4
Promethium 	(61) 	Pm-147 	2X10-3
Pm-149 	4X10- 4
Rhenium 	(75) 	Re-183 	6Xl0-3
Re-186 	9X10- 4
Re-188 	6X10- 4
Rhodium 	(45) 	Rh-103m 	1X10-l
Rh-105 	lXl0-3
1/ Values 	are given in Column I 	only for those materials normally
used as 	gases.
2/ μCi/g 	for 	solids.
2-A-3
2 	5
Complete printed PDF page 196
Column
II
Column 	Liquid
I 	and solid
Gas 	con- 	concen-
Element 	(atomic 	centration 	tration
number) 	Radionuclide 	μCi/ml _!/ 	μCi/ml 	2/
Rubidium 	(37) 	Rb-86 	7Xl0- 4
Ruthenium 	(44) 	Ru-97 	4X10- 3
Ru-103 	8x10-4
Ru-105 	lXl0- 3
Ru-106 	lXl0-4
Samarium 	(62) 	Sm-153 	8X10- 4
Scandium 	(21) 	Sc-46 	4X10- 4
Sc-47 	9X10- 4
Sc-48 	3X10-4
Selenium 	(34) 	Se-75 	3X10- 3
Silicon 	(14) 	Si-31 	9X10- 3
Silver 	(47) 	Ag-105 	lXl0- 3
Ag-llOm 	3X10- 4
Ag-111 	4X10- 4
Sodium 	(11) 	Na-24 	2x10- 3
Strontium 	(38) 	Sr-85 	lXl0- 3
Sr-89 	1x10-4
Sr-91 	7X10- 4
Sr-92 	7X10- 4
Sulfur 	(16) 	S-35 	9X10-8 	6X10- 4
Tantalum 	(73) 	Ta-182 	4X10- 4
Technetium 	(43) 	Tc-96m 	lXlQ-l
Tc-96 	lXl0- 3
Tellurium 	(52) 	Te-125m 	2X10- 3
Te-127m 	6X10- 4
Te-127 	3X10- 3
Te-129m 	3X10-4
Te-131m 	6X10- 4
Te-132 	3X10-4
Terbium 	(65) 	Tb-160 	4X10- 4
Thallium 	(81) 	Tl-200 	4X10- 3
Tl-201 	3X10- 3
Tl-202 	lXl0- 3
Tl-204 	lXl0- 3
Thulium 	(69) 	Tm-170 	5x10- 4
Tm-171 	5X10- 3
1/ Values 	are given in Column I 	only for those materials normally
used as 	gases.
2/ 	μCi/g 	for solids.
2-A-4
2 6 5 	, 	I
Complete printed PDF page 197
Element 	(atomic
number)
Tin 	(50)
Tungsten 	(Wolfram) 	(74)
Vanadium 	(23)
Xenon 	( 54)
Ytterbium 	(70)
Yttrium 	(39)
Zinc 	(30)
Zirconium 	(40)
Beta- and/or gamma-
emitting radioactive
material not listed
above with half-life
of less 	than 3 years.
Column
I
Gas 	con-
centration
Radionuclide 	μCi/ml ll
Sn-113
Sn-125
W-181
W-187
V-48
Xe-131m
Xe-133
Xe-135
Yb-175
Y-90
Y-91m
Y-91
Y-92
Y-93
Zn-65
Zn-69m
Zn-69
Zr-95
Zr-97
4X10- 6
3X10- 6
lXl0- 6
lXl0- 10
Column
II
Liquid
and solid
concen-
tration
μCi/ml 	2/
9X10- 4
2x10- 4
4X10-3
7X10- 4
3X10- 4
lXl0-3
2X10- 4
3X10-2
3X10- 4
6X10- 4
3X10- 4
lXl0-3
7X10- 4
2x10-2
6X10- 4
2X10- 4
lXl0-6
1/ Values 	are given in Column I 	only for those materials normally
used as gases.
~/ 	μCi/g 	for solids.
Note 1: 	Many radionuclides transform into other radionuclides.
In expressing the concentrations in Appendix A, 	the activity
stated is 	that of the parent radionuclide and takes into account
the radioactive decay products.
2-A-5
2 6 5
Complete printed PDF page 198
Note 2: 	Where there is involved a 	combination of radionuclides,
the limit for the combination should be derived as 	follows:
Determine for each radionuclide in the product the ratio between
the radioactivity concentration present in the product and the
exempt 	radioactivity concentration established in Appendix A for
the specific radionuclide when not in combination. 	The sum of
such ratios may not exceed "1".
Example: 	Concentration of Radionuclide A in Product+
Exempt concentration of Radionuclide A
Concentration of Radionuclide Bin Product 	<1
Exempt concentration of Radionuclide B
Note 3: 	To convert μCi/ml 	to SI units of megabecquerels per liter
multiply
the above values by 37.
Example: 	Zirconium 	(40) 	Zr-97 	(2x10- 4 μCi/ml multiplied by 37 is
equivalent
to 74 x 10- 4 MBq/1)
2-A-6
265
Complete printed PDF page 199
Appendix B of Subchapter 2
EXEMPT QUANTITIES 	(2/2/93)
Radioactive
Material
Antimony-122 	(Sb 122)
Antimony-124 	(Sb 124)
Antimony-125 	(Sb 125)
Arsenic-73 	(As 73)
Arsenic-74 	(As 74)
Arsenic-76 	(As 76)
Arsenic-77 	(As 77)
Barium-131 	(Ba 131)
Barium-133 	(Ba 133)
Barium-140 	(Ba 140)
Bismuth-210 	(Bi 210)
Bromine-82 	(Br 82)
Cadmium-109 	(Cd 109)
Cadmium-115m 	(Cd 115m)
Cadmium-115 	(Cd 115)
Calcium-45 	(Ca 45)
Calcium-47 	(Ca 47)
Carbon-14 	(C 14)
Cerium-141 	(Ce 141)
Cerium-143 	(Ce 143)
Cerium-144 	(Ce 144)
Cesium-129 	(Cs 129)
Cesium-131 	(Cs 131)
Cesium-134m 	(Cs 134m)
Cesium-134 	(Cs 134)
Cesium-135 	(Cs 135)
Cesium-136 	(Cs 136)
Cesium-137 	(Cs 137)
Chlorine-36 	(Cl 36)
Chlorine-38 	(Cl 38)
Chromium-51 	(Cr 51)
Cobalt-57 	(Co 57)
Cobalt-58m 	(Co 58m)
Cobalt-58 	(Co 58)
Cobalt-60 	(Co 60)
Copper-64 	(Cu 64)
Dysprosium-165 	(Dy 165)
Dysprosium-166 	(Dy 166)
Erbium-169 	(Er 169)
Erbium-171 	(Er 171)
Europium-152 	(Eu 152)9.2h
Europium-152 	(Eu 152)13 yr
Europium-154 	(Eu 154)
2-B-l
Micro-
curies
100
10
10
100
10
10
100
10
10
10
1
10
10
10
100
10
10
100
100
100
1
100
1,000
100
1
10
10
10
10
10
1,000
100
10
10
1
100
10
100
100
100
100
1
1
2 6 5
Complete printed PDF page 200
Radioactive
Material
Europium-155 	(Eu 155)
Fluorine-18 	(F 18)
Gadolinium-153 	(Gd 153)
Gadolinium-159 	(Gd 159)
Gallium-67 	(Ga 67)
Gallium-72 	(Ga 72)
Germanium-68 	(Ge 68)
Germanium-71 	(Ge 71)
Gold-195 	(Au 195)
Gold-198 	(Au 198)
Gold-199 	(Au 199)
Hafnium-181 	(Hf 181)
Holmium-166 	(Ho 166)
Hydrogen-3 	(H 3)
Indium-111 	(In 111)
Indium-113m 	(In 113m)
Indium-114m 	(In 114m)
Indium-115m 	(In 115m)
Indium-115 	(In 115)
Iodine-123 	(I 123)
Iodine-125 	(I 125)
Iodine-126 	(I 126)
Iodine-129 	(I 129)
Iodine-131 	(I 131)
Iodine-132 	(I 132)
Iodine-133 	(I 133)
Iodine-134 	(I 134)
