Ex parte Baxter International, Inc.

Ex parte Baxter International, Inc., 2009-006493 (March 18, 2010) [Considering expert testimony]Ptab Appeals

Gesamter Gesetzestext

UNITED STATES PATENT AND TRADEMARK OFFICE
____________
BEFORE THE BOARD OF PATENT APPEALS
AND INTERFERENCES
____________
Ex parte BAXTER INTERNATIONAL, INC.,
Patent Owner and Appellant
____________
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,4341
Technology Center 3900
____________
Decided: March 18, 2010
____________
Before MICHAEL R. FLEMING, Chief Administrative Patent Judge,
ALLEN R. MACDONALD, Vice Chief Administrative Patent Judge,
CAROL A. SPIEGEL, ROMULO H. DELMENDO, and
JEFFREY T. SMITH, Administrative Patent Judges.
DELMENDO, Administrative Patent Judge.
DECISION ON APPEAL
1 The patent under reexamination (hereinafter the “‘434 Patent”) issued to
Bruce A. Peterson, Michael E. Hogard, Harley D. Johnson, Thomas D.
Kelly, Jean M. Long, and William G. Preston, Jr. on September 21, 1993
from Application 07/688,174 filed on April 19, 1991.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
2
Baxter International, Inc., the owner of the patent under
reexamination, appeals under 35 U.S.C. §§ 134(b) and 306 from a final
rejection of claims 12-19 and 26-34 (Appeal Brief filed June 12, 2008,
hereinafter “App. Br.,” at 1; Final Office Action mailed December 14,
2007).2 We have jurisdiction under 35 U.S.C. §§ 134(b) and 306.
We AFFIRM.
STATEMENT OF THE CASE
This reexamination proceeding arose from a third-party request for ex
parte reexamination filed by John C. Phillips of Fish & Richardson, PC (San
Diego, California) on October 18, 2005. We have been informed that the
‘434 Patent is one of several patents involved in a patent infringement action
in which the United States Court of Appeals for the Federal Circuit recently
issued a decision on appeal of the District Court’s rulings. Fresenius USA,
Inc. v. Baxter Int’l, Inc., 582 F.3d 1288 (Fed. Cir. 2009). Additionally,
Appellant states that this appeal is related to an appeal in Reexamination
Control 90/007,730 involving United States Patent 6,284,131 B1 (‘131
Patent) (Appeal No. 2009-006498) (App. Br. 1).
Judges Spiegel, Delmendo, and Smith heard oral arguments on June
17, 2009, a written transcript of which was entered into the record on July
10, 2009. Subsequent to the oral arguments, the merits panel was expanded
2 See Patent Assignment Abstract of Title, Reel 006570, Frame 0001, which
was entered into the record of this proceeding as “Title Report” on October
20, 2005.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
3
to include MICHAEL R. FLEMING, Chief Administrative Patent Judge,
and ALLEN R. MACDONALD, Vice Chief Administrative Patent Judge.
The ‘434 Patent states that the invention relates to an improved kidney
dialysis machine including a touch screen user interface (col. 1, ll. 6-7; col.
8, ll. 19-20; appealed claims 12 and 26).
Claims 12, 26-28, and 30-34 on appeal read as follows:
12. In a method of operating a hemodialysis machine, an
improvement in a user/machine interfacing process, the
improvement comprising:
(a) providing a touch screen operably coupled to the
machine, the touch screen adapted to display an indicium
soliciting a user of the machine to touch a region on the screen
so as to select a machine-operation parameter;
(b) touching the region to select the machine operation
parameter;
(c) in response to said touching, invoking on the screen a
data-entry display associated with the machine-operation
parameter, the data entry display including a display of a
permissibly settable range of the machine-operation parameter,
so as to allow the user to select and enter a parametric value
associated with the machine-operation parameter within said
range; and
(d) touching the data-entry display to select and enter the
parametric value so as to cause the machine to operate in
conformance with the selected machine-operation parameter
and the entered parametric value.
26. A hemodialysis machine comprising:
(a) means for controlling a dialysate parameter selected
from a group consisting of dialysate temperature and dialysate
concentration, and means for delivering the dialysate to a
dialysate compartment of a hemodialyzer; and
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
4
(b) a user/machine interface operably coupled to said
dialysate-delivery means, the user/machine interface
comprising a touch screen adapted to display an indicium
corresponding to a parameter pertinent to operation of the
hemodialysis machine for performing hemodialysis and to
permit the user, by touching the indicium, to cause a change in
the parameter.
27. The hemodialysis machine according to claim 26
wherein the touch screen further provides an indicium
soliciting, from the user, programmed settings of a time-varying
machine-operating parameter.
28. The hemodialysis machine according to claim 26
further comprising means for delivering blood at a prescribed
flow rate from a patient, through a blood compartment of a
hemodialyzer, and back to the patient, wherein the touch screen
further provides a visual display of a parameter associated with
said means for delivering blood.
30. The hemodialysis machine according to claim 26
further comprising means for delivering an anticoagulant to a
patient. wherein the touch screen further provides an indicium
soliciting input from the user corresponding to a rate of
anticoagulant delivery.
31. The hemodialysis machine according to claim 26
wherein the touch screen further provides an indicium
soliciting, from the user, a programmed setting of an alarm
limit about the machine operating parameter.
32. In a method of operating a hemodialysis machine, an
improvement in a man/machine interfacing process used in
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
5
changing and setting a treatment parameter, the improvement
comprising:
(a) providing a touch screen interface with a parameter-
select indicium thereon;
(b) touching the parameter-select indicium;
(c) in response to said touching, invoking on the touch
screen a value-select indicium and a value-enter indicium;
(d) touching the value-select indicium to select a
parametric value corresponding to the treatment parameter;
(e) touching the value-enter indicium to enter the
parametric value; and
(f) in response to steps (b)-(e), comparing the entered
parametric value against present limits corresponding to a
permissible operational range for said parametric value and, if
the entered parametric value is within said limits, causing the
treatment parameter to operate in accordance with the entered
parametric value.