Iodine-135 	(I 135)
Iridium-192 	(Ir 192)
Iridium-194 	(Ir 194)
Iron-52 	(Fe 52)
Iron-55 	( Fe 55)
Iron-59 	(Fe 59)
Krypton-BS 	(Kr 85)
Krypton-87 	(Kr 87)
Lanthanum-140 	(La 140)
Lutetium-177 	(Lu 177)
Manganese-52 	(Mn 52)
Manganese-54 	(Mn 54)
Manganese-56 	(Mn 56)
Mercury-197m 	(Hg 197m)
Mercury-197 	(Hg 197)
Mercury-203 	(Hg 203)
Molybdenum-99 	(Mo 99)
Neodymium-147 	(Nd 147)
Neodymium-149 	(Nd 149)
Nickel-59 	(Ni 59)
2-B-2
Micro-
curies
10
1,000
10
100
100
10
10
100
10
100
100
10
100
1,000
100
100
10
100
10
100
1
1
0.1
1
10
1
10
10
10
100
10
100
10
100
10
10
100
10
10
10
100
100
10
100
100
100
100
2
Complete printed PDF page 201
Radioactive
Material·
Nickel-63 	(Ni 63)
Nickel-65 	(Ni 65)
Niobium-93m 	(Nb 93m)
Niobium-95 	(Nb 95)
Niobium-97 	(Nb 97)
Osmium-185 	(Os 	185)
Osmium-191m 	(Os 191m)
Osmium-191 	(Os 	191)
Osmium-193 	(Os 	193)
Palladium-103 	(Pd 103)
Palladium-109 	(Pd 109)
Phosphorus-32 	(P 32)
Platinum-191 	(Pt 191)
Platinum-193m 	(Pt 193m)
Platinum-193 	(Pt 193)
Platinum-197m 	(Pt 197m)
Platinum-197 	(Pt 197)
Polonium-210 	(Po 210)
Potassium-42 	(K 42)
Potassium-43 	(K 43)
Praseodymium-142 	(Pr 142)
Praseodymium-143 	(Pr 143)
Promethium-147 	(Pm 147)
Promethium-149 	(Pm 149)
Rhenium-186 	(Re 186)
Rhenium-188 	(Re 188)
Rhodium-103m 	(Rh 103m)
Rhodium-105 	(Rh 105)
Rubidium-Bl 	(Rb 81)
Rubidium-86 	(Rb 86)
Rubidium-87 	(Rb 87)
Ruthenium-97 	(Ru 97)
Ruthenium-103 	(Ru 103)
Ruthenium-105 	(Ru 105)
Ruthenium-106 	(Ru 106)
Samarium-151 	(Sm 151)
Samarium-153 	(Sm 153)
Scandium-46 	(Sc 46)
Scandium-47 	(Sc 47)
Scandium-48 	(Sc 48)
Selenium-75 	(Se 75)
Silicon-31 	(Si 31)
Silver-105 	(Ag 105)
Silver-ll0m 	(Ag 110m)
Silver-111 	(Ag 111)
Sodium-22 	(Na 22)
2-B-3
Micro-
curies
10
100
10
10
10
10
100
100
100
100
100
10
100
100
100
100
100
0.1
10
10
100
100
10
10
100
100
100
100
10
10
10
100
10
10
1
10
100
10
100
10
10
100
10
1
100
10
2265
Complete printed PDF page 202
Radioactive
Material
Sodium-24 	(Na 24)
Strontium-85 	(Sr 85)
Strontium-89 	(Sr 89)
Strontium-90 	(Sr 90)
Strontium-91 	(Sr 91)
Strontium-92 	(Sr 92)
Sulphur-35 	(S 35)
Tantalum-182 	(Ta 182)
Technetium-96 	(Tc 96)
Technetium-97m 	(Tc 97m)
Technetium-97 	(Tc 97)
Technetium-99m 	(Tc 99m)
Technetium-99 	(Tc 99)
Tellurium-125m 	(Te 125m)
Tellurium-127m 	(Te 127m)
Tellurium-127 	(Te 127)
Tellurium-129m 	(Te 129m)
Tellurium-129 	(Te 129)
Tellurium-131m 	(Te 131m)
Tellurium-132 	(Te 132)
Terbium-160 	(Tb 160)
Thallium-200 	(Tl 200)
Thallium-201 	(Tl 201)
Thallium-202 	(Tl 202)
Thallium-204 	(Tl 204)
Thulium-170 	(Tm 170)
Thulium-171 	(Tm 171)
Tin-113 	(Sn 113)
Tin-125 	(Sn 125)
Tungsten-181 	(W 181)
Tungsten-185 	(W 185)
Tungsten-187 	(W 187)
Vanadium-48 	(V 48)
Xenon-13lm 	(Xe 131m)
Xenon-133 	(Xe 133)
Xenon-135 	(Xe 135)
Ytterbium-175 	(Yb 175)
Yttrium-87 	(Y 87)
Yttrium-88 	(Y 88)
Yttrium-90 	(Y 90)
Yttrium-91 	(Y 91)
Yttrium-92 	(Y 92)
Yttrium-93 	(Y 93)
Zinc-65 	(Zn 65)
Zinc-69m 	(Zn 69m)
Zinc-69 	( Zn 69)
Zirconium-93 	(Zr 93)
2-B-4
Micro-
curies
10
10
1
0.1
10
10
100
10
10
100
100
100
10
10
10
100
10
100
10
10
10
100
100
100
10
10
10
10
10
10
10
100
10
1,000
100
100
100
10
10
10
10
100
100
10
100
1,000
10
265
Complete printed PDF page 203
Radioactive
Material
Zirconium-95 	(Zr 95)
Zirconium-97 	(Zr 97)
Any radioactive material
not listed above other than
alpha-emitting radioactive
material
Micro-
curies
10
10
0.1
Note 	1: 	For purposes 	of C.25(f) (5) (ii) 	where 	there 	is 	involved a
combination of radionuclides, 	the limit for the combination should
be derived as 	follows:
Determine 	the 	amount 	of 	each 	radionuclide 	possessed 	and
1,000 	times 	the 	amount 	in 	Appendix 	B 	for 	each 	of 	those
radionuclides 	when 	not 	in 	combination. 	The 	sum 	of 	the
ratios 	of those quantities may not exceed 1.
Example:
Amt. 	of 	Radionuclide 	A possessed 	+ Amt. 	of 	Radionuclide 	B
possessed::; 1
1000 	x 	Appendix 	B 	quantity 	1000 	x 	Appendix 	B
quantity for Radionuclide A 	for Radionuclide B
Note 	2: 	To 	convert 	micro curies 	(μCi) 	to 	SI
kilobecquerels 	(kBq), 	multiply the above values by 37. units 	of
Example: 	Zirconium-97 	(10 	μCi 	multiplied by 	37 	is 	equivalent 	to
370 	kBq).
2-B-5
265
Complete printed PDF page 204
Appendix A of Subchapter 4
PROTECTION FACTORS FOR RESPIRATORS 1 	(2/2/93)
Protection Factors 4 	Tested & Certified Equipment
Description
Administration
Modes 	Particu- 	Particu- 	National Institute for
lates 	lates, 	Occupational Safety and
I.AIR-PURIFYING RESPIRATORS
Facepiece, 	half-mask 7
Facepiece, 	full
Facepiece, 	half-mask
full, 	or hood
II.ATMOSPHERE-SUPPLYING
RESPIRATORS
NP
NP
pp
1. 	Air-line 	respirator
Facepiece, 	half-mask 	CF
Facepiece, 	half-mask 	D
Facepiece, 	full 	CF
Facepiece, 	full 	D
Facepiece, 	full 	PD
Hood 	CF
Suit 	CF
2. 	Self-contained
breathing apparatus
(SCBA)
Facepiece, 	full
Facepiece, 	full
Facepiece, 	full
Facepiece, 	full
III.COMBINATION RESPIRATORS
D
PD
RD
RP
10
50
1000
Any combination of
air-purifying and
atmosphere-supplying
respirators
Protection factor
for type and mode
of operation as
listed above
See next page 	for 	footnotes.
4-A-1
1000
5
2000
5
2000
8
9
50
10,00011
50
5, 000 12
tests 	for permissibility
30 CFR 11,
Subpart K.
30 CFR 11,
Subpart J.
30 CFR 11,
Subpart H.
30 CFR 11,
10
Sec. 	11. 63 (b).
265
Complete printed PDF page 205
FOOTNOTES
1. 	For 	use 	in 	the 	selection 	of 	respiratory 	protective
equipment to be used only where 	the 	contaminants 	have been
identified 	and 	the 	concentrations, 	or 	possible
concentrations, 	are known.
2. 	Only for 	shaven faces 	and where nothing interferes with the
seal 	of 	tight-fitting 	facepieces 	against 	the 	skin. 	Hoods
and suits 	are excepted.
3.
4.
The
CF=
D
NP
mode symbols are defined as 	follows:
continuous 	flow
demand
negative 	pressure, 	that 	is, 	hegative
inhalation phase 	during
PD= pressure demand, 	that is, 	always positive pressure
PP 	positive pressure
RD 	demand, 	recirculating or closed circuit
RP= pressure demand, 	recirculating or closed circuit
a. 	The 	protection 	factor 	is 	a 	measure 	of 	the 	degree 	of
protection 	afforded 	by 	a 	respirator, 	defined 	as 	the
ratio 	of 	the 	concentration 	of 	airborne 	radioactive
material 	outside 	the 	respiratory protective 	equipment
to 	that 	inside 	the 	equipment, 	usually 	inside 	the
facepiece, 	under 	conditions 	of use. 	It 	is 	applied to
the 	ambient 	airborne 	concentration 	to 	estimate 	the
concentrations 	inhaled by the 	wearer 	according 	to 	the
following formula:
Concentration inhaled= Ambient airborne concentration
Protection factor
b. 	The protection factors 	apply:
(i) 	Only for individuals trained in using respirators
and wearing properly 	fitted 	respirators 	that 	are
used and maintained under 	supervision 	in a 	well-
planned respiratory protective program.
(ii) 	For 	air-purifying 	respirators 	only 	when 	high
efficiency 	particulate 	filters, 	above 	99. 97%
removal 	efficiency by thermally 	generated 	0.3 /ffil
dioctyl 	phthalate 	(DOP) 	test 	or 	equivalent, 	are
used 	in 	atmospheres 	not 	deficient 	in 	oxygen 	and
not 	containing 	radioactive 	gas 	or 	vapor
respiratory hazards.
(iii)No 	adjustment 	is 	to 	be 	made 	for 	the 	use 	of
sorbents 	against 	radioactive material 	in 	the 	form 	of
gases 	or vapors.
(iv) 	For 	atmosphere-supplying 	respirators 	only 	when
supplied 	with 	adequate 	respirable 	air.
Respirable 	air 	shall 	be 	provided 	of 	the 	quality
and 	quantity 	required 	in 	accordance 	with 	the
National 	Institute 	for 	Occupational 	Safety 	and
Health 	and 	the 	Mine 	Safety 	and 	Health
Administration 	certification described 	in 	30 	CFR
11. 	Oxygen and air shall not be used in the same
apparatus.
4-A-2
5
Complete printed PDF page 206
5. 	Excluding 	radioactive 	contaminants 	that 	present 	an
absorption 	or 	submersion 	hazard. 	For 	tritium 	oxide,
approximately one-third of the 	intake 	occurs 	by absorption
through 	the 	skin 	so 	that 	an 	overall 	protection 	factor 	of
less 	than 	2 	is 	appropriate 	when 	atmosphere-supplying
respirators 	are used to 	protect 	against 	tritium oxide. 	If
the 	protection 	factor 	for 	respiratory protective 	equipment
is 	5, 	the 	effective protection 	factor 	for 	tritium is 	about
1. 4; 	with 	protection 	factors 	of 	10, 	the 	effective 	factor
for tritium oxide is about 1.7; 	and with protection factors
of 	100 	or more, 	the 	effective 	factor 	for 	tritium oxide 	is
about 	1.9. 	Air-purifying 	respirators 	are 	not 	suitable 	for
protection 	against 	tritium 	oxide. 	See 	also 	footnote 	9
concerning supplied-air suits.
6. 	Canisters 	and 	cartridges 	shall 	not be used beyond 	service-
life limitations.
7. 	Under-chin 	type 	only. 	This 	type 	of 	respirator 	is 	not
satisfactory 	for 	use 	where 	it 	might 	be 	possible, 	such 	as,
if an accident or emergency were 	to 	occur, 	for 	the 	ambient
airborne 	concentrations 	to 	reach 	instantaneous 	values
greater 	than 	10 	times 	the 	pertinent 	values 	in 	Table 	I,
Column 	3 	of 	Appendix 	B 	of 	Subchapter 	4. 	This 	type 	of
respirator is not suitable for protection against plutonium
or other high-toxicity materials. 	The mask is to be tested
for 	fit prior to use, 	each time it is donned.
8. 	a. 	Equipment 	shall 	be 	operated 	in 	a 	manner 	that 	ensures
that 	proper 	air 	flow-rates 	are 	maintained. 	A
protection factor of no more than 1000 may be utilized
for 	tested-and-certified 	supplied-air 	hoods 	when 	a
minimum 	air 	flow 	of 	6 	cubic 	feet 	per 	minute 	( 0 .17
m3 /min) 	is maintained and calibrated air line pressure
gauges 	or 	flow 	measuring 	devices 	are 	used. 	A
protection factor of up to 2000 may be used for 	tested
and 	certified 	hoods 	only 	when 	the 	air 	flow 	is
maintained 	at 	the 	manufacturer's 	recommended 	maximum
rate 	for 	the 	equipment, 	this 	rate 	is 	greater 	than 	6
cubic feet per minute 	(0.17 m3 /min) 	and calibrated air
line 	pressure 	gauges 	or 	flow 	measuring 	devices 	are
used.
b. 	The design of the 	supplied-air hood or helmet, 	with a
minimum flow 	of 	6 	cubic 	feet 	per minute 	(0.17 	m3 /min)
of 	air, 	may 	determine 	its 	overall 	efficiency 	and 	the
protection 	it 	provides. 	For 	example, 	some 	hoods
aspirate contaminated air into the breathing zone when
the 	wearer 	works 	with 	hands-over-head. 	This
aspiration 	may 	be 	overcome 	if 	a 	short 	cape-like
extension 	to 	the 	hood 	is 	worn 	under 	a 	coat 	or 	over-
alls. 	Other 	limitations 	specified 	by 	the 	approval
agency 	shall 	be 	considered 	before 	using 	a 	hood 	in
certain types 	of atmospheres. 	See footnote 9.
9. 	Appropriate protection 	factors 	shall 	be 	determined, 	taking
into account the design of the suit and its permeability to
the 	contaminant under 	conditions 	of use. 	There 	shall be a
standby 	rescue 	person 	equipped with 	a 	respirator 	or 	other
apparatus 	appropriate 	for 	the 	potential 	hazards 	and
communications 	equipment 	whenever 	supplied-air 	suits 	are
used.
4-A-3
265
Complete printed PDF page 207
10. 	No 	approval 	schedules 	are 	currently 	available 	for 	this
equipment. 	Equipment 	is 	to 	be 	evaluated 	by 	testing 	or 	on
the basis of reliable test information.
11. 	This 	type 	of 	respirator may provide 	greater protection and
be 	used 	as 	an 	emergency 	device 	in 	unknown 	concentrations
for 	protection 	against 	inhalation 	hazards. 	External
radiation 	hazards 	and 	other 	limitations 	to 	permitted
exposure, 	such 	as 	skin 	absorption, 	must 	be 	taken 	into
account in such circumstances.
12. 	Quantitative 	fit 	testing 	shall 	be 	performed 	on 	each
individual, 	and no more than 	0.02% 	leakage 	is 	allowed with
this 	type 	of apparatus. 	Perceptible outward leakage 	of 	gas
from this or any positive pressure self-contained breathing
apparatus 	is 	unacceptable 	because 	service 	life 	will 	be
reduced substantially. 	Special training in the use of this
type of apparatus 	shall be provided to the wearer.