33. In a method of operating a hemodialysis machine, an
improvement in a man/machine interfacing process used in
changing and setting a treatment parameter, the improvement
comprising:
(a) providing a touch screen interface with a parameter-
select indicium thereon;
(b) touching the parameter-select indicium;
(c) in response to said touching, invoking on the touch
screen a value-select indicium and a value-enter indicium;
(d) touching the value-select indicium to select a
parametric value corresponding to the treatment parameter;
(e) touching the value-enter indicium to enter the
parametric value; and
(f) in response to steps (b)-(e), invoking an alarm limit to
be set at a preset increment relative to the entered parametric
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
6
value so as to enable an alarm signal to be triggered whenever,
during subsequent operation of the hemodialysis machine, the
parametric value passes outside the alarm limit.
34. In a method of operating a hemodialysis machine, an
improvement in a user/machine interfacing process, the
improvement comprising:
(a) providing a touch screen operably coupled to the
machine, the touch screen adapted to display an indicium
soliciting a user of the machine to touch a region on the screen
so as to cause the screen to display a machine-operation
parameter;
(b) touching the region on the screen;
(c) in response to said touching, invoking on the screen
an analog display associated with the machine-operation
parameter, the analog display including a display of a
previously set value of the machine-operation parameter and
allowing the suer [sic, user] to select and enter a desired value
associated with the machine-operation parameter; and
(d) touching the data-entry display to select and enter the
desired value of the machine-operation parameter so as to cause
the machine to operate in conformance with the desired value.
(Claims App’x, App. Br. 95, 98-101.)
The Examiner relied upon the following as evidence of
unpatentability (Examiner’s Answer mailed September 26, 2008, hereinafter
“Ans.,” 4-22):
Lichtenstein 	4,370,983 	Feb. 1, 1983
Thompson 	4,710,166 	Dec. 1, 1987
Kerns 	4,756,706 	July 12, 1988
Rubalcaba 	4,898,578 	Feb. 6, 1990
CMS 08 – HANDBOOK F330311-F330379 (Fresenius AG 1988) (hereinafter
“CMS 08 Handbook”).
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
7
S ARNS
® 9000 P ERFUSION S YSTEM : O PERATORS MANUAL F298963-
F299092 (3M 1989) (hereinafter “Sarns 9000 Manual”).
In rebuttal to the Examiner’s obviousness conclusion as to the
combination of a touch screen user interface and a hemodialysis machine,
Appellant relies on the following evidence appended to the Appeal Brief
(Evidence App’x):
Ex. A: Rebuttal Expert Report of J. Dennis Bruner, ¶ 40.
Ex. B: Rebuttal Expert Report of Lee W. Henderson, ¶ 97.
Ex. C: Dep. Tr. of Ben J. Lipps, 38-39, 65-66, 80, 190.
Ex. D: Dep. Tr. of Martin Joseph Crnkovich, 10-11, 13-14.
Ex. E: Dep. Tr. of Thomas D. Kelly, 63-66.
Ex. F: Decl. of John Turner, ¶¶ 16, 18, 21-23.
Ex. G: Touch the Future, DrakeWillock, cover page only (date
unknown).
Ex. H: Drake Willock System 1000: The Future Generation (partially
illegible), The Althin Group (date unknown).
Ex. I: Tina – your new co-worker, Althin (date unknown).
Ex. J: Martin Crnkovich, June-July Hemodialysis R&D (publication
status or date unknown).
Ex. K: Christian Schlaeper and Michael Fasanella, 2008K Product
Launch, 3, 22 (2000).
Ex. L: The Highest Standards in Hemodialysis, Fresenius USA
(1998).
Ex. M: Step Up to the 2008K, Fresenius Medical Care (publication
date unknown).
Ex. P: Design and Development Plan: Phase One – Hemodialysis
Equipment (publication status and date unknown).
Ex. Q: Dep. Tr. of Martin J. Crnkovich, 29.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
8
The Examiner rejected the claims under 35 U.S.C. § 103(a) as
follows:
I. 	Claims 26-29 and 31 as unpatentable over the combined
teachings of CMS 08 Handbook, Lichtenstein, Sarns 9000
Manual, Rubalcaba, and Kerns (Ans. 4-12);
II. 	Claim 30 as unpatentable over the combined teachings of CMS
08 Handbook, Lichtenstein, Sarns 9000 Manual, Rubalcaba,
Kerns, and Thompson (Ans. 12-14);
III. 	Claims 12-18 as unpatentable over the combined teachings of
CMS 08 Handbook, Lichtenstein, Sarns 9000 Manual,
Rubalcaba, and Kerns (Ans. 14-16);
IV. 	Claim 19 as unpatentable over the combined teachings of CMS
08 Handbook, Lichtenstein, Sarns 9000 Manual, Rubalcaba,
and Kerns (Ans. 16-17);
V. 	Claim 32 as unpatentable over the combined teachings of CMS
08 Handbook, Lichtenstein, Sarns 9000 Manual, Rubalcaba,
and Kerns (Ans. 17-19);
VI. 	Claim 33 as unpatentable over the combined teachings of CMS
08 Handbook, Lichtenstein, Sarns 9000 Manual, Rubalcaba,
and Kerns (Ans. 19-20); and
VII. 	Claim 34 as unpatentable over the combined teachings of CMS
08 Handbook, Lichtenstein, Sarns 9000 Manual, Rubalcaba,
and Kerns (Ans. 20-22).
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
9
ISSUES
Obviousness of Combining a Touch Screen Interface
with a Hemodialysis Machine
The Examiner found that the CMS 08 Handbook and Lichtenstein
independently describe a computer-controlled hemodialysis machine that
differs from the claimed machine only in terms of the touch screen
user/machine interface limitation (Ans. 4-7; claim 26). To resolve this
difference, the Examiner relied on the Sarns 9000 Manual, Rubalcaba, and
Kerns, all of which disclose the use of a touch screen interface to control
critical operational variables, including fluids delivered to a patient, in a
medical device (Ans. 8-9). Relying on reasoning based on the same core
prior art references against similar claims in Ex parte Connell, Appeal No.