Note 1: 	Protection factors 	for 	respirators 	approved by the U.S.
Bureau 	of 	Mines 	and 	the 	National 	Institute 	for 	Occupational
Safety 	and 	Health, 	according 	to 	applicable 	approvals 	for
respirators 	for type and mode of use to protect against airborne
radionuclides, 	may be used to the extent that they do not exceed
the 	protection 	factors 	listed 	in 	this 	table. 	The 	protection
factors 	listed 	in 	this 	table 	may 	not 	be 	appropriate 	to
circumstances where 	chemical 	or 	other 	respiratory hazards 	exist
in 	addition 	to 	radioactive 	hazards. 	The 	selection 	and 	use 	of
respirators 	for 	such 	circumstances 	should 	take 	into 	account
applicable 	approvals 	of 	the 	U.S. 	Bureau 	of 	Mines 	and 	the
National Institute for Occupational Safety and Health.
Note 	2: 	Radioactive 	contaminants, 	for 	which 	the 	concentration
values 	in 	Table 	I, 	Column 	3 	of Appendix 	B of 	Subchapter 	4 	are
based on 	internal 	dose 	due 	to 	inhalation, 	may 	present 	external
exposure 	hazards 	at 	higher 	concentrations. 	Under 	these
circumstances, 	limitations 	on occupancy may have 	to be 	governed
by external dose limits.
4-A-4
2265
Complete printed PDF page 208
Appendix C of Subchapter 4
QUANTITIES 1 OF MATERIAL REQUIRING LABELING 	(2/2/93)
Radionuclide 	Quantity 	Radionuclide 	Quantity
( Ci)* 	( Ci)*
Hydrogen-3 	1,000 	Chromium-48 	1,000
Beryllium-7 	1,000 	Chromium-49 	1,000
Beryllium-10 	1 	Chromium-51 	1,000
Carbon-11 	1,000 	Manganese-51 	1,000
Carbon-14 	1,000 	Manganese-52m 	1,000
Fluorine-18 	1,000 	Manganese-52 	100
Sodium-22 	10 	Manganese-53 	1,000
Sodium-24 	100 	Manganese-54 	100
Magnesium-28 	100 	Manganese-56 	1,000
Aluminum-26 	10 	Iron-52 	100
Silicon-31 	1,000 	Iron-55 	100
Silicon-32 	1 	Iron-59 	10
Phosphorus-32 	10 	Iron-60 	1
Phosphorus-33 	100 	Cobalt-55 	100
Sulfur-35 	100 	Cobalt-56 	10
Chlorine-36 	10 	Cobalt-57 	100
Chlorine-38 	1,000 	Cobalt-58m 	1,000
Chlorine-39 	1,000 	Cobalt-58 	100
Argon-39 	1,000 	Cobalt-60m 	1,000
Argon-41 	1,000 	Cobalt-60 	1
Potassium-40 	100 	Cobalt-61 	1,000
Potassium-42 	1,000 	Cobalt-62m 	1,000
Potassium-43 	1,000 	Nickel-56 	100
Potassium-44 	1,000 	Nickel-57 	100
Potassium-45 	1,000 	Nickel-59 	100
Calcium-41 	100 	Nickel-63 	100
Calcium-45 	100 	Nickel-65 	1,000
Calcium-47 	100 	Nickel-66 	10
Scandium-43 	1,000 	Copper-60 	1,000
Scandium-44m 	100 	Copper-61 	1,000
Scandium-44 	100 	Copper-64 	1,000
Scandium-46 	10 	Copper-67 	1,000
Scandium-47 	100 	Zinc-62 	100
Scandium-48 	100 	Zinc-63 	1,000
Scandium-49 	1,000 	Zinc-65 	10
Titanium-44 	1 	Zinc-69m 	100
Titanium-45 	1,000 	Zinc-69 	1,000
Vanadium-47 	1,000 	Zinc-71m 	1,000
Vanadium-48 	100 	Zinc-72 	100
Vanadium-49 	1,000 	Gallium-65 	1,000
* To 	convert μCi 	to kBq, 	multiply the μCi value by 37.
4-C-l
265
Complete printed PDF page 209
Radionuclide 	Quantity 	Radionuclide 	Quantity
( Ci)* 	( Ci)*
Gallium-66 	100 	Krypton-Sl 	1,000
Gallium-67 	1,000 	Krypton-S3m 	1,000
Gallium-6S 	1,000 	Krypton-SSm 	1,000
Gallium-70 	1,000 	Krypton-SS 	1,000
Gallium-72 	100 	Krypton-S7 	1,000
Gallium-73 	1,000 	Krypton-SS 	1,000
Germanium-66 	1,000 	Rubidium-79 	1,000
Germanium-67 	1,000 	Rubidium-Slm 	1,000
Germanium-6S 	10 	Rubidium-Sl 	1,000
Germanium-69 	1,000 	Rubidium-S2m 	1,000
Germanium-71 	1,000 	Rubidium-S3 	100
Germanium-75 	1,000 	Rubidium-S4 	100
Germanium-77 	1,000 	Rubidium-S6 	100
Germanium-7S 	1,000 	Rubidium-S7 	100
Arsenic-69 	1,000 	Rubidium-SS 	1,000
Arsenic-70 	1,000 	Rubidium-S9 	1,000
Arsenic-71 	100 	Strontium-SO 	100
Arsenic-72 	100 	Strontium-Sl 	1,000
Arsenic-73 	100 	Strontium-S3 	100
Arsenic-74 	100 	Strontium-SSm 	1,000
Arsenic-76 	100 	Strontium-SS 	100
Arsenic-77 	100 	Strontium-S7m 	1,000
Arsenic-7S 	1,000 	Strontium-S9 	10
Selenium-70 	1,000 	Strontium-90 	0.1
Selenium-73m 	1,000 	Strontium-91 	100
Selenium-73 	100 	Strontium-92 	100
Selenium-75 	100 	Yttrium-S6m 	1,000
Selenium-79 	100 	Yttrium-S6 	100
Selenium-Slm 	1,000 	Yttrium-S7 	100
Selenium-Sl 	1,000 	Yttrium-SS 	10
Selenium-S3 	1,000 	Yttrium-90m 	1,000
Bromine-74m 	1,000 	Yttrium-90 	10
Bromine-74 	1,000 	Yttrium-91m 	1,000
Bromine-75 	1,000 	Yttrium-91 	10
Bromine-76 	100 	Yttrium-92 	100
Bromine-77 	1,000 	Yttrium-93 	100
Bromine-Som 	1,000 	Yttrium-94 	1,000
Bromine-SO 	1,000 	Yttrium-95 	1,000
Bromine-S2 	100 	Zirconium-S6 	100
Bromine-S3 	1,000 	Zirconium-SS 	10
Bromine-S4 	1,000 	Zirconium-S9 	100
Krypton-74 	1,000 	Zirconium-93 	1
Krypton-76 	1,000 	Zirconium-95 	10
Krypton-77 	1,000 	Zirconium-97 	100
Krypton-79 	1,000
* To convert μCi 	to kBq, 	multiply the μCi 	value by 37.
4-C-2
265
Complete printed PDF page 210
Radionuclide 	Quantity 	Radionuclide 	Quantity
( Ci)* 	( Ci)*
Niobium-88 	1,000 	Palladium-101 	1,000
Niobium-89m 	Palladium-103 	100
(66 min) 	1,000 	Palladium-107 	10
Niobium-89 	Palladium-109 	100
(122 min) 	1,000 	Silver-102 	1,000
Niobium-90 	100 	Silver-103 	1,000
Niobium-93m 	10 	Silver-104m 	1,000
Niobium-94 	1 	Silver-104 	1,000
Niobium-95m 	100 	Silver-105 	100
Niobium-95 	100 	Silver-106m 	100
Niobium-96 	100 	Silver-106 	1,000
Niobium-97 	1,000 	Sil ver-108m 	1
N.i::obium-98 	1,000 	Silver-llOm 	10
Molybdenum-90 	100 	Silver-111 	100
Molybdenum-93m 	100 	Silver-112 	100
Molybdenum-93 	10 	Silver-115 	1,000
Molybdenum-99 	100 	Cadmium-104 	1,000
Molybdenum-101 	1,000 	Cadmium-107 	1,000
Technetium-93m 	1,000 	Cadmium-109 	1
Technetium-93 	1,000 	Cadmium-113m 	0.1
Technetium-94m 	1,000 	Cadmium-113 	100
Technetium-94 	1,000 	Cadmium-115m 	10
Technetium-96m 	1,000 	Cadmium-115 	100
Technetium-96 	100 	Cadmium-117m 	1,000
Technetium-97m 	100 	Cadmium-117 	1,000
Technetium-97 	1,000 	Indium-109 	1,000
Technetium-98 	10 	Indium-ll0m
Technetium-99m 	1,000 	(69.lm) 	1,000
Technetium-99 	100 	Indium-11O
Technetium-101 	1,000 	(4. 9h) 	1,000
Technetium-104 	1,000 	Indium-111 	100
Ruthenium-94 	1,000 	Indium-112 	1,000
Ruthenium-97 	1,000 	Indium-113m 	1,000
Ruthenium-103 	100 	Indium-114m 	10
Ruthenium-105 	1,000 	Indium-115m 	1,000
Ruthenium-106 	1 	Indium-115 	100
Rhodium-99m 	1,000 	Indium-116m 	1,000
Rhodium-99 	100 	Indium-117m 	1,000
Rhodium-100 	100 	Indium-117 	1,000
Rhodium-l0lm 	1,000 	Indium-119m 	1,000
Rhodium-101 	10 	Tin-110 	100
Rhodium-102m 	10 	Tin-111 	1,000
Rhodium-102 	10 	Tin-113 	100
Rhodium-103m 	1,000 	Tin-117m 	100
Rhodium-105 	100 	Tin-119m 	100
Rhodium-106m 	1,000 	Tin-121m 	100
Rhodium-107 	1,000 	Tin-121 	1,000
Palladium-100 	100
* To convert μCi 	to kBq, 	multiply the μCi value by 37.
4-C-3
265
Complete printed PDF page 211
Radionuclide 	Quantity 	Radionuclide 	Quantity
( Ci)* 	( Ci)*
Tin-123m 	1,000 	Tellurium-133 	1,000
Tin-123 	10 	Tellurium-134 	1,000
Tin-125 	10 	Iodine-120m 	1,000
Tin-126 	10 	Iodine-120 	100
Tin-127 	1,000 	Iodine-121 	1,000
Tin-128 	1,000 	Iodine-123 	100
Antimony-115 	1,000 	Iodine-124 	10
Antimony-116m 	1,000 	Iodine-125 	1
Antimony-116 	1,000 	Iodine-126 	1
Antimony-117 	1,000 	Iodine-128 	1,000
Antimony-118m 	1,000 	Iodine-129 	1
Antimony-119 	1,000 	Iodine-130 	10
Antimony-120 	Iodine-131 	1
(16ml 	1,000 	Iodine-132m 	100
Antimony-120 	Iodine-132 	100
(5. 76d) 	100 	Iodine-133 	10
Antimony-122 	100 	Iodine-134 	1,000
Antimony-124m 	1,000 	Iodine-135 	100
Antimony-124 	10 	Xenon-120 	1,000
Antimony-125 	100 	Xenon-121 	1,000
Antimony-126m 	1,000 	Xenon-122 	1,000
Antimony-126 	100 	Xenon-123 	1,000
Antimony-127 	100 	Xenon-125 	1,000
Antimony-128 	Xenon-127 	1,000
(10.4ml 	1,000 	Xenon-129m 	1,000
Antimony-128 	Xenon-131m 	1,000
(9.Olh) 	100 	Xenon-133m 	1,000
Antimony-129 	100 	Xenon-133 	1,000
Antimony-130 	1,000 	Xenon-135m 	1,000
Antimony-131 	1,000 	Xenon-135 	1,000
Tellurium-116 	1,000 	Xenon-138 	1,000
Tellurium-121m 	10 	Cesium-125 	1,000
Tellurium-121 	100 	Cesium-127 	1,000
Tellurium-123m 	10 	Cesium-129 	1,000
Tellurium-123 	100 	Cesium-130 	1,000
Tellurium-125m 	10 	Cesium-131 	1,000
Tellurium-127m 	10 	Cesium-132 	100
Tellurium-127 	1,000 	Cesium-134m 	1,000
Tellurium-129m 	10 	Cesium-134 	10
Tellurium-129 	1,000 	Cesium-135m 	1,000
Tellurium-131m 	10 	Cesium-135 	100
Tellurium-131 	100 	Cesium-136 	10
Tellurium-132 	10 	Cesium-137 	10
Tellurium-133m 	100 	Cesium-138 	1,000
* To convert μCi 	to kBq, 	multiply the μCi 	value by 37.
4-C-4
265
Complete printed PDF page 212
Radionuclide
Barium-126
Barium-128
Barium-131m
Barium-131
Barium-133m
Barium-133
Barium-135m
Barium-139
Barium-140
Barium-141
Barium-142
Lanthanum-131
Lanthanum-132
Lanthanum-135
Lanthanum-137
Lanthanum-138
Lanthanum-140
Lanthanum-141
Lanthanum-142
Lanthanum-143
Cerium-134
Cerium-135
Cerium-137m
Cerium-137
Cerium-139
Cerium-141
Cerium-143
Cerium-144
Praseodymium-136
Praseodymium-137
Praseodymium-138m
Praseodymium-139
Praseodymium-142m
Praseodymium-142
Praseodymium-143
Praseodymium-144
Praseodymium-145
Praseodymium-147
Neodymium-136
Neodymium-138
Neodymium-139m
Neodymium-139
Neodymium-141
Neodymium-147
Neodymium-149
Neodymium-151
Quantity
( Ci)*
1,000
100
1,000
100
100
100
100
1,000
100
1,000
1,000
1,000
100
1,000
10
100
100
100
1,000
1,000
100
100
100
1,000
100
100
100
1
1,000
1,000
1,000
1,000
1,000
100
100
1,000
100
1,000
1,000
100
1,000
1,000
1,000
100
1,000
1,000
Radionuclide
Promethium-141
Promethium-143
Promethium-144
Promethium-145
Promethium-146
Promethium-147
Promethium-148m
Promethium-148
Promethium-149
Promethium-150
Promethium-151
Samarium-141m
Samarium-141
Samarium-142
Samarium-145
Samarium-146
Samarium-147
Samarium-151
Samarium-153
Samarium-155
Samarium-156
Europium-145
Europium-146
Europium-147
Europium-148
Europium-149
Europium-150
(12.62h)
Europium-150
(34.2y)
Europium-152m
Europium-152
Europium-154
Europium-155
Europium-156
Europium-157
Europium-158
Gadolinium-145
Gadolinium-146
Gadolinium-147
Gadolinium-148
Gadolinium-149
Gadolinium-151
Gadolinium-152
Gadolinium-153
Gadolinium-159
* To 	convert μCi 	to kBq, 	multiply the μCi 	value by 37.