2003-02353 (BPAI Aug. 28, 2003), the Examiner concluded that the
collective teachings of the prior art references would have prompted one of
ordinary skill in the art to incorporate a touch screen user interface into a
conventional hemodialysis machine (Ans. 10-12, 14-16, and 22-23).
Furthermore, the Examiner found Appellant’s rebuttal arguments and
evidence unpersuasive (Ans. 22-61).
3 Ex parte Connell was an appeal from the rejected claims of application
09/711,240. Application 09/711,240, filed November 13, 2000 and now
abandoned, is a continuation of application 09/067,922, filed April 28, 1998
and now abandoned, which is a continuation of application 08/479,688, filed
June 7, 1995 and now U.S. Patent 5,744,027, which is a divisional of
application 08/122,047, filed September 14, 1993 and now U.S. Patent
5,486,286, which is a divisional of application 07/688,174, filed April 14,
1991 and now U.S. Patent 5,247,434, which is the patent under
reexamination in this appeal.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
10
Appellant contends that the Sarns 9000 Manual, Rubalcaba, and
Kerns do not teach a centralized control (App. Br. 16-17). Appellant also
asserts that the testimonies of certain witnesses demonstrate that the use of a
touch screen interface in a hemodialysis machine would have been
confronted with skepticism and is therefore unobvious (App. Br. 28-31).
Appellant further urges that it has produced evidence of secondary
considerations of nonobviousness including commercial success and
copying (App. Br. 73-90).
Thus, the principal dispositive issues are:
(1) Has the Examiner erred in concluding that a person of ordinary
skill in the art would have been prompted to couple a touch screen
user/machine interface with a hemodialysis machine in view of the
collective teachings of the prior art references?
(2) Has the Examiner erred in finding that the relied-upon evidence is
insufficient to confer nonobviousness?
Claim 26
The Examiner found that the teachings of CMS 08 Handbook and/or
Lichtenstein describe, or suggest, to one of ordinary skill in the art, element
(a) of claim 26 (“means for controlling a dialysate parameter selected from a
group consisting of dialysate temperature and dialysate concentration, and
means for delivering the dialysate to a dialysate compartment of a
hemodialyzer”) (Ans. 5-6, 25-27, 38-39). First, the Examiner found that the
CMS 08 Handbook “provides all necessary equipment for hemodialysis
including various motorized pumps (Na, K, HCO 3, and UFR), pump lines,
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
11
dialyzer and computer control (microprocessor) devices for the delivery,
setting and control of dialysate concentration and ultrafiltration rate . . . ”
(Ans. 38). Second, the Examiner found that Lichtenstein, which discloses
manual control of various flow rates, teaches temperature sensors, heaters,
coolers, and microprocessor equipment, as well as motorized pumps, pump
lines, and microprocessor control (Ans. 6, 39). Together with additional
findings concerning the use of a touch screen user interface in other similar
medical devices, the Examiner concluded that “it would have been prima
facie obvious to one of ordinary skill in the art . . . to have modified the
CMS08/A2008 dialysis machine or the dialysis machine of Lichtenstein, to
utilize a touch screen . . . to control the central processing unit
microcomputer and machine (including various pumps, lines etc.) for
delivery of treated or treatment fluids to a patient” (Ans. 8-10).
Appellant contends that the means for controlling a dialysate
temperature and means for controlling a dialysate concentration, both as
specified in claim 26, require: a microprocessor, a heater, and a temperature-
sensing device; and a microprocessor and concentrate pump, respectively
(App. Br. 37). According to Appellant, the Examiner “no where identifies
corresponding structures in the CMS08 Handbook that are necessary to
meet” these limitations of element (a) (id.). With respect to Lichtenstein,
Appellant argues that the reference does not suggest any particular input
device for operational parameters and that, unlike the claimed invention,
“control is preferably automatically controlled by the microcomputer” (App.
Br. 41).
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
12
Thus, the issues are:
(3) Has the Examiner erred in finding that the CMS 08 Handbook or
Lichtenstein describes element (a) of claim 26?
(4) Has the Examiner erred in finding that Lichtenstein teaches user-
initiated control of operational parameters?
Claim 27
The Examiner found that the sodium, potassium, and bicarbonate
concentrations and the ultrafiltration rate described in the CMS 08
Handbook are “programmed settings of time-varying machine-operating
parameter[s],” as recited in claim 27 (Ans. 44-45). According to the
Examiner, “the CMS08 [Handbook] explicitly exemplifies the same time
variable parameters during hemodialysis, see particular settings for sodium,
potassium and bicarbonate concentration[s] and ultrafiltration rate” (Ans.
45).
Appellant, on the other hand, contends that the CMS 08 Handbook
does not describe the claim limitation because the prior art device “does not
actively ‘solicit’ information related to the setting of a time-varying
hemodialysis parameter by requesting an input” (App. Br. 49-50).
Thus, the issue is:
(5) Has the Examiner erred in finding that the CMS 08 Handbook
describes a “time-varying machine-operating parameter,” as recited in claim
27?
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
13
Claims 28 and 29
The Examiner found that the CMS 08 Handbook describes a
microcomputer control that operates in concert with components (e.g., blood
pumps) of the A2008 hemodialysis machine, as required by the means-plus-
function language “means for delivering blood at a prescribed flow rate” of
claim 28 (Ans. 46). Additionally, the Examiner found that Lichtenstein also
describes the disputed claim limitation (Ans. 46-49).
Appellant asserts “that the blood pump in the A2008 [hemodialysis
machine] is initially set manually (CMS08 Handbook, F330337) and is not
controlled by the user interface on the CMS08 unit” (App. Br. 52). As to
Lichtenstein, Appellant argues that while Lichtenstein discloses
computerized control of blood flow, it does not suggest a visual display for
the automatic, computer-controlled operation of the function at a prescribed
flow rate (App. Br. 53).
Thus, the issue is:
(6) Has the Examiner erred in finding that either the CMS 08
Handbook or Lichtenstein describes a “means for delivering blood at a
prescribed flow rate,” as recited in claim 28?