4-C-5
Quantity
( Ci)*
1,000
100
10
10
1
10
10
10
100
1,000
100
1,000
1,000
1,000
100
1
100
10
100
1,000
1,000
100
100
100
10
100
100
1
100
1
1
10
" ""	l.UU
100
1,000
1,000
10
100
0.001
100
10
100
10
100
265
Complete printed PDF page 213
Radionuclide 	Quantity 	Radionuclide 	Quantity
( Ci)* 	( Ci)*
Terbium-147 	1,000 	Ytterbium-162 	1,000
Terbium-149 	100 	Ytterbium-166 	100
Terbium-150 	1,000 	Ytterbium-167 	1,000
Terbium-151 	100 	Ytterbium-169 	100
Terbium-153 	1,000 	Ytterbium-175 	100
Terbium-154 	100 	Ytterbium-177 	1,000
Terbium-155 	1,000 	Ytterbium-178 	1,000
Terbium-156m 	Lutetium-169 	100
(5.Oh) 	1,000 	Lutetium-170 	100
Terbium-156m 	Lutetium-171 	100
(24. 4h) 	1,000 	Lutetium-172 	100
Terbium-156 	100 	Lutetium-173 	10
Terbium-157 	10 	Lutetium-174m 	10
Terbium-158 	1 	Lutetium-174 	10
Terbium-160 	10 	Lutetium-176m 	1,000
Terbium-161 	100 	Lutetium-176 	100
Dysprosium-155 	1,000 	Lutetium-177m 	10
Dysprosium-157 	1,000 	Lutetium-177 	100
Dysprosium-159 	100 	Lutetium-178m 	1,000
Dysprosium-165 	1,000 	Lutetium-178 	1,000
Dysprosium-166 	100 	Lutetium-179 	1,000
Holmium-155 	1,000 	Hafnium-170 	100
Holmium-157 	1,000 	Hafnium-172 	1
Holmium-159 	1,000 	Hafnium-173 	1,000
Holmium-161 	1,000 	Hafnium-175 	100
Holmium-162m 	1,000 	Hafnium-177m 	1,000
Holmium-162 	1,000 	Hafnium-178m 	0.1
Holmium-164m 	1,000 	Hafnium-179m 	10
Holmium-164 	1,000 	Hafnium-180m 	1,000
Holmium-166m 	1 	Hafnium-181 	10
Holmium-166 	100 	Hafnium-182m 	1,000
Holmium-167 	1,000 	Hafnium-182 	0.1
Erbium-161 	1,000 	Hafnium-183 	1,000
Erbium-165 	1,000 	Hafnium-184 	100
Erbium-169 	100 	Tantalum-172 	1,000
Erbium-171 	100 	Tantalum-173 	1,000
Erbium-172 	100 	Tantalum-174 	1,000
Thulium-162 	1,000 	Tantalum-175 	1,000
Thulium-166 	100 	Tantalum-176 	100
Thulium-167 	100 	Tantalum-177 	1,000
Thulium-170 	10 	Tantalum-178 	1,000
Thulium-171 	10 	Tantalum-179 	100
Thulium-172 	100 	Tantalum-180m 	1,000
Thulium-173 	100 	Tantalum-180 	100
Thulium-175 	1,000 	Tantalum-182m 	1,000
* To convert μCi 	to kBq, 	multiply the μCi value by 37.
4-C-6
265
Complete printed PDF page 214
Radionuclide 	Quantity 	Radionuclide 	Quantity
( Ci)* 	( Ci)*
Tantalum-182 	10 	Iridium-188 	100
Tantalum-183 	100 	Iridium-189 	100
Tantalum-184 	100 	Iridium-190m 	1,000
Tantalum-185 	1,000 	Iridium-190 	100
Tantalum-186 	1,000 	Iridium-192m
Tungsten-176 	1,000 	(1.4m) 	10
Tungsten-177 	1,000 	Iridium-192
Tungsten-178 	1,000 	(73. Sd) 	1
Tungsten-179 	1,000 	Iridium-194m 	10
Tungsten-181 	1,000 	Iridium-194 	100
Tungsten-185 	100 	Iridium-195m 	1,000
Tungsten-187 	100 	Iridium-195 	1,000
Tungsten-188 	10 	Platinum-186 	1,000
Rhenium-177 	1,000 	Platinum-188 	100
Rhenium-178 	1,000 	Platinum-189 	1,000
Rhenium-181 	1,000 	Platinum-191 	100
Rhenium-182 	Platinum-193m 	100
(12. 7h) 	1,000 	Platinum-193 	1,000
Rhenium-182 	Platinum-195m 	100
(64.Oh) 	100 	Platinum-197m 	1,000
Rhenium-184m 	10 	Platinum-197 	100
Rhenium-184 	100 	Platinum-199 	1,000
Rhenium-186m 	10 	Platinum-200 	100
Rhenium-186 	100 	Gold-193 	1,000
Rhenium-187 	1,000 	Gold-194 	100
Rhenium-188m 	1,000 	Gold-195 	10
Rhenium-188 	100 	Gold-198m 	100
Rhenium-189 	100 	Gold-198 	100
Osmium-180 	1,000 	Gold-199 	100
Osmium-181 	1,000 	Gold-200m 	100
Osmium-182 	100 	Gold-200 	1,000
Osmium-185 	100 	Gold-201 	1,000
Osmium-189m 	1,000 	Mercury-193m 	100
Osmium-19lm 	1,000 	Mercury-193 	1,000
Osmium-191 	100 	Mercury-194 	1
Osmium-193 	100 	Mercury-195m 	100
Osmium-194 	1 	Mercury-195 	1,000
Iridium-182 	1,000 	Mercury-197m 	100
Iridium-184 	1,000 	Mercury-197 	1,000
Iridium-185 	1,000 	Mercury-199m 	1,000
Iridium-186 	100 	Mercury-203 	100
Iridium-187 	1,000
* To convert μCi 	to kBq, 	multiply the μCi value by 37.
4-C-7
265
Complete printed PDF page 215
Radionuclide
Thallium-194m
Thallium-194
Thallium-195
Thallium-197
Thallium-198m
Thallium-198
Thallium-199
Thallium-201
Thallium-200
Thallium-202
Thallium-204
Lead-195m
Lead-198
Lead-199
Lead-200
Lead-201
Lead-202m
Lead-202
Lead-203
Lead-205
Lead-209
Lead-210
Lead-211
Lead-212
Lead-214
Bismuth-200
Bismuth-201
Bismuth-202
Bismuth-203
Bismuth-205
Bismuth-206
Bismuth-207
Bismuth-210m
Bismuth-210
Bismuth-212
Bismuth-213
Bismuth-214
Polonium-203
Polonium-205
Polonium-207
Polonium-210
Astatine-207
Astatine-211
Radon-220
Radon-222
Francium-222
Quantity
( Ci)*
1,000
1,000
1,000
1,000
1,000
1,000
1,000
1,000
1,000
100
100
1,000
1,000
1,000
100
1,000
1,000
10
1,000
100
1,000
0.01
100
1
100
1,000
1,000
1,000
100
100
100
10
0.1
1
10
10
100
1,000
1,000
1,000
0.1
100
10
1
1
100
Radionuclide
Francium-223
Radium-223
Radium-224
Radium-225
Radium-226
Radium-227
Radium-228
Actinium-224
Actinium-225
Actinium-226
Actinium-227
Actinium-228
Thorium-226
Thorium-227
Thorium-228
Thorium-229
Thorium-230
Thorium-231
Thorium-232
Thorium-234
Thorium-natural
Protactinium-227
Protactinium-228
Protactinium-230
Protactinium-231
Protactinium-232
Protactinium-233
Protactinium-234
Uranium-230
Uranium-231
Uranium-232
Uranium-233
Uranium-234
Uranium-235
Uranium-236
Uranium-237
Uranium-238
Uranium-239
Uranium-240
Uranium-natural
Neptunium-232
Neptunium-233
Neptunium-234
Neptunium-235
Neptunium-236
(1. 15E+5)
* To 	convert μCi 	to kBq, 	multiply the μCi 	value by 37.
4-C-8
Quantity
( Ci)*
100
0.1
0.1
0.1
0.1
1,000
0.1
1
0.01
0.1
0.001
1
10
0.01
0.001
0.001
0.001
100
100
10
100
10
1
0.1
0.001
1
100
100
0.01
100
0.001
0.001
0.001
0.001
0.001
100
100
1,000
100
100
100
1,000
100
100
0.001
65
Complete printed PDF page 216
Radionuclide
Neptunium-236
(22. 5h)
Neptunium-237
Neptunium-238
Neptunium-239
Neptunium-240
Plutonium-234
Plutonium-235
Plutonium-236
Plutonium-237
Plutonium-238
Plutonium-239
Plutonium-240
Plutonium-241
Plutonium-242
Plutonium-243
Plutonium-244
Plutonium-245
Americium-237
Americium-238
Americium-239
Americium-240
Americium-241
Americium-242m
Americium-242
Americium-243
Americium-244m
Americium-244
Americium-245
Americium-246m
Americium-246
Curium-238
Curium-240
Curium-241
Any alpha-emitting
radionuclide not
listed above or
mixtures 	of alpha
emitters 	of unknown
composition
Quantity
( Ci)*
1
0.001
10
100
1,000
10
1,000
0.001
100
0.001
0.001
0.001
0.01
0.001
1,000
0.001
100
1,000
100
1,000
100
0.001
0.001
10
0.001
100
10
1,000
1,000
1,000
100
0.1
1
0.001
Radionuclide
Curium-242
Curium-243
Curium-244
Curium-245
Curium-246
Curium-247
Curium-248
Curium-249
Berkelium-245
Berkelium-246
Berkelium-247
Berkelium-249
Berkelium-250
Californium-244
Californium-246
Californium-248
Californium-249
Californium-250
Californium-251
Californium-252
Californium-253
Californium-254
Einsteinium-250
Einsteinium-251
Einsteinium-253
Einsteinium-254m
Einsteinium-254
Fermium-252
Fermium-253
Fermium-254
Fermium-255
Fermium-257
Mendelevium-257
Mendelevium-258
Any radionuclide
Quantity
( Ci)*
0.01
0.001
0.001
0.001
0.001
0.001
0.001
1,000
100
100
0.001
0.1
10
100
1
0.01
0.001
0.001
0.001
0.001
0.1
0.001
100
100
0.1
1
0.01
1
1
10
1
0.01
10
0.01
other than alpha-
emitting radionuclides
not listed above, 	or
mixtures of beta
emitters of unknown
composition 	0.01
* To 	convert μCi 	to kBq, 	multiply the μCi 	value by 37.
4-C-9
2 6 5
Complete printed PDF page 217
Radionuclide 	Quantity
( Ci)*
Radionuclide 	Quantity
( Ci)*
NOTE: 	Where 	there 	is 	involved 	a 	combination 	of 	radionuclides 	in 	known
amounts, 	the 	limit 	for 	the 	combination 	shall 	be 	derived 	as 	follows:
determine, 	for 	each 	radionuclide 	in 	the 	combination, 	the 	ratio 	between 	the
quantity 	present 	in 	the 	combination 	and 	the 	limit 	otherwise 	established 	for
the specific radionuclide when not in combination. 	The sum of such ratios 	for
all 	radionuclides in the combination may not exceed "1" -- that is, 	unity.
1 The 	quantities 	listed 	above 	were 	derived 	by 	taking 	1/l0th 	of 	the 	most
restrictive 	ALI 	listed 	in 	Table 	I, 	Columns 	1 	and 	2, 	of 	Appendix 	B 	to
Subchapter 	4, 	rounding 	to 	the 	nearest 	factor 	of 	10, 	and 	constraining 	the
values 	listed between 	37 	Bq and 37 MBq 	(0.001 	and 1,000 μCi). 	Values 	of 	3.7
MBq 	(100 μCi) 	have been assigned for 	radionuclides 	having a 	radioactive 	half-
life in excess 	of E+9 	years, 	except 	rhenium, 	37 MBq 	(1,000 μCi), 	to 	take into
account their low specific activity.
* To convert μCi 	to kBq, 	multiply the μCi value by 37.
4-C-10
265
Complete printed PDF page 218
Appendix D of Subchapter 4
RADIATION SYMBOL 	(8/19/98)
Cross-hatched area is magenta, 	or purple, 	or black. 	The background is yellow.
60°
1 I 	I
~j.:+A+-: 	:
I 	I I A 	I
: 	:.r 2 	I
I_...'+- 	:
~SA.;..·---►..:
4-D-1
265
Complete printed PDF page 219
'11-45-63
4-D-l
265
Complete printed PDF page 220
Appendix A of Subchapter 5
SUBJECTS FOR INSTRUCTION OF RADIOGRAPHER TRAINEES 	(2/2/93)
Training to qualify individuals as 	radiographer trainees 	shall be
presented on a 	formal basis and shall include the following
subjects:
I.
II.
III.
IV.
v.
Fundamentals of Radiation Safety
A. 	Characteristics of radiation
B. 	Units 	of radiation dose and quantity of radioactivity
C. 	Significance of radiation dose
1. 	Radiation protection standards
2. 	Biological effects of radiation
3. 	Case histories of radiography accidents
D. 	Levels 	of radiation from sources of radiation
E. 	Methods 	of controlling radiation dose
1. 	Working time
2. 	Working distances
3. 	Shielding
Radiation Detection Instrumentation to be Used
A. 	Use of radiation survey instruments
1. 	Operation
2. 	Calibration
3. 	Limitations
B. 	Survey techniques
C. 	Use of personnel monitoring equipment
1. 	Film badges
2. 	Thermoluminescent dosimeters 	{TLD)
3. 	Pocket dosimeters
4. 	Alarm ratemeter
The Requirements of Pertinent Federal and State Regulations
The Registrant's Written Operating and Emergency
Procedures
Radiographic Equipment to be Used
A. 	Remote handling equipment
B. 	Operation and control of radiographic exposure devices
and sealed sources, 	including pictures or models of
source assemblies 	{pigtails)
C. 	Storage and transport containers, 	source changers
D. 	Operation and control of x-ray equipment
E. 	Collimators
5-A-1
2265
Complete printed PDF page 221
Appendix A of Subchapter 6
TIME-TEMPERATURE CHART 	(2/2/93)
Thermometer 	Minimum Developing Time
Reading 	(minutes)
(degrees)
C 	F
26. 7 	80 	2
26.1 	79 	2
25.6 	78 	2 1/2
25.0 	77 	2 1/2
24.4 	76 	3
23.9 	75 	3
23.3 	74 	3 1/2
22.8 	73 	3 1/2
22.2 	72 	4
21.7 	71 	4
21.1 	70 	4 1/2
20.6 	69 	4 1/2
20.0 	68 	5
19.4 	67 	5 1/2
18.9 	66 	5 1/2
18.3 	65 	6
17.8 	64 	6 1/2
17.2 	63 	7
16.7 	62 	8
16.1 	61 	8 1/2
15.6 	60 	9 1/2
6-A-1
2265
Complete printed PDF page 222
Appendix B of Subchapter 6
TEMPERATURE-IMMERSION CHART 	(2/2/93)
Developer Temperature 	Minimum Immersion
Time*
oc 	OF 	Seconds
35.5 	96 	19
35 	95 	20
34.5 	94 	21
34 	93 	22
33.5 	92 	23
33 	91 	24
32 	90 	25
31.5 	89 	26
31 	88 	27
30.5 	87 	28
30 	86 	29
29.5 	85 	30
*Immersion time only, 	no crossover time included.
6-B-1
2265
Complete printed PDF page 223
Appendix C of Subchapter 6
HALF-VALUE LAYER 	(7/25/96)
X-Ray Tube Voltage
(kilovolt peak)
Design 	Measured
Operating 	Potential
Range 	(kVp)
Below 51 	30
40
50
51 to 70 	51
60
70
Above 70 	71
80
90
100
llO
120
130
140
150
Minimum Half-Value Layer
(millimeters of aluminum)
Dental Intraoral 	All Other
Manufactured Before 	Diagnostic
8/1/74 	& On or After 	X-Ray
12/1/80 	Systems
N/A 	0.3
N/A 	0.4
1.5 	0.5
1.5 	1.2
1.5 	1. 3
1.5 	1.5
2.1 	2.1
2.3 	2.3
2.5 	2.5
2.7 	2.7
3.0 	3.0
3.2 	3.2
3.5 	3.5
3.8 	3.8
4.1 	4.1
6-C-l
265
Complete printed PDF page 224
Appendix A of Subchapter 13
SUBJECTS TO BE INCLUDED IN TRAINING COURSES
FOR LOGGING SUPERVISORS 	(2/2/93)
I. 	Fundamentals of Radiation Safety
A. 	Characteristics of radiation
B. 	Units 	of radiation dose and quantity of radioactivity
C. 	Significance of radiation dose
1. 	Radiation protection standards
2. 	Biological effects 	of radiation dose
D. 	Levels of radiation from sources of radiation
E. 	Methods of minimizing radiation dose
1. 	Working time
2. 	Working distances
3. 	Shielding
F. 	Radiation 	safety 	practices 	including 	prevention 	of
contamination and methods of decontamination
II. 	Radiation Detection Instrumentation to be Used
A. 	Use of radiation survey instruments
1. 	Operation
2. 	Calibration
3. 	Limitations
B. 	Survey techniques
C. 	Use of personnel monitoring equipment
III. 	Equipment to be Used
A. 	Handling equipment
B. 	Sources of radiation
C. 	Storage and control of equipment
D. 	Operation and control of equipment
IV. 	The Requirements 	of Pertinent Federal and State Regulations
V. 	The Licensee's Written Operating and Emergency Procedures
VI. 	The Licensee's Recordkeeping Procedures
13-A-1
2265
Complete printed PDF page 225
13-A-2
265
Complete printed PDF page 226
Appendix A of Subchapter 14
INFORMATION ON RADIATION SHIELDING
REQUIRED FOR PLAN REVIEWS 	(2/2/93)
I. 	ALL THERAPEUTIC RADIATION MACHINES
A. 	Basic facility information including: 	name, 	telephone
number and department registration number of the
individual responsible for preparation of the
shielding plan; 	name and telephone number of the
facility supervisor; 	and the street address 	including
room number of the external beam radiation therapy
facility. 	The plan should also indicate whether this
is a 	new structure or a modification to existing
structure(s).