Claim 30
The Examiner found that Thompson discloses the administration of a
secondary fluid such as an anticoagulant in administration systems operating
under the control of electronic instrumentation (Ans. 13). The Examiner
then concluded that one of ordinary skill in the art would have found it
obvious to deliver an anticoagulant to a patient undergoing hemodialysis
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
14
with the CMS 08 unit for the purpose of preventing coagulation or blood
clotting and to control the delivery of the anticoagulant with a touch screen
user interface (Ans. 14).
Appellant contends that “Thompson does not contain any disclosure
relevant to ‘an indicium soliciting from the user’ the anticoagulant delivery
rate, though the rate can apparently be set using the device” (App. Br. 55).
Thus, the issue is:
(7) Has the Examiner erred in reaching the conclusion that a person
having ordinary skill in the art would have found it obvious to control
anticoagulant delivery rate with a touch screen user interface, as required by
claim 30?
Claim 31
The Examiner found that the CMS 08 Handbook, Lichtenstein, and
the Sarns 9000 Manual teach the use of an alarm when critical parameters
(e.g., conductivity and blood ultrafiltration rate) fall outside preset limits and
concluded that a person having ordinary skill in the art would have found it
obvious to set these limits by using a touch screen user interface (Ans. 12,
50-51).
Appellant argues that neither the CMS 08 Handbook nor Lichtenstein
discloses a touch screen user interface and that the touch screen in the Sarns
9000 Manual does not solicit the user to enter an alarm limit (App. Br. 56-
57).
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
15
Thus, the issue is:
(8) Has the Examiner erred in reaching the conclusion that a person
having ordinary skill in the art would have found it obvious to set an alarm
in the CMS 08 hemodialysis machine of a critical parameter through a touch
screen user interface, as required by claim 31?
Claim 32
The Examiner found that Rubalcaba teaches the concept of comparing
values entered into a touch screen user interface with preset limits and if the
values are within the preset limits, causing the parameter to be entered, and
the machine to be operated in accordance with the set parameter (Ans. 18-
19).
Appellant contends that Rubalcaba does not disclose element (f) of
claim 32 because, unlike the claimed invention, Rubalcaba discloses
automatic calculation of a fourth variable after three other variables are
entered by the user (App. Br. 58).
Thus, the issue is:
(9) Has the Examiner erred in finding that Rubalcaba discloses
element (f) of claim 32 because claim 32 excludes Rubalcaba’s automatic
calculation?
Claim 33
The Examiner found that the Sarns 9000 Manual teaches the concept
of monitoring a parameter and sounding an alarm and stopping operation
when the parameter falls outside preset limits and concluded that it would
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
16
have been obvious for one of ordinary skill in the art to apply this concept to
the hemodialysis machine described in the CMS 08 Handbook (Ans. 19-20).
Appellant argues that the prior art does not teach element (f) of claim
33, including the “preset increment” limitation (App. Br. 59-60).
Thus, the issue is:
(10) Has the Examiner erred in finding that the Sarns 9000 Manual
teaches the concept of monitoring a parameter and sounding an alarm and
stopping operation when the parameter falls outside preset limits?
Claim 34
The Examiner found that Rubalcaba teaches the concept of the use of
“a memory function whereby the values may be stored for later recall, and
the entered values may be displayed” (Ans. 21). According to the Examiner,
Rubalcaba teaches that when the user returns to the titration calculator
screen, the previously entered values are displayed (id.).
Appellant contends that Rubalcaba “is indefinite as to whether the
touch screen automatically displays a previously set value or if the user must
retrieve the previously set value from the memory storage” (App. Br. 61).
Thus, the issue is:
(11) Has the Examiner erred in finding that Rubalcaba discloses the
automatic display of a previously entered value for user convenience, as
required by element (c) of claim 34?
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
17
Claims 12-18
The Examiner concluded that the collective teachings of the CMS 08
Handbook, Lichtenstein, the Sarns 9000 Manual, Rubalcaba, and Kerns
would have led a person having ordinary skill in the art to arrive at a method
of operating a hemodialysis machine, as recited in claim 12 (Ans. 14-16).
Appellant contends that the Examiner’s conclusion is in error because:
(i) the Sarns 9000 Manual does not teach element (c) of claim 12 (App. Br.
64); (ii) “the cited reference fails to disclose a touch screen that centralizes
controls in a single locations [sic]” (id. at 68); (iii) “[n]one of the references
teach or suggest an apparatus that will minimize the display of the data-entry
mechanism” (id. at 70); and (iv) Rubalcaba and Kerns teach a numeric
keypad that is not “invoked” by touching the desired parameter (id. at 73).
Thus, the issues are:
(12) Has the Examiner erred in finding that the prior art teaches
element (c) of claim 12?
(13) Has the Examiner erred in reaching the obviousness conclusion
as to claim 12 because the claim requires a touch screen that centralizes
controls in a single location and the prior art fails to disclose such a
limitation?
(14) Has the Examiner erred in reaching the obviousness conclusion
as to claim 12 because the claim requires the minimization of the display of
a data-entry mechanism and the prior art fails to disclose such a limitation?
(15) Has the Examiner erred in reaching the obviousness conclusion
because Rubalcaba and Kerns do not teach a numeric keypad that is
“invoked” by touching a desired parameter?
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
18
FINDINGS OF FACT (“FF”)
1. 	The CMS 08 Handbook describes an ancillary section of an
A2008 dialysis treatment machine (F330324).
2. 	The CMS 08 Handbook states that the volumetric dosing of up
to three (3) different electrolyte concentrations via stepper
motor driven line pumps into the A2008 dialysate mixing
system can be programmed (using a control unit) in ten minute
time intervals (F330320, F330324, F330327, F330328).
3. 	In particular, the CMS 08 Handbook identifies the ultrafiltration
rate, the Na concentration of the dialysate, the K concentration
of the dialysate, and the bicarbonate concentration of the
dialysate as “time depend[e]nt parameters” and that these
parameters may be programmed by using a keyboard
(F330324).
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
19
4. 	A figure in section 2.4 (F330328) of the CMS 08 Handbook is
reproduced below:
The figure reproduced above is said to be a block diagram of
the CMS 08 and A2008 combination.