B. 	All wall, 	floor, 	and ceiling areas 	struck by the
useful beam shall have primary barriers.
C. 	Secondary barriers shall be provided in all wall,
floor, 	and ceiling areas not having primary barriers.
II. 	THERAPEUTIC RADIATION MACHINES UP TO ONE HUNDRED FIFTY kV
(PHOTONS ONLY)
In addition to the requirements listed in Section I 	above,
therapeutic radiation machine facilities which produce only
photons with a maximum energy less than or equal to one
hundred fifty kV shall submit shielding plans which contain,
as a minimum, 	the following additional information:
A. 	Equipment specifications, 	including the manufacturer
and model number of the therapeutic radiation machine,
as well as 	the maximum technique factors.
B. 	Maximum design workload for the facility including
total weekly radiation output, 	expressed in gray 	(rad)
or air kerma at one meter, 	total beam-on time per day
or week, 	the average treatment time per patient, 	along
with the anticipated number of patients 	to be treated
per day or week.
C. 	A facility blueprint/drawing indicating: 	scale 0.25
inch= one foot is typical; 	direction of North; 	normal
location of the therapeutic radiation machine's
radiation port(s); 	the port's travel and traverse
limits; 	general direction(s) 	of the useful beam;
locations 	of any windows and doors; 	and the location
of the therapeutic radiation machine control panel.
If the control panel is located inside the external
beam radiation therapy treatment room, 	the location of
the operator's booth shall be noted on the plan and
the operator's station at the control panel shall be
behind a protective barrier sufficient to ensure
compliance with subchapter 4.
D. 	The structural composition and thickness 	or
lead/concrete equivalent of all walls, 	doors,
partitions, 	floor, 	and ceiling of the room(s)
concerned.
E. 	The type of occupancy of all adjacent areas 	inclusive
of space above and below the room(s) 	concerned. 	If
there is an exterior wall, 	show distance to the
closest area(s) 	where it is likely that individuals
may be present.
F. 	At least one example calculation which shows 	the
methodology used to determine the amount of shielding
14-A-1
2265
Complete printed PDF page 227
required for each physical condition 	(ie: 	primary and
secondary/leakage barriers, 	restricted and
unrestricted areas, 	entry door(s)) 	and shielding
material in the facility.
(1) 	If commercial software is used to generate
shielding requirements, 	please also identify the
software used and the version/ 	revision date.
(2) 	If the software used to generate shielding
requirements is not in the open literature,
please also submit quality control sample
calculations to verify the result obtained with
the software.
III. 	THERAPEUTIC RADIATION MACHINES OVER ONE HUNDRED FIFTY kV
In addition to the requirements listed in Section I 	above,
therapeutic radiation machine facilities which produce
photons with a maximum energy in excess 	of one hundred fifty
kV and/or electrons and/or protons or other subatomic
particles 	shall submit shielding plans which contain, 	as a
minimum, 	the following additional information:
A. 	Equipment specifications including the manufacturer
and model number of the therapeutic radiation machine,
and gray 	(rad) 	at the isocenter and the energy(s) 	and
type(s) 	of radiation produced 	(ie: 	photon, 	electron).
The source to isocenter distance shall be specified.
B. 	Maximum design workload for the facility including
total weekly radiation output expressed in gray 	(rad)
at one meter, 	total beam-on time per day or week, 	the
average treatment time per patient, 	along with the
anticipated number of patients to be treated per day
or week.
C. 	Facility blueprint/drawing including both floor plan
and elevation views indicating relative orientation of
the therapeutic radiation machine, 	scale 0.25 inch=
one 	foot is typical, 	type(s), 	thickness and minimum
density of shielding material(s), 	direction of North,
the locations and size of all penetrations 	through
each shielding barrier 	(ceiling, 	walls and floor), 	as
well as details of the door(s) 	and maze.
D. 	The structural composition and thickness 	or concrete
equivalent of all walls, 	doors, 	partitions, 	floor, 	and
ceiling of the room(s) 	concerned.
E. 	The type of occupancy of all adjacent areas 	inclusive
of space above and below the room(s) 	concerned. 	If
there is an exterior wall, 	show distance to the
closest area(s) 	where it is likely that individuals
may be present.
F. 	Description of all assumptions 	that were in shielding
calculations including, 	but not limited to, 	design
energy 	(ie., 	room may be designed for six MV unit
although only a 	four MV unit is 	currently proposed,
work-load, 	presence of integral beam-stop in unit,
occupancy and use(s) 	of adjacent areas, 	fraction of
time that useful beam will intercept each permanent
barrier 	(walls, 	floor and ceiling) 	and "allowed"
radiation exposure in both restricted and unrestricted
areas.
G. 	At least one example calculation which shows the
methodology used to determine the amount of shielding
required for each physical condition 	(ie., 	primary and
secondary/leakage barriers, 	restricted and
14-A-2
2265
Complete printed PDF page 228
unrestricted areas, 	small angle scatter, 	entry door(s)
and maze) 	and shielding material in the facility.
(1) 	If commercial software is used to generate
shielding requirements, 	please also identify the
software used and the version/ 	revision date.
(2) 	If the software used to generate shielding
requirements is not in the open literature,
please also submit quality control sample
calculations to verify the result obtained with
the software.
IV. 	NEUTRON SHIELDING
In addition to the requirements listed in Section III above,
therapeutic radiation machine facilities which are capable
of operating above ten MV shall submit shielding plans which
contain, 	as a minimum, 	the following additional information:
A. 	The structural composition, 	thickness, 	minimum density
and location of all neutron shielding material.
B. 	Description of all assumptions 	that were used in
neutron shielding calculations including, 	but not
limited to, 	neutron spectra as 	a 	function of energy,
neutron fluence rate, 	absorbed dose and dose
equivalent 	(due to neutrons) 	in both restricted and
unrestricted areas.
C. 	At least one example calculation which shows the
methodology used to determine the amount of neutron
shielding required for each physical condition 	(ie.,
restricted and unrestricted areas, 	entry door(s) 	and
maze) 	and neutron shielding material utilized in the
facility.
(1) 	If commercial software is used to generate
shielding requirements, 	please also identify the
software used and the version/ 	revision date.
(2) 	If the software used to generate shielding
requirements is not in the open literature,
please also submit quality control sample
calculations to verify the 	result obtained with
the software.
D. 	The method(s) 	and instrumentation which will be used
to verify the adequacy of all neutron shielding
installed in the facility.
V. 	REFERENCES
A. 	NCRP Report 49, 	"Structural Shielding Design and
Evaluation for Medical Use of X Rays and Gamma Rays of
Energies Up to 10 MeV" 	(1976).
B. 	NCRP Report 51, 	"Radiation Protection Design
Guidelines 	for 0.1-100 MeV Particle Accelerator
Facilities" 	(1977).
C. 	NCRP Report 79, 	"Neutron Contamination from Medical
Electron Accelerators" 	(1984).
14-A-3
2265
Complete printed PDF page 229