5. 	Lichtenstein discloses a computer-controlled medical care
system (e.g., hemodialysis) in which “[t]he adjustment of the
various flow rates can be effected either manually by an
operator or automatically where control means are provided”
(col. 5, l. 34 to col. 6, l. 19; col. 6, ll. 54-56).
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
20
6. 	Lichtenstein teaches that the microcomputer can be any
conventional computer suitable for the disclosed purposes and
may include “a display means for indicating the sensed
parameters and/or any needed adjustments in the operating
conditions to maintain the procedure on schedule while keeping
the physiological status of the patient within predetermined
limits” (col. 8, ll. 1-34).
7. 	The ‘434 Patent Specification states:
Exemplary are parameters that are programmed to
change over time (so-called profiled parameters).
In this class are the sodium concentration of the
dialysate solution, the bicarbonate concentration of
the dialysate, kT/V, and the ultrafiltration rate.
(Col. 9, ll. 32-36.)
8. 	Thompson teaches that a secondary fluid, such as an
anticoagulant, may be administered to a patient in
administration systems operated under the control of electronic
instrumentation either on a gravity delivery or a positive
pressure basis (col. 1, ll. 9-25).
9. 	The Sarns 9000 Manual teaches the concept of using a touch
screen interface to control various operational variables in a
perfusion system (F298966-67, F298971-72, F298978,
F298987-88, F299010-11, F299014, F299054-59).
10. 	The Sarns 9000 Manual further teaches the concept of using
alarms to alert the operator of a perfusion system when certain
monitored parameters (e.g., pressure for the arterial and
cardioplegia pumps) fall outside preset limits and to stop the
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
21
pumps or display messages if pressures, temperatures, flow
rates, or other alarm parameters exceed preset limits (F298987-
88).
11. 	Rubalcaba teaches the use of a touch screen interface to control
operational variables in a drug delivery system (Abstract;
Figures 3 and 5-9; col. 1, ll. 20-38; col. 8, l. 52 to col. 12, l. 12).
12. 	Rubalcaba further teaches the concept of comparing an entered
value of an operating parameter in a drug infusion system to
preset values and, if outside the preset range, displaying the
entered value along with an indication that the entered value is
invalid (col. 6, ll. 1-24).
13. 	Rubalcaba also teaches (col. 6, ll. 25-32):
According to another aspect of the invention, there
is provided means for storing the values that are
assigned and calculated for the parameters. In the
illustrated embodiment, the CPU circuitry 32
functions to serve this purpose. It includes suitable
programming and sufficient memory to
accomplish storage function so that the various
values of the parameters need not be reentered.
14. 	Rubalcaba teaches that the system calculates “dosage, infusion
rate, weight of drug, or volume of solution from the remaining
ones of these parameters using one of two formulae” (col. 4, ll.
34-57).
15. 	The Sarns 9000 Manual discloses an indicium for a pulsatile
screen that may be accessed through a “MAIN SCREEN”
(F299053-54).
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
22
16. 	The Sarns 9000 Manual teaches that the pulsatile screen
includes a “FLOW display” showing the average flow of an
arterial pump and permitting an increase or decrease by
touching up or down arrows (F299055).
17. 	Sarns 9000 teaches: “Touch the display area, use the
increase/decrease switches to adjust the value, and touch the
display area again” (F299055).
18. 	Kerns teaches the use of a touch screen interface to control
operational parameters in an infusion pump system (Figures 7,
8, 11-18; col. 6, ll. 7-20; col. 11, ll. 39-51).
19. 	The Rau publication, discussed in Fresenius, 582 F.3d at 1301,
was found by our reviewing court, the United States Court of
Appeals of the Federal Circuit, to teach the use of a touch
screen interface on a medical device such as an anesthesia
delivery system.
20. 	The Rau publication was also found by the court to suggest the
use of such a touch screen interface in a hemodialysis machine.
Fresenius, 582 F.3d at 1301.
21. 	Dr. Rau, the author of the Rau publication, testified in court
“that his publication ‘gave numerous examples of where you
can use [touch screen technology] beneficially’ and stated that
‘one of the examples which [he] mentioned was hemodialysis
and a hemodialysis machine.” Fresenius, 582 F.3d at 1301.
22. 	Expert witness Dr. Robert C. Phares also testified in court as to
“the ease and prevalence of ‘integrating a touch screen into
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
23
some kind of computer-controlled machine’ and opined that as
of the relevant filing date, it would not have been difficult for
one to [of ordinary skill in the art] to integrate a touch screen
interface into a hemodialysis machine.” Fresenius, 582 F.3d at
1301.
23. 	Lipps testified that 20 years prior to the date of his testimony
(September 26, 2005), touch screens were “highly unreliable”
and not economical (Ex. C at 38).
24. 	Crnkovich testified that he “used a couple of touch screens
during the early ‘80s, and . . . did not have a good experience
with them” (Ex. D, 10:19-21).
25. 	Kelly testified that, in 1988, he had heard from others that
touch screen interfaces posed cost and reliability concerns (Ex.
E, 63-66).
26. 	Turner declares that in proposing the release of “2008K”
hemodialysis machine, Fresenius listed the touch screen as the
first among the “state-of-the art features” (Ex. F, ¶ 18).
27. 	Turner further testified that Fresenius sought to compete with
Althin, another manufacturer of units with a touch screen
interface (Ex. F, ¶¶ 21-23).
PRINCIPLES OF LAW
On appeal to this Board,
An appellant may attempt to overcome an examiner’s
obviousness rejection on appeal to the Board by submitting
arguments and/or evidence to show that the Examiner made an
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
24
error in either (1) an underlying finding of fact upon which the
final conclusion of obviousness was based, or (2) the reasoning
used to reach the legal conclusion of obviousness. Similarly, the
applicant may submit evidence of secondary considerations of
non-obviousness. See Kahn, 441 F.3d at 985-86 (“On appeal to
the Board, an applicant can overcome a rejection by showing
insufficient evidence of prima facie obviousness or by rebutting
the prima facie case with evidence of secondary indicia of
nonobviousness.”) (quoting In re Rouffet, 149 F.3d 1350, 1355
(Fed. Cir. 1998), overruled in part on other grounds, KSR, 550
U.S. at 422).
Ex parte Frye, ___ USPQ2d ___ (BPAI March 1, 2010) (precedential)
http://des.uspto.gov/Foia/ReterivePdf?system=BPAI&flNm=fd2009006013-
02-26-2010-1. See also 37 C.F.R. § 41.37(c)(1)(vii).
“During reexamination, as with original examination, the PTO must
give claims their broadest reasonable construction consistent with the
specification . . . . Therefore, we look to the specification to see if it provides
a definition for claim terms, but otherwise apply a broad interpretation.” In
re ICON Health and Fitness, Inc., 496 F.3d 1374, 1379 (Fed. Cir. 2007).
This longstanding principle is based on the notion that “during patent
prosecution when claims can be amended, ambiguities should be recognized,
scope and breadth of language explored, and clarification imposed.” In re
Zletz, 893 F.2d 319, 321 (Fed. Cir. 1989). That is, a patent applicant has the
opportunity and responsibility to remove any ambiguity in claim term
meaning by amending the application. “Only in this way can uncertainties
of claim scope be removed, as much as possible, during the administrative
process.” Id. at 322. “[A]s applicants may amend claims to narrow their
scope, a broad construction during prosecution creates no unfairness to the
applicant or patentee.” ICON Health, 496 F.3d at 1379.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
25
The Supreme Court of the Unites States explained that “[w]hen a
work is available in one field of endeavor, design incentives and other
market forces can prompt variations of it, either in the same field or a
different one.” KSR Int’l Co. v. Teleflex, Inc., 550 U.S. 398, 417 (2007).
According to the Court, “if a technique has been used to improve one device,
and a person of ordinary skill in the art would recognize that it would
improve similar devices in the same way, using the technique is obvious
unless its actual application is beyond his or her skill.” Id.
ANALYSIS
Concurrent Court Proceedings
In Fresenius, our reviewing court reversed the district court’s decision
to overturn a jury’s verdict of obviousness as to claims 1-3 and 13-16 of the
‘131 Patent and held that substantial evidence supported a conclusion that
the subject matter “directed to a hemodialysis machine integrated with a
touch screen user interface” would have been obvious to a person having
ordinary skill in the art. Fresenius, 582 F.3d at 1300-1302. As to secondary
considerations (e.g., commercial success of an embodiment of the patents in
suit), our reviewing court stated that substantial evidence supported the
jury’s finding that Appellant’s evidence was insufficient. Fresenius, 582
F.3d at 1302.
The court found, however, that Fresenius (the alleged infringer) failed
to prove invalidity by clear and convincing evidence with respect to the
asserted ‘434 Patent claims (i.e., claims 26-31). Fresenius, 582 F.3d at
1299. Specifically, the court stated that “Fresenius neither identified the
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
26
structure in the specification that corresponds to the means for delivering
dialysate nor compared it to the structures present in the prior art.” Id. at
1299. In addition, the court held that “Fresenius failed to present any
evidence that the required stepper motor structure [in claim 30] existed in
the prior art.” Id. at 1300.
Although claims 26-31 were not proven invalid in court, a lower
standard of proof and the broadest reasonable interpretation standard of
claim construction apply at the PTO and therefore the agency is not bound
by the court’s determination. Ethicon v. Quigg, 849 F.2d 1422, 1429 (Fed.
Cir. 1988) (“[T]he Board and the courts take different approaches in
determining invalidity and on the same evidence could quite correctly come
to different conclusions.”); In re Swanson, 540 F.3d 1368, 1377 (Fed. Cir.
2008) (explaining the different standards of proof and claim construction
used in reexamination relative to civil litigation in district court).
Obviousness of Combining a Touch Screen Interface
with a Hemodialysis Machine
We share the Examiner’s well-stated position that the collective
teachings of the applied prior art references would have prompted a person
of ordinary skill in the art to couple a touch screen interface with a
hemodialysis machine. Both CMS 08 Handbook and Lichtenstein identify a
need to control certain parameters pertinent to the operation of a
hemodialysis machine (FF 1-6). The Sarns 9000 Manual, Kerns, and
Rubalcaba all disclose the concept of controlling critical operational
parameters in medical devices, which are similar to those disclosed in either
the CMS 08 Handbook or Lichstenstein in terms of requiring the precise
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
27
delivery of fluids into a human body, by a touch screen user/machine
interface (FF 9-18). The interrelatedness of the prior art teachings, coupled
with the expectation that a technique used to improve one device would also
improve other similar devices, amply support the Examiner’s obviousness
conclusion. KSR, 550 U.S. at 417-18.
While Appellant urges that the Sarns 9000 Manual, Kerns, and
Rubalcaba do not teach centralized control, we have not been directed to any
language in the appealed claims that require centralized control. In re Self,
671 F.2d 1344, 1348 (CCPA 1982) (“Many of appellant's arguments fail
from the outset because, as the solicitor has pointed out, they are not based
on limitations appearing in the claims.”).
We also find no error in the Examiner’s refusal to credit the
testimonies of certain experts as supporting Appellant’s skepticism or lack
of a reasonable expectation of success argument. As pointed out by the
Examiner (Ans. 35-36), these experts testified on the knowledge of a person
of ordinary skill in the art at a time period significantly before the relevant
filing date (FF 23-25). We have not been directed to any evidence
indicating that further advances in touch screen technology did not occur up
to the relevant filing date. Moreover, we have not been directed to any
evidence that these witnesses, before giving their testimonies, considered the
teachings of the applied prior art references, which plainly disclose the
successful incorporation of touch screen interfaces into critical medical
systems. Indeed, Appellant has admitted that “not a single witness from
either Baxter or Fresenius had even heard of the Sarns machine” (App. Br.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
28
26). That admission reveals the lack of any substantial value in the relied-
upon evidence vis-a-vis Appellant’s argument.
Appellant relies primarily on the Turner Report as demonstrating
commercial success or copying (FF 26-27). We, like the Examiner (Ans.
54-57), find the relied-upon evidence, including the Turner Report,
unpersuasive. The evidence fails to conclusively establish that commercial
success, if any, was directly attributable to the touch screen interface. In re
Huang, 100 F.3d 135, 140 (Fed. Cir. 1996) (“Even assuming that Huang had
sufficiently demonstrated commercial success, that success is relevant in the
obviousness context only if there is proof that the sales were a direct result
of the unique characteristics of the claimed invention – as opposed to other
economic and commercial factors unrelated to the quality of the patented
subject matter.”); Asyst Technologies., Inc. v. Emtrak, Inc., 544 F.3d 1310,
1316 (Fed. Cir. 2008) (“While the evidence shows that the overall system
drew praise as a solution to a felt need, there was no evidence that the
success of the commercial embodiment . . . was attributable to the . . . only
material difference between [the prior art] and the patented invention.”).
The obviousness of combining a touch screen interface with a
hemodialysis machine is buttressed by Fresenius, which upheld a jury
determination that the incorporation of a touch screen in a hemodialysis
machine would have been obvious to a person of ordinary skill in the art
over similar prior art. Here, Appellant has urged in the present appeal that
we should defer to the district court’s ruling because “[a]part from the
Thompson reference, the evidence here is the same as that examined by the
District Court” (App. Br. 11; Reply Br. 8; FF 19-22). Because Appellant has
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
29
repeatedly relied on the district court proceeding in support of
nonobviousness, we discern no unfairness in subjecting Appellant to any
adverse consequence that may arise from such reliance in the current
reexamination, which is conducted under the preponderance of the evidence
standard – a standard that is substantially lower than the clear and
convincing standard used in district court.
Claim 26
It is undisputed that the structure corresponding to the recited “means
for controlling . . . dialysate concentration” includes, at a minimum, a
microprocessor and a concentrate pump (App. Br. 37; Ans. 5-6, 25-27, 38-
39). Contrary to Appellant’s belief, the Examiner did identify the structures
in the CMS 08 Handbook that correspond to the recited means-plus-function
language (Ans. 5-6, 25-27, 38-39). Indeed, the CMS 08 Handbook describes
the control of Na, K, and HCO3 flow rates using a control unit that may be
programmed in ten minute intervals (i.e., a microprocessor) and a pump
driven by a stepper motor (FF 1-4). Thus, with respect to the CMS 08
Handbook, Appellant failed to demonstrate any reversible error on the part
of the Examiner. Additionally, Appellant did not assert that Lichtenstein
lacks any structure corresponding to element (a) of claim 26 (App. Br. 40-
43). Thus, we cannot reverse the Examiner’s rejection of claim 26 on the
basis of the recited means-plus-function language in claim 26, element (a).
We are not persuaded by Appellant’s contention that Lichtenstein
lacks any teaching with respect to any particular input device for operational
parameters, i.e., user-initiated control of an operational parameter (App. Br.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
30
40-43). As pointed out by the Examiner (Ans. 6), Lichtenstein plainly
teaches that the adjustment of the flow rates can be effected manually, which
would necessarily require user input (FF 5).
In light of these factual findings, we affirm the rejection of claim 26.
Claim 27
Appellant’s argument that the CMS 08 does not teach the setting of a
time-varying hemodialysis parameter is factually incorrect (App. Br. 49-50).
Like the ‘434 Patent Specification, the CMS 08 Handbook explicitly teaches
that Na concentration of the dialysate, K concentration of the dialysate,
bicarbonate concentration of the dialysate, and ultrafiltration rate are time-
dependent hemodialysis parameters that can be adjusted or controlled by the
user (FF 3 and 7).
Thus, we uphold the rejection of claim 27.
Claims 28 and 29
Appellant does not dispute that the CMS 08/A2008 combination
discloses a blood pump, which together with the microprocessor, constitutes
one of two components corresponding to the recited “means for delivering
blood at a prescribed flow rate” (App. Br. 52). Rather, Appellant’s position
appears to be that the control unit described in the CMS 08 Handbook does
not control the blood pump and therefore the rejection cannot stand (id.).
We cannot agree. The Examiner’s view is that a person having
ordinary skill in the art would have found it obvious to control the blood
pump with the CMS 08 microcomputer control because the reference
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
31
“makes clear that the computer control device is functional in concert with
the A2008 device” (Ans. 46; FF 2 and 3). This position is reasonable.
Appellant has not shown why the Examiner’s reasoning is flawed and
therefore we discern no reversible error.
Turning to Lichtenstein, Appellant’s argument that Lichtenstein does
not disclose “a visual display of a parameter associated with said means for
delivering blood” (claim 28) is also unpersuasive (App. Br. 53). As noted
by the Examiner (Ans. 6-7), Lichtenstein teaches the use of a microcomputer
including a display means (FF 6). Appellant failed to address this teaching
in any meaningful way and therefore we see no error in the Examiner’s
reasoning.
Claim 30
We are not persuaded by Appellant’s argument that Thompson does
not contain any disclosure relevant to “an indicium soliciting input from the
user corresponding to a rate of anticoagulant delivery rate” (claim 30; App.
Br. 55). Here, Appellant does not contest the Examiner’s finding that
Thompson teaches that an anticoagulant may be administered to a patient
undergoing medical treatment involving a system operated under the control
of electronic instrumentation (FF 8). Given that it was known that operating
parameters may be entered by soliciting input from a user (FF 2 and 3), we
detect no error in the Examiner’s reasoning based on the collective teachings
of the prior art references.
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
32
Claims 31 and 33
We agree with the Examiner that the Sarns 9000 Manual would have
provided a reason for one of ordinary skill in the art to include an
alarm/shutoff system for association with the CMS 08/A2008 operation
parameters in order to improve safety (FF 10). While Appellant argues that
the prior art does not solicit the user to enter an alarm limit (App. Br. 56-57),
the Sarns 9000 Manual describes the setting of alarm limits and therefore the
entry of such limits would have been obvious to a person having ordinary
skill in the art.
Claim 32
Again, we find ourselves in agreement with the Examiner (Ans. 18-
19). In this case, Rubalcaba plainly teaches that if the entered value is not
within preset limits, the value will be displayed along with an indication that
it is invalid, together with an indication of the valid range (FF 12). While
Appellant argues that Rubalcaba teaches automatic calculation of a fourth
variable after entry of three other variables (FF 14), that fact does not
demonstrate error because claim 32 does not exclude automatic calculation.
Thus, the invention recited in appealed claim 32 cannot be distinguished
over the prior art on this basis. In re Self, 671 F.2d at 1348.
Claim 34
Appellant contends that Rubalcaba is unclear “whether the touch
screen automatically displays a previously set value or if the user must
retrieve the previously set value from memory storage” (App. Br. 61). This
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
33
argument is unpersuasive. As found by the Examiner (Ans. 21, 52),
Rubalcaba teaches “storage function so that various values of the parameters
need not be reentered” (FF 13). That teaching would have led a person
having ordinary skill in the art to incorporate the memory function for each
parameter.
Claims 12-18
Appellant has not offered any additional argument in support of the
separate patentability of any claim (App. Br. 72-73). Accordingly, we
confine our discussion to claim 12. See 37 C.F.R. § 41.37(c)(1)(vii).
Regarding element (c) of claim 12, Appellant urges that the up and
down arrows in the Sarns 9000 Manual “are not invoked when the user seeks
to enter a value for the flow rate; they are already on the pulsatile screen . . .
.” (App. Br. 64). A similar argument is advanced against Rubalcaba and
Kerns (App. Br. 73).
Appellant’s arguments are unpersuasive because, as pointed out by the
Examiner (Ans. 52), it “fails to recognize that the user [of the Sarns 9000
Manual] must first enter the pulsatile screen from the Main Screen by
touching the Pulse indicium” (Id.; FF 15-17). Moreover, Appellant has not
explained how the difference in graphical indicia on a computer display, if
any, constitutes a nonobvious advance over the prior art.
We also find no merit in Appellant’s contention that “the cited
reference fails to disclose a touch screen that centralizes controls in a single
locations [sic]” or “minimize the display of the data-entry mechanism”
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
34
(App. Br. 68, 70). Appellant has failed to direct us to any language in claim
12 that recites such limitations. In re Self, 671 F.2d at 1348.
Moreover, with respect to centralized control, a person having
ordinary skill in the art is presumed to have some common sense and thus
would have found it obvious to consolidate the control into a single system
for ease and convenience. KSR, 550 U.S. at 421 (“A person of ordinary skill
is also a person of ordinary creativity, not an automaton.”); Perfect Web
Techs., Inc. v InfoUSA, Inc., 587 F.3d 1324, 1329 (Fed. Cir. 2009) (“KSR
expanded the sources of information for a properly flexible obviousness
inquiry to include . . . the background knowledge, creativity, and common
sense of the person of ordinary skill.”).
CONCLUSION
On this record, we conclude that there is no error in the Examiner’s
factual findings and conclusions that:
a person of ordinary skill in the art would have been prompted to
couple a touch screen user/machine interface with a hemodialysis machine
in view of the collective teachings of the prior art references;
Appellant’s relied-upon evidence is insufficient to confer
nonobviousness;
the CMS 08 Handbook or Lichtenstein describes element (a) of claim
26;
Lichtenstein teaches user-initiated control of operational parameters;
the CMS 08 Handbook describes a “time-varying machine-operating
parameter,” as recited in claim 27;
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
35
either the CMS 08 Handbook or Lichtenstein describes a “means for
delivering blood at a prescribed flow rate,” as recited in claim 28;
a person having ordinary skill in the art would have found it obvious
to control anticoagulant delivery rate with a touch screen user interface, as
required by claim 30;
a person having ordinary skill in the art would have found it obvious
to set an alarm in the CMS 08 hemodialysis machine of a critical parameter
through a touch screen user interface, as required by claim 31;
Rubalcaba discloses element (f) of claim 32;
the Sarns 9000 Manual teaches the concept of monitoring a parameter
and sounding an alarm and stopping operation when the parameter falls
outside preset limits;
Rubalcaba discloses the automatic display of a previously entered
value for user convenience, as required by element (c) of claim 34;
the Sarns 9000 Manual teaches or would have suggested element (c)
of claim 12; and
one of ordinary skill in the art would have found the subject matter of
claim 12 obvious over the applied prior art references.
DECISION
The Examiner’s decision to reject:
I. 	Claims 26-29 and 31 as unpatentable over the combined
teachings of CMS 08 Handbook, Lichtenstein, Sarns
9000 Manual, Rubalcaba, and Kerns;
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
36
II. 	Claim 30 as unpatentable over the combined teachings of
CMS 08 Handbook, Lichtenstein, Sarns 9000 Manual,
Rubalcaba, Kerns, and Thompson;
III. 	Claims 12-18 as unpatentable over the combined
teachings of CMS 08 Handbook, Lichtenstein, Sarns
9000 Manual, Rubalcaba, and Kerns;
IV. 	Claim 19 as unpatentable over the combined teachings of
CMS 08 Handbook, Lichtenstein, Sarns 9000 Manual,
Rubalcaba, and Kerns;
V. 	Claim 32 as unpatentable over the combined teachings of
CMS 08 Handbook, Lichtenstein, Sarns 9000 Manual,
Rubalcaba, and Kerns;
VI. 	Claim 33 as unpatentable over the combined teachings of
CMS 08 Handbook, Lichtenstein, Sarns 9000 Manual,
Rubalcaba, and Kerns; and
VII. 	Claim 34 as unpatentable over the combined teachings of
CMS 08 Handbook, Lichtenstein, Sarns 9000 Manual,
Rubalcaba, and Kerns,
is affirmed.
Requests for extensions of time in this ex parte reexamination
proceeding are governed by 37 C.F.R. § 1.550(c). See 37 C.F.R. § 41.50(f).
AFFIRMED
Appeal 2009-006493
Reexamination Control 90/007,751
Patent 5,247,434
37
bim
FOR PATENT OWNER:
DAVID L. CAVANAUGH
WILMER CUTLER PICKERING HALE AND DORR, LLP
1875 PENNSYLVANIA AVENUE, NW
WASHINGTON, DC 20006
FOR THIRD-PARTY REQUESTER:
JOHN C. PHILLIPS
FISH & RICHARDSON, PC
12390 EL CAMINO REAL
SAN DIEGO, CA 92130